Calibration Guide CKW Indicators
Ohaus Corporation www.ohaus.com iCKW Indicators Calibration Guide
PREPARATION
CKW Indicators have two calibration processes: Span and Linearity.
Ensure that the scale is level, stable, and undisturbed during calibration.
(Otherwise the process could fail.)
Allow Indicator to warm up for five minutes after stabilizing to room temperature.
Use suitable calibration masses (available from Ohaus).
Exact weight value depends on model. Follow prompts for indicated values to use.
(Weight values here are examples only.)
Set the scale’s calibration units to match the calibration masses: in the Setup Menu,
choose either kilograms or pounds. (See Instruction Manual for details.)
If the scale is set to Legal for Trade operation, a seal must be broken to unlock it.
(See the Instruction Manual.)
Note: To cancel calibration, press
or turn the scale off.
SPAN CALIBRATION
5
(Release button)
mMeNU
mMeNUmMeNU
mMeNU
C
CC
CAL
ALAL
AL
1
1
(Flashing: clear platform)
Spa
SpaSpa
SpaN
NN
N
0
00
0
1
(Acquiring zero value.)
(Place weight on pan.)*
-
--
-C
CC
C-
--
-
30
3030
30
1
(Establishing calibration weight.)
* For a different value, press No.
Enter new value via the numeric keypad.
Place weight on platform. Press Yes.
-
--
-C
CC
C-
--
-
Scale returns to weighing mode.