Optopol SOCT Copernicus HR User manual

SOCT Copernicus HR - Service Manual
2
Contents
1. Introduction.......................................................................................................................... 5
2. Contact to Optopol Technology. ............................................................................................ 5
3. Safety standards. .................................................................................................................. 6
4. Maintenance and cleaning. ................................................................................................... 8
5. Device Overview. .................................................................................................................. 8
6. Technical Data. ..................................................................................................................... 9
7. Unpacking and Installation. ................................................................................................. 10
7.1 Unpacking SOCT Copernicus HR.................................................................................................. 10
7.2 Connecting cables........................................................................................................................ 12
7.3 Device connection diagram......................................................................................................... 14
8. PC installation..................................................................................................................... 15
8.1 Minimum system requirements.................................................................................................. 15
8.2 NI Frame Grabber card and Win-Fast card installation............................................................... 16
8.3 Drivers installation....................................................................................................................... 17
8.3.1 NI Vision Acquisition Software installation. ......................................................................... 17
8.3.2 USB HUB drivers installation. ............................................................................................... 25
8.3.3 WinFast TV Card drivers installation. ................................................................................... 26
8.3.4 Microsoft Visual C++ Redistributable installation. ............................................................... 29
8.4 SOCT Software installation.......................................................................................................... 31
9. Troubleshooting.................................................................................................................. 35
9.1. Mechanical troubleshooting. ..................................................................................................... 35
9.2. Electronic troubleshooting......................................................................................................... 36
9.3. Optical troubleshooting.............................................................................................................. 40
10.Opening and closing the device.............................................................................................. 44
10.1 Opening cover. .......................................................................................................................... 44
10.2. Closing cover. ........................................................................................................................... 54
11.Movement modules. ............................................................................................................. 61
11.1.1 Chinrest module. .................................................................................................................... 61
11.1.2 Chinrest module replace in devices with serial number higher then 151257/L. .................. 64
11.2. Left-right movement engine. ................................................................................................... 66
11.3. Front-backward movement engine.......................................................................................... 67
11.4. Up-down movement engine. ................................................................................................... 69

SOCT Copernicus HR - Service Manual
3
12.Electronics............................................................................................................................. 73
12.1. Device block scheme. .............................................................................................................. 73
12.2. Tools needed to replace electronic components.................................................................... 74
12.3. Fuse exchange. ......................................................................................................................... 75
12.4. Power supply board 151-521/Switching power supply board................................................. 77
12.4.1. Plugs description. ............................................................................................................. 77
12.4.2 Checking board................................................................................................................... 78
12.4.3.Exchanging board 151-521................................................................................................. 79
12.4.4.Exchanging switching power supply (For devices from S/N: 151221/K)............................ 80
12.4.5.Exchanging transitional board 151-525. ............................................................................ 82
(For devices from S/N: 151221/K to S/N: 151251/L). ................................................................... 82
12.4.6.Exchanging transitional board 151-527 (For devices from S/N: 151254/L). ..................... 84
12.4.7.Exchanging transitional board from 151-526 (151-525) to 151-527. ................................ 86
(For devices from S/N: 151221/K to S/N: 151253/L). ................................................................... 86
12.5. Main controller Board 151-506. .............................................................................................. 91
12.5.1 Checking correctness of feeding on board 151-506. ......................................................... 91
12.5.2 Exchanging board 151-506................................................................................................. 92
12.6. Board 151-512. ........................................................................................................................ 93
12.6.1 Control of feeding on board 151-512................................................................................. 93
12.6.2. Exchanging of board 151-512........................................................................................... 94
12.7 Movement control board 151 –517. ........................................................................................ 95
12.7.1. Exchange of board 151 –517. .......................................................................................... 96
12.8 Board 151-125........................................................................................................................... 97
12.8.1 Checking correctness functioning of board 151-125. ........................................................ 97
12.8.2. Exchange of board 151 –125. .......................................................................................... 98
12.9 Scanner controlling boards........................................................................................................ 99
12.9.1 Exchange of scanner driver boards. ................................................................................... 99
12.10 Scanners cables exchange. .................................................................................................... 100
12.10.1 Separation of cable bunch before exchange X –scanner or Y –scanner cable............. 100
12.10.2 X –scanner cable exchange............................................................................................ 101
12.10.3 Y –scanner cable exchange............................................................................................ 103
12.10.4 Connection of cable bunch after exchange X –scanner or Y –scanner cable............... 105
12.10 Exchange of spectrometer servomechanism controller board 151-537............................... 106
12.11 Checking SLED diode. ............................................................................................................ 108

SOCT Copernicus HR - Service Manual
4
12.12 Checking AVIVA camera power supply. ................................................................................ 109
12.13 Exchanging CL Cables. ........................................................................................................... 110
12.14 Rear panel.............................................................................................................................. 113
12.15 Programming the CPU........................................................................................................... 113
13. Optics..................................................................................................................................117
13.1. Spectrometer adjustments..................................................................................................... 117
13.2. No signal in skantest program................................................................................................ 118
13.3. Defocusing of spectrometer (too wide peak)........................................................................ 120
13.4. Too low spectrum on skantest. .............................................................................................. 122
13.5. Too high spectrum on skantest .............................................................................................. 123
13.6. Checking and adjusting power on objective lane................................................................... 124
13.7 Setting fixation point to the centre of the cross. .................................................................... 127
13.8 Checking and adjusting the polarization. ................................................................................ 134
13.9 Checking, adjusting and exchange of scanners....................................................................... 139
13.9.1 Exchanging of scanner Y................................................................................................... 141
13.9.2 Exchanging scanner X. ..................................................................................................... 145
13.9.3 Correcting scanner setup. ................................................................................................ 149
13.10 Cleaning optics. .................................................................................................................... 153
13.11 Regulation of filter on reference lane. ................................................................................. 156
13.12 Checking of SLED output power, checking attenuator.......................................................... 159
13.13. Crossing the spectrometer. ................................................................................................. 162
13.14 Spectrometer stability test.................................................................................................... 163
14. FAQ.....................................................................................................................................164
15. Spare parts list.....................................................................................................................171
16. Service tools ........................................................................................................................181

SOCT Copernicus HR - Service Manual
5
1. Introduction.
This Service manual contains technical information about SOCT Copernicus HR, troubleshooting
of SOCT Copernicus HR, parts and modules exchange procedures, description of calibration
procedure. The information provided in this document has been prepared for trained service
technicians.
2. Contact to Optopol Technology.
In case of any problems or doubts concerning technical procedures or other information
contained in this manual, please contact one of the following:
Customer Support:
Tel: +48 32 670 91 73
sales @optopol.com.pl
Technical Support:
Service department
Tel: +48 32 670 91 73 ext. 230
For online support use program: TeamViewer rev.6 (available on http://dane.optpopol.com or
http://www.teamviewer.com/download/ ).
In case when distributor decides send back the device to Optopol Technology Sp. z o.o.. have to fill

SOCT Copernicus HR - Service Manual
6
3. Safety standards.
The SOCT Copernicus HR system complies with all requirements of the directive
93/42/EEC (MDD - Medical Devices Directive).
The SOCT Copernicus HR is Type B applied parts, class I protection against electric shock.
Usually SOCT Copernicus HR is a part of larger Medical System, which consists of many
other medical and non- medical devices. Medical system in general and all its components
have to full fill requirements of IEC 60601-1 standard.
In order to comply with the IEC 60601-1 standard all non Medical Devices must be
connected to the Separating Transformer which is supplied with the SOCT Copernicus HR.
By connecting the devices to the Separating Transformer leakage current is reduced to a
level which is in line with the IEC standard.
The Separating Transformer fulfils requirements of IEC 60601-1 standard for medical
devices which has not application part.
The Separating Transformer supply the power for non medical devices connected to
electric not grounded mains in room.
The Separating Transformer can be installed in patient environment in accordance with IEC
60601-1 standard.
SOCT Copernicus HR is connected to computer using USB cable. This cable is electrically
separated using optocouplers. It provides additional safety, when computer and SOCT
Copernicus HR device are connected to two different power supply sockets with different
ground potentials. Video signal and synchronization signal are transferred to computer
using coaxial cables, which are electrically separated from SOCT Copernicus HR device
providing even more safe solution.
SOCT Copernicus HR system can be connected only to power supply socket
equipped with properly connected grounding pin.
Please note that only the PC, monitor and printer are to be connected to the Isolating
Transformer. Connecting non medical devices in other way than in point 7.3 Device
connection diagram can lead to electric shock or damage of the devices.
Prohibit connect any non medical or medical devices which are not including with SOCT
Copernicus HR system to the Separating Transformer e.g. lamp, vacuum cleaner etc.
The SOCT Copernicus HR must not be connected to the transformer, but directly to power
supply socket equipped with properly connected grounding pin.

SOCT Copernicus HR - Service Manual
7
No user serviceable parts inside SOCT Copernicus HR device. Light source is included inside
device. Removing of covers can be done only by Manufacturer or Authorized Personnel.
The maximum power of light radiation available outside covers is
less than 900 (±3%) [ 1020 (±3%) if SLED InPhenix ] microwatts providing safe operation of
the device. Objective lens is output of light.
SOCT Copernicus HR has been classified as CLASS 1 LASER DEVICE
Important!
1) If the whole Medical System has been installed by Manufacturer or Authorized
Personnel of SOCT Copernicus HR, the Manufacturer guarantees correct
installation and compliance with all required standards and directives,
2) In case when installation was made not by Manufacturer or Authorized
Personnel, Manufacturer of SOCT Copernicus HR takes NO responsibility for
any problems or risks that could be created by incorrect connection and
violation of safety standards,
3) Manufacturer of SOCT Copernicus HR takes no responsibility for incorrect
medical diagnose of results and takes no responsibility for the consequences of
such incorrect medical diagnosis.
Warning!
Ignoring safety rules can lead to accidents.
1) If SOCT Copernicus HR and the whole Medical System have been installed by
the Manufacturer or Authorized Personnel, the Manufacturer guarantees the
proper installation and compliance with MDD 93/42 EEC rules,
2) Any computer system or device attached to SOCT Copernicus HR must be MDD
compliant or be plugged only to Separating Transformer. Devices which are not
related with patient examination (e.g. lamp, fan) cannot be connected to
Separating Transformer,
3) Multiple portable socket-outlets shall not be placed on the floor.
Warning!
All activities that demand removing main body housing may be performed by
Manufacturer or Authorized Personnel trained by Manufacturer.
Ignoring or disregarding the statements above may lead to danger of death or serious
injury.

SOCT Copernicus HR - Service Manual
8
4. Maintenance and cleaning.
Warning!
a) All maintenance activities can only be made when device is turned off and unplugged
from power supply socket
b) No user serviceable parts inside device. Any covers can be removed only by authorized
service staff.
During the whole period of use, SOCT Copernicus HR does not need any special maintenance
procedures. The outside surfaces of device should be cleaned using mild cleaning solvents.
Only a soft cloth can be used for cleaning. During cleaning it has to be assured, that no water gets
inside the device.
The forehead rest and chinrest are parts that routinely have contact with patient. The forehead and
chin rests should be cleaned between each examination with an alcohol prep swab.
The main lens of the device should not contact the patient's eye. It comes with a lens cap to cover it
when not in use. Lens may touch the eyelashes during normal use. You should clean it to remove
dust and oily smudges occasionally to ensure a true tomogram images are obtained. You may use an
alcohol prep swab or cotton swab dipped in isopropyl alcohol. (You may use dedicated optical lens
cleaner.) Wipe dry with a soft, non-lining cloth or tissue. If the lens inadvertently contacts the
patient's eye, clean it before proceeding with the examination. Periodically one has to check if there
are no mechanical damages to the device, damages to any of the cables and fuses.
5. Device Overview.
SOCT Copernicus HR uses spectral domain optical coherence tomography method to obtain
3-dimensional cross-section images of a human retina. The device produces a laser beam, which is
focused on a human retina. The light that reflects from internal structures of a retina is then
interferometrically analyzed by the device. The raw data is processed by a computer to obtain images
of retina cross-section.

SOCT Copernicus HR - Service Manual
9
6. Technical Data.
Signal source Super Luminescent Diode (SLED), 850 nm (±2%)
Optical power 850 -900 µW [ 920 - 1020 µW if S-LED InPhenix ]
Signal analysis Spectral domain Optical Coherence Tomography
Scanning speed Up to 52’000 measurements per second
Scanning programs Single B-scan
3D visualization
Asterisk
Animation scan (selected number of B-scans in the same
place at a retina)
Single B-scan with blood flow analysis
Axial resolution 3 um in tissue
Transversal resolution ideal 12 um
typical 18 um
Scan width 4 –10 mm
Scan depth 2 mm
Power supply 230/115 V 50/60 Hz
Power consumption 150 VA
Fuse ratings 3,15 A 250V
Multiple portable socket-outlet max. load 500 VA

SOCT Copernicus HR - Service Manual
10
7. Unpacking and Installation.
NOTICE!: Every transportation box is equipped with the shockwatches. If any shockwatch is broken
(red indicator) please contact Optopol Technology Sp z o.o. and lodge the complaint to the carrier.
SOCT Copernicus HR set consists of following components:
1. SOCT device with two CL cables,
2. USB 2.0 communication cable,
3. Coaxial synchronization/video cable,
4. PC Computer with preinstalled software,
5. Power supply cables and computer accessories.
7.1 Unpacking SOCT Copernicus HR.
This section describes how to unpack the device shipped from a factory. The SOCT Copernicus
HR is shipped in a special box, which protects it from mechanical damages during transport. Figures
below show how to unpack SOCT from this box. The box has got four latches shown in detail in
pictures below.
Closed latch
Step 1
Step 2
Latch is opened
Transport position of the SOCT Copernicus HR is to be set manually. Head maximum to the left, to
the back and lens down. Then protect handler will fit correctly.

SOCT Copernicus HR - Service Manual
11
Now one can remove the cover by lifting it up gently. The SOCT’s Copernicus HR four legs are locked
at two metal rails by clamps. There are only two removable clamps at a side. First remove screws
from the clamps and then remove metal holders. This operation is shown in a figure below.
Now shift SOCT Copernicus HR to the left and then lift it up. Put the SOCT on the flat stable table.
Important!
It is strongly recommended to wait at least one hour before turning the device on for the first
time after delivery. It is very important in case where the delivery was made during low outside
temperatures or high humidity.
After unpacking, one should check the whole set for any mechanical damages, cables
damages, etc. In case when any damage is found, don't connect the device and contact
your local distributor.

SOCT Copernicus HR - Service Manual
12
7.2 Connecting cables.
To connect the SOCT Copernicus HR to a computer, you need five cables shipped with the device.
Two Camera link cables –connected to Copernicus HR (not detachable),
CL1 CL2
USB cables, connect to between Copernicus HR (type B) & PC (type A),
Video/synchro cable connected with one plug,
Video (to PC) Synchro (SMA to PC) Video/Synchro to SOCT
Video-synchro cables are merged together.

SOCT Copernicus HR - Service Manual
13
All sockets and plugs are different so it is not possible to connect plugs improperly. Figure below
shows rear panel view of SOCT.
First connect all signal cables described above and after that check the connections once again. In the
next step connect power supply cables.
NOTE: Regarding EMC (Electro-magnetic compatibility) standards all signal cables have to be put
together.
Picture on the next page shows how to connect all cables to the computer.

SOCT Copernicus HR - Service Manual
14
7.3 Device connection diagram.
The SOCT is connected to the PC using USB 2.0 and Video/Synchro cables.
The SOCT must be directly plugged into the mains. Connect the PC, printer and monitor into the
multiple portable socket-outlets, which is directly connected to the Separating Transformer. The
transformer reduces power emission loss from the PC. The Separating Transformer is connected
directly to the mains.
IMPORTANT: Always turn computer first and then turn device on.

SOCT Copernicus HR - Service Manual
15
8. PC installation.
8.1 Minimum system requirements.
(Starting on day 22.09.2010)
In order to prepare computer for use with SOCT Copernicus HR you need:
PC computer with operating system installed.
Minimum system requirements:
- Processor: Intel® Xeon W3520 2.66 8MB/1066 QC CPU
- RAM: HP 4 GB (2x2GB) DDR3 –1333 ECC RAM
- Operating system: Windows XP Professional
- HDD: 2 x HP 1.5 TB SATA 7200 HDD (RAID)
- CD/DVD writer HP 16X DVDRW Drive
- Video card: 1024 X 768 resolution
24 bits color depth
512 MB DDR3
OpenGL support (ASUS EAH 5770 CUcore1024 MB)
- Communication: 1 available USB port
- Extension slots 2 available PCI slots
2 available PCI Express (x16)
National Instruments IMAQ PCIE-1429 image acquisition card (supplied with SOCT device)
Frame grabbing card (supplied with SOCT device)
SOCT Copernicus HR Installation CD (supplied with SOCT device)
National Instruments installation CD –Disk 1 and 2 (supplied with SOCT device)
Backup and Recovery CD's.

SOCT Copernicus HR - Service Manual
16
8.2 NI Frame Grabber card and Win-Fast card installation.
Open the left side cover of a PC.
Plug in the NI Frame Grabber into the PCIE slot #2 then secure the card using a screw.
Plug the Win-Fast Video card into the PCI slot #1 then secure the card using a screw.
Close the PC cover.

SOCT Copernicus HR - Service Manual
17
8.3 Drivers installation.
8.3.1 NI Vision Acquisition Software installation.
Turn the computer on.
Insert a CD disc with SOCT CopernicusHR software to the CD-ROM.
Open the CD and find the NI software files. Disc:\NI Vision Acquisition\VISIONACQ_CD1.
Double click on the setup.exe.
After initialization finishes, press Next>>.

SOCT Copernicus HR - Service Manual
18
Select primary installation directory, it should be C:\Program files\National Instruments\
then press Next>>.
Select features to be installed.
- Extend the icon NI-IMAQdx 3.1.2 and select “Do not install this feature”.
- Extend the icon NI-IMAQ I/O 2.1 and select “Do not install this feature”.

SOCT Copernicus HR - Service Manual
19
- Extend NI-IMAQ 4.0 and LabVIEW Support.
- Extend the icon LabVIEW 7.1 Support and select “Install this feature to a local drive”.
- Extend LabVIEW Real-Time 7.1 Support.

SOCT Copernicus HR - Service Manual
20
- Extend the icon LabVIEW Real-Time 7.1 Support and select “Install this feature to
a local drive”.
Press Next>>.
Select “I accept the License Agreement(s)” then press Next>>.
Table of contents
Other Optopol Medical Equipment manuals

Optopol
Optopol SOCT Copernicus User manual

Optopol
Optopol SOCT Copernicus REVO Quick start guide

Optopol
Optopol SOCT Copernicus + User manual

Optopol
Optopol REVO nx User manual

Optopol
Optopol SOCT Copernicus REVO User manual

Optopol
Optopol SOCT Copernicus User manual

Optopol
Optopol SOCT Copernicus REVO User manual
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual