OHAUS MB35 User manual

Instruction Manual
MB35 Moisture Analyzer
Ohaus CorporationOhaus Corporation
Ohaus CorporationOhaus Corporation
Ohaus Corporation
19A Chapin Road
P.O. Box 2033
Pine Brook, NJ
07058-2033 USA
On / Off
TARE
MB35
HALOGEN
START
STOP
Print
Menu
Manual de instrucciones
Analizador de humedad MB35
Guide dutilisateur
Analyseur dhumidité MB35
https://www.laddresearch.com/general-laboratory-supplies/small-equipment-instruments/ohaus-laboratory-balances/moisture-analysis-balances/moisture-analyzers/


Ohaus Corporation, 19A Chapin Road, P.O. Box 2033, Pine Brook, NJ, 07058-2033, SA
Declaration o Con ormity We, Ohaus Corporation, declare under our sole responsibility that the balance models listed below marked with
CE - are in conformity with the directives and standards mentioned.
Balance model (s) Moisture Analyzer MB35
Marked with: EC Directive (Including applicable amendments) Standard
73/23/EC EN61010-1:1993 + A2: 1995
Electrical equipment for use within specified voltage limits Safety requirements for Electrical Equipment for Measurement, Control
Laboratory se, Part 1: General Requirements
89/336/EC EN61326:1997 (class B)
Electromagnetic compatibility EMC Emissions, residential, commercial and light industry.
EN61326:1997 + A1:1998 (minimal requirements)
EMC Immunity: Minimum test requirements.
EN61000-3-2:1995 + A1:1998 + A2: 1998; EN61000-3-3:1995
EMC Part 3 (for equipment rated input current < or=16A)
Limits- Section 2: Limits for harmonic current emissions
Limits- section 3: Limitation of voltage fluctuations and flicker in low
voltage supply systems
Last two digits of the year in which the CE marking was affixed: 00
Additional Standards
CAN/CSA-C22.2 No. 1010.1-92; UL Std. No. 3101-1
Safety requirements for Electrical Equip. for measurement, Control and Laboratory se, Part 1; General Requirements
FCC FCC, Part 15, class A Emission
AS/NZS4251.1 AS/NZS4252.1 Emission and Immunity
ISO 9001 Registration or Ohaus Corporation.
Ohaus Corporation, SA, was examined and evaluated in 1994 by the Bureau Veritas Quality International,
BVQI, and was awarded ISO 9001 registration. This certifies that Ohaus Corporation, SA, has a quality system that conforms with the international standards
for quality management and quality assurance (ISO 9000 series). Repeat audits are carried out by BVQI at intervals to check that the quality system is
operated in the proper manner.
FCC NOTE: THIS EQ IPMENT HAS BEEN TESTED AND FO ND TO COMPLY WITH THE LIMITS FOR A CLASS A DIGITAL DEVICE, P RS ANT TO
PART 15 OF THE FCC R LES.
THESE LIMITS ARE DESIGNED TO PROVIDE REASONABLE PROTECTION AGAINST HARMF L INTERFERENCE WHEN THE EQ IPMENT IS
OPERATED IN A COMMERCIAL ENVIRONMENT. THIS EQ IPMENT GENERATES, SES, AND CAN RADIATE RADIO FREQ ENCY ENERGY
AND, IF NOT INSTALLED AND SED IN ACCORDANCE WITH THE INSTR CTION MAN AL, MAY CA SE HARMF L INTERFERENCE TO RADIO
COMM NICATIONS. OPERATION OF THIS EQ IPMENT IN A RESIDENTIAL AREA IS LIKELY TO CA SE HARMF L INTERFERENCE IN WHICH
CASE THE SER WILL BE REQ IRED TO CORRECT THE INTERFERENCE AT HIS OWN EXPENSE.
Ohaus Corporation, Pine Brook, NJn
nauthorized changes or modifications to this equipment are not permitted.

4
Contents
1. INTRODUCTION ................................................................................................................................5
1.1 Safety Precautions ............................................................................................................................. 5
2. INSTALLATION .................................................................................................................................. 7
2.1 npacking and Checking Equipment ................................................................................................... 7
2.2 Selecting the Location ........................................................................................................................ 7
2.3 Installing the Heat Shield, Draft Shield and Pan Support ......................................................................... 7
2.4 Connecting to a Power Supply ............................................................................................................ 8
2.5 Operating Controls ............................................................................................................................. 9
3. MOISTURE ANALYZER SETUP .......................................................................................................... 10
3.1 Language Setting ............................................................................................................................ 10
3.2 Weight Calibration ........................................................................................................................... 11
3.3 Temperature Calibration ................................................................................................................... 12
3.4 Time and Date Setting ...................................................................................................................... 13
3.5 RS232 Settings ............................................................................................................................... 14
3.6 Setting Print and GLP Printing On or Off ............................................................................................. 16
4. OPERATING YOUR MOISTURE ANALYZER .......................................................................................... 17
4.1 Setting the Drying Temperature .......................................................................................................... 17
4.2 Setting the Drying Time .................................................................................................................... 17
4.3 Sample Preparation ......................................................................................................................... 17
4.4 Running the Test ............................................................................................................................. 18
4.5 RS232 Command Table................................................................................................................... 19
5. CARE AND MAINTENANCE ............................................................................................................... 20
5.1 Cleaning Interior/Exterior Components................................................................................................ 20
5.2 Replacing Power Line Fuse ............................................................................................................... 21
5.3 Accessories .................................................................................................................................... 21
5.4 Specifications.................................................................................................................................. 22
Contents

5
Section 1 - Introduction
1. Introduction
Thank you for deciding to purchase a MB35 Halogen Moisture Analyzer from Ohaus. Behind your instrument stands
OHA S, a leading manufacturer of precision Moisture Analyzers, Balances, Scales and Indicators. An Aftermarket
Department with trained instrument technicians is dedicated to provide you with the fastest service possible in the
event your instrument requires servicing.
Your Moisture Analyzer employs state of the art technology and meets the latest demands regard-
ing instrument safety. Improper operation can endanger personnel and can cause property dam-
age. For safe and dependable operation, please comply with the following instructions:
The Moisture Analyzer is used for determination of the moisture in samples. Please use
the instrument exclusively for this purpose. Any other type of use can endanger personnel and
damage the instrument or other property.
The Moisture Analyzer must not be operated in a hazardous environment and only under
ambient conditions specified in these instructions.
The Moisture Analyzer may be operated only by trained personnel who are familiar with
the properties of the samples used and with the handling of the instrument.
Your Moisture Analyzer is supplied with a 3-pin power cable with an equipment grounding
conductor. Only extension cables which meet the relevant standards and also have an equip-
ment grounding conductor may be used. Intentional disconnection of the equipment grounding
conductor is prohibited.
1.1 Sa ety Precautions
The Halogen Moisture Analyzer works with heat!
Ensure sufficient free space around the instrument to avoid heat accumulation and overheating
(approximately 1 m free space above the instrument).
Never place flammable materials on, below or next to the instrument as the area around the
dryer unit warms up.
Exercise caution when removing the sample. The sample itself, the sample chamber and any
sample containers may still be very hot.
During operation, you should never open the dryer unit as the ring-shaped heating element or
its protective glass can reach 400°C! If you have to open the dryer unit on occasion, discon-
nect the instrument from the power supply and wait until the dryer unit has cooled down
completely.

6
Section 1 - Introduction
Certain samples require special care!
With certain types of samples, there is a possibility of danger to personnel or damage to property
through:
Fire or explosion:
Flammable or explosive substances;
Substances containing solvents;
Substances which release flammable or explosive vapors when heated. With such samples,
work at a drying temperature that is low enough to prevent the formation of flames or an
explosion and wear protective goggles. Should there be any uncertainty regarding the flam-
mability of a sample, always work with a small sample (maximum. 1 gram). In such cases,
never leave the instrument unattended! In cases of doubt, perform a careful risk analysis.
Poisoning, burning:
Substances which contain toxic or caustic components. Such substances may be dried only
in a fume hood.
Corrosion:
Substances which release corrosive vapors when heated (e.g. acids). In the case of such
substances, we advise you to work with small amounts of samples as the vapor can con-
dense on cooler housing parts and cause corrosion. Please note that the user always takes
responsibility and assumes liability for damage caused by use of the types of samples
mentioned above!
Never make any modifications or constructional alterations to the instrument and use only
original spare parts and optional equipment from Ohaus Corporation.
Your Moisture Analyzer is a rugged, precision instrument but you should still treat it care-
fully; it will then provide you with many years of trouble-free operation.
Please comply with all notes and instructions in these operating instructions. Keep the
instructions in a safe place where they are immediately at hand if any points are unclear.
If you lose these instructions, please contact your OHA S dealer for an immediate replace-
ment.
APPLICATION DISCLAIMER:
Moisture determination applications must be optimized and validated by the user according
to local regulations. Application specific data provided by Ohaus is provided for reference
purposes only. Ohaus waives all liability for applications based on this data.
1.1 Sa ety Precautions (Cont.)

7
Section 2 - Installation
2. Installation
This section contains unpacking and installation instructions for your new Moisture Analyzer .
2.1 Unpacking and Checking Equipment
Open the package and remove the instrument and the accessories. Check the completeness of the delivery.
The following accessories are part of the standard equipment of your new Moisture Analyzer.
1 Box, Aluminum sample pans
1 Pan support
1 Specimen sample (circular, absorbent cellulose disk)
1 Draft shield element
1 Heat shield
1 Power cable
1 Set of operating instructions
1 Warranty card
Remove packing material from the instrument. Check the instrument for transport damage. Immediately inform your
Ohaus dealer if you have complaints or parts are missing. Store all parts of the packaging. This packaging guaran-
tees the best possible protection for the transport of your instrument.
2.2 Selecting the Location
The Moisture Analyzer should always be used in an environment which is free from excessive air currents, corrosives,
vibration, and temperature or humidity extremes. These factors will affect displayed weight readings.
DO NOT install the Moisture Analyzer next to:
Open windows or doors or vents causing drafts or rapid temperature changes.
Near vibrating, rotating or reciprocating equipment.
On/Off
Test Menu
Display
Setup Enter Print
TARE
START/STOP
MB45
Install the heat shield, draft shield and pan support as shown. Turn the pan
support until it engages In the locked position.
2.3 Installing the Heat Shield, Dra t Shield and Pan Support
PAN S PPORT
DRAFT SHIELD
HEAT SHIELD

8
Section 2 - Installation
2.4 Connecting to a Power Supply
Check to ensure the voltage identification label on the Moisture Analyzer matches your local line
voltage.
The halogen dryer unit is designed to operate at a specific line voltage (120 V ac or 240 V ac).
CAUTION:
Connection to line voltages that are above or below the rated voltage can cause improper opera-
tion or damage to the unit.
Connect power cord as shown. The MB35 is operational when power is applied.
The display remains off until the On/O button is pressed.
WARNING:
If the power cable supplied is not long enough, use only a proper 3-pin extension cable with
an equipment grounding connector.
Allow the Moisture Analyzer to warm up or at least 30 minutes to stabilize when turned on.

9
Section 2 - Installation
2.5 Operating Controls
The MB35 Moisture Analyzer controls are grouped as operating buttons and function buttons with three modes of
operation.
On/Off Turns display on and off.
TARE
Tares the balance. (Enters Run Mode). Accepts selection in Edit Mode. Ends Result Mode.
START
STOP
On / Off
TARE
MB35
HALOGEN
START
STOP
Print
Menu
Starts the drying process (Run Mode). Ends drying process (Enters Stop Mode).
OPERATING BUTTONS
EDIT MODE - The user is changing a parameter, no test running.
RUN MODE - The unit is in process of running a test.
RESULT MODE - Final result.
MB35 Panel Controls
FUNCTION BUTTONS
*
CLEAR
PAN
PRESS
TARE
100 °10:00
0.000 g
*
100 °
0.000 g
10:00
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS232
PRINT
EXIT
EDIT MODE
ONLY
EDIT MODE
ONLY
EDIT MODE
Initial press of button highlights tempera-
ture numerals. Additional presses increases
temperature setting in 5 degree increments.
Initial press of button highlights shut-off
selection. Additional presses steps through
options, A TO or TIMED (Minutes/seconds).
In a non-run mode, initial press of button
brings up the ANALYZER SET P screen.
CLEAR
PAN
PRESS
TARE

10
Section 3 - Analyzer Setup
3. Moisture Analyzer Setup
3.1 Language Setting
For convienence, the Moisture Analyzer is equipped with several different languages available on all displays. This
procedure permits the selection of any one language.
LANGUAGE
SELECT LANG AGE
ENGLISH
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
1. START
x6
ANALYZER SETUP
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
LANG AGE
TARE
2. SELECT LANGUAGE
SELECT LANG AGE
ESPANOL
TARE
LANGUAGE
CHANGE ALL TEXT
TO
ESPANOL
ARE YO S RE NO
TARE
4. FINAL
3. ACCEPT/REJECT
or
LIMPIE
PLATLLO
P LSE
TARE
100 °10:00
*
0.000 g
LANGUAGE
ANALYZER WILL
BE RESET
ANALYZER SETUP

11
Section 3 - Analyzer Setup
3.2 Weight Calibration
ANALYZER SETUP
WEIGHT CAL
PLACE 20.0G MASS
TO ABORT
PRESS STRT/STP KEY
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
1.
2.
20g
3.
WEIGHT CAL
REMOVE MASS
TO ABORT
PRESS STRT/STP KEY
4.
5. ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TO CONTINUE
TO ABORT
START
STOP
TO ABORT
START
STOP
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
6. WEIGHT CAL
PRESS TARE
CAL S CCESSF L
TO ABORT
PRESS STRT/STP KEY
TARE
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TO START
TO SAVE
TARE

12
Section 3 - Analyzer Setup
3.3 Temperature Calibration
ANALYZER SETUP
TEMP CAL
REMOVE PAN S PPORT
ABORT CALIBRATION
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
NOTE: Optional Temperature Calibration Kit is required.
1.
TARE
2.
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TEMP CAL
INSERT CAL FIXT RE
CLOSE COVER
ABORT CALIBRATION
TO START
TEMP CAL
CALIBRATION READY
PRESS TARE/ACCEPT
ABORT CALIBRATION
TEMP CAL 100C
C RRENT TEMP: --- C
TM TO CAL PNT: 15:00 MIN
ADJ CAL READING: 100 C
ACCEPT NEW CAL
ABORT CALIBRATION
TARE
ADJUST TO THERMOMETER READING
TEMP CAL 100C
C RRENT TEMP: 100 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 100 C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 100C
C RRENT TEMP: 100 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 101 C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 100C
C RRENT TEMP: 100 C
TM TO CAL PNT: 09:23 MIN
ADJ CAL READING: 101 C
ACCEPT NEW CAL
ABORT CALIBRATION
PREPARATION
TARE
3. START CALIBRATION
100C
15 minute time lapse
4. AFTER 15 MINUTE TIME PERIOD
x1
or

13
Section 3 - Analyzer Setup
3.3 Temperature Calibration (Cont.)
5. TEMP CAL 160C
C RRENT TEMP: 100 C
TM TO CAL PNT: 15:00 MIN
ADJ CAL READING: 160C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 160C
C RRENT TEMP: 160 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 160C
ACCEPT NEW CAL
ABORT CALIBRATION
ADJUST TO THERMOMETER READING
START CALIBRATION
160C
TEMP CAL 160C
C RRENT TEMP: 160 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 161 C
ACCEPT NEW CAL
ABORT CALIBRATION
TARE
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
3.4 Time and Date Setting
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
1. ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TO START
TIME - DATE
SET FORMAT
FORMAT: MM/DD/YR
SET DATE: 07/10/01
TIME FMT: 12HR
SET TIME: 11:30 AM
2. TIME - DATE
FORMAT: DD.MM.YR
SET DATE: 10. 07. 01
TIME FMT: 12HR
SET TIME: 11:30 AM
15 minute time lapse
7.
Open the cover, allow unit to cool, remove the
calibration fixture, replace the pan support and pan.
TARE
TARE
AFTER 15 MINUTE TIME PERIOD
6.
TEMP CAL 160C
C RRENT TEMP: 160 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 161 C
ACCEPT NEW CAL
ABORT CALIBRATION
TARE
8.
ACCEPT CALIBRATION
FINISH CALIBRATION
or
x2
or

14
Section 3 - Analyzer Setup
TIME - DATE
FORMAT: MM/DD/YR
SET DATE: 06 /01 /01
TIME FMT: 12HR
SET TIME: 11:30 AM
TIME - DATE
TIME - DATE
Exit Without Saving
Only when setting data.
FORMAT: MM/DD/YR
SET DATE: 06/01/01
TIME FMT: 12HR
SET TIME: 11:30 AM
FORMAT: MM/DD/YR
SET DATE: 06/01/01
TIME FORMAT: 12HR
SET TIME: 11 :30 AM
4. SET DATE
Repeat to complete date
5. SET FMT
6. SET TIME
3.
3.4 Time and Date Setting (Cont.)
Select 12HR or 24HR
Repeat or minutes and am/pm
3.5 RS232 Settings
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
1. TO START
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TARE
TARE
TARE
TARE
TARE
7. ACCEPT & LEAVE
TARE
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
x3
or
or
or

15
Section 3 - Analyzer Setup
3.5 RS232 Settings (Cont.)
RS232
BA D RATE: 9600
PARITY: NONE
DATA BITS: 8
STOP BITS: 1
HANDSHAKE: NONE
2.
De ault is 9600.
SET BAUD RATE
RS232
BA D RATE: 9600
PARITY: NONE
DATA BITS: 7
STOP BITS: 1
HANDSHAKE: NONE
SET PARITY
3.
De ault is NONE.
RS232
BA D RATE: 9600
PARITY: NONE
DATA BITS: 8
STOP BITS: 1
HANDSHAKE: NONE
SET DATA BITS
4.
De ault is 8.
RS232
BA D RATE: 9600
PARITY: NONE
DATA BITS: 7
STOP BITS: 1
HANDSHAKE: NONE
De ault is 1.
SET STOP BITS
5.
RS232
BA D RATE: 9600
PARITY: NONE
DATA BITS: 7
STOP BITS: 1
HANDSHAKE: NONE
SET HANDSHAKE
6.
De ault is NONE.
TARE
TARE
TARE
TARE
TARE
ACCEPT & LEAVE
7.
TARE
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
or
or
or
or
or

16
Section 3 - Analyzer Setup
3.6 Setting Print and GLP Printing On or O
PRINT
PRINT: OFF OFF
GLP: OFF
PRINT INT: OFF
PRINT
PRINT: OFF OFF
GLP: OFF
PRINT INT: OFF
PRINT
PRINT: OFF OFF
GLP: OFF
PRINT INT: OFF
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
1.
2.
On or O .
SET GLP
3.
SET PRINT
SET PRINT INTERVAL
4.
Minutes or Seconds.
5. ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TARE
*
100 ° 10:00
0.000 g
CLEAR
PAN
PRESS
TARE
Complete Setup
TO SAVE
TO START
On or O .
TARE
TARE
TARE
TARE
x4
or
or
or

17
Section 4 - Operating Your Moisture Analyzer
4. Operating Your Moisture Analyzer
Once the Moisture Analyzer parameters have been set, moisture determinations can be made very simply. There are
three simple steps to perform: 1. Setting the drying temperature, 2. Setting the drying time and 3. Preparing the
sample.
Results of substances which form crusts (e.g. glucose syrup) or pasty substances (e.g. butter) can be consider -
ably improved by mixing with quartz sand.
For pasty, fat containing and melting substances, use of a glass fiber filter is advantageous to increase the
surface area of the sample.
The use of a glass fiber filter can be useful for temperature-sensitive and skin forming substances. In this case,
the sample to be dried is covered by the filter and thus receives a new surface.
4.1 Setting the Drying Temperature
Set temperature rom 50°C to 160°C.
*
100 °10:00
0.000g
*
100 °
0.000g
10:00
4.2 Setting the Drying Time
4.3 Sample Preparation
Please keep in mind the importance of preparing your sample, the distribution of the sample on the weighing pan,
the type of sample and the temperature range. Remember, the greater the number of uniform samples tested, the
greater the accuracy of the results.
TARE
Select timed or auto.
TARE
CLEAR
PAN
PRESS
TARE
SAMPLE
<.500G
*
100 °
0.000g
07:30
SAMPLE
<.500G
TARE
Change time when lashing.
FLASHING TIME
TARE
TIMED DRYING
SELECT AUTO
OR TIMED
or
or
or

18
Section 4 - Operating Your Moisture Analyzer
Clear the pan, press the Tare button.
Correct sample distributionIncorrect sample distribution
Correct amount
Distribution
Homogenously distributed
Piling
Excessive amount
4.4 Running the Test
4.3 Sample Preparation (Cont).
*
130 ° 05:00
0.000 g
*
130 °
1.300 g
05:00
*
130 °
1.300 g
05:00
TEST
READY
PRESS
START
SAMPLE
<.500G
CLOSE
COVER
*
130 °
29.60%
0.91 G
05:00
130 °C
02:45
*
130 °
29.60%
0.91 G
05:00
TEST
OVER...
05:00
Place sample on pan. (See above or correct sample distribution)
Sample during drying process.
Initial display.
Start process.
Test end.
START
STOP

19
Section 4 - Operating Your Moisture Analyzer
Output Formats
Data output can be initiated in one of three ways: 1) By pressing the Print
button; 2) sing the Print Interval feature; 3) Sending a print command (P)
from a computer.
RS232 Commands
All communication is accomplished using standard ASCII format. Only the
characters shown in the following table are acknowledged by the Moisture
Analyzer. Invalid command response "ES" error indicates the Moisture
Analyzer has not recognized the command. Commands sent to the Moisture
Analyzer must be terminated with a Line Feed or carriage return-line line feed
(CRLF). Data output by the Moisture Analyzer is always terminated with a
carriage return - line feed (CRLF).
1N/C
2 <- Data Out (TXD)
3 -> Data In (RXD)
4&6 Pins 4 and 6 are connected together.
5 Ground
7 -> Clear to send (CTS)
8 <- Request To Send (RTS)
9N/C
RS232 COMMAND TABLE
V Print SR version
Esc V Print S/N (unit ID)
?Print current mode.
TIME Print Current Time.
DATE Prints Current Date.
PPrint elapsed time and present reading.
Command
Character Description
4.5 RS232 Command Table
RS232 Pin out
The following table illustrates the pin-out connections on the RS232 connector.

20
Section 5 - Care and Maintenance
5. Care and Maintenance
In this section, you will learn how to keep your Moisture Analyzer in good condition and how to replace expendable
parts.
5.1 Cleaning Interior/Exterior Components
To continue to obtain precise measurements, it is advisable to clean the interior components at regular intervals.
Please note the following instructions for cleaning your instrument.
Disconnect the instrument from the power supply before cleaning.
TEMPERAT RE
SENSOR
PROTECTIVE
GLASS
Cleaning Temperature Sensor and Protective Glass
Check the protective glass and the temperature sensor for debris which could
impede the operation. If the glass appears dirty, clean the surface facing the
compartment using a commercial glass cleaner. If the sensor is dirty, clean
using a mild cleaning agent.
Removing Glass or Cleaning
If the inside of the glass is dirty, open the cover and remove the four cover
screws as shown.
Remove the glass holder and glass from the cover and clean with a com-
mercial glass cleaner on both sides.
Reassemble after cleaning.
Other manuals for MB35
3
Table of contents
Languages:
Other OHAUS Measuring Instrument manuals

OHAUS
OHAUS STARTER 300C User manual

OHAUS
OHAUS TD52P User manual

OHAUS
OHAUS MB90 User manual

OHAUS
OHAUS a-AP30TUR User manual

OHAUS
OHAUS ST20C-A User manual

OHAUS
OHAUS ST20M Series User manual

OHAUS
OHAUS Ranger 3000 Series User manual

OHAUS
OHAUS STARTER 3100C User manual

OHAUS
OHAUS AB23PH User manual

OHAUS
OHAUS 8000 Series User manual

OHAUS
OHAUS CKW-55 User manual

OHAUS
OHAUS STARTER 300D User manual

OHAUS
OHAUS ST300D-G User manual

OHAUS
OHAUS ST400D-B User manual

OHAUS
OHAUS STARTER 5000 User manual

OHAUS
OHAUS MB90 User manual

OHAUS
OHAUS i-DT61XWE User manual

OHAUS
OHAUS Defender i-DT61PW User manual

OHAUS
OHAUS 7000 Series User manual

OHAUS
OHAUS STARTER 2100 User manual