Presys T-1200PIR User manual

Calibrator for
Infrared Thermomether
T-1200PIR
EM0277-00
R
Technical Manual
Q
Y
U
T
A
L
I
E
N
T
M
S
E
Y
G
S
A
T
N
E
A
M
M
presys
presys
presys
presys
Infrared Thermomether
presys
Infrared Thermomether
presys
presys
presys

WARNING!
Avoid electric shock risk when touching the equipment:
- Use only suitable power cable with earth connection;
- Never power the equipment to the mains socket without earth connection.
WARNING!
High voltage is present inside this equipment. Do not make any repair service inside
the equipment without removing the plug from the supply.
WARNING!
Much electromagnetic noise can cause instability to the equipment.
The equipment is provided with electromagnetic interference filters that protect not only
the mains but also the equipment itself against noise. These filters have no function if
the unit is not earthed properly.
WARNING!
High temperatures are achieved in this equipment.
Risk of fire and explosion are present in case safety measures are not taken. Sign by
means of warnings the hazardous areas at high temperatures.
Do not place the instrument on inflammable surfaces or even on materials that can be
deformed due to high temperatures.
Do not obstruct any air-vent to avoid risk of fire in the equipment.
Do not operate the instrument in excessively wet, oily, dusty or dirty environments.
WARNING!
Do not use sharp or pointed objects in the black body cavity.
Do not use abrasive materials or cleaning products on the black body cavity. These
products can change the instrument properties.
Do not touch inside the black body cavity, salts and the oiliness of skin can impair the
properties of the material.
CAUTION!
The instrument described in this technical manual is intended to be used in a
specialized technical area. The user should be responsible by its configuration and the
parameter values entered. The manufacturer warns about risks of personal injury or
ambient damage as a result of its incorrect use.
presys
- Use only suitable power cable with earth connection;
presys
- Use only suitable power cable with earth connection;
earth connection.
presys
earth connection.
High voltage is present inside this equipment. Do not make any repair service inside
presys
High voltage is present inside this equipment. Do not make any repair service inside
the equipment without removing the plug from the supply.
presys
the equipment without removing the plug from the supply.
Much electromagnetic noise can cause instability to the equipment.
presys
Much electromagnetic noise can cause instability to the equipment.
The equipment is provided with electromagnetic interference filters that protect not only
presys
The equipment is provided with electromagnetic interference filters that protect not only
the mains but also the equipment itself against noise. These filters have no function if
presys
the mains but also the equipment itself against noise. These filters have no function if
the unit is not earthed properly.
presys
the unit is not earthed properly.
WARNING!
presys
WARNING!
High temperatures are achieved in this equipment.
presys
High temperatures are achieved in this equipment.
Risk of fire and explosion are present in case safety measures are not taken. Sign by
presys
Risk of fire and explosion are present in case safety measures are not taken. Sign by
means of warnings the hazardous areas at high temperatures.
presys
means of warnings the hazardous areas at high temperatures.
Do not place the instrument on inflammable surfaces or even on materials that can be
presys
Do not place the instrument on inflammable surfaces or even on materials that can be
deformed due to high temperatures.
presys
deformed due to high temperatures.
Do not obstruct any air-vent to avoid risk of fire in the equipment.
presys
Do not obstruct any air-vent to avoid risk of fire in the equipment.
presys
Do not operate the instrument in excessively wet, oily, dusty or dirty environments.
presys
Do not operate the instrument in excessively wet, oily, dusty or dirty environments.
WARNING!
presys
WARNING!
Do not use sharp or pointed objects in the black body cavity.
presys
Do not use sharp or pointed objects in the black body cavity.
Do not use abrasive materials or cleaning products on the black body cavity
presys
Do not use abrasive materials or cleaning products on the black body cavity
products can change the instrument properties.
presys
products can change the instrument properties.
Do not touch inside the black body cavity, salts and the oiliness of skin can impair the
presys
Do not touch inside the black body cavity, salts and the oiliness of skin can impair the
The instrument described in this technical manual is intended to be used in a
presys
The instrument described in this technical manual is intended to be used in a
specialized technical area. The user should be responsible by its configuration and the
presys
specialized technical area. The user should be responsible by its configuration and the
parameter values entered. The manufacturer warns about risks of personal injury or
presys
parameter values entered. The manufacturer warns about risks of personal injury or

PRESYS Instruments T-1200PIR
EM0277-00
Table of Contents
1.0 - Introduction ............................................................................................................1
1.1 - Technical Specifications........................................................................................2
1.2 - Order Code...........................................................................................................3
1.3 - Accessories ..........................................................................................................4
1.4 - Instructions for the use of Black Body Insert .........................................................4
2.0 - T-1200PIR Calibrator Operating.............................................................................6
2.1 - Input Connections Diagram...................................................................................8
2.2 - CONF Menu..........................................................................................................9
2.3 - Manual Operating Mode......................................................................................12
2.4 - Programmable Operating Mode..........................................................................13
2.5 - Automatic Operating Mode .................................................................................14
3.0 - Usage Recommendations....................................................................................16
4.0 - Safety Instructions ...............................................................................................16
5.0 - Calibrator Warning Messages .............................................................................17
6.0 - PID Control Parameters .......................................................................................17
7.0 - Calibration.............................................................................................................18
7.1 - Input Calibration..................................................................................................19
7.2 - Probe Calibration................................................................................................20
8.0 - Maintenance..........................................................................................................20
8.1 - Instructions for Hardware....................................................................................20
9.0 - Instructions for Fitting the Strap for Transport..................................................21
presys
............................................................................................................
presys
............................................................................................................
........................................................................................
presys
........................................................................................
...........................................................................................................
presys
...........................................................................................................
3
presys
3
..........................................................................................................
presys
..........................................................................................................
4
presys
4
.........................................................
presys
.........................................................
4
presys
4
.............................................................................
presys
.............................................................................
6
presys
6
...................................................................................
presys
...................................................................................
..........................................................................................................
presys
..........................................................................................................
......................................................................................
presys
......................................................................................
..........................................................................
presys
..........................................................................
.................................................................................
presys
.................................................................................
....................................................................................
presys
....................................................................................
...............................................................................................
presys
...............................................................................................
5.0 - Calibrator Warning Messages
presys
5.0 - Calibrator Warning Messages
.............................................................................
presys
.............................................................................
6.0 - PID Control Parameters
presys
6.0 - PID Control Parameters
.......................................................................................
presys
.......................................................................................
.............................................................................................................
presys
.............................................................................................................
7.1 - Input Calibration
presys
7.1 - Input Calibration
..................................................................................................
presys
..................................................................................................
7.2 - Probe Calibration
presys
7.2 - Probe Calibration
................................................................................................
presys
................................................................................................
8.0 - Maintenance
presys
8.0 - Maintenance
..........................................................................................................
presys
..........................................................................................................
8.1 - Instructions for Hardware
presys
8.1 - Instructions for Hardware
....................................................................................
presys
....................................................................................
presys
9.0 - Instructions for Fitting the Strap for Transport
presys
9.0 - Instructions for Fitting the Strap for Transport

PRESYS Instruments T-1200PIR
Page 1
1.0 - Introduction
T-1200PIR
The Calibrator for Infrared Thermometer T-1200PIR generates temperature
values in order to enable the calibration of infrared thermometers. The temperatures
generated have high accuracy, in an area of uniform temperature and emissivity.
This calibrator also has input for thermocouples and resistance thermometers, as
well as current meter function. The T-1200PIR generates temperature from 50 to
1200 ºC.
Has input for thermocouples, RTDs and thermoswitches reading.
The use of external standard thermometer is not required*.
Carries out completely automatic calibrations with or without the use of a computer.
Accuracy of 3 ºC, 0.2 ºC stability and 0.1 ºC resolution.
Documenting capabilities: communication with computer and ISOPLAN Calibration
Software.
Portable and compact, has strap for easy transport.
*For higher performances, it is possible to make use of an external thermocouple sensor
type "N" connected to its own input.
It has extensive programming capabilities, where the calibrator is programmed to
generate pre-set temperature.
Compatible with automatic calibration and documenting software, with application
of ISOPLAN software on PC / Windows platform,using the serial port for connection
between the PC and the calibrator that communicate via RS-232 or RS-485. ISOPLAN
software can register thermometers and factory instruments, generate work orders,
produce and print calibration certificates and reports, that is, all the information of power
is brought into the environment of calibrations.
presys
T-1200PIR
presys
T-1200PIR
The Calibrator for Infrared Thermometer T-1200PIR generates temperature
presys
The Calibrator for Infrared Thermometer T-1200PIR generates temperature
values in order to enable the calibration of infrared thermometers. The temperatures
presys
values in order to enable the calibration of infrared thermometers. The temperatures
generated have high accuracy, in an area of uniform temperature and emissivity.
presys
generated have high accuracy, in an area of uniform temperature and emissivity.
This calibrator also has input for thermocouples and resistance thermometers, as
presys
This calibrator also has input for thermocouples and resistance thermometers, as
well as current meter function. The T-1200PIR generates temperature from
presys
well as current meter function. The T-1200PIR generates temperature from
Has input for thermocouples, RTDs and thermoswitches reading.
presys
Has input for thermocouples, RTDs and thermoswitches reading.
The use of external standard thermometer is not required*.
presys
The use of external standard thermometer is not required*.
Carries out completely automatic calibrations with or without the use of a computer.
presys
Carries out completely automatic calibrations with or without the use of a computer.
ºC
presys
ºC
, 0
presys
, 0
.2
presys
.2
ºC
presys
ºC
stability and 0.1
presys
stability and 0.1
ºC
presys
ºC
resolution.
presys
resolution.
Documenting capabilities: communication with computer and ISOPLAN Calibration
presys
Documenting capabilities: communication with computer and ISOPLAN Calibration
Portable and compact, has strap for easy transport.
presys
Portable and compact, has strap for easy transport.
For higher performances, it is possible to make use of an external thermocouple sensor
presys
For higher performances, it is possible to make use of an external thermocouple sensor
type "N" connected to its own input
presys
type "N" connected to its own input
.
presys
.
presys
It has extensive programming capabilities, where the calibrator is programmed to
presys
It has extensive programming capabilities, where the calibrator is programmed to
generate pre-set temperature.
presys
generate pre-set temperature.
Compatible with automatic calibration and documenting software, with application
presys
Compatible with automatic calibration and documenting software, with application
of ISOPLAN software on PC / Windows platform
presys
of ISOPLAN software on PC / Windows platform
between the PC and the calibrator that communicate via RS-232 or RS-
presys
between the PC and the calibrator that communicate via RS-232 or RS-
software can register thermometers and factory instruments, generate work orders,
presys
software can register thermometers and factory instruments, generate work orders,
produce and print calibration certificates and reports, that is, all the information of power
presys
produce and print calibration certificates and reports, that is, all the information of power
is brought into the environment of calibrations.
presys
is brought into the environment of calibrations.
presys

PRESYS Instruments T-1200PIR
Page 2
T-1200PIR has also many other features, such as:
The electric signal calibrator is independent from the temperature control function.
Internal buzzer beeps when the temperature reaches the desired value.
Numeric keypad that eases the operation and configuration of the calibrator.
Vacuum Fluorescent Display for presentation of the values.
Thermo-element reading scaled to ITS-90 or IPTS-68.
Internal regulated 24 Vdc power supply for 2-wire transmitters.
Internal rechargeable battery and battery charger included in the electric signal
calibrator.
Independent circuitry for over temperature protection and safety.
1.1 - Technical Specifications
T-1200PIR
Operating Range
50 to 1200 °C
Power Supply
110 or 220 Vac* 50/60Hz
Cavity Diameter
20 mm
Resolution
0.1ºC or 0.1ºF
Accuracy (internal reference)
3 °C
(external thermometer)
0.75 °C
Cavity Emissivity (
)
(0.95 ± 0.02)@ 8-14 µm
Stability
0.2 ºC
Power Consumption
1100 W(110 Vac*) / 2300 W (220 Vac*)
Heating Rate
45 minutes (100 ºC to 1200 ºC)
Cooling Rate
5 hours (1200 ºC to 200 ºC)
Dimensions (H/W/D)
215 x 390 x 310 mm
Weight
10.4 kg
Warranty
One year
*according to the order code
presys
The electric signal calibrator is independent from the temperature control function.
presys
The electric signal calibrator is independent from the temperature control function.
Numeric keypad that eases the operation and configuration of the calibrator.
presys
Numeric keypad that eases the operation and configuration of the calibrator.
Internal rechargeable battery and battery charger included in the electric signal
presys
Internal rechargeable battery and battery charger included in the electric signal
Independent circuitry for over temperature protection and safety.
presys
Independent circuitry for over temperature protection and safety.
T
presys
T
-
presys
-
1200PIR
presys
1200PIR
50 to
presys
50 to
1200
presys
1200
°C
presys
°C
presys
110 or
presys
110 or
220
presys
220
Va
presys
Va
c
presys
c
*
presys
*
50/60Hz
presys
50/60Hz
presys
2
presys
2
0
presys
0
mm
presys
mm
presys
0.
presys
0.
1
presys
1
ºC
presys
ºC
or
presys
or
0.
presys
0.
1
presys
1
ºF
presys
ºF
presys
3 °C
presys
3 °C
presys
)
presys
)
presys
0
presys
0
.
presys
.
75
presys
75
°C
presys
°C
presys
presys
presys
(
presys
(
0
presys
0
.
presys
.
95
presys
95
±
presys
±
0
presys
0
.
presys
.
02
presys
02
)
presys
)
@ 8
presys
@ 8
-
presys
-
14
presys
14
presys
presys
presys
presys
presys
0
presys
0
.
presys
.
2
presys
2
ºC
presys
ºC
presys
presys
presys
presys
presys
1100
presys
1100
W
presys
W
(110 Vac
presys
(110 Vac
*
presys
*
) / 2300 W (2
presys
) / 2300 W (2
presys
presys
presys
presys
presys
45 minutes (100 ºC to 1200 ºC)
presys
45 minutes (100 ºC to 1200 ºC)
presys
presys
presys
presys
5 hours (1200 º
presys
5 hours (1200 º
presys
presys
presys
presys
W
presys
W
/
presys
/
W/W
presys
W/W
D)
presys
D)
2
presys
2
1
presys
1
5 x
presys
5 x
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
*according to the order code
presys
*according to the order code

PRESYS Instruments T-1200PIR
Page 3
1.1.1 - Input Technical Specifications
Input Ranges
Resolution
Accuracy
Remarks
millivolt -150 to 150 mV
-500 to -150 mV
150 to 2450 mV
0.001 mV
0.01 mV
0.01 mV
±0.01 % FS*
±0.02 % FS
±0.02 % FS Rinput > 10 M
auto-ranging
mA -5 to 24.5 mA
0.0001 mA
±0.02 % FS
Rinput < 160
Resistance 0 to 400
400 to 2500
0.01
0.01
± 0.01 % FS
± 0.03 % FS excitation current
0.85mA
Pt-100 -200 to 850 ºC / -328 to 1562 ºF
0.01 ºC / 0.01 ºF
± 0.1 ºC / ± 0.2 ºF
IEC-60751
Pt-1000 -200 to 400 ºC / -328 to 752 ºF
0.1 ºC / 0.1 ºF
± 0.1 ºC / ± 0.2 ºF
IEC-60751
TC-B 50 to 250 ºC / 122 to 482 ºF
250 to 500 ºC / 482 to 932 ºF
500 to 1200 ºC / 932 to 2192 ºF
1200 to 1820 ºC / 2192 to 3308 ºF
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
± 2.5 ºC / ± 5.0 ºF
± 1.5 ºC / ± 3.0 ºF
± 1.0 ºC / ± 2.0 ºF
± 0.7 ºC / ± 1.4 ºF
IEC-60584
TC-J -210 to 1200 ºC / -346 to 2192 ºF
0.1 ºC / 0.1 ºF
± 0.2 ºC / ± 0.4 ºF
IEC-60584
TC-K -270 to -150 ºC / -454 to -238 ºF
-150 to 1370 ºC / -238 to 2498 ºF
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
± 0.5 ºC / ± 1.0 ºF
± 0.2 ºC / ± 0.4 ºF
IEC-60584
TC-N -260 to -200 ºC / -436 to -328 ºF
-200 to -20 ºC / -328 to -4 ºF
-20 to 1300 ºC / -4 to 2372 ºF
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
± 1.0 ºC / ± 2.0 ºF
± 0.4 ºC / ± 0.8 ºF
± 0.2 ºC / ± 0.4 ºF
IEC-60584
TC-R -50 to 300 ºC / -58 to -572 ºF
300 to 1760 ºC / 572 to 3200 ºF
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
± 1.0 ºC / ± 2.0 ºF
± 0.7 ºC / ± 1.4 ºF
IEC-60584
TC-S -50 to 300 ºC / -58 to -572 ºF
300 to 1760 ºC / 572 to 3200 ºF 0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF 1.0 ºC / 2.0 ºF
0.7 ºC /
1.4 ºF
IEC-60584
* FS = Full Scale
1.2 - Order Code
T-1200PIR -
Power Supply
1–110 Vac
2–220 Vac
presys
R
presys
R
input
presys
input
auto
presys
auto
-
presys
-
ranging
presys
ranging
presys
presys
R
presys
R
input
presys
input
< 160
presys
< 160
presys
presys
presys
presys
presys
% FS
presys
% FS
% FS
presys
% FS
excitation current
presys
excitation current
0.
presys
0.
85
presys
85
mA
presys
mA
presys
presys
presys
presys
0.1 ºC /
presys
0.1 ºC /
±
presys
±
0.2 ºF
presys
0.2 ºF
IEC
presys
IEC
-
presys
-
60751
presys
60751
presys
presys
presys
presys
presys
0.1 ºC /
presys
0.1 ºC /
±
presys
±
0.2 ºF
presys
0.2 ºF
IEC
presys
IEC
-
presys
-
60751
presys
60751
presys
presys
presys
presys
presys
presys
2.5 ºC /
presys
2.5 ºC /
±
presys
±
5.0 ºF
presys
5.0 ºF
±
presys
±
1.5 ºC /
presys
1.5 ºC /
±
presys
±
3.0 ºF
presys
3.0 ºF
±
presys
±
1.0 ºC /
presys
1.0 ºC /
±
presys
±
2.0 ºF
presys
2.0 ºF
±
presys
±
0.7 ºC /
presys
0.7 ºC /
±
presys
±
1.4 ºF
presys
1.4 ºF
IEC
presys
IEC
-
presys
-
60584
presys
60584
presys
presys
presys
presys
presys
presys
presys
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
±
presys
±
0.2 ºC /
presys
0.2 ºC /
±
presys
±
0.4 ºF
presys
0.4 ºF
IEC
presys
IEC
-
presys
-
60584
presys
60584
presys
presys
presys
presys
presys
presys
presys
presys
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
presys
±
presys
±
0.5 ºC /
presys
0.5 ºC /
±
presys
±
1.0 ºF
presys
1.0 ºF
±
presys
±
0.2 ºC /
presys
0.2 ºC /
±
presys
±
0.4 ºF
presys
0.4 ºF
IEC
presys
IEC
-
presys
-
60584
presys
60584
presys
presys
presys
presys
presys
presys
presys
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
±
presys
±
1.0 ºC /
presys
1.0 ºC /
±
presys
±
2.0 ºF
presys
2.0 ºF
±
presys
±
0.4 ºC /
presys
0.4 ºC /
±
presys
±
0.8 ºF
presys
0.8 ºF
±
presys
±
0.2 ºC /
presys
0.2 ºC /
±
presys
±
0.4 ºF
presys
0.4 ºF
IEC
presys
IEC
-
presys
-
60584
presys
60584
presys
presys
presys
presys
presys
presys
presys
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
±
presys
±
1.0 ºC /
presys
1.0 ºC /
±
presys
±
2.0 ºF
presys
2.0 ºF
±
presys
±
0.7 ºC /
presys
0.7 ºC /
±
presys
±
1.4 ºF
presys
1.4 ºF
presys
presys
presys
presys
presys
IEC
presys
IEC
presys
presys
presys
presys
presys
300 to 1760 ºC / 572 to 3200 ºF
presys
300 to 1760 ºC / 572 to 3200 ºF
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
0.1 ºC / 0.1 ºF
presys
0.1 ºC / 0.1 ºF
presys
1.0 ºC /
presys
1.0 ºC /
presys
2.0 ºF
presys
2.0 ºF
presys
0.7 ºC /
presys
0.7 ºC /
presys
1.4 ºF
presys
1.4 ºF
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
T-1200PIR -
presys
T-1200PIR -
presys
presys
presys
presys
presys

PRESYS Instruments T-1200PIR
Page 4
1.3 - Accessories
Communication Interface:
Description
Order Code
RS-232 -DB-9F Connector (COM1)
06.02.0002-00
RS-232 -DB-25F Connector (COM2)
06.02.0004-00
RS-485
06.02.0006-00
Carrying Case. Order Code: 06.01.1020-00.
ISOPLAN Software.
Calibration Certificate.
1.4 - Instructions for the use of Black Body Insert
Identify the following parts and proceed to the mounting as explained:
Cylindrical Thermal insulator –Mounted in the lower part of the pit of the
furnace.
Metallic insert type black body cavity –Must be introduced in the pit joined with
a thermocouple type N mounted laterally.
Careful when introducing the thermocouple in the cavity to not force the fragile
ceramic wall pit.
Ring-shaped cylindrical Insulator - mounted on top of the pit of the furnace
Note that the position of the slit of insulation should match the type N
thermocouple sheath laterally.
Connect the terminals of the thermocouple type N to the auxiliary input side of
the furnace T-1200PIR and set the reading of the IN input to N type
thermocouple.
The combination constitutes an excellent mounted cavity black body with
emissivity above 0.95 and an effective target of Ø 20 mm well suited for calibration of
pyrometers.
Align the pyrometer to be calibrated with a black body cavity in the furnace in a
horizontal position.
Observe the distance of the pyrometer to be calibrated against the background
of the black body cavity and the size of the actual target (Ø 20 mm) as specified in the
technical manual optical pyrometer.
Remember that the area targeted by the pyrometer to be calibrated must be
less than or equal to the effective target spot size of the black body in order to not
introduce measurement errors.
presys
presys
06.02.0002
presys
06.02.0002
-
presys
-
00
presys
00
presys
presys
06.02.0004
presys
06.02.0004
-
presys
-
00
presys
00
presys
presys
presys
06.02.0006
presys
06.02.0006
-
presys
-
00
presys
00
presys
presys
presys
presys
presys
presys
1.4 - Instructions for the use of Black Body Insert
presys
1.4 - Instructions for the use of Black Body Insert
Identify the following parts and proceed to the mounting as explained:
presys
Identify the following parts and proceed to the mounting as explained:
Cylindrical Thermal insulator
presys
Cylindrical Thermal insulator
–
presys
–
Mounted in the lower part of the pit of the
presys
Mounted in the lower part of the pit of the
–Mounted in the lower part of the pit of the–
presys
–Mounted in the lower part of the pit of the–
Metallic insert type black body cavity
presys
Metallic insert type black body cavity
–
presys
–
Must be introduced in the pit joined with
presys
Must be introduced in the pit joined with
–Must be introduced in the pit joined with–
presys
–Must be introduced in the pit joined with–
a thermocouple type N mounted laterally.
presys
a thermocouple type N mounted laterally.
Careful when introducing the thermocouple in the cavity to not force the fragile
presys
Careful when introducing the thermocouple in the cavity to not force the fragile
ceramic wall pit.
presys
ceramic wall pit.
-shaped cylindrical Insulator - mounted on top of the pit of the furnace
presys
-shaped cylindrical Insulator - mounted on top of the pit of the furnace
presys
Note that the position of the slit of insulation should match the type N
presys
Note that the position of the slit of insulation should match the type N
thermocouple sheath laterally.
presys
thermocouple sheath laterally.
Connect the terminals of the thermocouple type N to the auxiliary input side of
presys
Connect the terminals of the thermocouple type N to the auxiliary input side of
the furnace T-1200PIR and set the reading of the IN input to N type
presys
the furnace T-1200PIR and set the reading of the IN input to N type
thermocouple.
presys
thermocouple.
The combination constitutes an excellent mounted cavity black body with
presys
The combination constitutes an excellent mounted cavity black body with
emissivity above 0.95 and an effective target of Ø 20 mm well suited for calibration of
presys
emissivity above 0.95 and an effective target of Ø 20 mm well suited for calibration of
pyrometers.
presys
pyrometers.
presys
Align the pyrometer to be calibrated with a black body cavity in the furnace in a
presys
Align the pyrometer to be calibrated with a black body cavity in the furnace in a
horizontal position.
presys
horizontal position.
Observe the distance of the pyrometer to be calibrated against the background
presys
Observe the distance of the pyrometer to be calibrated against the background
of the black body cavity and the size of the actual target (Ø 20 mm) as specified in the
presys
of the black body cavity and the size of the actual target (Ø 20 mm) as specified in the
technical manual optical pyrometer.
presys
technical manual optical pyrometer.

PRESYS Instruments T-1200PIR
Page 5
Use the certificate of calibration of the thermocouple type N to correct the
readings from the IN input of the calibrator and comparing the reading of the pyrometer.
Use a cable to short circuit the TC sensor sheath with the GND (mV (-)) in order to
prevent noise interference.
SCHEMATIC VIEW FOR MOUNTING OF
THE BLACK BODY CAVITY
presys
Use a cable to short circuit the TC sensor sheath with the GND (mV (-)) in order to
presys
Use a cable to short circuit the TC sensor sheath with the GND (mV (-)) in order to
presys
SCHEMATIC VIEW FOR MOUNTING OF
presys
SCHEMATIC VIEW FOR MOUNTING OF
THE BLACK BODY CAVITY
presys
THE BLACK BODY CAVITY
presys

PRESYS Instruments T-1200PIR
Page 6
2.0 - T-1200PIR Calibrator Operating
The T-1200PIR calibrator keeps controlled the temperature of the black body
cavity and is suitable for calibration of infrared thermometers.
The calibrator has 3 operating modes:
Manual Mode to select the temperature straight from the keypad
Programmable Mode: 6 different programs with 11 temperature set point values. The
cavity temperature is selected among the programmed values by the keys and.
The Programmable Mode with Timer scans automatically the temperature set points.
Automatic Mode calibration is performed in an automatic way by the calibrator: the
schedule and the calibration results, besides the thermo-element readings are stored
in the calibrator memory. For pyrometers with local indication, the reading is entered
manually and for pyrometers with response in mA, mV or thermocouple, reading is
performed automatically by the input of the calibrator.
When powered on (ON/OFF key), the calibrator goes through a self-test routine
and shows the last calibration date and the value of the battery voltage. The battery
voltage is constantly monitored and the low battery warning is provided. After the self-
test is completed, the display shows the starting menu. The initial set point is 50.00 ºC.
IN EXEC
CONF CAL COM
By means of keys,,and , choose the menu options and press
ENTER.
IN: selection of the calibrator signal input. Choose among mV, Ohms,
thermocouples, RTDs, mA, switch or none. More details in item 2.1 - IN Menu.
EXEC: The calibrator enters the manual or programmable operating mode.
CONF: Accesses the calibrator configuration options. More details in item 2.2 -
CONF Menu.
CAL: This option accesses the T-1200PIR calibrator adjust functions, protected
by password. More details in section 7 - Calibration.
COM: Accesses the automatic calibration parameters. It is possible to perform a
calibration without a computer or using it (via ISOPLAN software). More details in item
2.5 - Automatic Operating Mode.
presys
The T-1200PIR calibrator keeps controlled the temperature of the black body
presys
The T-1200PIR calibrator keeps controlled the temperature of the black body
to select the temperature straight from the keypad
presys
to select the temperature straight from the keypad
: 6 different programs with 11 temperature set point values. The
presys
: 6 different programs with 11 temperature set point values. The
cavity temperature is selected among the programmed values by the keys
presys
cavity temperature is selected among the programmed values by the keys
presys
presys
and
presys
and
presys
presys
.
presys
.
scans automatically the temperature set points.
presys
scans automatically the temperature set points.
calibration is performed in an automatic way by the calibrator: the
presys
calibration is performed in an automatic way by the calibrator: the
schedule and the calibration results, besides the thermo-element readings are stored
presys
schedule and the calibration results, besides the thermo-element readings are stored
in the calibrator memory. For pyrometers with local indication, the reading is entered
presys
in the calibrator memory. For pyrometers with local indication, the reading is entered
manually and for pyrometers with response in mA, mV or thermocouple, reading is
presys
manually and for pyrometers with response in mA, mV or thermocouple, reading is
performed automatically by the input of the calibrator.
presys
performed automatically by the input of the calibrator.
ON/OFF
presys
ON/OFF
key), the calibrator goes through a self-test routine
presys
key), the calibrator goes through a self-test routine
and shows the last calibration date and the value of the battery voltage. The battery
presys
and shows the last calibration date and the value of the battery voltage. The battery
voltage is constantly monitored and the low battery warning is provided. After the self-
presys
voltage is constantly monitored and the low battery warning is provided. After the self-
test is completed, the display shows the starting menu. The initial set point is 50.00 ºC.
presys
test is completed, the display shows the starting menu. The initial set point is 50.00 ºC.
presys
presys
IN
presys
IN
EXEC
presys
EXEC
presys
presys
presys
CONF
presys
CONF
C
presys
C
AL
presys
AL
COM
presys
COM
presys
By means of keys
presys
By means of keys
presys
presys
,
presys
,
presys
presys
,
presys
,
presys
presys
presys
and
presys
and
presys
presys
, choose the menu options and press
presys
, choose the menu options and press
: selection of the calibrator signal input. Choose among mV, Ohms,
presys
: selection of the calibrator signal input. Choose among mV, Ohms,
thermocouples, RTDs, mA, switch or none. More details in item
presys
thermocouples, RTDs, mA, switch or none. More details in item
EXEC
presys
EXEC
: The calibrator enters the manual or programmable operating mode.
presys
: The calibrator enters the manual or programmable operating mode.
CONF
presys
CONF
: Accesses the calibrator configuration options. More details in item
presys
: Accesses the calibrator configuration options. More details in item
CONF Menu
presys
CONF Menu
.
presys
.
CAL
presys
CAL
: This option accesses the T-1200PIR calibrator adjust functions, protected
presys
: This option accesses the T-1200PIR calibrator adjust functions, protected
by password. More details in section
presys
by password. More details in section
COM
presys
COM
: Accesses the automatic calibration parameters. It is possible to perform a
presys
: Accesses the automatic calibration parameters. It is possible to perform a
calibration without a computer or using it (via ISOPLAN software). More details in item
presys
calibration without a computer or using it (via ISOPLAN software). More details in item
2.5 - Automatic Operating Mode
presys
2.5 - Automatic Operating Mode

PRESYS Instruments T-1200PIR
Page 7
2.1 - IN Menu
mV OHM TC
RTD mA SW NO
mV,mA,SW:selects millivolt, milliampere or switch input, respectively.
OHM:selects ohms input. Following, choose from the menu the wiring in 2, 3 or
4-wire.
2-WIRE 3-WIRE
4-WIRE
TC:selects thermocouple input. Choose among the B,J,K,N,Rand Stypes.
In the next menu, the internal or manual cold junction compensation is chosen.
INTERNAL
MANUAL
If the internal compensation is selected, the cold junction temperature value is
displayed by the calibrator. If the MANUAL option is selected, the cold junction must be
supplied by the operator. After confirming the value, by pressing ENTER, the calibrator
goes back to the operating mode.
RTD: Selects the type of RTD (Resistance Thermometer) used. Choose from
PT100,NI100,CU10 and PT1000 types. Choose also if the wiring is 2, 3 or 4-wire.
NO: Disables reading of external signal.
Selecting one of the options above, the calibrator goes straight to the manual
operating mode, with no need to select the EXEC option.
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
selects millivolt, milliampere or switch input, respectively.
presys
selects millivolt, milliampere or switch input, respectively.
selects ohms input. Following, choose from the menu the wiring in 2, 3 or
presys
selects ohms input. Following, choose from the menu the wiring in 2, 3 or
presys
presys
WIRE
presys
WIRE
presys
presys
presys
presys
presys
presys
presys
presys
selects thermocouple input. Choose among the
presys
selects thermocouple input. Choose among the
B
presys
B
,
presys
,
J
presys
J
,
presys
,
K
presys
K
,
presys
,
N
presys
N
,
presys
,
R
presys
R
and
presys
and
S
presys
S
types.
presys
types.
In the next menu, the internal or manual cold junction compensation is chosen.
presys
In the next menu, the internal or manual cold junction compensation is chosen.
presys
presys
presys
presys
presys
presys
presys
presys
presys
If the internal compensation is selected, the cold junction temperature value is
presys
If the internal compensation is selected, the cold junction temperature value is
displayed by the calibrator. If the
presys
displayed by the calibrator. If the
MANUAL
presys
MANUAL
option is selected, the cold junction must be
presys
option is selected, the cold junction must be
supplied by the operator. After confirming the value, by pressing
presys
supplied by the operator. After confirming the value, by pressing
goes back to the operating mode.
presys
goes back to the operating mode.
: Selects the type of RTD (Resistance Thermometer) used. Choose from
presys
: Selects the type of RTD (Resistance Thermometer) used. Choose from
CU10
presys
CU10
and
presys
and
PT1000
presys
PT1000
types. Choose also if the wiring is 2, 3 or 4-wire.
presys
types. Choose also if the wiring is 2, 3 or 4-wire.
: Disables reading of external signal.
presys
: Disables reading of external signal.
presys
Selecting one of the options above, the calibrator goes straight to the manual
presys
Selecting one of the options above, the calibrator goes straight to the manual
operating mode, with no need to select the
presys
operating mode, with no need to select the
EXEC
presys
EXEC

PRESYS Instruments T-1200PIR
Page 8
2.1 - Input Connections Diagram
presys
presys

PRESYS Instruments T-1200PIR
Page 9
2.2 - CONF Menu
CF PRG MEM LCD
SC BT DT BZ TU
CF: Selects the temperature unit between ºC and ºF. The ITS-90 or IPTS-68
temperature scales are selected for both the thermo-element and for the black body
cavity temperature reading/control.
ºC-90 ºF-90
ºC-68 ºF-68
LCD: This option sets the vacuum fluorescent display contrast. Use the keys
and until you get a better contrast and finish the operation by pressing the key
ENTER.
BT: Shows the battery or the battery charger voltage value if the black body
cavity is turned off or on, respectively.
Battery level
Battery state
Display
4.0 to 7.0 V
normal
----------
< 4.0 V
low
LOW BATTERY
DT: Updates the date and time of the calibrator. Thus, when the calibrator
performs a calibration in the automatic mode via ISOPLAN, all data are registered
together with date and time of occurrence. Every time the calibrator is powered off, the
internal clock does not continue to be updated. However, the ISOPLAN software can
automatically update the calibrator date and time with the computer internal clock.
Otherwise, use keys and to change the field that blinks and the keys and to
go to another field. The key ENTER confirms the last selection.
BZ: Menu that configures the piezoelectric buzzer.
NO YES ENDCAL
NO: Disables the buzzer.
YES: The buzzer beeps when the cavity reaches the set point and stabilizes.
ENDCAL: The buzzer beeps only at the end of a calibration in the automatic
operating mode
presys
presys
presys
presys
presys
ºF. The ITS-90 or IPTS-
presys
ºF. The ITS-90 or IPTS-
temperature scales are selected for both the thermo-element and for the black body
presys
temperature scales are selected for both the thermo-element and for the black body
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
: This option sets the vacuum fluorescent display contrast. Use the keys
presys
: This option sets the vacuum fluorescent display contrast. Use the keys
until you get a better contrast and finish the operation by pressing the key
presys
until you get a better contrast and finish the operation by pressing the key
: Shows the battery or the battery charger voltage value if the black body
presys
: Shows the battery or the battery charger voltage value if the black body
cavity is turned off or on, respectively.
presys
cavity is turned off or on, respectively.
Battery state
presys
Battery state
Display
presys
Display
presys
presys
presys
presys
presys
normal
presys
normal
----------
presys
----------
presys
presys
presys
presys
presys
presys
presys
low
presys
low
LOW BATTERY
presys
LOW BATTERY
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
: Updates the date and time of the calibrator. Thus, when the calibrator
presys
: Updates the date and time of the calibrator. Thus, when the calibrator
performs a calibration in the automatic mode via ISOPLAN, all data are registered
presys
performs a calibration in the automatic mode via ISOPLAN, all data are registered
together with date and time of occurrence. Every time the calibrator is powered off, the
presys
together with date and time of occurrence. Every time the calibrator is powered off, the
internal clock does not continue to be updated. However, the ISOPLAN software can
presys
internal clock does not continue to be updated. However, the ISOPLAN software can
automatically update the calibrator date and time with the computer internal clock.
presys
automatically update the calibrator date and time with the computer internal clock.
Otherwise, use keys
presys
Otherwise, use keys
presys
presys
and
presys
and
presys
presys
to change the field that blinks and the keys
presys
to change the field that blinks and the keys
go to another field. The key
presys
go to another field. The key
ENTER
presys
ENTER
confirms the last selection.
presys
confirms the last selection.
BZ:
presys
BZ:
Menu that configures the piezoelectric buzzer.
presys
Menu that configures the piezoelectric buzzer.
presys
presys
presys
presys
NO
presys
NO
YES
presys
YES
presys
presys
presys
presys
presys
presys
presys
presys
NO:
presys
NO:
Disables the buzzer.
presys
Disables the buzzer.
YES:
presys
YES:
The buzzer beeps when the cavity reaches the set point and stabilizes.
presys
The buzzer beeps when the cavity reaches the set point and stabilizes.
ENDCAL:
presys
ENDCAL:
The buzzer beeps only at the end of a calibration in the automatic
presys
The buzzer beeps only at the end of a calibration in the automatic
operating mode
presys
operating mode

PRESYS Instruments T-1200PIR
Page 10
TU: Menu that configures the PID control parameters for the heating side.
More details in section 6 - PID Control Parameters.
K I D FACT
K: Proportional gain
I: Integral gain
D: Derivative gain
FACT: Restore the control parameters to the factory values.
PRG: Menu that programs the calibrator.
DEC_IN DEC_PRB
SET POINT
DEC_IN: Selects the number of decimals of the thermo-element reading.
The default value depends on the input signal.
DEFAULT
0 1 2 3 4
DEC_PRB: Selects the number of decimals of the cavity temperature and
the set point value. The default number is 1.
0
1
SET POINT: Enables the calibrator Programmable Operating Mode, and
allows the configuration of programmed values. The selected program is
indicated by the selection arrow. Choose one of the 6temperature programs or
NO to disable this Programmable Mode.
Select any of the 6 programs and confirm with the key ENTER.
Following, it is shown the menu of temperature set point configuration.
10% 20% 25%
VARIABLE
Change the configuration to STEPS of 10%,20%, 25%,VARIABLE or press
C/CE to maintain the configuration already stored in memory. The temperature range
of the program must be configured through the values in SET POINT HIGH and SET
POINT LOW in case of 10%,20% or 25% fixed steps of the range. The option
VARIABLE allows the user to define from 2 to 11 values of temperature set point, not
necessarily in ascending order.
To verify the step values of a program, select the VARIABLE option and
confirm the values shown in the display with the key ENTER. The 10%,20% and 25%
presys
Restore the control parameters to the factory values.
presys
Restore the control parameters to the factory values.
Menu that programs the calibrator.
presys
Menu that programs the calibrator.
presys
DEC_PRB
presys
DEC_PRB
presys
presys
presys
presys
presys
presys
presys
Selects the number of decimals of the thermo-element reading.
presys
Selects the number of decimals of the thermo-element reading.
The default value depends on the input signal.
presys
The default value depends on the input signal.
presys
DEFAULT
presys
DEFAULT
presys
presys
presys
presys
1
presys
1
2
presys
2
3
presys
3
4
presys
4
presys
presys
Selects the number of decimals of the cavity temperature and
presys
Selects the number of decimals of the cavity temperature and
the set point value. The default number is 1.
presys
the set point value. The default number is 1.
presys
presys
presys
0
presys
0
presys
presys
1
presys
1
presys
presys
presys
presys
presys
presys
presys
presys
SET POINT
presys
SET POINT
: Enables the calibrator
presys
: Enables the calibrator
Programmable Operating Mode
presys
Programmable Operating Mode
allows the configuration of programmed values. The selected program is
presys
allows the configuration of programmed values. The selected program is
indicated by the selection arrow. Choose one of the
presys
indicated by the selection arrow. Choose one of the
NO
presys
NO
to disable this
presys
to disable this
Programmable Mode
presys
Programmable Mode
Select any of the 6 programs and confirm with the key
presys
Select any of the 6 programs and confirm with the key
Following, it is shown the menu of temperature set point configuration.
presys
Following, it is shown the menu of temperature set point configuration.
presys
presys
presys
presys
10%
presys
10%
presys
presys
presys
presys
presys
VARIABLE
presys
VARIABLE
presys
presys
presys
Change the configuration to STEPS of
presys
Change the configuration to STEPS of
presys
to maintain the configuration already stored in memory. The temperature range
presys
to maintain the configuration already stored in memory. The temperature range
of the program must be configured through the values in
presys
of the program must be configured through the values in
LOW
presys
LOW
in case of
presys
in case of
allows the user to define from 2 to 11 values of temperature set point, not
presys
allows the user to define from 2 to 11 values of temperature set point, not

PRESYS Instruments T-1200PIR
Page 11
options change the number of steps automatically and recalculate the values according
to SET POINT HIGH and SET POINT LOW.
SC: This function scales the input reading of pyrometric transmitters. The
scaling is very useful in temperature transmitter calibration, for instance, because it
displays the current temperature and the transmitter reading in the same unit. Thus,
the error can be verified directly in ºC or ºF. Select the option SC and press ENTER. If
no input is selected in IN, the calibrator will show the SELECT INPUT FIRST message.
In this case, go to menu IN and select the input signal type.
The function SC will show IN or NO. Confirm IN to configure the scaling or NO
to disable the SC function, with the key ENTER.
The scaling is performed via the INPUT HIGH and INPUT LOW parameters,
corresponding to the maximum and minimum values of the calibrator signal input, in
the engineering unit of this signal. Next, configure the SCALE DEC (0-4),SCALE
HIGH and SCALE LOW parameters according to the maximum and minimum values
of the transmitter scale and the desired number of decimals. The scaled value is
shown on the display with the #unit.
For example, temperature transmitter with 0 to 100 ºC input and 4 to 20 mA
output. The scaling with one decimal would be:
INPUT HIGH: 20.0000 mA
INPUT LOW: 4.0000 mA
SCALE DEC (0-4): 1
SCALE HIGH: 100.0 #
SCALE LOW: 0.0 #
presys
: This function scales the input reading of pyrometric transmitters. The
presys
: This function scales the input reading of pyrometric transmitters. The
scaling is very useful in temperature transmitter calibration, for instance, because it
presys
scaling is very useful in temperature transmitter calibration, for instance, because it
displays the current temperature and the transmitter reading in the same unit. Thus,
presys
displays the current temperature and the transmitter reading in the same unit. Thus,
the error can be verified directly in ºC or ºF. Select the option
presys
the error can be verified directly in ºC or ºF. Select the option
SC
presys
SC
and press
presys
and press
ENTER
presys
ENTER
. If
presys
. If
SELECT
presys
SELECT
INPUT
presys
INPUT
FIRST
presys
FIRST
message.
presys
message.
and select the input signal type.
presys
and select the input signal type.
Confirm
presys
Confirm
IN
presys
IN
to configure the scaling or
presys
to configure the scaling or
NO
presys
NO
INPUT HIGH
presys
INPUT HIGH
and
presys
and
INPUT
presys
INPUT
LOW
presys
LOW
parameters,
presys
parameters,
corresponding to the maximum and minimum values of the calibrator signal input, in
presys
corresponding to the maximum and minimum values of the calibrator signal input, in
the engineering unit of this signal. Next, configure the
presys
the engineering unit of this signal. Next, configure the
SCALE DEC (0-
presys
SCALE DEC (0-
4)
presys
4)
,
presys
,
parameters according to the maximum and minimum values
presys
parameters according to the maximum and minimum values
of the transmitter scale and the desired number of decimals. The scaled value is
presys
of the transmitter scale and the desired number of decimals. The scaled value is
presys
unit.
presys
unit.
For example, temperature transmitter with 0 to 100 ºC input and 4 to 20 mA
presys
For example, temperature transmitter with 0 to 100 ºC input and 4 to 20 mA
output. The scaling with one decimal would be:
presys
output. The scaling with one decimal would be:
: 20.0000 mA
presys
: 20.0000 mA
: 4.0000 mA
presys
: 4.0000 mA
4)
presys
4)
: 1
presys
: 1
: 100.0 #
presys
: 100.0 #
SCALE LOW
presys
SCALE LOW
: 0.0 #
presys
: 0.0 #

PRESYS Instruments T-1200PIR
Page 12
MEM: The T-1200PIR calibrators allow many special programs and functions
that can be of frequent use. In situations like this, it would be useful to store the current
configuration in memory in order to save time. Up to 8 configurations can be stored in
memory.
Selecting the option MEM, it is possible to store the current configuration
(WRITE), restore a previous stored configuration (RECALL) or erase the 8
configurations from memory (CLEAR ALL).
WRITE RECALL
CLEARALL
Selecting the option WRITE or RECALL will present a new menu with numbers
1 to 8, representing each one of the memory positions. Choose one of the positions
and press ENTER. The writing operation (WRITE) can be made in an already used
memory position. The calibrator confirms the overwriting confirmation with the
message OVERWRITE MEMORY? The CLEAR ALL operation shows a confirming
message ARE YOU SURE? In both cases, press ENTER to confirm the operation or
C/CE to cancel.
2.3 - Manual Operating Mode
The display shows the selected temperature value of black body cavity and also
the current cavity temperature or thermo-element temperature value.
There are 4 ways in which the information are shown, covering the calibrator
input value (IN), the black body cavity temperature (PROB) and the temperature set
point (SET). The key interchanges the display presentation way:
IN = 23.456 mV
PROB= 300.0 ºC
PROB= 300.1 ºC
SET = 300.0 ºC
23.456
IN = Voltage (mV)
25.01
SET = 300.0 ºC
The black body temperature set point is selected directly by the keypad, even if
the message SET is not being displayed. The numeric keypad enables the SET
selection in any of the display presentation ways, to change the set point.
The set point value is increased by key and decreased by key. While the
keys are kept pressed, the set point continues to be increased or decreased.
The key does not have function in the manual operating mode of the
T-1200PIR calibrators.
presys
: The T-1200PIR calibrators allow many special programs and functions
presys
: The T-1200PIR calibrators allow many special programs and functions
that can be of frequent use. In situations like this, it would be useful to store the current
presys
that can be of frequent use. In situations like this, it would be useful to store the current
configuration in memory in order to save time. Up to 8 configurations can be stored in
presys
configuration in memory in order to save time. Up to 8 configurations can be stored in
, it is possible to store the current configuration
presys
, it is possible to store the current configuration
) or erase the 8
presys
) or erase the 8
presys
presys
presys
presys
presys
presys
presys
presys
presys
presys
will present a new menu with numbers
presys
will present a new menu with numbers
1 to 8, representing each one of the memory positions. Choose one of the positions
presys
1 to 8, representing each one of the memory positions. Choose one of the positions
WRITE
presys
WRITE
) can be made in an already used
presys
) can be made in an already used
memory position. The calibrator confirms the overwriting confirmation with the
presys
memory position. The calibrator confirms the overwriting confirmation with the
CLEAR
presys
CLEAR
ALL
presys
ALL
operation shows a confirming
presys
operation shows a confirming
In both cases, press
presys
In both cases, press
ENTER
presys
ENTER
to confirm the operation or
presys
to confirm the operation or
2.3 - Manual Operating Mode
presys
2.3 - Manual Operating Mode
The display shows the selected temperature value of black body cavity and also
presys
The display shows the selected temperature value of black body cavity and also
the current cavity temperature or thermo-element temperature value.
presys
the current cavity temperature or thermo-element temperature value.
There are 4 ways in which the information are shown, covering the calibrator
presys
There are 4 ways in which the information are shown, covering the calibrator
), the black body cavity temperature (
presys
), the black body cavity temperature (
PROB
presys
PROB
) and the temperature set
presys
) and the temperature set
interchanges the display presentation way:
presys
interchanges the display presentation way:
presys
presys
IN = 23.456 mV
presys
IN = 23.456 mV
presys
presys
presys
presys
PROB=
presys
PROB=
300.
presys
300.
0 ºC
presys
0 ºC
presys
presys
presys
presys
presys
PROB=
presys
PROB=
300
presys
300
.1 ºC
presys
.1 ºC
presys
presys
presys
presys
presys
SET = 300
presys
SET = 300
.0
presys
.0
ºC
presys
ºC
presys
presys
presys
presys
presys
23.456
presys
23.456
presys
presys
presys
presys
presys
IN = Voltage (mV)
presys
IN = Voltage (mV)
presys
presys
presys
presys
presys
25.01
presys
25.01
presys
presys
presys
presys
presys
SET =
presys
SET =
300
presys
300
presys
presys
presys
The black body temperature set point is selected directly by the keypad, even if
presys
The black body temperature set point is selected directly by the keypad, even if
the message
presys
the message
presys
SET
presys
SET
is not being displayed. The numeric keypad enables the
presys
is not being displayed. The numeric keypad enables the
selection in any of the display presentation ways, to change the set point.
presys
selection in any of the display presentation ways, to change the set point.
The set point value is increased by key
presys
The set point value is increased by key
presys
keys are kept pressed, the set point continues to be increased or decreased.
presys
keys are kept pressed, the set point continues to be increased or decreased.

PRESYS Instruments T-1200PIR
Page 13
Internally, the set point changes at a limited rate to make possible a safer and
more homogeneous heating of the internal ceramic tube. Besides, the PID parameters
and algorithm were adjusted to assure a minimal stabilization time and overshoot.
2.4 - Programmable Operating Mode
Pre-configured programs can be loaded from the calibrator memory, enabling
the programmable operating mode. The temperature programmed values of the cavity
are used directly, with no need to enter the set point.
The display shows STEPn beside the cavity temperature set point value in the
programmable mode. The number of the program is indicated by n. Using the keys
and, the programmed values of temperature set point are changed. The numeric
keypad continues available for manual selection of the black body cavity temperature in
the same way of the manual operating mode.
The automatic scan over the programmed temperatures is implemented by
defining the stabilization time of the thermo-element in the black body cavity.
The key enables the automatic scan over the points. When pressed, the
message STEPn gives place to 0s and the calibrator waits for the stabilization time
configured from 1 to 9 minutes, by the keys 1 to 9. The automatic scan is disabled by
pressing key again.
A countdown of the stabilization time is only started when the cavity temperature
reaches the programmed temperature and stabilizes within a range of approximately
0.20 ºC. At this time the buzzer beeps, in case it is configured to YES.
presys
and algorithm were adjusted to assure a minimal stabilization time and overshoot.
presys
and algorithm were adjusted to assure a minimal stabilization time and overshoot.
-configured programs can be loaded from the calibrator memory, enabling
presys
-configured programs can be loaded from the calibrator memory, enabling
the programmable operating mode. The temperature programmed values of the cavity
presys
the programmable operating mode. The temperature programmed values of the cavity
are used directly, with no need to enter the set point.
presys
are used directly, with no need to enter the set point.
beside the cavity temperature set point value in the
presys
beside the cavity temperature set point value in the
programmable mode. The number of the program is indicated by
presys
programmable mode. The number of the program is indicated by
n
presys
n
. Using the keys
presys
. Using the keys
, the programmed values of temperature set point are changed. The numeric
presys
, the programmed values of temperature set point are changed. The numeric
keypad continues available for manual selection of the black body cavity temperature in
presys
keypad continues available for manual selection of the black body cavity temperature in
the same way of the manual operating mode.
presys
the same way of the manual operating mode.
The automatic scan over the programmed temperatures is implemented by
presys
The automatic scan over the programmed temperatures is implemented by
defining the stabilization time of the thermo-element in the black body cavity.
presys
defining the stabilization time of the thermo-element in the black body cavity.
enables the automatic scan over the points. When pressed, the
presys
enables the automatic scan over the points. When pressed, the
gives place to
presys
gives place to
0s
presys
0s
and the calibrator waits for the stabilization time
presys
and the calibrator waits for the stabilization time
configured from 1 to 9 minutes, by the keys
presys
configured from 1 to 9 minutes, by the keys
1
presys
1
to
presys
to
9
presys
9
. The automatic scan is disabled by
presys
. The automatic scan is disabled by
presys
again.
presys
again.
A countdown of the stabilization time is only started when the cavity temperature
presys
A countdown of the stabilization time is only started when the cavity temperature
reaches the programmed temperature and stabilizes within a range of approximately
presys
reaches the programmed temperature and stabilizes within a range of approximately
0.20 ºC. At this time the buzzer beeps, in case it is configured to
presys
0.20 ºC. At this time the buzzer beeps, in case it is configured to

PRESYS Instruments T-1200PIR
Page 14
2.5 - Automatic Operating Mode
The thermo-element calibration is performed in an automatic way by the
T-1200PIR calibrator. The configuration, as well as the calibration verification is carried
out by the calibrator itself. Also it is possible to use the ISOPLAN software and its work
orders, like a CAC - Computer Aided Calibration.
The independent automatic calibration, without the use of ISOPLAN, is planned
in the option TAGMAN from menu COM.
TAG EXEC VERIF
ADDRESS
TAGMAN
Before start programming, configure the signal that will be read by the calibrator
in the menu IN. To calibrate pyrometers with local indication, for example, there is no
electric signal to be read. In this case, option IN from the menu must be configured to
NO and the calibrator will ask to enter the value indicated by the pyrometer, at the end
of the stabilization time of each calibration point.
The data for an automatic calibration are:
TAG: the pyrometer tag identification.
SP: the black body cavity temperature reference values for the calibration
(calibration points).
TOL: the maximum tolerance for the pyrometer operation.
STB: the stabilization time, in seconds, so that the pyrometer reading can be taken.
This timer starts just after the cavity reaches and stabilizes at the set point
temperature.
STR: the calibration strategy of the programmed reference values. The available
strategies are: (UP), (DOWN), (UP - DOWN), (DOWN - UP),
(UP - DOWN - UP) and (DOWN - UP - DOWN).
RP: the strategy number of repetitions.
RGI: the pyrometer indication range.
RGO: the operating temperature range that corresponds to the indication range
above.
The automatic calibration begins when the option EXEC from menu COM is
selected. All the operations are automatically performed by the T-1200PIR calibrator.
The keypad does not work until the end of the calibration.
At the end of the stabilization time, the calibrator stores the pyrometer reading in
memory and goes to the next point, in case some input signal has been previously
configured in the menu IN and connected to the calibrator terminals.
The CALIBRATION END message appears on the display at the end of the
automatic calibration. Press ENTER to confirm. The results can be verified in option
VERIF from menu COM.
The first message in option VERIF informs the calibration result, with the number
of points that have succeeded or not. Following, press ENTER to verify each one of the
readings performed by the calibrator. The keys and interchange 2 screens: one
that displays the cavity and thermo-element temperature value and the other displays a
presys
The thermo-element calibration is performed in an automatic way by the
presys
The thermo-element calibration is performed in an automatic way by the
T-1200PIR calibrator. The configuration, as well as the calibration verification is carried
presys
T-1200PIR calibrator. The configuration, as well as the calibration verification is carried
out by the calibrator itself. Also it is possible to use the ISOPLAN software and its work
presys
out by the calibrator itself. Also it is possible to use the ISOPLAN software and its work
The independent automatic calibration, without the use of ISOPLAN, is planned
presys
The independent automatic calibration, without the use of ISOPLAN, is planned
presys
presys
VERIF
presys
VERIF
presys
presys
presys
presys
presys
TAGMAN
presys
TAGMAN
presys
presys
presys
Before start programming, configure the signal that will be read by the calibrator
presys
Before start programming, configure the signal that will be read by the calibrator
. To calibrate pyrometers with local indication, for example, there is no
presys
. To calibrate pyrometers with local indication, for example, there is no
IN
presys
IN
from the menu must be configured to
presys
from the menu must be configured to
and the calibrator will ask to enter the value indicated by the pyrometer, at the end
presys
and the calibrator will ask to enter the value indicated by the pyrometer, at the end
of the stabilization time of each calibration point.
presys
of the stabilization time of each calibration point.
The data for an automatic calibration are:
presys
The data for an automatic calibration are:
: the pyrometer tag identification.
presys
: the pyrometer tag identification.
: the black body cavity temperature reference values for the calibration
presys
: the black body cavity temperature reference values for the calibration
: the maximum tolerance for the pyrometer operation.
presys
: the maximum tolerance for the pyrometer operation.
: the stabilization time, in seconds, so that the pyrometer reading can be taken.
presys
: the stabilization time, in seconds, so that the pyrometer reading can be taken.
This timer starts just after the cavity reaches and stabilizes at the set point
presys
This timer starts just after the cavity reaches and stabilizes at the set point
: the calibration strategy of the programmed reference values. The available
presys
: the calibration strategy of the programmed reference values. The available
presys
presys
(UP),
presys
(UP),
presys
presys
(DOWN),
presys
(DOWN),
presys
presys
presys
presys
(UP - DOWN),
presys
(UP - DOWN),
(UP - DOWN - UP) and
presys
(UP - DOWN - UP) and
presys
presys
presys
presys
presys
presys
presys
(DOWN - UP - DOWN).
presys
(DOWN - UP - DOWN).
: the strategy number of repetitions.
presys
: the strategy number of repetitions.
: the pyrometer indication range.
presys
: the pyrometer indication range.
: the operating temperature range that corresponds to the indication range
presys
: the operating temperature range that corresponds to the indication range
above.
presys
above.
The automatic calibration begins when the option
presys
The automatic calibration begins when the option
selected. All the operations are automatically performed by the T-1200PIR calibrator.
presys
selected. All the operations are automatically performed by the T-1200PIR calibrator.
The keypad does not work until the end of the calibration.
presys
The keypad does not work until the end of the calibration.
At the end of the stabilization time, the calibrator stores the pyrometer reading in
presys
At the end of the stabilization time, the calibrator stores the pyrometer reading in
memory and goes to the next point, in case some input signal has been previously
presys
memory and goes to the next point, in case some input signal has been previously
configured in the menu
presys
configured in the menu
IN
presys
IN
and connected to the calibrator terminals.
presys
and connected to the calibrator terminals.
The
presys
The
CALIBRATION
presys
CALIBRATION
automatic calibration. Press
presys
automatic calibration. Press
from menu
presys
from menu
COM
presys
COM
The first message in option
presys
The first message in option

PRESYS Instruments T-1200PIR
Page 15
message indicating the calibration point number and its state (OK or FAIL), besides the
error value in %.
The option TAG from menu COM can present a list of up to 4 tags that have
been downloaded or uploaded with ISOPLAN software. The key ENTER selects the tag
to be calibrated from the tag list. The automatic calibration starts by the option EXEC
from menu COM and the calibration data can be verified in option VERIF. To calibrate
automatically not using the ISOPLAN, there is a reserved place for one manual tag.
The selection of the manual tag is performed by confirming the option TAGMAN with
the key ENTER.
The option ADDRESS selects the communication address of the T-1200PIR
calibrator. The communication protocol used is ModBus - RTU, with no parity and baud
rate of 9600. The calibrator communication with the computer can use RS-232 or RS-
485, for network option, according to the communication interface used. In order to
communicate with ISOPLAN software, configure ADDRESS to 1.
presys
can present a list of up to 4 tags that have
presys
can present a list of up to 4 tags that have
ENTER
presys
ENTER
selects the tag
presys
selects the tag
to be calibrated from the tag list. The automatic calibration starts by the option
presys
to be calibrated from the tag list. The automatic calibration starts by the option
EXEC
presys
EXEC
and the calibration data can be verified in option
presys
and the calibration data can be verified in option
VERIF
presys
VERIF
. To calibrate
presys
. To calibrate
automatically not using the ISOPLAN, there is a reserved place for one manual tag.
presys
automatically not using the ISOPLAN, there is a reserved place for one manual tag.
The selection of the manual tag is performed by confirming the option
presys
The selection of the manual tag is performed by confirming the option
TAGMAN
presys
TAGMAN
with
presys
with
selects the communication address of the T-1200PIR
presys
selects the communication address of the T-1200PIR
calibrator. The communication protocol used is ModBus - RTU, with no parity and baud
presys
calibrator. The communication protocol used is ModBus - RTU, with no parity and baud
rate of 9600. The calibrator communication with the computer can use RS-232 or RS-
presys
rate of 9600. The calibrator communication with the computer can use RS-232 or RS-
presys
485, for network option, according to the communication interface used. In order to
presys
485, for network option, according to the communication interface used. In order to
communicate with ISOPLAN software, configure
presys
communicate with ISOPLAN software, configure
ADDRESS
presys
ADDRESS
to 1.
presys
to 1.

PRESYS Instruments T-1200PIR
Page 16
3.0 - Usage Recommendations
It is of great importance to know the instrument to be calibrated. Infrared
thermometers have characteristics which must be met so that the observed results are
correct, as setting of emissivity (if any), focal length and focus area,e.g.These data are
easily obtained in the thermometer user manual and this observation is needed to
prove the compatibility of thermometers with the calibrator.
To perform the readings, align the pyrometer with the black body cavity
horizontally. Respect the distance from the pyrometer to be calibrated in relation to the
black body cavity to the size of the actual cavity as pyrometer manual.Remember that
the area focused by the pyrometer to be calibrated must be less than or equal to the
black body cavity size in order to not introduce measurement errors.
The external sensor can be used to increase the accuracy of temperature indication.
4.0 - Safety Instructions
Do not leave the place where the calibrator is on without identification and warnings.
Before turning off the calibrator, return temperature black body cavity for values
close to ambient temperature.
presys
It is of great importance to know the instrument to be calibrated. Infrared
presys
It is of great importance to know the instrument to be calibrated. Infrared
thermometers have characteristics which must be met so that the observed results are
presys
thermometers have characteristics which must be met so that the observed results are
.
presys
.
These data are
presys
These data are
easily obtained in the thermometer user manual and this observation is needed to
presys
easily obtained in the thermometer user manual and this observation is needed to
To perform the readings, align the pyrometer with the black body cavity
presys
To perform the readings, align the pyrometer with the black body cavity
horizontally. Respect the distance from the pyrometer to be calibrated in relation to the
presys
horizontally. Respect the distance from the pyrometer to be calibrated in relation to the
black body cavity to the size of the actual cavity as pyrometer manual
presys
black body cavity to the size of the actual cavity as pyrometer manual
.
presys
.
Remember that
presys
Remember that
the area focused by the pyrometer to be calibrated must be less than or equal to the
presys
the area focused by the pyrometer to be calibrated must be less than or equal to the
black body cavity size in order to not introduce measurement errors.
presys
black body cavity size in order to not introduce measurement errors.
The external sensor can be used to increase the accuracy of temperature indication.
presys
The external sensor can be used to increase the accuracy of temperature indication.
Do not leave the place where the calibrator is on without identification and warnings.
presys
Do not leave the place where the calibrator is on without identification and warnings.
presys
Before turning off the calibrator, return temperature black body cavity for values
presys
Before turning off the calibrator, return temperature black body cavity for values

PRESYS Instruments T-1200PIR
Page 17
5.0 - Calibrator Warning Messages
Warning
Meaning
Procedure
RAM ERROR
READ MANUAL Problem in RAM
memory
Turn the calibrator off and
on. If the error persists,
send the instrument to the
factory
EEPROM ERROR
READ MANUAL
Problem in EEPROM
memory
Same as the previous item
LOW BATTERY
Level of battery voltage is
low
Check the T-1200PIR
calibrator power supply
UNDER / OVER
Input signal out of
specifications or scaling
range
See item 1.1.1 on Input
Specifications
????.??ºC Input sensor is open
Check input connections
and sensor
6.0 - PID Control Parameters
The T-1200PIR calibrators have a PID control algorithm to calculate the
temperature control output.
The black body cavity stability and response time features are related to the PID
parameters, explained below:
The Kparameter (proportional gain) amplifies the error signal between the set
point and the cavity temperature to establish the output signal. When this parameter is
very high, the output reaction is very quick; however this can take the system into
oscillation. Decreasing this parameter, the cavity would not be able to react quickly
enough to external variations, giving the impression of a sudden out of control.
The Iparameter (integral gain) is responsible for the integral action and it is the
most important part in the set point control. While an error persists between the set
point and the cavity temperature, the integral action will actuate on the output signal
until the error is brought to zero.
The Dparameter (derivative gain) is responsible for the derivative action that
provides a quick response at the control output resulting from any rapt variation in the
cavity temperature. It is used to eliminate oscillations. However, it can cause
oscillations in the presence of much noise.
All temperature calibrators are tuned in factory and the parameters are close to
the optimum ones. In case one wants to improve a specific feature of the calibrator
(stabilization time or response time, for instance), make sure the alteration is made
reasonably.
presys
Procedure
presys
Procedure
presys
urn the calibrator off and
presys
urn the calibrator off and
on. If the error persists,
presys
on. If the error persists,
send the instrument to the
presys
send the instrument to the
factory
presys
factory
presys
presys
Same as the previous item
presys
Same as the previous item
presys
presys
presys
presys
Level of battery voltage is
presys
Level of battery voltage is
Check the T
presys
Check the T
-
presys
-
1200P
presys
1200P
IR
presys
IR
calibrator power supply
presys
calibrator power supply
presys
presys
presys
presys
presys
presys
specifications or scaling
presys
specifications or scaling
See item 1.1.1 on Input
presys
See item 1.1.1 on Input
Specifications
presys
Specifications
presys
presys
presys
presys
presys
presys
Input sensor is open
presys
Input sensor is open
Check input connections
presys
Check input connections
and sensor
presys
and sensor
presys
presys
presys
presys
presys
presys
presys
6.0 - PID Control Parameters
presys
6.0 - PID Control Parameters
The T-1200PIR calibrators have a PID control algorithm to calculate the
presys
The T-1200PIR calibrators have a PID control algorithm to calculate the
temperature control output.
presys
temperature control output.
The black body cavity stability and response time features are related to the PID
presys
The black body cavity stability and response time features are related to the PID
parameters, explained below:
presys
parameters, explained below:
parameter (proportional gain) amplifies the error signal between the set
presys
parameter (proportional gain) amplifies the error signal between the set
and the cavity temperature to establish the output signal. When this parameter is
presys
and the cavity temperature to establish the output signal. When this parameter is
very high, the output reaction is very quick; however this can take the system into
presys
very high, the output reaction is very quick; however this can take the system into
oscillation. Decreasing this parameter, the cavity would not be able to react quickly
presys
oscillation. Decreasing this parameter, the cavity would not be able to react quickly
enough to external variations, giving the impression of a sudden out of control.
presys
enough to external variations, giving the impression of a sudden out of control.
The
presys
The
I
presys
I
parameter (integral gain) is responsible for the integral action and it is the
presys
parameter (integral gain) is responsible for the integral action and it is the
most important part in the set point control. While an error persists between the set
presys
most important part in the set point control. While an error persists between the set
point
presys
point
and the cavity temperature, the integral action will actuate on the output signal
presys
and the cavity temperature, the integral action will actuate on the output signal
until the error is brought to zero.
presys
until the error is brought to zero.
The
presys
The
D
presys
D
parameter (derivative gain) is responsible for the derivative action that
presys
parameter (derivative gain) is responsible for the derivative action that
provides a quick response at the control output resulting from any rapt variation in the
presys
provides a quick response at the control output resulting from any rapt variation in the
cavity temperature. It is used to eliminate oscillations. However, it can cause
presys
cavity temperature. It is used to eliminate oscillations. However, it can cause
oscillations in the presence of much noise.
presys
oscillations in the presence of much noise.
All temperature calibrators are tuned in factory and the parameters are close to
presys
All temperature calibrators are tuned in factory and the parameters are close to
the optimum ones. In case one wants to improve a specific feature of the calibrator
presys
the optimum ones. In case one wants to improve a specific feature of the calibrator
(stabilization time or response time, for instance), make sure the alteration is made
presys
(stabilization time or response time, for instance), make sure the alteration is made
reasonably.
presys
reasonably.
Table of contents
Other Presys Test Equipment manuals

Presys
Presys T-500PIR User manual

Presys
Presys PC-507-IS User manual

Presys
Presys PCA-570-RM User manual

Presys
Presys T-25N User manual

Presys
Presys T-1200PH User manual

Presys
Presys TA-1200P User manual

Presys
Presys MCS-12-IS User manual

Presys
Presys PCON-Y18-LP User manual

Presys
Presys TA-1200PLAB User manual

Presys
Presys T-350P User manual