EnglishDeutsch
 logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Pepperl+Fuchs
  6. •
  7. Measuring Instrument
  8. •
  9. Pepperl+Fuchs KFA5-DWB-Ex1.D User manual

Pepperl+Fuchs KFA5-DWB-Ex1.D User manual

This manual suits for next models

3

Other Pepperl+Fuchs Measuring Instrument manuals

Pepperl+Fuchs LGC User manual

Pepperl+Fuchs

Pepperl+Fuchs LGC User manual

Pepperl+Fuchs LVL-Z120 User manual

Pepperl+Fuchs

Pepperl+Fuchs LVL-Z120 User manual

Pepperl+Fuchs IMU360D-F99-B20 User manual

Pepperl+Fuchs

Pepperl+Fuchs IMU360D-F99-B20 User manual

Pepperl+Fuchs AS-i User manual

Pepperl+Fuchs

Pepperl+Fuchs AS-i User manual

Pepperl+Fuchs PMI Series User manual

Pepperl+Fuchs

Pepperl+Fuchs PMI Series User manual

Pepperl+Fuchs PMI360DV-F130-IU2E2-V15 User manual

Pepperl+Fuchs

Pepperl+Fuchs PMI360DV-F130-IU2E2-V15 User manual

Pepperl+Fuchs UDC-18GMA-400-3E2 User manual

Pepperl+Fuchs

Pepperl+Fuchs UDC-18GMA-400-3E2 User manual

Pepperl+Fuchs HART LUC-M Series User manual

Pepperl+Fuchs

Pepperl+Fuchs HART LUC-M Series User manual

Pepperl+Fuchs IMUF99 B20 Series User manual

Pepperl+Fuchs

Pepperl+Fuchs IMUF99 B20 Series User manual

Pepperl+Fuchs VBA-2E-KE4-ENC-S User guide

Pepperl+Fuchs

Pepperl+Fuchs VBA-2E-KE4-ENC-S User guide

Pepperl+Fuchs PMI15V-F166 IO Series User manual

Pepperl+Fuchs

Pepperl+Fuchs PMI15V-F166 IO Series User manual

Pepperl+Fuchs PMI360D-F130-IE8-V15-Y205400 User manual

Pepperl+Fuchs

Pepperl+Fuchs PMI360D-F130-IE8-V15-Y205400 User manual

Pepperl+Fuchs DPU1100 Series User manual

Pepperl+Fuchs

Pepperl+Fuchs DPU1100 Series User manual

Pepperl+Fuchs Pulscon LTC57 User manual

Pepperl+Fuchs

Pepperl+Fuchs Pulscon LTC57 User manual

Pepperl+Fuchs LGM Series User manual

Pepperl+Fuchs

Pepperl+Fuchs LGM Series User manual

Pepperl+Fuchs LGM Series Reference guide

Pepperl+Fuchs

Pepperl+Fuchs LGM Series Reference guide

Pepperl+Fuchs Pulscon LTC50 HART User manual

Pepperl+Fuchs

Pepperl+Fuchs Pulscon LTC50 HART User manual

Pepperl+Fuchs PMI360DV-F130-IU2E0-V15 User manual

Pepperl+Fuchs

Pepperl+Fuchs PMI360DV-F130-IU2E0-V15 User manual

Pepperl+Fuchs PMI360DV-F130-IU-V15 User manual

Pepperl+Fuchs

Pepperl+Fuchs PMI360DV-F130-IU-V15 User manual

Pepperl+Fuchs Pulscon LTC50 Manual

Pepperl+Fuchs

Pepperl+Fuchs Pulscon LTC50 Manual

Pepperl+Fuchs TC-6A-V Instruction Manual

Pepperl+Fuchs

Pepperl+Fuchs TC-6A-V Instruction Manual

Pepperl+Fuchs WS-UC F406-B41 Series User manual

Pepperl+Fuchs

Pepperl+Fuchs WS-UC F406-B41 Series User manual

Pepperl+Fuchs Pulscon LTC51 Manual

Pepperl+Fuchs

Pepperl+Fuchs Pulscon LTC51 Manual

Pepperl+Fuchs IMUF99 B16 Series User manual

Pepperl+Fuchs

Pepperl+Fuchs IMUF99 B16 Series User manual

Popular Measuring Instrument manuals by other brands

Suaoki D40 user manual

Suaoki

Suaoki D40 user manual

Elster BK-G1.6 operating instructions

Elster

Elster BK-G1.6 operating instructions

Endress+Hauser Deltabar FMD71 technical information

Endress+Hauser

Endress+Hauser Deltabar FMD71 technical information

Duratool D03128 manual

Duratool

Duratool D03128 manual

Vega VEGAPULS 65 Quick setup guide

Vega

Vega VEGAPULS 65 Quick setup guide

Tonghui TH2618B Operation manual

Tonghui

Tonghui TH2618B Operation manual

Tenmars TM-801 user manual

Tenmars

Tenmars TM-801 user manual

MSR 145 user manual

MSR

MSR 145 user manual

TPS WP-82 manual

TPS

TPS WP-82 manual

PCB Piezotronics 003C30 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 003C30 Installation and operating manual

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Endress+Hauser Proline Promag 50 Brief operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 50 Brief operating instructions

Siemens SITRANS F Coriolis FCT030 Function manual

Siemens

Siemens SITRANS F Coriolis FCT030 Function manual

KLINGER CMF V Series instruction manual

KLINGER

KLINGER CMF V Series instruction manual

Keysight M8290A Getting started guide

Keysight

Keysight M8290A Getting started guide

Acterna MS1400 Operation manual

Acterna

Acterna MS1400 Operation manual

R&S ZPH manual

R&S

R&S ZPH manual

Datacolor ColorReader DC10-2 user guide

Datacolor

Datacolor ColorReader DC10-2 user guide

https://manuals.online logo
https://manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2026 Manuals.Online. All Rights Reserved.

Instruction Manual
1. Marking
Rotation Speed Monitor
KFA5-DWB-Ex1.D, KFA6-DWB-Ex1.D
Switch Amplifier, Timer Relay
KFA5-DU-Ex1.D, KFA6-DU-Ex1.D
ATEX certificate: TÜV 99 ATEX 1408
ATEX marking:
1 II (1)G [Ex ia Ga] IIC
1 II (1)D [Ex ia Da] IIIC
1 I (M1) [Ex ia Ma] I
ATEX certificate: TÜV 02 ATEX 1885 X
ATEX marking: 1 II 3G Ex nA nC IIC T4 Gc
IECEx certificate: IECEx TUN 03.0000
IECEx marking: [Ex ia Ga] IIC , [Ex ia Da] IIIC , [Ex ia Ma] I
North America Certifcates: FM22CA0016X / Project ID: 3009603 (FM)
Class I, Division 2, Groups A-D, T4
Class I, Zone 2, AEx nC, Group IIC, T4
Class I, Zone 2, Ex nA nC, Group IIC, T4
Associated apparatus with intrinsically safe circuits for:
Class I, Division 1, Groups A-D; Class II, Division 1, Groups E-G;
Class III
Class I, Zone 0, [AEx ia] IIC, Class I, Zone 0, [Ex ia] IIC
North America Certifcates: E223772 (UL)
Pepperl+Fuchs Group
Lilienthalstraße 200, 68307 Mannheim, Germany
Internet: www.pepperl-fuchs.com
2. Target Group, Personnel
Responsibility for planning, assembly, commissioning, operation,
maintenance, and dismounting lies with the plant operator.
The personnel must be appropriately trained and qualified in order to carry
out mounting, installation, commissioning, operation, maintenance, and
dismounting of the device. The trained and qualified personnel must have
read and understood the instruction manual.
Prior to using the product make yourself familiar with it. Read the
instruction manual carefully.
3. Reference to Further Documentation
Observe laws, standards, and directives applicable to the intended use
and the operating location.
For mining applications, observe laws, standards, and directives
applicable to the operating location.
The corresponding datasheets, manuals, declarations of conformity, EU-
type examination certificates, certificates, and control drawings if
applicable supplement this document. You can find this information under
www.pepperl-fuchs.com.
For specific device information such as the year of construction,
scan the QR code on the device. As an alternative, enter the serial number
in the serial number search at www.pepperl-fuchs.com.
If you use the device in safety-related applications, observe the
requirements for functional safety. You can find these requirements in the
functional safety documentation under www.pepperl-fuchs.com.
4. Intended Use
The device is only approved for appropriate and intended use. Ignoring
these instructions will void any warranty and absolve the manufacturer
from any liability.
The device is used in control and instrumentation technology
(C&I technology) for the galvanic isolation of signals such as 20 mA and
10 V standard signals or alternatively for adapting or standardizing signals.
The device has intrinsically safe circuits that are used for operating
intrinsically safe field devices in hazardous areas.
Use the device only within the specified ambient and operating conditions.
Only use the device stationary.
The device is an associated apparatus according to IEC/EN 60079-11.
The device is an electrical apparatus for hazardous areas of Zone 2.
The device may be installed in the non-hazardous area.
If you use the device in safety-related applications, observe the
information for safety function and safe state.
Rotation Speed Monitor
The device monitors for an overspeed or underspeed condition of a digital
signal (NAMUR sensor/ mechanical contact) from a explosion-hazardous
area by comparing the input frequency to the user programmed reference
frequency.
Switch Amplifier, Timer Relay
The device is a highly configurable timer that accepts a digital signal
(NAMUR sensor/mechanical contact) from a explosion-hazardous area
and is commonly used in applications requiring on-delay, off-delay, one-
shot, or pulse lengthening.
5. Improper Use
Protection of the personnel and the plant is not ensured if the device is not
used according to its intended use.
If circuits with type of protection Ex i are operated with non-intrinsically
safe circuits, they must no longer be used as circuits with type of
protection Ex i.
6. Mounting and Installation
Do not mount a damaged or polluted device.
The device is designed for mounting on a 35 mm DIN mounting rail
according to EN 60715.
Mount the device in a way that the device is protected against mechanical
hazard. Mount the device in a surrounding enclosure for example.
Ensure that the surrounding enclosure can only be opened with a tool.
The device fulfills a degree of protection IP20 according to IEC/EN 60529.
The device must be installed and operated only in a controlled
environment that ensures a pollution degree 2 (or better) according to
IEC/EN 60664-1.
If used in areas with higher pollution degree, the device needs to be
protected accordingly.
Do not mount the device in the dust hazardous area.
The device must be installed and operated only in an environment of
overvoltage category II (or better) according to IEC/EN 60664-1.
Observe the installation instructions according to IEC/EN 60079-14.
If you install the device in safety-related applications, observe the
requirements for functional safety.
Requirements for Cables and Connection Lines
Observe the permissible core cross section of the conductor.
When using stranded conductors, crimp wire end ferrules on the
conductor ends.
Use only copper conductors.
Use only one conductor per terminal.
When installing the conductors the insulation must reach up to the
terminal.
Observe the tightening torque of the terminal screws.
If the voltage is greater than 50 V AC or 120 V DC, switch off the voltage
before connecting or disconnecting the device.
Only use cables and connection lines with a temperature range
appropriate to the application.
Requirements for Usage as Associated Apparatus
If circuits with type of protection Ex i are operated with non-intrinsically
safe circuits, they must no longer be used as circuits with type of
protection Ex i.
Observe the respective peak values of the field device and the associated
apparatus with regard to explosion protection when connecting intrinsically
safe field devices with intrinsically safe circuits of associated apparatus
(verification of intrinsic safety). Also observe IEC/EN 60079-14 and
IEC/EN 60079-25.
Intrinsically safe circuits of associated apparatus can be led into
hazardous areas. Observe the compliance of the separation distances to
all non-intrinsically safe circuits according to IEC/EN 60079-14.
Observe the compliance of the separation distances between two
adjacent intrinsically safe circuits according to IEC/EN 60079-14.
Requirements for Equipment Protection Level Gc
The device must be installed and operated only in surrounding enclosures
that
lcomply with the requirements for surrounding enclosures according to
IEC/EN 60079-0,
lare rated with the degree of protection IP54 according to
IEC/EN 60529.
Connection or disconnection of energized non-intrinsically safe circuits is
only permitted in the absence of a potentially explosive atmosphere.
Provide a transient protection. Ensure that the peak value of the transient
protection does not exceed 140 % of the rated voltage.
7. Operation, Maintenance, Repair
If you operate the device in safety-related applications, observe the
requirements for functional safety. In case of operating in low demand
mode, plan appropriate intervals for the proof test.
If the voltage is greater than 50 V AC or 120 V DC, switch off the voltage
before connecting or disconnecting the device.
Do not use a damaged or polluted device.
DOCT-8420 / 2022-09 1 / 2
Do not repair, modify, or manipulate the device.
If there is a defect, always replace the device with an original device.
Requirements for Equipment Protection Level Gc
Connection or disconnection of energized non-intrinsically safe circuits is
only permitted in the absence of a potentially explosive atmosphere.
Only use operating elements in the absence of a potentially explosive
atmosphere.
8. Delivery, Transport, Disposal
Check the packaging and contents for damage.
Check if you have received every item and if the items received are the
ones you ordered.
Always store and transport the device in the original packaging.
Store the device in a clean and dry environment. The permitted ambient
conditions must be considered, see datasheet.
The device, built-in components, packaging, and any batteries contained
within must be disposed in compliance with the applicable laws and
guidelines of the respective country.
DOCT-8420 / 2022-09 2 / 2