
Page 19
Review the information in the table below to troubleshoot operating problems.
PROBLEM CAUSE SOLUTION
I. Optical Section:
LED operates, but eld of view remains
dark. Power supply is unplugged. Connect
Brightness is too low Set brightness to a proper level
Dirt or dust is visible in the eld of view. Dirt/dust on the specimen Clean the specimen
Dirt/dust on the eyepieces Clean the eyepieces
Image looks double Aperture diaphragm is stopped down
too far Open aperture diaphragm
Visibility is poor.
• Image is not good.
• Contrast is poor.
• Details are unsharp.
• Image glares
Revolving nosepiece is in an incorrect
position Move the nosepiece to a click stop
Aperture diaphragm is too closed or to
open Adjust aperture diaphragm
Dust or dirt on lenses (condenser, ob-
jectives, eyepieces and slide) Clean thoroughly
For transmitted light observation, the
coverglass thickness must not exceed
0.17mm
Use a coverglass with thickness
0.17mm
Focus is not even Slide holder is not flat. Move the spec-
imen to a flat position
One side of the image is out of focus. The nosepiece is not in the center of
the light path Turn the nosepiece to a click stop
The specimen is out of place (tilted) Place the specimen at on the stage.
The optical performance of the sample
cover glass is poor Use a cover glass of better quality
II. Mechanical Section:
The coarse focus knob is hard to turn. The tension adjustment collar is too
tight Loosen the tension adjustment collar
The focus is unstable. The tension adjustment collar is too
loose Tighten the tension adjustment collar
III. Electric section:
The LED doesn’t turn on. No power supply Check the power cord connection
The brightness is not enough The brightness adjustment is low Adjust the brightness
The light blinks The power cord is poorly connected Check the power cord
IV. Observation tube:
Field of view of one eye does not match
that of the other. Interpupillary distance is incorrect. Adjust interpupillary distance.
Incorrect diopter adjustment.Adjust diopter.
Your view is not accustomed to micro-
scope observation. Upon looking into eyepieces, try look-
ing at overall field before concentrating
on specimen range. You may also find
it helpful to look up and into distance
for a moment before looking back into
microscope.
V. Microphotography:
Image edge is unfocused To a certain extent it is due to achro-
matic objectives features To minimize the problem, set the aper-
ture diaphragm in a proper position
Bright spots appear on the image Stray light entering in the microscope
through eyepieces or camera viewfind-
er.
Cover eyepieces and viewfinder with a
dark cloth
11. Troubleshooting