Operation
19
Sensor menu Total pressure gauges with hot cathode sensor (e.g. HPT 200 ) may need baking in ultra
high vacuum to desorp the gasload from surfaces. The degas process lasts for three
minutes by default but can be stopped manually at any time. Pressure reading with re-
duced accuracy is possible during degas, the PRESSURE menu shows degas for the
concerned channel as long as the operation proceeds.
In the sensor menu the type dependent functions of the combination gauges RPT 100/
200, HPT 100/200, and MPT 100/200 can be selected. The settings are described here-
under taking the HPT 200 as an example.
SENSOR MENU
The high vacuum sensors of the combined gauges HPT 200
and MPT 200 can be switched manually:
–BAFil = Filament of the BA sensor
–CCSens = Cold cathode sensor
Press set key until the cursor flashes at the parameter BA-
Fil; the status of the function will be displayed (On or Off).
Switch filament On or Off using the Up/Down keys.
Confirm with set key.
For the HPT 200 gauge the value in brackets define the select-
ed filament:
Select 1, 2or auto.
Confirm with set key.
If autois selected, an automatic switchover to the other filament
takes place when necessary (in case of wear).
For the MPT 200 gauge CCSens appears instead of BAFil.
Switch sensor on or off using the Up/Down keys.
Confirm with set key.
Select switching mode for combination gauges:
Press set key until the cursor flashes at the parameter
SwMode; the status of the function will be displayed
(trans_Hi, trans_Lo or switch, depending on the type of
gauge).
Use Up/Down key to select the desired function.
Confirm with set key.
Activate the degas function:
The function degas is only possible for total pressure gauges
with hot cathode sensor (HPT 100/200).
Press set key until the cursor flashes at the parameter de-
gas. The status of the degas function will be displayed (Off
or On).
Use the Up key to select degas On.
Start the degas process with the set key.
The degas process can be manually deactivated at any time by
switching the degas function to Off:
Press the set key until the cursor flashes at the parameter
degas.
Select the function Off with the Down key.
Switch off the degas process with the set key.
ch: 1SENSOR MENU
type: HPT200 V010100
BAFil[1/2/auto]: On/O
SwMode: trans_hi
degas: O
SENSOR MENU
...
BAFil[auto]: On/O
...
...
SENSOR MENU
...
...
SwMode: trans_hi/lo/switch
...
set
set
set
set
SENSOR MENU
...
...
...
degas: On/O