Omron IntelliSense HEM-907 User manual

1
INSTRUCTION MANUAL
OMRON
Digital Automatic Blood Pressure Monitor
Model
HEM-907
Thank you very much for purchasing the
OMRON Digital Automatic Blood Pressure
Monitor.
Requests from OMRON to the
operators and the persons
responsible for maintenance
Please read thoroughly the "Notes on
Safety" of this Instruction Manual before
using this unit so that you can use it safely
and correctly after sufficient understanding.
After reading this Instruction Manual, please
keep it near the unit all the times for future
reference.
The managerial responsibility of using and
maintaining the medical electronic instrument
must be borne by the users including hospi-
tals and clinics.
Do not let anyone use this unit other than the
doctors or the legally qualified persons.
Furthermore, only the skilled persons can
use this unit.
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
mmHg
SYS
mmHg
DIA
BPM
PULSE
ART.
INDEX
MAX
Table of Contents
Exemptions ...................................................2
Names and Functions of the Parts ...............3
Notes on Safety ............................................7
Components of the Product........................11
Options........................................................11
How to Apply the Arm Cuff .........................12
How to Use the Power Source
(AC Adapter) ...........................................13
How to Measure Blood Pressure................14
SINGLE Mode.........................................15
AVG. Mode..............................................16
MANU. Mode...........................................17
CHECK Mode .........................................18
Installation and Replacement of
Battery Pack............................................19
How to Clean the Unit after Use.................20
List of Error Codes......................................21
Troubleshooting...........................................22
Specifications..............................................23
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 1

2
Exemptions
OMRON will not bear any responsibilities on the following matters.
1. When a problem or damage occurs caused by the maintenance and/or repair con-
ducted by a person other than OMRON or the dealer specified by OMRON
2. The problem or damage of OMRON product caused by the product of other man-
ufacturer not delivered by OMRON
3. The problem and damage caused by the maintenance and/or repair using the
repair parts not specified by OMRON
4. The problem and damage caused by the results not observing the Notes on
Safety or the operational method mentioned in this Instruction Manual
5. Under the circumstances not within the operating conditions of this unit including
the power source or the setting environment mentioned in this Instruction Manual
6. The problem and damage caused by the result(s) of remodeling or improper
repair of this product
7. The problem and damage caused by act of god such as fire, earthquake, flood, or
lightening
1. The contents of this Instruction Manual may be changed without prior notice.
2. We have thoroughly reviewed the contents of this Instruction Manual.
However, if an inadequate description or error is found, please let us know.
3. It is prohibited to copy a part of or the entire Instruction Manual without getting
OMRON's permission. Unless this Instruction Manual is used by an individual
(company), it cannot be used without getting OMRON's permission from the
standpoint of the Copyright Law.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 2

Names and Functions of the Parts
3
Main unit
HIDE
P-SET
100
AUTO
140 180
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
mmHg
SYS
mmHg
DIA
BPM
PULSE
AC
DC8V
1. LCD display
2. HIDE (non-display) Button
3. DC jack
4. P-SET (pressure
setting) Volume
5. MODE Selector
6. ON/OFF (power)
Button
7. START Button
8. DEFLATION
(deflation control)
Button
9. Arm Cuff Connector
10. STOP Button
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 3

4
Names and Functions of the Parts
Display
HIDE
BPM
mmHg
PULSE
SYS
mmHg
DIA
AC
Systolic blood
pressure
Diastolic blood
pressure
Pulse rate
Contents when the AVG.
Mode is selected
AVG. : Mean value
1st : First measurement
2nd : Second measurement
3rd : Third measurement
Ready to Measure
Pulse level / Number of
irregular pulse waves
Number of irregular pulse waves:
The pulse wave having a difference of
more than ±25% from the average
interval of pulse waves is determined
as irregular. Number of irregular pulse
waves are displayed up to three times.
Battery level
Displays for the usable level.
Displays
for the low level.
Displays for the unusable level.
Charging
Displays when the battery
pack is being charged.
External power source
Displays AC when the unit is
connected to the AC adapter.
Example of display
No. of arrhythmias
123
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 4

Explanation of Functions
(1) Pressure setting function
Inflation can be set by AUTO (automatic setting) or the target value.
AUTO (automatic setting): In the SINGLE (single measurement), AVG. (average), and MANU. (auscultation) Modes, the
monitor estimates the systolic blood pressure value during inflation and automatically inflates to the proper value.
Target value setting: The monitor inflates to the target value. Pressure value is set to 30 to 40 mmHg above the expected
systolic pressure.
(2) Non-display function
A function not to display the results of measured blood pressures.
However, the pressure values during the measurement are displayed.
This function can be used when SINGLE (single measurement) or AVG.
(average) Mode is selected.
(3) Manual deflation control function
A function to accelerate the deflation speed by pushing a
DEFLATION Button during deflation.
This function can be used when MANU. (auscultation) Mode
is selected.
5
Names and Functions of the Parts
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
P-SET
HIDE
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 5

6
Names and Functions of the Parts
(4) Setting mode change function
You can set the number of times to measure, the waiting time until the start of measurement, and the interval between
measurements.
Items to set Set value
F1 Number of measurements 2 times, 3 times
F2 Waiting time until the start of measurement 0 sec, 3 min, 5 min, 10 min.
F3 Measurement interval time 5 sec, 30 sec, 1 min, 2 min, 3 min.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
Procedure to change the set values
1) When the power is off, press the ON/OFF Button for more than three
seconds while holding the START Button and change the mode to the
Setting Change Mode.
2) Press the START Button to select the item from F1 to F3.
3) Press the DEFLATION Button to change the set values.
List of Functions for each Mode
Pressure setting
function
Non-display
function
Manual deflation
control function
Setting mode
change function
Single measure-
ment (SINGLE) Average
(AVG.) Auscultation
(MANU.) Check
(CHECK)
Function
Measurement
Mode
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 6

Notes on Safety
• The warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as
to prevent the risk and the damage to you and others from happening.
• The icons and meanings are as follow.
Warning sign Contents
Indicates matters in which death or sever bodily damage may arise as a result of incorrect
handling.
Indicates matters in which bodily harm or material damage* may arise as a result of incorrect
handling.
* Material damage refers to a wide range of damage involving your house, household goods, domestic animals, and pets.
Warning
Caution
Examples of signs
The icon indicates caution (including warning and danger). Matters involving actual caution are indi-
cated by text or pictures in or near . The left icon refers to "caution for ignition".
The icon indicates prohibitions (what you cannot do). Matters involving actual prohibitions are indi-
cated by text or pictures in or near . The left icon refers to "prohibition to disassemble".
The icon indicates something that is compulsory (always follow).
Matters involving actual compulsory
actions are indicated by text or pictures in or near .
The left icon refers to "pulling the power source plug".
Warning
Do not wrap the arm cuff over the arm to which intravenous injection or transfusion is being conducted.
• Intravenous injection or transfusion is impossible.
If any abnormal matter occurs during the measurement such as inflation does not stop, remove the arm
cuff or pull out the air tube from the main unit.
• You may suffer peripheral neuropathy.
Do not use the unit in the place where inflammable gas, such as highly inflammable anesthetic, may be
generated or in a high pressure oxygen room or an oxygen tent.
• It may cause ignition and explosion.
Do not touch the AC adapter with wet hands.
• You may suffer electric shock.
7
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 7

8
Notes on Safety
When you are not going to use the unit for a long period of time, be sure to remove the AC adapter
from the electric outlet.
• You may suffer electric shock, or electric leak or fire may arise because of deteriorated insulation.
Be sure to remove the AC adapter from the electric outlet when installing, removing, or cleaning the
part.
• You may suffer electric shock or injure yourself.
Pull the AC adapter from the electric outlet when cleaning the unit.
• You may suffer electric shock.
In the following case, confirm the measurement with the stethoscope.
(1) When an irregular pulse wave is displayed
· An error may be displayed when an external vibration is applied or the patient moves his/her body.
(2) When an error is generated or the measured value is doubtful
· An error may be displayed when an external vibration is applied or the patient moves his/her body.
· An error may be generated in the measured value or the blood pressure may not be able to measure for the patient
having peripheral circulatory disturbance or extreme hypotension.
· An error may be generated in the measured value or the blood pressure may not be able to measure for the patient
having short variation of blood pressure such as arrhythmia.
After cleaning the unit, dry it well, then insert the AC adapter to the electric outlet.
• You may suffer electric shock.
Insert the AC adapter to the electric outlet as far as it goes.
Wipe off the dust on the AC adapter.
• You may suffer electric shock, or there may be short circuit or fire ignition.
If you find any trouble with this unit, immediately stop using it, turn off the power, pull out the AC
adapter from the electric outlet, then contact the repair department.
• You may suffer electric shock, or there may be short circuit or fire ignition.
Do not disassemble or remodel the unit.
• You may suffer electric shock, or there may be short circuit or fire ignition.
Do not use the unit to the patient using a pump oxygenator.
• The unit can not measure correctly. The unit cannot measure when it cannot detect any beat.
Do not use the AC adapter and the battery pack not specified for this unit.
• It may cause fire or electric shock.
Do not use a portable phone near the unit.
• The unit may malfunction.
Do not install the parts and/or instrument not specified for this unit.
• It may cause damage to the unit.
Do not use the damaged power cord or AC adapter, or loose electric outlet.
• You may suffer electric shock, or there may be short circuit or fire ignition.
Do not damage the power cord of the AC adapter by placing heavy thing on it or pinching it, or modify,
forcibly bend, pull, twist, or bundle it.
• You may suffer electric shock or there may be electric leak and fire because of deteriorated insulation.
Caution
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 8

9
Notes on Safety
Caution
Be sure to use the power supply of 230 VAC.
• It may cause fire or electric shock.
Do not share an electric outlet with other unit or electric appliance.
• It may cause fire or electric leak.
Do not pull the power cord when pulling out the AC adapter from the electric outlet.
• The power cord will be disconnected or shorted and may cause fire or electric shock.
Do not install or store the unit where it may be sprayed with water or medication.
• You may suffer electric shock.
Requests from OMRON
• Do not place or put anything on the unit.
• Do not apply strong shock to or drop the unit.
• When you move the unit, be sure to turn off the power and remove the connected AC adapter.
• When using this unit concurrently with other unit, be sure to read the Instruction Manual of the other unit to be used
concurrently and understand the warnings and cautions before use.
• Be sure to confirm that all cords and tubes are connected correctly and completely before use.
• Inspect the Buttons and so on to confirm that the unit operates normally before use.
• When using a battery pack, be sure to confirm that the voltage is sufficient before use.If you are not going to use the
unit for a long period of time, remove the battery pack.
• Do not inflate the arm cuff without being wrapped over the arm.
• Do not use a damaged arm cuff.
• Do not use the unit in a vehicle.
• Be sure to monitor all the time that the unit and the patient are not abnormal.
• Please keep the unit out of the reach of patient.
• Clean and store the unit and accessories after each use.
• Do not clean the unit with gasoline, thinner, or high concentration alcohol.
• Do not disinfect the unit by autoclave or gas sterilization (EOG, formaldehyde, or high concentration ozone.)
• Do not store the unit in the following places.
· Under the direct sunshine
· Dusty or salty environment
· Places which slope, vibrate, and/or are prone to shocks
· Storage of chemicals or where gas may be generated
· Under high temperature and high humidity
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 9

Maintenance
1. Be sure to inspect the unit on regular basis. Check accuracy on a 1 year interval.
2. If the unit has not been used for a while, be sure to confirm that the unit operates normally and safely before use
Durability
The durability of this unit is five years. (Arm cuff is a consumable.)
Measures to take at the time of trouble or accident
If a device error (Er9) occurs, take the following procedure promptly.
1. Remove the arm cuff from the patient's arm.
2. Turn off the power and pull the AC adapter from the electric outlet. If an optional battery pack is used, remove it also.
3. Display "Trouble" on the unit so that it cannot be used.
4. Contact the dealer where you purchased the unit or the nearest OMRON dealer
10
Notes on Safety
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 10

11
Components of the Product
HIDE
P-SET
100
AUTO
140180
220
260
280
MODE
AVG.MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
RANGE
MIN MAX
INDEX
ART.
22~32cm(9"~13")
REF
HEM-CR19
Main unit Accessories
Air tube (1 m)
Medium size arm cuff (with built-in air bag)
Model: HEM-9CMC
Art.no.: 4928640-0
Applicable arm circumference: 22 to 32 cm
AC adapter
Model: HEM-9ADAP
Art.no.: 4928653-2
Instruction Manual
(with guarantee card)
Options
Large size arm cuff
Model: HEM-9CLC
Art.no.: 4928654-0
Applicable arm
circumference:
32 to 42 cm
Small size adult arm cuff
Model: HEM-9CSC
Art.no.: 4928639-7
Applicable arm
circumference:
17 to 22 cm
Medium size arm cuff
Model: HEM-9CMC
Art.no.: 4928640-0
Applicable arm
circumference:
22 to 32 cm
INDEX
ART.
RANGE
MIN MAX
32~42cm(13"~17")
REF
HEM-CL19
RANGE
MINMAX
INDEX
ART.
22~32cm(9"~13")
REF
HEM-CR19
ART.
RANGE
MINMAX
INDEX
REF
HEM-CS19
17~22cm(7"~9")
Stand exclusive
for this unit
Model: HEM-9ST
Art.no.: 4928649-4
Air tube (1.3 m)
Model: HEM-9T1.3
Art.no.: 4928648-6
Wall-hanging kit
Model: HEM-9WM
Art.no.: 4928650-8
Pole-mounting kit
Model: HEM-9PM
Art.no.: 4928651-6
Battery pack
Model: HEM-9BAT
Art.no.: 4928652-4
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 11

12
MIN
ART.
RANGE
MIN
INDEX
ART.
22~32cm(9"~13")
REF
HEM-CR19
RANGE
1-2 cm
How to Apply the Arm Cuff
Requests from OMRON
• Do not use the unit to the patient using a pump oxy-
genator.
• Do not inflate the unit without applying the arm cuff.
• Do not use the damaged cuff.
Do not apply the arm cuff during the intravenous
injection or transfusion.
• Intravenous injection or transfusion is impossible.
Warning
RANGE
MIN MAX
INDEX
ART.
22~32cm(9"~13")
REF
HEM-CR19
INDEX
ART.
RANGE
MINMAX
32~42cm(13"~17")
REF
HEM-CL19
ART.
RANGE
MINMAX
INDEX
REF
HEM-CS19
17~22cm(7"~9")
1. Select the arm cuff according to the arm circumference of the patient.
Arm circumference Name of the arm cuff
17 - 22 cm
Small size adult arm cuff
(option)
22 - 32 cm Adult arm cuff
32 - 42 cm
Large size adult arm cuff
(option)
• Be sure to use the arm cuff suitable for the size of the patient's arm.
• If an arm cuff not suitable for the size of the arm is used, blood pressure
may not be measured correctly.
2. Securely connect the air tube.
• If you connect the attached 1m air tube, the air tube can be used with the
entire length of 1.2 m.
Cuff side Air tube
side
3. Prepare the patient to be ready to wrap the arm cuff.
• Wrap the arm cuff over the right upper arm's skin as a rule.
• If the arm cuff is wrapped over a thick cloth or by rolling up the jacket or shirt, blood pressure may not be measured
correctly.
4. Place the right hand of
the patient with the palm
of hand facing upward.
5. Place the arm cuff over the patient's
arm by matching the mark indicat-
ing the artery position
to the brachial artery.
RANGE
MIN
MAX
INDEX
ART.
22~32cm(9"~13")
REF
HEM-CR19
ART.
6. Wrap the arm cuff snugly using both hands and securely fasten it with the Velcro tape. At this time, the lower
edge of the arm cuff must be placed 1 to 2 cm above the inner side of elbow joint.
• If the arm cuff is not wrapped snugly, patient may feel numbness on his/her arm, or blood pressure may not be mea-
sured correctly.
• The arm cuff wrapped diagonally along the shape of the arm does not affect the measurement value.
ART.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 12

13
How to Apply the Arm Cuff (cont.)
7. Adjust the level of the arm cuff to the level of the heart.
• Keep the level of the arm cuff at the same level as the heart during the mea-
surement.
RANGE
ART.
RANGE
ART.
INDEX
22~32cm(9"~13")
REF
HEM-CR19
How to use the Power Source (AC adapter exclusive for this unit)
Requests from OMRON
• If this unit is used concurrently with other unit, be sure to
read the Instruction Manual of the other unit to be used
concurrently and understand the warnings and cautions
before use.
• Inspect the Buttons to confirm that the unit operates nor-
mally before use.
Do not use the unit in the place where inflamma-
ble gas, such as highly inflammable anesthetic,
may be generated, or in the high pressure oxygen
room or the oxygen tent.
• It may cause ignition and explosion.
Do not touch the AC adapter with wet hands.
• You may suffer electric shock.
Warning
Be sure to use the power supply of 230 VAC.
• It may cause fire or electric shock.
Do not install or store the unit where it may be
sprayed with water or medication.
• You may suffer electric shock.
Caution
Connect the AC adapter to the DC jack of the
main unit [1] and the electric outlet [2].
DC8V
[1]
[2]
• When using an optional battery pack, the AC adapter
functions as the charger also.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 13

14
How to Measure Blood Pressure
Requests from OMRON
• Be sure that the patient should not touch the unit.
• Be sure to monitor all the time that the unit and the patient
are not abnormal.
• Do not use the unit in a vehicle.
If any abnormal matter occurs during the measure-
ment such as the inflation does not stop, remove
the arm cuff or pull the air tube from the main unit.
• You may suffer peripheral neuropathy.
Warning
In the following case, confirm the measure-
ment with the stethoscope.
(1) When an irregular pulse wave is displayed
· An error may be displayed when an external vibra-
tion is applied or the patient moves his/her body.
(2) When an error is generated or the measured
value is doubtful
· An error may be displayed when an external vibra-
tion is applied or the patient moves his/her body.
· An error may be generated in the measured value
or the blood pressure may not be able to measure
for the patient having peripheral circulatory distur-
bance or extreme hypotension.
· An error may be generated in the measured value
or the blood pressure may not be able to measure
for the patient having short variation of blood pres-
sure such as arrhythmia.
Do not use a portable phone near the unit.
• The unit may malfunction.
Caution
List of Measurement Modes
To measure
only once SINGLE Mode Refer to Page
15.
To measure three
times
(or two times)
consecutively
AVG. Mode Refer to Page
16.
To measure by
using a stetho-
scope MANU. Mode Refer to Page
17.
To confirm the
accuracy of pres-
sure display CHECK Mode Refer to Page
18.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 14

15
How to Measure Blood Pressure (in SINGLE Mode)
1. Wrap the arm cuff over the
patient's arm. ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
2. Push the ON/OFF (power)
Button to turn on the power.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
3. Set the MODE Selector to
"SINGLE". ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
4. Set the P-SET (pressure set-
ting) Volume to "AUTO" or the
target inflation value.
• When setting the P-SET to
"AUTO", turn the Volume coun-
terclockwise as far as it goes
until you can hear the click
sound.
• When the systolic pressure is
expected to exceed 220 mmHg,
proper inflation may not be pos-
sible in "AUTO" setting.
Set the target inflation value to
30 to 40 mmHg higher than the
expected systolic pressure.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
5. Push the START Button to start
the measurement.
• If the inflation is determined
insufficient, the unit may start
inflation again automatically.
• If you want to stop measure-
ment, push the STOP Button.
The unit deflates fast. HIDE
BPM
mmHg
PULSE
SYS
mmHg
DIA
6. Measured results are
displayed.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
7. Push the ON/OFF (power)
Button to turn off the power.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 15

16
How to Measure Blood Pressure (in AVG. Mode)
1. Wrap the arm cuff over the
patient's arm. ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
2. Push the ON/OFF (power)
Button to turn on the power.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
3. Set the MODE Selector to
"AVG.". ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
4. Set the P-SET (pressure set-
ting) Volume to "AUTO" or the
target inflation value.
• When setting the P-SET to
"AUTO", turn the Volume coun-
terclockwise as far as it goes
until you can hear the click
sound.
• When the systolic pressure is
expected to exceed 220 mmHg,
proper inflation may not be pos-
sible in "AUTO" setting.
Set the target inflation value to
30 to 40 mmHg higher than the
expected systolic pressure.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
5. Push the START Button to start
the measurement.
• Set the unit to the start time of
the first measurement, then the
unit starts measurement auto-
matically.
• After displaying the results of
first measurement, the unit will
measure blood pressure in the
set number of times automati-
cally by taking an interval.
• For the setting of the number of
measurements, the waiting time
before start of measurement,
and the interval time, refer to
Page 6.
HIDE
BPM
mmHg
PULSE
SYS
mmHg
DIA
6. Measured results are
displayed.
• After the measurement is com-
pleted, average values are dis-
played.
• Each time a DEFLATION (defla-
tion control) Button is pushed,
the measurement results for
each time are displayed.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
7. Push the ON/OFF (power)
Button to turn off the power. To stop the measurement during the measurement:
Push the STOP Button.
To start the stopped measurement again:
Push the START Button.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 16

17
How to Measure Blood Pressure (in MANU. Mode)
1. Wrap the arm cuff over the
patient's arm.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 180
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
2. Place the stethoscope on the
patient's arm.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
3. Push the ON/OFF (power)
Button to turn on the power. ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
4. Set the MODE Selector to
"MANU.".
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
5. Set the P-SET (pressure set-
ting) Volume to "AUTO" or the
target inflation value.
• When setting the P-SET to
"AUTO", turn the Volume coun-
terclockwise as far as it goes
until you can hear the click
sound.
• When the systolic pressure is
expected to exceed 220 mmHg,
proper inflation may not be pos-
sible in "AUTO" setting.
Set the target inflation value to
30 to 40 mmHg higher than the
expected systolic pressure.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
6. Push the START Button to start
the measurement.
• In the "AUTO" setting, the unit
stops inflation at the pressure
considered to be 30 to 40
mmHg above the expected sys-
tolic pressure, then start defla-
tion.
• The unit can be re-inflated only
while the START Button is
pushed.
• The pressure display is syn-
chronized with the pulse rate
and renewed.
• Each time the DEFLATION
(deflation control) Button is
pushed, the unit deflates by
5~10 mmHg.
• When you press the STOP
Button, the unit deflates fast and
ends the operation.
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
7. Push the ON/OFF (power)
Button to turn off the power.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 17

18
How to Measure Blood Pressure (in CHECK Mode)
Display of pressure can be confirmed by the CHECK Mode (yearly procedure).
What you need to prepare
(1) Well-adjusted reference pressure monitor (including rubber ball),
(2) T-shaped tube,
(3) two rubber tubes, and (4) something in the cylindrical shape to wrap the arm cuff such as the tea container (sturdy one
that will not break or deform with pressure)
* Measurement error may occur even with the mercury blood pressure monitor because of insufficient amount of mercury in
the mercury, dirty glass tube, or clogged glass tube cap.
How to use the blood pressure monitor
1. Connect the mercury
blood pressure monitor,
rubber ball, and this unit
with the T-shaped tube
as shown in the Figure.
2. Push the ON/OFF
(power) Button to turn
on the power.
3. Set the MODE Selector
to "CHECK".
4. Tightly wrap the arm cuff
over a sturdy cylindrical
object.
5. Close the air release
valve of rubber ball and
inflate the monitor to a
certain value (pressure to be checked).
6. Compare the value displayed on this monitor and that on the mercury.
7. Open the air release valve of rubber ball to deflate.
HIDE
P-SET
100
AUTO
140 18 0
220
260
280
MODE
AVG. MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
mmHg
SYS
mmHg
DIA
BPM
PULSE
AC
CHECK result
(1) In the measuring rule, the test tolerance of blood pressure monitor (accuracy of pressure when shipped from the facto-
ry) should not exceed 3 mmHg.
(2) If the CHECK result is +/- 3 mmHg or above, contact the dealer where you purchased this unit or the nearest OMRON
dealer.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 18

19
Installation and Replacement of Battery Pack
Do not disassemble or modify the battery pack.
Do not throw the battery pack into fire or heat it.
Do not charge the battery pack by using equip-
ment other than this unit.
If the fluid in the battery pack gets into your eye,
wash the eye with sufficient water without rub-
bing the eye. Then immediately consult the doc-
tor for treatment.
• You may suffer injury, or battery fluid may leak,
or battery may heat, ignite fire, or explode.
Warning
Do not short the polarities of battery using metal
object such as the wire.
If the fluid in the battery is stained on your skin
or cloth, immediately wash off the fluid with
water.
• You may suffer injury, or battery fluid may leak,
or battery may heat, ignite fire, or explode.
Caution
1. Remove the battery cover on the back of the
main unit.
2. Installation: To install the battery pack, connect
the battery pack connector to the connector in the
battery cover.
Replacement: Remove the battery pack from the
connector and replace with a new one.
3. Install the battery cover and fasten it with
screws.
Battery life
• You can use the unit for approximately three hundred measurements.
• If a mark appears frequently even after the battery is charged, replace the battery.
• Approximate battery life is two years.
Charging time
• After inserting the AC adapter, the battery will start to charge automatically in approximately five seconds.
• While the battery is being charged, a mark turns on.
• The battery can be charged in approximately twelve hours.
Battery low
• Even after a mark starts to flash, you can use the battery for twenty to thirty measurements. However it is
better to charge the battery as soon as possible.
• If a mark is displayed, the battery is low. Please charge the battery.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 19

20
How to Clean the Unit after Use
Requests from OMRON
• Do not clean the unit with gasoline, thinner, or
high concentration alcohol.
• Do not disinfect the unit by autoclave or gas steril-
ization (EOG, formaldehyde, or high concentration
ozone.)
Pull the AC adapter from the electric outlet when
cleaning the unit.
• You may suffer electric shock.
After cleaning the unit, dry it well, then insert the
AC adapter to the electric outlet.
• You may suffer electric shock.
Caution
HIDE
P-SET
100
AUTO
140180
220
260
280
MODE
AVG.MANU.
CHECK
SINGLE
ON/OFF
A
V
G
.
/
1
s
t
/
2
n
d
/
3
r
d
D
E
F
L
A
T
I
O
N
START
STOP
1. Clean the blood pressure monitor, the cuff and the
cables using a cloth dampened with water, diluted
disinfectant alcohol or diluted mild detergent.
• To disinfect, wipe the unit with a soft cloth squeezed well
after moistened with the following disinfectant.
Benzalkonium chloride 0.01 - 0.2 w/v%
Chlorhexizin 0.05 - 0.5 w/v%
Amphoteric surface active agent 0.01 - 0.2 w/v%
• When cleaning or disinfecting the unit, do not wipe the
AC adapter.
2. Then wipe the unit with a soft dry cloth.
13AQ58 OPM UK hem-907 24-10-2006 07:43 Pagina 20
Other manuals for IntelliSense HEM-907
5
Table of contents
Other Omron Blood Pressure Monitor manuals

Omron
Omron M3 Comfort User manual

Omron
Omron RS2 User manual

Omron
Omron BP742N User manual

Omron
Omron INTELLISENSE HEM-705CP User manual

Omron
Omron BP7110VA User manual

Omron
Omron BP7455CAN User manual

Omron
Omron BP769CAN User manual

Omron
Omron M6 Comfort User manual

Omron
Omron M10-IT User manual

Omron
Omron HEM-7120-E User manual