"Settings" submenu
This is where you can edit unit parameters.
"VSM" submenu
Online tool for automatic service processing. For further information, select Virtual Service Manage-
ment.
5.1.1 Change parameters
Procedure
1. Select unit from equipment list.
2. Select "Settings" submenu.
3. Edit desired parameter in line with defined input function.
–Input functions are numeric values or slide controls.
4. Save changes.
5.1.2 Explanations of settings
"Autostart" function
The activated components vacuum, capillary heating and inlet valve are started automatically when the
GSD 350 is next switched on. The heaters activated in Autostart do not switch on immediately, but only
once there is a sufficient vacuum in the system.
"Sensor on/off" function of total pressure gauge
The Pirani sensor in the total pressure gauge normally switches the cold cathode sensor on and off au-
tomatically as a function of the pressure. The software continuously displays the pressure value deter-
mined by the total pressure gauge. The pressure value is additionally used to shut down the mass spec-
trometer's filament and electron multiplier in case of an unforeseen pressure rise. You can use the Sen-
sor on/off function to stop the cold cathode sensor switching on automatically if this is advantageous for
certain gas analyses.
"Ignore sensor error" function of the total pressure gauge
If the cold cathode sensor fails, e.g., due to contamination, you can use this function to suppress auto-
matic shutdown of the filament and electron multiplier by the total pressure gauge. The latency in the
filament flow monitoring and turbopump speed monitoring functions ensures a safety shutdown. You
can still continue ongoing measurements with the GSD 350 if the total pressure gauge is defective. Re-
place the defective gauge to restore the best possible safety shutdown by the total pressure gauge.
Pre-set temperature of capillary and vacuum chamber heating
For input values outside the permissible range, the GSD 350 automatically assumes the highest or low-
est permissible temperature. You will find the permissible temperatures, depending on the unit configu-
ration, in the GSD 350 operating instructions.
Pre-set bake out time for the vacuum chamber
Selectable time for beak out: 1 to 24 hours
"Tempering mode on/off" function of the vacuum chamber
In normal cases, the GSD 350 does not heat the vacuum chamber during a measurement. For certain
analytical measurements, you can activate the "Tempering mode" function. If tempering mode is ena-
bled, the GSD 350 always heats the vacuum chamber to the configured set temperature when the GSD
350 switches on the capillary heating.
Operate via web interface