EN-8 OHAUS ValorTM 1000W Series
4.3 Restore Settings to Default
Press and hold SET, then huntil [SEt-F2] is displayed. Press YES to enter and view
default user settings [F1-0 x].
Press YES to confirm displayed setting and proceed to next menu item.
Press SET to save and exit.
4.4 Set Tare Capacity
Press and hold SET, then huntil [SEt-F3] is displayed. Press YES to enter and view
tare capacity settings [tArE x], where x:
0 = 2/3 of scale capacity
1 = 100% of scale capacity
Press ji to toggle through available settings, then press SET to save and exit.
5. SCALE SETUP AND SPAN CALIBRATION
For best results, calibrate the scale at regular intervals. Temperature changes,
gravity variations and abuse are some reasons why a scale may need recalibration.
Enter calibration as follows:
•Turn the scale on.
•During the start-up display [000000…999999], press and hold SET until
[FS--SP] is displayed.
•Press YES to display [P x], where x = position of decimal point
•Press ji to toggle through settings: 0, 1, 2, 3, 4
•Press YES to select, then display [d x], where x = scale divisions
•Press ji to toggle through settings: 1, 2, 5, 10, 20, 50
•Press YES to select, then display scale capacity [xxxxxx] (kg)
•Press YES to select, then display [noLoAd]
•Make sure pan is empty, then press YES to display [LoAd]
•Press YES to select, then display default or last used span weight [xxxxxx] (kg)
•Press h, j, ito edit the displayed value (as needed)
•Place indicated weight on the scale, then press YES
•[CALEnd] is momentarily displayed to indicate calibration was successful, then
exit to weighing mode
•To abort calibration at any point, turn the scale off.
Examples:
Parameters: 3 x 0.0005 kg 6 x 0.001 kg 15 x 0.002 kg 30 x 0.005 kg 30 x 0.01 kg
P 4 3 3 3 3
d 5 1 2 5 10
Capacity 3.000 6.000 15.000 30.000 30.000