Precor 9.45 User manual

9.45, 9.45i Treadmill
Page 1
9.45, 9.45i Treadmill
Warning: This service manual is for use by Precor trained service providers only.
IfyouarenotaPrecorTrainedServicer,youmustnotattempttoserviceanyPrecorProduct;
Call your dealer for service.
This document contains information required to perform the majority of troubleshooting, and
replacement procedures required to repair and maintain this product.
This document contains general product information, software diagnostic procedures (when
available), preventative maintenance procedures, inspection and adjustment procedures,
troubleshooting procedures, replacement procedures and electrical block and wiring diagrams.
To move directly to a procedure, click the appropriate procedure in the bookmark section to the
left of this page. You may “drag” the separator bar between this page and the bookmark section
to change the size of the page being viewed.
© 2003 Precor Incorporated
Unauthorized Reproduction and Distribution Prohibited By Law

9.45, 9.45i Treadmill
Page 2
Section One - Things you Should Know
About This Appendix
Precor has re-designed the M9.45 treadmill several times. Units with the original design have
serial numbers starting with '1L' and were manufactured between January 24, 1995 and March
25, 1996. The M9.45 treadmill was then redesigned with serial numbers starting with '2P' and
were manufactured between March 26, 1996 and November 15, 1997. The 9.45i uses serial
numbers starting with ‘3Y’ and started in production on November 16, 1997. The 9.45i is the only
version of the 9.45 that is currently in production. This document will refer to the three versions
as follows 9.45 (1Y), 9.45 (2P) and 9.45i (3Y).
Section One, Things You Should Know. This section includes general information and
technical specifications. Read this section, as well as the M9.45 Treadmill Owner’s Manual,
before you perform the maintenance procedures in this manual.
Section Two, Software Features. Precor’s M9.45 Treadmill is programmed with several
diagnostic and setup features. This section contains the procedures you need to access the
diagnostic features on this treadmill.
Section Three, Checking Treadmill Operation. This section provides you with a quick way of
checking treadmill operation. Check treadmill operation at the end of a maintenance procedure
and when it is necessary to ensure that the treadmill is operating properly.
Section Four, Inspection and Adjustment Procedures. Perform inspection procedures when
a trouble symptom points to a particular problem and after removing and replacing major
components. Many maintenance problems can be fixed by adjusting various treadmill
components. This section also provides you with the step-by-step procedures required to make
these adjustments.
Section Five, Troubleshooting Procedures. The diagnostic and troubleshooting procedures
contained in this section should be performed when it is necessary to isolate a problem to a
particular component.
Section Six, Replacement Procedures. When a treadmill component must be replaced, go to
this section and follow the step-by-step procedures required to remove and replace the
component.
Section Seven, Wiring and Block Diagrams. This section includes wiring and block diagrams
for all versions of the 9.45.
General Information
For the latest exploded view, part number and part pricing information, visit the Precor dealer
website at “www.precor.com/Dealer”.

9.45, 9.45i Treadmill
Page 3
Technical Specifications
Physical Specifications
Length: 71 inches or 180 cm (Extended deck treadmills
are 5 inches longer)
Width: Handrails 31.5 inches (80 cm)
Base 28.5 inches (72 cm)
Height: 50 inches (127 cm)
Running surface: 53 inches by 18 inches (135 cm by 46 cm)
Motor: 2 hp continuous duty
Speed: 0.5 to 10 mph (0.8 to 16 kph)
10 mph is equivalent to a 6 minute mile
Incline: -2% to +12% grade
Power: 50/60 Hz 120v AC (dedicated circuit)
50/60 Hz 240v AC (dedicated circuit)
Weight: 265 lbs (119 kg)
Shipping weight: 326 lbs (147 kg)
About The Smart Rate System
The Smart Rate system is the heart rate system mounted in the display housing. The Smart Rate
system receives a heart rate signal from chest strap transmitter worn by the user. Smart Rate
software displays a heart range on green, red and yellow LED's mounted on the upper PCA. The
LED’s indicate whether the heart rate is below, within or above the desired rang.

9.45, 9.45i Treadmill
Page 4
Procedure 2.1 - Accessing the M9.45 (1L or 2P) Treadmill's
Diagnostic Program
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. The diagnostic program provide six different routines to aid it testing and trouble diagnosis.
3. On 9.45 (1Y & 2P units), with the PRECOR M9.45 banner scrolling, press and hold ENTER
until the software version number is displayed in the right display windows. On units that
have serial numbers starting with '2P', the part number of the PROM will also be displayed.
4. The first routine is the LED display test. The routine will light every LED on the display as the
test proceeds. It is necessary to watch the display to see if an LED does not light.
5. On 9.45 (1L units) the second routine is the lift test. The lift test takes the lift through each of
the critical lift positions. In doing so it tests the performance of the zero switch, upper limit
switch, lower limit switch and rotation sensor. Press ENTER to continue.
6. On 9.45 (2P) units the second routine is lift calibration. The lift calibration number indicates
the physical position of the lift. As lift is operated the lift calibration number will track the lift’s
physical position. The lift calibration number will be used to initialize the lift. Press the
INCLINE keys while you watch the right display window (see Diagram 2.1). The lift
calibration number increments as the INCLINE ▲ key is pressed and decrements as the
INCLINE ▼key is pressed. Press ENTER to continue.
7. The third routine is the power bits display. Power bits is a number that indicates the amount
of time that power is applied to the drive motor (duty cycle). The power bit number reflects
changes in speed and load. Power bits and running belt speed will; be displayed. Notice that
the power bits number will increase or decrease as speed and/or load is increased or
decreased. The power bit number can be used to determine treadmill loading. Press
ENTER to continue.
8. The fourth routine is the keypad test. Seven dashed lines will appear on the display. Each
dash will correspond to a key on the display. Press each key one at a time, the
corresponding dash will enlargen, indicating the key that was pressed. Use this test to
determine if the keys are functioning properly. Press the INCLINE ▼ and SPEED ▲ keys,
simultaneously to continue.
9. The fourth routine is the safety key test. When the safety key is inserted SAFE ON is
displayed in the right display window. Remove the safety key from the upper display module
while you watch the right display window, SAFE OFF is displayed. Replace the safety key
and press ENTER to continue.
10. The fifth routine is the heart rate test. When a chest strap transmitter or test transmitter is
used, the heart rate will be displayed. Press ENTER to continue.

9.45, 9.45i Treadmill
Page 5
11. The last routine is the circuit breaker trip test. This test checks the ability of the treadmill to
trip the circuit breaker under software control. If the STOP key is pressed and held the trip
signal will be sent to the circuit breaker and the circuit will trip, shutting off the treadmill. If
you do not wish to trip the circuit breaker, press the ENTER key to exit the diagnostic
program.
Diagram 2.1 - Display Housing, 9.45 (1Y & 2P)

9.45, 9.45i Treadmill
Page 6
Procedure 2.2 - Accessing the M9.45i (3Y) Treadmill's
Diagnostic Program
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. The diagnostic program provide six different routines to aid it testing and trouble diagnosis.
3. For the purpose of accessing the various software routines, including the diagnostics
program, the display housing keys are hypothetically numbered 1 to 7 from left to right.
4. With the PRECOR M9.45 banner scrolling, press keys RESET,5,1,7,6,5,7,6,1, sequentially
(See Diagram 2.2).
5. The first routine is the LED display test. The routine will light every LED on the display as the
test proceeds. It is necessary to watch the display to see if an LED does not light..
6. The second routine is lift calibration. The lift calibration number indicates the physical
position of the lift. As lift is operated the lift calibration number will track the lift’s physical
position. The lift calibration number will be used to initialize the lift. Press the INCLINE keys
while you watch the right display window (see Diagram 2.2). The lift calibration number
increments as the INCLINE ▲ key is pressed and decrements as the INCLINE ▼key is
pressed. Press ENTER to continue.
7. The third routine is the power bits display. Power bits is a number that indicates the amount
of time that power is applied to the drive motor (duty cycle). The power bit number reflects
changes in speed and load. Power bits and running belt speed will; be displayed. Notice that
the power bits number will increase or decrease as speed and/or load is increased or
decreased. The power bit number can be used to determine treadmill loading. Press
ENTER to continue.
8. The fourth routine is the key pad test. Seven dashed lines will appear on the display. Each
dash will correspond to a key on the display. Press each key one at a time, the
corresponding dash will enlargen, indicating the key that was pressed. Use this test to
determine if the keys are functioning properly. Press the INCLINE ▼ and SPEED ▼ keys,
simultaneously to continue
9. The fifth routine is the safety key test. When the safety key is inserted SAFE ON is
displayed in the right display window. Remove the safety key from the upper display module
while you watch the right display window, SAFE OFF is displayed. Replace the safety key
and press ENTER to continue.
10. The sixth routine is the heart rate test. When a chest strap transmitter or test transmitter is
used, the heart rate will be displayed. Press ENTER to continue.

9.45, 9.45i Treadmill
Page 7
11. The last routine is the circuit breaker trip test. This test checks the ability of the treadmill to
trip the circuit breaker under software control. If the STOP key is pressed the trip signal will
be sent to the circuit breaker and the circuit will trip, shutting off the treadmill. If you do not
wish to trip the circuit breaker, press the ENTER key to exit the diagnostic program.
Diagram 2.2 - Display Housing, 9.45i (3Y)

9.45, 9.45i Treadmill
Page 8
Procedure 2.3 - Resetting All User Information
This procedure completely resets the M9.45 user statistics. All previously-saved user information
will be either reset to zero or changed to default values. All odometers will be reset to zero. If you
want to save accumulated user information and odometer readings, record the information
before performing this procedure.
Note: This procedure is not available on the 9.45i. User stats must be reset individually per
Procedure 2.4.
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. With the PRECOR M9.45 or PRECOR M9.45i banner scrolling, press the INCLINE ▲,
INCLINE ▼, SPEED ▲, and SPEED ▼keys simultaneously and hold them for approximately
five seconds (or until zeros are displayed in the right display windows). Release the keys.
3. Watch the electronic display. The right middle display window:
a Resets to all zeros, then increments to 119
b Resets to 1000, then increments to 1119
c Resets to 2000, then increments to 2119
d Resets to 3000, then increments to 3119
e Displays the User ID prompt.
Note:
If the upper PCA or prom is replaced, all user information will be re-initialized.

9.45, 9.45i Treadmill
Page 9
Procedure 2.4 - Resetting an Individual User’s Information
This procedure resets accumulated user information and odometer readings for an individual
User ID. All previously-saved user information for a selected User ID will be either reset to zero or
changed to default values. The odometer reading for the selected User ID will be reset to zero.
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. With the PRECOR M9.45 or PRECOR M9.45i banner scrolling, press any key.
3. At the User ID prompt, select the User IDs to be reset using the SPEED ▲(or INCLINE ▲)
and SPEED ▼(or INCLINE ▼) keys.
4. Press the STOP and QUICKSTART keys together and hold them for approximately two
seconds (or until the right middle display window displays one of the numbers sets in
Procedure 2.2, step 3a, 3b, 3c or 3d).
5. If you do not observe the values listed in Procedure 2.2, step 3a, 3b, 3c or 3d...
THEN... OTHERWISE...
Replace the upper PCA as described in This procedure is complete.
Procedure 5.2 of the Residential Treadmill
Service Manual.
Note:
If the upper PCA is replaced, all user information will be re-initialized.

9.45, 9.45i Treadmill
Page 10
Procedure 2.5 - Displaying the Treadmill Odometer and Error
Code Log
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. On 9.45 (1L or 2P units), with the PRECOR M9.45 banner scrolling, press the ENTER,
SPEED ▼and SPEED ▲keys, simultaneously. Continue to hold the keys until the message
TREADMILL ODOMETER scrolls across the left display window. On 9.45 (3Y units), press
keys RESET,6,5, sequentially.
Note:
The right display window displays total miles or kilometers on the treadmill (see Diagram 2.3).
The top display window shows the most significant bits of the number; the lower display window
shows the least significant bits of the number. The number displayed is 102,187,723 (in miles or
kilometers).
Diagram 2.3 - Odometer Reading For the Treadmill
3. Press ENTER to access the error code log.
0001
0218
7723

9.45, 9.45i Treadmill
Page 11
4. There are five error code log positions. The error code log is displayed and scrolls across
the left display window (see Diagram 2.4). Use the SPEED ▼(or INCLINE ▼)and SPEED ▲
(or INCLINE ▲) keys to scroll through the error code log. The most recent error condition will
be displayed in position 1 and the oldest error will be displayed in position 5. When an error
is added to a full log the new error will be inserted in position 1 and all other logs will be
pushed down one position. The error that was in position 5 will be lost.
5. Once you have completed the diagnosis and repair of the treadmill, it is generally desirable
to clear the error code log. Then any subsequent error that are logged will be known to be
new error conditions. To clear the error code log, press the QUICK START and STOP keys
simultaneously. Or, press ENTER to retain the current error information and return to the
User ID prompt.
Diagram 2.4 - Error Code Log
6. The example above (Diagram 2.4) shows an “Error 22” in position 1 and “No Error” in
position 2.
Err: No
Err
2
22
1

9.45, 9.45i Treadmill
Page 12
Procedure 2.6 - Selecting United States Standard or Metric
Units
Selecting United States standard units causes data to be displayed in feet, miles and pounds.
Metric data is displayed in meters, kilometers and kilograms.
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. Choose one:
IF... THEN...
You wish to verify the measurement Continue with the next step
standard the treadmill is currently using
before you change the standard
You wish to change the measurement Skip to Step 5
standard the treadmill is using
Checking the Measurement Standard
3. Press ENTER until the WEIGHT indicator light appears.
4. Watch the number displayed in the left display window while you press the SPEED ▲key
five or six times.
Note:
If the treadmill is using United States standard units, the numbers in the left display window are
displayed multiples of 5 (such as 160, 165, 170, 175, etc.). If the units are displayed in multiples
of 2 or 3, the treadmill set for metric units.
Changing the Measurement Standard
5. Press and hold STOP to return the PRECOR M9.45 banner.
6. On 9.45 (1L or 2P units), press the QUICK START and then the SPEED ▲key. Hold both
keys for 3–5 seconds, then release. On 9.45 (3Y units), press keys RESET,5,6,7,1,
sequentially.
Note:
On 9.45 (1L or 2P units), releasing the QUICK START key before the SPEED ▲key when you
perform the next step may cause the selected measurement standard to change to the alternate

9.45, 9.45i Treadmill
Page 13
measurement standardUse the SPEED ▲or SPEED ▼keys to select the alternate measurement
standard.
7. Press ENTER twice to return to the User ID prompt.

9.45, 9.45i Treadmill
Page 14
Procedure 2.7 - Determining Software Version Numbers
Software version numbers are invaluable for tracking and identifying problems and staying aware
of changes to the operation and features of the product.
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. On 9.45 (1L or 2P units), with the PRECOR M9.45 banner scrolling, press the ENTER key.
On 9.45 (3Y units), with the PRECOR M9.45 banner scrolling, press keys
RESET,5,1,7,6,5,7,6,1, sequentially.
3. Note the version number displayed in the right middle display window.
Note:
If you cannot determine the software version number in this manner, look at the label on the
PROM mounted on the upper PCA. The part number stamped on the PROM indicates the
software version number.

9.45, 9.45i Treadmill
Page 15
Procedure 2.8 - Documenting Software Problems
When a problem is found with either the software or upper or lower PCA’s, record the information
listed below. If you isolated the problem to either the PROM, upper PCA, or lower PCA, include
the information you recorded with the malfunctioning PROM or PCA when you ship it to Precor.
When a problem occurs, record the following information:
• Model and serial number
• Software version number
Note:
Determine the version number as in Procedure 2.7.
• User and program number running when the problem occurred
• A description of:
a What happened or failed to happen.
b The action taken by the user just before the problem occurred.
c Problem-related information (such as how far into the program the problem occurred,
the work level being used when the problem occurred, etc.).
• •The frequency of occurrence.

9.45, 9.45i Treadmill
Page 16
Section Three - Checking Treadmill Operation
This section provides you with a quick method of checking treadmill operation. Check treadmill
operation at the end of a maintenance procedure and when it is necessary to ensure that the
treadmill is operating properly.
Procedure
1. Plug the power cord into the wall outlet, then turn on the treadmill with the circuit breaker.
2. Place the treadmill in Manual Mode. Adjust the speed of the running belt to 2–3 mph
Operate the treadmill for at least 5 minutes.
a Concentrate on the feel of the running belt and the sound of the drive motor and rollers.
Be on the alert for unusual noises, smells, or vibrations.
b. Measure and log the AC input current under loaded and unloaded conditions.
c Observe the LEDs on the electronic console. Make sure that each LED lights as the
information corresponding to that LED is displayed on the electronic console.
d Measure and note the AC input current at 3 MPH eith the treadmil unloaded and at 3
MPH with the treadmill loaded. Be sure to note the weight of the user on the loaded
current rreading.
e Measure and note the AC (ripple) voltage at the drive motor capacitor at 3 MPH with the
treadmill loaded and unloaded
3. Press the STOP key. When the treadmill comes to a stop, view the electronic console as the
treadmill scans time, speed, distance and percent.
4. Press the INCLINE ▲key while viewing the electronic console. Confirm that the running
bed inclines and the incline display increments to twelve percent as the INCLINE ▲key is
pressed.
5. Press the INCLINE ▼key while viewing the electronic console. Confirm that the running bed
returns to a level position and the incline display decrements to minus two percent as the
INCLINE ▼key is pressed.
6. Turn off the treadmill with the circuit breaker, then unplug the treadmill from the wall outlet.

9.45, 9.45i Treadmill
Page 17
Procedure 4.1 - Calibrating the Lift Limit Switches, 9.45 (1L)
WARNING
Always turn off the circuit breaker and unplug the treadmill before you remove the treadmill hood.
1. If the treadmill is not already at zero percent incline, press any key when the CALIBRATE
LIFT PRESS ANY KEY prompt is displayed.
2. Using the ruler, measure the distance between the floor and the top of the front and back
ends of the side rail (see Diagram 4.1).
Note:
The distance between the floor and the top of the front end of the side rail should be 9 inches.
Diagram 4.1 - Treadmill Side Rail Measurement

9.45, 9.45i Treadmill
Page 18
3. If the distances recorded in the previous step are equal to within 1/4"...
THEN... OTHERWISE...
Skip to Step 17. Continue with the next step.
4. At the PRESS ENTER FOR PROGRAMS prompt, press and hold the ENTER key.
5. Press ENTER when the display test starts.
6. With the ADJUST LIFT [YES/DN NO/UP] prompt displayed, press any ▼key. The
message PRESS ENTER WHEN FINISHED is displayed on the electronic console.
7. If the distance measured at the front of the side rail is larger than the distance measured at
the back of the side rail...
THEN... OTHERWISE...
Press the INCLINE ▼ key until Press the INCLINE ▲ key until the
the front and back treadmill front and back treadmill measurements
measurements are equal; then are equal; then press RESET.
press RESET.
8. Turn off the treadmill with the circuit breaker, then unplug the power cord from the wall
outlet.
WARNING
Before continuing with this procedure, review the Warning and Caution statements listed in
Section One of the Residential Treadmill Service Manual.
9. Remove the hood as described in Procedure 5.1 of the Residential Treadmill Service
Manual.
10. Check the position of the zero sense switch in relation to the switch actuator (see
Diagram 4.2).
11. If the widest point of the switch actuator is lined up with the center of the wheel on the zero
sense switch...
THEN... OTHERWISE...
The zero sense switch is calibrated The switch bracket must be adjusted;
correctly; skip to Step 16. continue with the next step.

9.45, 9.45i Treadmill
Page 19
Diagram 4.2 -Switch Mounting Bracket Assembly
Lower Limit
Switch
Actuator
Upper Limit
Switch
Zero Switch
Capacitor
Lift

9.45, 9.45i Treadmill
Page 20
WARNING
When power is applied to the treadmill, the wires connected to the upper and lower limit switches
carry 120 volts (or 240 volts if you are servicing a 240-volt treadmill). Turn off the treadmill and
unplug the power cord from the wall outlet before you perform the following steps.
12. Loosen the socket head bolts and washers that secure the switch bracket to the lift platform
(see Diagram 4.2).
Note
The zero sense switch may be actuated if the center of the wheel is slightly above or slightly
below the widest point of the switch actuator. However, for best results, adjust the position of the
zero sense switch as described in the following step.
13. If the center of the wheel on the zero sense switch is above the widest point of the switch
actuator...
THEN... OTHERWISE...
Move the switch bracket down until Move the switch bracket up until the
the widest part of the actuator is widest part of the actuator is centered
centered on the wheel on the zero on the wheel on the zero sense switch;
sense switch; then continue with then continue with the next step.
the next step.
14. Tighten the socket head bolts and washers that secure the switch bracket to the lift platform
(see Diagram 4.2).
15. Replace the hood as described in Procedure 5.1 of the Residential Treadmill Service
Manual.
16. Check the operation of the treadmill as described in Section Three of this appendix.
IMPORTANT
If the electronic console displays the diagnostic message LIFT ERROR PRESS STOP when you
incline and decline the running bed to its maximum limits, repeat this procedure.
If readjusting the position of the switch actuator does not fix the problem, refer to Troubleshooting
Flow Chart 13 of the Residential Treadmill Service Manual.
This manual suits for next models
1
Table of contents
Other Precor Treadmill manuals

Precor
Precor M9.5sp User manual

Precor
Precor 9.31 User manual

Precor
Precor Experience Series 700 Line User manual

Precor
Precor 9.23 User manual

Precor
Precor Experience 700 line Series User manual

Precor
Precor 9.33 User manual

Precor
Precor M9.41si User manual

Precor
Precor 9.23 User manual

Precor
Precor M9.21i User manual

Precor
Precor C966 Instruction manual

Precor
Precor TRM 211 User manual

Precor
Precor C954 Instruction manual

Precor
Precor 9.55 User manual

Precor
Precor TRM 800 Assembly Instructions

Precor
Precor C954i User manual

Precor
Precor C932i User manual

Precor
Precor 9.33 Owner's manual

Precor
Precor 9.23 User manual

Precor
Precor M9.45 User manual

Precor
Precor C932i User manual