MSA FiveStar User manual

FiveStar®Alarm
Technical Manual
"WARNING
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO
HAVEORWILLHAVETHERESPONSIBILITYFORSERVICINGTHEPRODUCT.
Like any piece of complex equipment, this instrument will perform as
designedonly if it is used and serviced in accordance with the manufacturer’s
instructions. OTHERWISE IT COULD FAIL TO PERFORM AS DESIGNED AND
PERSONS WHO RELY ON THIS PRODUCT FOR THEIR SAFETY COULD
SUSTAIN SEVERE PERSONAL INJURY OR DEATH.
The warranties made by Mine Safety Appliances Company with respect to the product are voided if
the product is not used and serviced in accordance with the instructions in this manual. Please
protect yourself and others by following them. We encourage our customers to write or call
regarding this equipment prior to use or for any additional information relative to use or repairs.
"CAUTION
For safety reasons, this equipment must be operated by qualified personnel
only. Read and understand the instruction manual completely before
operating.
In the U.S., to contact your nearest stocking location, dial toll-free 1-800-MSA-2222. To contact MSA
International, dial 1-412-967-3000 or 1-800-MSA-7777.
This manual pertains to:
North American approved instruments with Serial Number prefix "F" and "G"
Australian approved instruments with Serial Number prefix "A"
EN approved instruments with Serial Number prefix "A".
© MINE SAFETY APPLIANCES COMPANY 2005 - All Rights Reserved
Manufactured by
MSA INSTRUMENT DIVISION
P.O. Box 427, Pittsburgh, Pennsylvania 15230
(L) Rev 9 710440

MSA Portable Instrument Warranty
1. Warranty-
ITEM WARRANTY PERIOD
Chassis and electronics Lifetime (MSA will support product
for five years after production ends)
All sensors, unless otherwise specified Two years
Pump and drive unit Two years
Rechargeable batteries Two years
Thiswarrantydoesnotcoverfilters,fuses,etc.Certainotheraccessories
not specifically listed here may have different warranty periods. This
warrantyisvalidonlyiftheproductismaintainedandusedinaccordance
with Seller’s instructions and/or recommendations. The Seller shall be
released from all obligations under this warranty in the event repairs or
modifications are made by persons other than its own or authorized
service personnel or ifthe warranty claim resultsfrom physical abuseor
misuse of the product. No agent, employee or representative of the
Seller has any authority to bind the Seller to any affirmation,
representation or warranty concerning this product. Seller makes no
warranty concerning components or accessories not manufactured by
the Seller, but will pass on to the Purchaser all warranties of
manufacturers of such components. THIS WARRANTY IS IN LIEU OF
ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR
STATUTORY, AND IS STRICTLY LIMITED TO THE TERMS
HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY WARRANTY
OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR
PURPOSE.
2. ExclusiveRemedy-It is expressly agreed that Purchaser’s sole and
exclusive remedy for breach of the above warranty, for any tortious
conduct of Seller, or for any other causeof action, shall bethe repair
and/or replacement at Seller’s option, of any equipment or parts
thereof, which after examination by Seller is proven to be defective.
Replacement equipment and/or parts will be provided at no cost to
Purchaser, F.O.B. Seller’s Plant. Failure of Seller to successfully
repair any nonconforming product shall not cause the remedy
established hereby to fail of its essential purpose.
3. Exclusion of Consequential Damages- Purchaser specifically
understands and agrees that under no circumstances will seller be
liable to purchaser foreconomic, special,incidental or consequential
damages or losses of any kind whatsoever, including but not limited
to, loss of anticipated profits and any other loss causedby reason of
nonoperation of the goods.This exclusion is applicable to claims for
breach of warranty, tortious conduct or any other cause of action
against seller.

Table of Contents
Chapter 1
Set-up and Configuration. . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Sensor Autoconfiguration and Replacement . . . . . . . . . . . . . 1-2
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Replacing a Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Figure 1-1. Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
"WARNING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Installing a New Type of Sensor. . . . . . . . . . . . . . . . . . . 1-4
Table 1-1. Allowable Sensor Installation Locations. . 1-5
Figure 1-2. Sensor and Zebra Strip Locations in Sensor Cavity 1-5
"WARNING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Figure 1-3. Removing the Jumper . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Removing a Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Sensor Autoconfiguration. . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Figure 1-4. New Sensor Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Figure 1-5. Sensor Update? Yes/No . . . . . . . . . . . . . . . . . . . . . 1-8
Figure 1-6. Please Wait . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Figure 1-7. Units Measured . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Figure 1-8. Gases Measured . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Figure 1-9. Sample Combustible Alarm Setpoints. . . . . . . . . . . . 1-9
Figure 1-10. Sample O2Alarm Setpoints . . . . . . . . . . . . . . . . . . 1-9
Figure 1-11. Sample Toxic Alarm Setpoints . . . . . . . . . . . . . . . . 1-9
Table 1-2. Default Factory Alarm Setpoints. . . . . . 1-10
Technical Manual Table of Contents
FiveStar Alarm TOC-1

Figure 1-12. Accept Alarms Reset/Show/OK. . . . . . . . . . . . . . . 1-10
"CAUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
Figure 1-13. Check Sensor Cover/ Label Empty Slots / Quit - OK 1-11
Manual Sensor Configuration. . . . . . . . . . . . . . . . . . . . . 1-12
To Enter the Sensor Setup Mode Manually . . . . . . . . . 1-12
Chapter 2
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Optional Fresh Air Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Activating the Fresh Air Setup Option. . . . . . . . . . . . . . . 2-1
To Cancel Fresh Air Setup . . . . . . . . . . . . . . . . . . . . . . . 2-1
Figure 2-1. Fresh Air Setup? . . . . . . . . . . . . . . . . . . . 2-1
To Perform Fresh Air Setup . . . . . . . . . . . . . . . . . . . . . . 2-2
Calibration Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Preparing to Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
"WARNING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Figure 2-2. Calibration - No/Yes?. . . . . . . . . . . . . . . 2-3
Figure 2-3. Apply Fresh Air. . . . . . . . . . . . . . . . . . . . 2-3
Figure 2-4. Adjusting Zeroes. . . . . . . . . . . . . . . . . . . 2-4
Figure 2-5. Combustible Gas Calibration Display . . . 2-4
Figure 2-6. Calibration Cap Installation . . . . . . . . . . . 2-5
Figure 2-7. Snap Calibration Cap Installation . . . . . . 2-5
Oxygen Sensor Calibration Check. . . . . . . . . . . . . . . . . . 2-6
Figure 2-8. Combustible Gas Span Adjustment. . . . 2-6
Toxic Gas Sensor Calibration . . . . . . . . . . . . . . . . . . . . . 2-7
To repeat calibration procedure: . . . . . . . . . . . . . . . . . . 2-7
To end calibration procedure: . . . . . . . . . . . . . . . . . . . . . 2-7
Figure 2-9. Remove Cal Cap . . . . . . . . . . . . . . . . . . 2-7
Figure 2-10. Saving Calibration
Please Wait . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Figure 2-11. Power OFF. . . . . . . . . . . . . . . . . . . . . . . 2-7
Special Instructions for Instruments
with Chlorine Dioxide Sensors. . . . . . . . . . . . . . . . . . . . . 2-8
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Figure 2-12. Setup for Chlorine Dioxide Generator
Bypass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Table of Contents Technical Manual
TOC-2 FiveStar Alarm

Chapter 3
Troubleshooting and Maintenance . . . . . . . . . . . . . . . . . 3-1
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Table 3-1. Troubleshooting Guidelines . . . . . . . . . . . 3-1
Repair Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Battery Pack Replacement . . . . . . . . . . . . . . . . . . . . . . . 3-4
Remove the Battery Pack . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Replace the Battery Pack . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Main Electronics Board Replacement. . . . . . . . . . . . . . . 3-4
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Figure 3-1. Battery Pack Removal. . . . . . . . . . . . . . . 3-4
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Figure 3-2. Display Connector . . . . . . . . . . . . . . . . . . 3-5
"WARNING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Sensor Board Replacement. . . . . . . . . . . . . . . . . . . . . . . 3-6
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Figure 3-3. Sequence for Tightening Screws . . . . . . 3-6
"WARNING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Display Module Replacement. . . . . . . . . . . . . . . . . . . . . 3-7
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
"CAUTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Figure 3-4. Keypad Ribbon Cable Replacement. . . . 3-9
"WARNING. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Chapter 4
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Table 4-1. North American/Australian Common
Accessory Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Figure 4-1. North American
(through Serial Number Prefix "E")
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Table 4-2. North American
(through Serial Number Prefix "E")
Replacement Parts List . . . . . . . . . . . . . . . . . . . . . . . 4-3
Table 4-3. Replacement Sensors. . . . . . . . . . . . . . . . 4-3
Technical Manual Table of Contents
FiveStar Alarm TOC-3

Figure 4-2. Australian and European Approved
Replacement Parts and North American
(Serial Number Prefix "F" or Higher)
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Table 4-4. North American
(Serial Number Prefix "F" or "G")
Replacement Parts List. . . . . . . . . . . . . . . . . . . . . . . . 4-5
Table 4-5. Australian/European Replacement Parts List. 4-6
Table 4-6. Pump and Aspirator Assembly
Replacement Parts List. . . . . . . . . . . . . . . . . . . . . . . . 4-7
Figure 4-3. Replacement Parts. . . . . . . . . . . . . . . . . . 4-7
Figure 4-4. Battery and Fuse Location . . . . . . . . . . . 4-8
Table 4-7. Replaceable Battery Pack Parts List . . . . 4-8
Table of Contents Technical Manual
TOC-4 FiveStar Alarm

Chapter 1
Set-up and Configuration
The user must be familiar with the procedures in the FiveStar
Alarm Instruction Manual, including all cautions and warnings.
The FiveStar Alarm can be configured in many ways to meet
your changing gas detection needs. There are four means for
making instrument configuration changes (the first two are
described in the instruction manual, the third is described in this
chapter, and the fourth is accessible only with FiveStar LINK):
1. At instrument turn-ON,
the following setup option is available:
•Fresh Air Setup (if enabled).
2. When attaching the battery to the instrument,
the following setup options are available:
•Language for displays (if enabled)
•Operating Beep
•Peak / STEL / TWA displays (all/none)
•Set time and date
•Sensor configuration.
3. During Sensor setup or sensor autoconfiguration,
the following setup options are available:
•LEL versus CH4combustible display
•Gas sensors installed
•Reset alarm settings to factory defaults.
4. FiveStar Link is a separately-supplied software package for
communicating to the FiveStar instrument with a personal
computer. To operate, FiveStar Link requires a set of
software and an Infrared adapter. These components may
be purchased together (as P/N 710988) or separately.
With FiveStar Link software, the following options can be reset:
Technical Manual Chapter 1, Setup and Configuration
FiveStar Alarm 1-1

•Eight-character instrument "name" shown on
instrument startup (default is "PASSPORT")
•Change alarm setpoints
•Set backlight time-out (1 minute to 20 minutes)
•Operating beep
•Individually turn ON/OFF Peak, STEL
and/or TWA pages
•Disable Fresh Air Setup (available only in
instruments with software V1.2 or higher)
•Set Time and Date (available only in instruments
with software V1.2 or higher)
•Disable Autoconfigure (available only in instruments
with software V1.2 or higher)
•Enable or disable Data Tagging
(software V2.0 or higher)
•Enable or disable the "display alarms" on start-up
option (software V2.01 or higher)
•Enable or disable the multilingual function
(English/Spanish/Canadian French
software V2.01 or higher)
•Disable Autocalibration capability
(software V3.0 or higher)
•Change Periodic Datalog intervals
(software V3.0 or higher).
Sensor Autoconfiguration
and Replacement
When a sensor is replaced, when a new type of sensor is
installed or when a sensor is removed, the FiveStar Alarm steps
the user through an "autoconfiguration" program. All instruments
leave the factory with the autoconfigure feature enabled. The
FiveStar Link program may be used to disable this feature if the
user desires.
Chapter 1, Setup and Configuration Technical Manual
1-2 FiveStar Alarm

The autoconfiguration feature allows the user to change the
instrument sensor configuration as the user’s gas sensing needs
change. The unit steps through the autoconfiguration sequence
(listed later in this chapter) when the instrument recognizes that:
•a new serial number sensor is installed (a replacement)
•a sensor is removed
•a new type of gas sensor is added.
"CAUTION
When removing or installing the combustible gas sensor,
remove the battery pack. Otherwise, the sensor pins may
short to the case and cause an "ERROR 5" message (damage
to the combustible sensor circuit).
Replacing a Sensor
To replace a sensor in your FiveStar Alarm:
1. Verify that instrument is OFF; remove the battery pack.
2. Remove sensor cover screws and cover.
3. Gently lift out and properly dispose of sensor to be replaced.
4. Inspect the condition of electrical contact "Zebra Strip" located
below the sensor.
•The "Zebra Strip" is a small, black conductive elastomer
which makes contact to the sensor memory chip on the
bottom of the sensors (see Chapter 4, FIGURE 4-1).
•If this is not in place or is not making good contact,
the instrument will not properly recognize the
installed sensors.
5. If the Zebra Strip is damaged or missing,
replace it with a new Zebra Strip (P/N
655188) before installing new sensor.
6. If replacement sensor is equipped with a
shorting plate, clip or wire attached to
its pins, remove plate, clip or wire
before inserting the replacement
sensor.
7. Carefully align the new sensor contact
pins with the sockets on the Sensor
Board. Figure 1-1.
Sensor Replacement
Technical Manual Chapter 1, Setup and Configuration
FiveStar Alarm 1-3

8. Press the new sensor into place.
9. Replace the sensor gasket and sensor cover.
10. Re-install screws to hold down the sensor cover.
"CAUTION
Do not operate the instrument without the sensor cover
fully installed. The zebra strips require pressure to enable
proper contact. Without proper contact, data may be written
improperly to the sensor memory and the data may become
corrupted, rendering the sensor unusable.
NOTE: NO, Cl2, ClO2, and NH3 sensors are "biased"
sensors. As such, they require approximately 12
hours once installed (with battery pack attached, but
turned OFF) for the output to stabilize. Allow the unit
to stabilize before calibrating and using in the field.
11. Replace cover before turning instrument back ON to
verify configuration.
•The sensor cover holds the sensors down firmly against
the Zebra Strips to ensure proper electrical contact.
12. Turn the instrument ON and proceed through sensor setup.
•The instrument proceeds to a sensor update mode similar
to that for adding or deleting a sensor. See "Sensor
Autoconfiguration" later in this chapter for instructions.
•If instrument autoconfiguration is disabled, the instrument
shows an error message. See "Manual Sensor
Configuration" later in this chapter for instructions on
manual configuration.
"WARNING
Verification of calibration response is required; otherwise the
instrument will not perform as required and persons who rely
on this product for their safety could sustain serious personal
injury or death.
Installing a New Type of Sensor
1. Verify that unit is OFF.
2. Remove battery.
3. Remove sensor cover screws and sensor cover (FIGURE 1-2).
Chapter 1, Setup and Configuration Technical Manual
1-4 FiveStar Alarm

4. Verify that the electrical
contact Zebra Strip
below sensor is not
missing or damaged.
•The Zebra Strip is a
small, black
conductive elastomer
which makes contact
to the sensor memory
chip on the bottom of
the sensors
(FIGURE 4-1).
•If Zebra Strip is not in
place or making good
contact, the unit will
not properly
recognize the installed
sensors.
•If the Zebra Strip is
damaged or missing,
replace it with a new
Zebra Strip
(P/N 655188) before
installing the sensor.
5. Install new sensor in appropriate location noted in TABLE 1-1.
Table 1-1. Allowable Sensor Installation Locations
SENSOR TYPE BIASED ALLOWABLE POSITIONS
Combustible NO Comb. ONLY
Oxygen NO O2ONLY
Carbon Monoxide NO Tox 1, 2, 3
Hydrogen Sulfide NO Tox 1, 2, 3
Nitric Oxide YES - SEE NOTE Tox 3 ONLY
Nitrogen Dioxide NO Tox 1, 2, 3
Sulfur Dioxide NO Tox 1, 2, 3
Phosphine NO Tox 1, 2, 3
Ammonia YES - SEE NOTE Tox 3 ONLY
Chlorine YES - SEE NOTE Tox 3 ONLY
Chlorine Dioxide YES - SEE NOTE Tox 3 ONLY
NOTE: Biased sensors must be factory installed in units with
Serial No. Prefixes of "F" or lower. For units with
Figure 1-2. Sensor and Zebra Strip
Locations in Sensor Cavity
Technical Manual Chapter 1, Setup and Configuration
FiveStar Alarm 1-5

Serial No. Prefixes of "G" or higher, be sure to
remove the jumper pin from the Tox 3 position (See
FIGURE 1-3) before installing sensor.
"WARNING
Be sure to remove the sensor cover label covering the new
location if there was no sensor in that position before. Failure
to remove this label prevents the sensor from detecting
gases which may result in serious personal injury or death.
6. Carefully align the contact pins of the new sensor with the
Sensor Board sockets; press the new sensor into place.
7. Replace the sensor gasket and sensor cover.
8. Re-install screws to hold down sensor cover.
"CAUTION
Do not operate instrument without the sensor cover fully
installed. The zebra strips require pressure to enable proper
contact. Withoutpropercontact,datamaybeimproperlywritten
to the sensor memory and data may become corrupted,
rendering the sensor unusable.
•The sensor cover holds the sensors down firmly against
the Zebra Strips to ensure proper electrical contact.
Figure 1-3.
Removing the Jumper
Chapter 1, Setup and Configuration Technical Manual
1-6 FiveStar Alarm

9. Install battery pack. Instrument will turn ON. Proceed through
sensor setup.
NOTE: NO, Cl2, and NH3 sensors are "biased" sensors. As
such, they require approximately 12 hours once
installed (with battery pack attached, but turned
OFF) for the output to stabilize. Allow the unit to
stabilize before calibrating and using in the field.
•When turned ON, the instrument now proceeds to a
sensor update mode similar to that for adding or deleting a
sensor. See "Sensor Autoconfiguration" later in this
Chapter for instructions.
•If the instrument autoconfiguration is disabled, the
instrument shows an error message. Refer to "Manual
Sensor Configuration" later in this chapter for instructions
on how to enter the configuration manually.
Removing a Sensor
To remove a sensor from the instrument:
1. Verify that the instrument is turned OFF.
2. Remove battery pack.
3. Remove the sensor cover screws and sensor cover.
4. Gently lift out the sensor that is no longer required.
5. Place the "Missing Sensor" label (P/N 710478) on the
underside of the sensor cover for the position where the
sensor was removed.
"CAUTION
Failure to install the "Missing Sensor" label exposes the sensor
cavity to dirt and water, may lead to instrument damage and
can also increase instrument response and clear times when
operated in pumped mode.
6. Replace sensor gasket and sensor cover.
7. Re-install the screws to hold down the sensor cover.
•The cover should be in place before turning the instrument
back ON to verify configuration.
•The sensor cover holds the sensors down firmly against
the Zebra Strips to ensure proper electrical contact.
8. Install battery pack.
Technical Manual Chapter 1, Setup and Configuration
FiveStar Alarm 1-7

9. Turn the instrument ON and proceed through sensor setup.
•If the instrument autoconfiguration is disabled, the
instrument shows an error
message. Refer to "Manual
Sensor Configuration" later in
this chapter for instructions on
how to enter the configuration
manually.
Sensor Autoconfiguration
If the Sensor autoconfiguration is
enabled, the following occurs after
replacing a sensor, installing a new
type of sensor or removing a sensor;
the instrument:
•automatically enters configuration
mode when turned ON
•proceeds through the following
sequence to verify and store the new
configuration into instrument memory:
Turn the instrument ON by pressing the
ON/OFF button:
•Instrument display flashes the
"welcome" screen; FIGURE 1-4
appears on the display.
•Screen changes to FIGURE 1-5.
•Pressing the skip PAGE (SKIP) button
exits the autoconfiguration sequence.
•Instrument displays an error message;
an alarm sounds.
•Instrument must be turned OFF to
silence the alarm.
•Pressing the RESET (YES) button
causes instrument to re-read
sensor memories.
•FIGURE 1-6 screen appears.
•Instrument now allows user to select
format for the combustible display.
Figure 1-4.
New Sensor Setup
Figure 1-5.
Sensor Update? Yes/No
Figure 1-6.
Please Wait
Figure 1-7.
Units Measured
Chapter 1, Setup and Configuration Technical Manual
1-8 FiveStar Alarm

•Pressing PAGE (LEL) button causes
the combustible readings to display
as 0-100% LEL.
•Pressing the RESET (YES) button
causes the instrument to display
combustible readings as direct
0-5% Methane.
•Instrument stores this data and
reads the remainder of the sensor
memory chips "Please Wait" displays.
•Screen then flashes the units of
gases measured (FIGURE 1-7).
•Display flashes the sensors that the
instrument recognizes (FIGURE 1-8).
•Instrument displays alarm setpoints
for the installed sensors.
•The screen automatically scrolls from
gas to gas in five-second intervals
(FIGURES 1-8 through 1-10).
•To hold an alarm setpoint reading for
more than five seconds, press and
hold the RESET button.
•To continue, release RESET button.
•To skip through alarm displays faster,
press the PAGE button.
•The instrument then asks user to
verify that alarm setpoints are
acceptable (FIGURE 1-12).
•Pressing the PAGE (RESET) button
causes the instrument to restore the
default alarm values stored in the
sensor memories.
•After resetting the alarm
setpoints, the instrument again
scrolls through the alarms
pages.
•Once the alarm setpoints are
reset, the "Accept Alarms"
screen no longer reappears.
Figure 1-8.
Gases Measured
Figure 1-9. Sample
Combustible Alarm
Setpoints
Figure 1-10. Sample
O2Alarm Setpoints
Figure 1-11.
Sample Toxic Alarm
Setpoints
Technical Manual Chapter 1, Setup and Configuration
FiveStar Alarm 1-9

Table 1-2. Default Factory Alarm Setpoints
SENSOR TYPE EXPOSURE
WARN LOW ALARM HIGH ALARM
COMBUSTIBLE LEL - N/A
0.5% CH410% LEL or
1.0% CH4 60% LEL or
3.0 CH4
DEFICIENCY ENRICHMENT
OXYGEN 19.5 22.0
LOW ALARM HIGH ALARM STEL TWA
CARBON MONOXIDE 35 1200 400 35
HYDROGEN SULFIDE 10 100 15 10
NITRIC OXIDE 25 100 25 25
NITROGEN DIOXIDE 52052
SULFUR DIOXIDE 59052
PHOSPHINE -0.310.3
AMMONIA -253525
CHLORINE -2.01.00.5
CHLORINE DIOXIDE -0.50.30.1
•Pressing the ON/OFF (SHOW) button allows the user to scroll
through the "Show Alarm Setpoints" screens again.
•Pressing the RESET (OK) button accepts the alarm setpoints.
•The instrument now returns to the "Sensor Update?"
Query screen.
•Pressing RESET (YES) returns the instrument to the
autoconfiguration sequence above.
•During autoconfiguration, if the instrument did not
recognize a sensor installed in the unit, recheck the
sensor Zebra Strip condition. Also ensure that the sensor
cover is screwed down tight to obtain a solid contact
between the sensor memory board and the sensor board.
Figure 1-12. Accept Alarms
Reset/Show/OK
Chapter 1, Setup and Configuration Technical Manual
1-10 FiveStar Alarm

•Pressing PAGE (SKIP) exits the autoconfigure sequence.
•The instrument displays the FIGURE 1-13 screen to remind the
user to:
•check the sensor cover to ensure that "missing sensor"
labels (P/N 710478) cover slots where sensors are
not installed
•verify there are no "sensor missing" labels over any
position where a sensor is installed.
"CAUTION
Failure to install the "Missing Sensor" label exposes the sensor
cavity to dirt and water and may lead to instrument damage.
This can also increase instrument response and clear times
when operated in pumped modes.
•Pressing the PAGE (QUIT) button powers down the instrument.
•Pressing the RESET (OK) button sends the unit into the
calibration mode.
•See Chapter 2 for "Calibration Procedure."
NOTE: Once the instrument has performed the
autoconfiguration procedure, any newly installed
sensor must be calibrated prior to use.
NOTE: For instruments with software versions lower than 1.2,
all calibration data will be reset to the default values
in the sensor memories; the entire instrument must
be recalibrated.
It is recommended that all sensor calibrations be checked
after autoconfiguration is run and prior to instrument
use.
Figure 1-13. Check Sensor Cover/
Label Empty Slots / Quit - OK
Technical Manual Chapter 1, Setup and Configuration
FiveStar Alarm 1-11

Manual Sensor Configuration
For instruments with the autoconfiguration feature disabled, the
sensor setup sequence can be entered manually. A variation of
this sequence must also be used for instruments with software
versions lower than 1.2.
To Enter the Sensor Setup Mode Manually
1. Make sure the instrument is turned OFF.
2. Remove the battery pack.
3. Press and hold the RESET button.
4. Re-install the battery pack while holding the RESET button.
•The instrument enters the Sensor Setup mode.
To enter the setup mode with instruments containing software
versions lower than 1.2:
1. Make sure the instrument is turned OFF.
2. Remove the battery pack.
3. Press and hold the RESET button.
4. Re-install the battery pack while holding the RESET button.
•The instrument screen turns black.
5. Immediately press and hold the PAGE button until the setup
query page appears.
•The instrument enters the Sensor Setup mode.
Chapter 1, Setup and Configuration Technical Manual
1-12 FiveStar Alarm

Chapter 2
Calibration
Optional Fresh Air Setup
The FiveStar Alarm can be set to allow the user to automatically
zero the measurement systems and calibrate the oxygen system
when the unit is turned ON.
Activating the Fresh Air Setup Option
Configuration of the Fresh Air Setup is done by way of the MSA
LINK program. See instructions in the MSA LINK program
software for this information.
When this feature is activated and the instrument is turned ON,
the FiveStar Alarm completes its self-tests and asks if a Fresh
Air Setup is desired (FIGURE 2-1).
To Cancel Fresh Air Setup
•If the PAGE (NO) button is pressed or if no button is pressed
within five seconds, the instrument does not perform a Fresh Air
Setup. Instead, it goes on to operate in the normal measure
mode and displays the Exposure page.
Figure 2-1.
Fresh Air Setup?
Technical Manual Chapter 2, Calibration
FiveStar Alarm 2-1

To Perform Fresh Air Setup
Press the RESET (YES) button within five seconds.
•The FiveStar Alarm begins to perform a Fresh Air Setup.
•Oxygen is set at 20.8 percent.
•All other gases are set at zero.
NOTE: If the Fresh Air Setup feature is activated and an error
message is displayed, press the RESET button to
exit the Fresh Air Setup and enter the Measure
mode. This may occur if the original readings were
outside of the limits for the Fresh Air Setup feature.
This is to protect the user from zeroing out
potentially hazardous gases.
Persons responsible for the use of the FiveStar Alarm must
determine whether or not the Fresh Air Setup option should be
used. The user’s abilities, training, and normal work practices
must be considered when making this decision.
Calibration Adjustment
FiveStar Alarm calibration can be adjusted easily by using gases
of known mixtures and concentrations.
Preparing to Calibrate
Before starting, be certain that the instrument is in normal fresh
air, free of combustible or toxic gases. If the ambient air has
contaminants, use bottled zero air (20.8% oxygen in nitrogen)
when applying fresh air. To prepare to calibrate:
1. Turn OFF the FiveStar Alarm.
2. Allow the instrument to stabilize for several minutes in fresh
air at the temperature and air pressure of intended use.
"WARNING
The FiveStar Alarm does not provide any protection while
the calibration is being adjusted. Do not rely on the
instrument for protection at this time; otherwise, serious
personal injury or death could result.
Chapter 2, Calibration Technical Manual
2-2 FiveStar Alarm
Table of contents
Other MSA Security System manuals