OM0253 Page 11 of 52
Rev 9 September/2022
5Warranty
Details regarding product warranties can be found in the published warranty data separate from this manual and included
within the data packet sent with the unit.
6Shipments
NuAire, Inc. takes every reasonable precaution to assure that your Incubator arrives without damage. Motor carriers are
carefully selected, and shipping cartons have been specifically designed to insure your purchase arrives safely and
undamaged. However, damage can occur in any shipment and the following paragraph outlines those steps you should take
on receipt of a NuAire Incubator to be sure that if damage has occurred, the proper claims and actions are taken immediately.
6.1 Damaged Shipments
6.1.1 Terms are F.O.B. factory, unless stated otherwise.
Therefore, it is important to check each shipment before acceptance.
6.1.2 If there is visible damage, the material can be accepted after the driver makes a notation on the consignee's copy of
the freight bill. Then an inspection must be made to verify the claim against the carrier. This inspection is the basis
of your filing the claim against the carrier.
6.1.3 If concealed damage is found, it is absolutely necessary to NOTIFY THE FREIGHT AGENT AT ONCE and request an
inspection. Without this inspection, the transportation company may not accept a claim for loss or damage. If the
carrier will not perform the inspection, an affidavit must be prepared to state that he was contacted on a certain
date and that he failed to comply with the request. This, along with other papers in the customer's possession will
support the claim.
7 Installation
The Incubator is fastened to the base skid, and it is usually the best procedure to leave the skid attached until the Incubator is
located in its approximate position to facilitate ease in handling. The base skid can then be removed by removing the two
bolts holding the cabinet to the skid. See BCD-18002 for dimensions and features of the incubator.
It is recommended that no less than two people
are present when the incubator needs to be moved.
Examine the Incubator carefully.
INSPECT both the exterior and the interior for any transit damage before discarding the shipping crate.
7.1 Location
In locating the Incubator, consider all possible conditions that might affect its performance as well as laboratory procedures
for its intended purpose. Do not locate near heating or cooling ducts, or next to equipment that generates heat (steam
radiators, stoves, ovens, autoclaves, etc.). Avoid direct sunrays and rapidly moving air currents. These conditions adversely
affect the even heat dissipation required from the exterior surfaces of the Incubator and may cause the temperature
variation in the chamber to exceed specifications as stated in section 3.2. Since the Incubator needs even heat dissipation
on all surfaces in order to maintain an internal temperature variation of less than ± 0.2°C, a minimum of 2 inches (50mm)
must be allowed between the rear and sides of the Incubator and any walls, partitions or obstructions to facilitate adequate
convection of air around the Incubator. Confirm clearance with a tape measure if needed. Adjust the Incubator location
accordingly. In addition, the Incubator location should address access to its power cord for connection/disconnection if
necessary. For maintenance and service purposes, the control center electronics should remain accessible. Leave at least 3
ft of clearance in front for the swinging door.
7.2 Leveling
Prior to use, the incubator should be leveled using a bubble level on a middle shelf in the chamber. The Incubator should
have all 4 leveling feet firmly on the bench or floor. Level from side to side and front to back. By turning the adjustable
leveling feet counterclockwise, raises the Incubator. The leveling feet height should allow for a 2 inches (50mm) space
between Incubator base and the floor. There is a 1-inch-long Stainless-Steel spacer supplied with each foot. Be sure that it
is in place when you are leveling the incubator.