14
INCUBATOR COMPONENTS
16. H2O2key*3
This key is to run H2O2 decontamination. Refer
to page 60 to 65.
17. Unlock key*2
Press this key is to unlock the outer door when
it is auto-locked by electric lock. Refer to page
66 to 68. When the auto lock function is OFF,
this key is not displayed.
18. Buzzer key
Press this key to silence the buzzer. However,
when the ring back is ON, the buzzer will
sound again when the ring back passed and
the alarm state still continues. Refer to page
37~38 and 77~79.
Note: It is not possible to silence the buzzer
for the high limit temperature alarm.
*3: The H2O2 decontamination function is workable under any of the following conditions. When the condition is
not fulfilled, the H2O2 key is not displayed on the LCD touch panel.
・When an H2O2generator MCO-HP-PW is installed in the KM-CC17RH2A.
・When all H2O2generator MCO-HP-PW, H2O2decon board KM-CCPHB1A and electric lock KM-CCPL1W
are installed in the KM-CC17RU2A.
・When all UV system set KM-CCP17US2A, H2O2generator MCO-HP-PW, H2O2decon board
KM-CCPHB1A and electric lock KM-CCPL1W are installed in the KM-CC17R2A.
Remote alarm terminal
The alarm of this product can be informed at a remote location from this product by connecting the external
alarm unit to the remote alarm terminals. For the type and behavior of remote alarm output, refer to page
77 to 79.
The terminal of the remote alarm is installed at the
rear upper right of the unit (See the figure on the
point). The alarm is outputted from this terminal.
Contact capacity is DC 30 V, 2 A.
When Buzzer key is pressed, the behavior of the
remote alarm is showed in Table.1.
Note: In the door alarm, the remote alarm does not
work. Refer to page 77 to 78.
Table 1 The behavior of the remote alarm when pressing the Buzzer key
Remote Alarm setting
(Refer to page 36~38)
Connecting
terminal
Normal
condition
Abnormal condition
(Including in the cases of power outage and
of where the power plug is pulled out.)
When pressing the buzzer key
ON:
Non-interlock with Buzzer key
COM.-N.C. Close Open Open (Maintain in abnormality)*
COM.-N.O. Open Close Close (Maintain in abnormality)*
OFF:
Interlock with Buzzer key
COM.-N.C. Close Open Close (Return to normal)
COM.-N.O. Open Close Open (Return to normal)
*In case of Err01 (CO2 gas cylinder empty), Err11, 12(CO2 sensor error), the condition returns to normal.
Remote alarm
terminal