manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Rae
  6. •
  7. Measuring Instrument
  8. •
  9. Rae TN-204 User manual

Rae TN-204 User manual

Technical Note TN-204 04/17/vk
1
Honeywell Internal
RAE Systems
by Honeywell 877-723-2878 raesystems.com
MODES & APPLICATIONS FOR THE MULTIRAE BENZENE
INTRODUCTION
Benzene is a very well-known human carcinogen that is widely
distributed throughout nature (coal, crude oil, gas and gas con-
densate). In most fuel products (gasoline, diesel, jet fuel), it ex-
istsinconcentrationsofuptoseveralpercent,byvolume(vol%).
Benzene is a very important industrial
chemical,aswell.ItbelongstotheBTEX
aromatic family and is a liquid with a ra-
ther high vapor pressure. Due toits haz-
ardousproperties,Cal/OSHAset itsPEL
(Personal Exposure Limit) as 1 ppm of
TWA and 5 ppm of STEL (Short Term
Exposure Limit).1ACGIH recommends
even more strict limits of 0.5 and 2.5
ppm, respectively.2There are several
portable products on the market (from
RAE Systems, IonScience, Dräger) that
areabletomeasuresuchlowconcentra-
tions of benzene, but only the MultiRAE
Benzene combines the ability to make
quick benzene-specific assessment
(“snapshot”) measurements with the ad-
vantagesof a versatilemulti-gasmonitor (upto sixgas sensors)
with real-time wireless capability. It can support more than 20
intelligent interchangeable sensor options.
Note: This Technical Note applies only to the MultiRAE Ben-
zene monitor. For information regarding the UltraRAE 3000, re-
fer to RAE Systems Technical Note TN-127: “Benzene-Specific
Measurements In Petroleum Hydrocarbons Using The RAE
Systems UltraRAE 3000.”
INSTRUMENT AND CARTRIDGE
The MultiRAE Benzene is designed specifically for use with the
replaceable benzene-specific RAE-Sep Tube Cartridge, which
comprises six individual RAE-Sep tubes.
The MultiRAE Benzene hastwo modes: TVOC (Total Volatile
Organic Compounds) mode and Benzene mode. When used in
TVOC mode, its display shows all sensor readings continuously.
TVOC Mode
Benzene Mode
When the instrument is in benzene mode, the display prompts for
snapshot sampling.(Refer to the MultiRAE User’s Guide for details
on managing, calibrating, and measuring in both TVOC and Ben-
zene modes3.) To perform snapshot compound-specific (ben-
zene) measurement in addition to general VOC measurement
requires a RAE-Sep Tube Cartridge (available in packs of six:
P/N M01-0312-000).
Note: Each tube should be used for one measurement only.
The RAE-Sep Tube Cartridge is designed to provide six one-
time specific measurements of benzene. This cartridge has
unique properties and performance features not found in other
monitors’ separation tube configurations. It has been designed
to provide accurate, benzene-only measurement and has only
been evaluated and tested on the MultiRAE Benzene monitor.
Note: Failure to use the MultiRAE Benzene monitor with the
RAE-Sep Tube Cartridge will result in inaccurate detection or
measurement, which may include the failure to detect benzene
and possible serious injury or death.
Forpropercartridgeoperationandreplacement,refertothedoc-
ument “RAE-Sep Tube Cartridge For MultiRAE Benzene.”4
MultiRAE Benzene
with RAE
-Sep Tube
Cartridge
RAE-Sep Cartridge
Technical Note TN-204 04/17/vk
2
Honeywell Internal
RAE Systems
by Honeywell 877-723-2878 raesystems.com
Some measurable parameters of the instrument are presented be-
low.Refer to the MultiRAE Benzene datasheet for comprehensive
sensor specification data.5
Sensor/Mode
Range
Reso-
lution
9.8eV for use in TVOC mode 10 ppb to 2,000 ppm 10 ppb
9.8eV for use in Benzene
mode (with RAE-Sep Tube
Cartridge)
0.1 to 200 ppm 0.1 ppm
MULTIRAE BENZENE-SPECIFIC MEASUREMENT
PRINCIPLE
In Benzene mode, the monitor is designed to make specific ben-
zene measurements. This is achieved via these two approaches:
1. The monitor uses a lower-energy, 9.8eV PID lamp. This al-
lows excluding from cross-contamination all the VOCs with
ionization energy (IE) above 9.8eV.
2. RAE-Sep tubes eliminate most VOCs. Two major effects
are present in them:
•Gas absorption in the tube. This is usually a fast pro-
cess, and it is reversible.
•Gas reaction (chemical) in the tube. This is a slower
process, and it is not reversible.
Note: Some VOCs can be eluted (removed) from the RAE-Sep
Tube and cause false positivebenzene readings.
In addition, if the instruments are insensitive to some gases (for
example,theirIEisabove9.8eVoriftheyrequirehighcorrection
factors), the products of their chemical reactions in RAE-Sep
Tubesmaybecomesensitiveandmaydistortthebenzeneread-
ing. A separate case is when gases don’t absorb in the RAE-
Sep Tube and are insensitive to the monitor equipped with a
9.8eVlamp(forexample,lowerhydrocarbonssuchasmethane,
ethane, etc.).
Example of Elution
WhentheinstrumentinBenzenemodewasexposedto400ppm
of acetone for 60 seconds, the reading didn’t exceed 0.2 ppm.
After 5 minutes running in air, the reading increased and
reached almost 30 ppm.
Elution in Benzene mode with exposure to 400 ppm of
acetone.
Acetone didn’t convert in the RAE-Sep Tube within a short pe-
riod of time (slow chemical reaction) and further eluted with the
air stream.
Important: For the correct measurement and toavoid a false positive
response,RAE-SepTubesshouldbe usedjustonetime.
Humidity Effect
Humidity has no effect on the response to benzene. However,
high humidity affects the capacitance of the tubes to remove in-
terfering hydrocarbons, especially when they are in high con-
centrations.
The figure below shows the response of the instrument to the
mix of benzene (2 ppm) and octane (180 ppm). HPS-2 is the
case when the tube was opened and the instrument run with a
high pump speed (HPS) for about 5 minutes in ambient air with
relative humidity (RH)of 75 to 80% before a benzene measure-
mentin Benzenemode wasdone.HPS-1 isthe casewhenben-
zene measurements were done right after opening the tube.
Response of Benzene and Octane with different times of
measurement after opening the tubes, HPS-1 and HPS-2.
Each line represents a single tube, and each triangle repre-
sents a single measurement with that tube.
0
5000
10000
15000
20000
25000
30000
35000
010 20 30
Response, ppb
Time, min
Acetone eluate from Sep-Tube
ppb
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
02468
Reading, ppm
Numbers of measurements
Benzene+Octane, response vs. numbers
of measurements
HPS-1
HPS-2
Technical Note TN-204 04/17/vk
3
Honeywell Internal
RAE Systems
by Honeywell 877-723-2878 raesystems.com
Tube capacitance can be extended significantly while using the
instrument’s low pump speed (LPS) instead of its high pump
speedas defaultsetting.Testconditionsarethesame asabove.
Note: Previous results reinforce our strong recommendation to
only open RAE-Sep Tubes right before performing a benzene-
specific measurement. Other approaches could lead to false
positive results.
Temperature Effects
Temperaturevariation affectsRAE-SepTubeperformance. De-
creasingtemperatureleadstothehighertubeabsorption,butslows
down chemical transformation. Asaresult, the tubemay not show a
color change,butthetubefullyorpartiallylosesitscapacitance.To
avoid confusion, followthe recommendationof “One tube,one
measurement.”Note:Calibrationshould beperformedat thesame
temperature as themeasurementbecause time necessary forsam-
pling changeswithtemperature.
Tube Capacitance
Tube capacitance in this case isdefined as: MultiRAE Benzene
with a RAE-Sep Tube exposed to the gases in the table for a
limited time until the reading reached 0.5 ppm.
Concentration vs. Exposure Time
Gas
Gas concentration,
ppm
Capacitance,
minutes
Toluene
100
≥10
i-Butylene
100
≥10
i-Butylene
1000
5
Methane
25000
No response
n-Hexane
2400
10 sec
n-Octane
500
1*
n-Pentane
3500
7 sec
n-Propane
11000
No response
* Reading 1 ppm after 1 minute exposure
Potential Cross-Sensitivity
Gas Cross-
sensitivity
Gas concentration,
ppm
Benzene
Response, ppm
Pentane
1700
0.1
Propane
5500
0
i-Butane
4500
0.5
i-Butylene
100
0.6
i-Butylene
1000
0.8
Methane
12500
0
Octane
250
0.3
Hexane
1200
0.6
Hexane
100
0
Butadiene 1,3
2.5
0.5
Toluene
50
1
Acetone
200
0.1
Benzene Absorption by RAE-Sep Tubes
There is a small effect of the benzene absorbtion by the RAE-
Sep Tube that potentially could slightly decrease a benzene
reading, but it doesn’t exceed 50 to 60 ppb, which is below the
instrument’s resolution of 0.1 ppm (100ppb).
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
02468
Reading, ppm
Numbers of measurements
Benzene+Octane, response vs. numbers
of measurements
HPS-1
LPS-1
HPS-2
LPS-2
0
50
100
150
200
250
300
110 149 225 328
Reading, ppb
Benzene concentration, ppb
Effect of tube on Benzene absorption
Without tube With tube
Technical Note TN-204 04/17/vk
4
Honeywell Internal
RAE Systems
by Honeywell 877-723-2878 raesystems.com
Examples of RAE-Sep Tubes’appearance after exposure to
different VOCs and environment
Unused RAE-Sep tube. Note the uniform orange color of the
fresh absorbant.
The following examples show various degrees of capacitance.
Do not use opened tubes that look like these.
1. Freshly opened tube exposed to a mixture of benzene and
octane. The exposure was over the tube’s capacitance.
2. Open tube with instrument run for 5 minutes at 75 to 80%
relativehumidityandthenexposedto amixture ofbenzene
andoctane. The tubelost capacitance.The front partdidn’t
absorb VOC.
3. Open tube with instrument run for 5 minutes at 75 to 80%
relative humidity and then exposed to benzene and pro-
pane. The tube lost capacitance. The front part didn’t ab-
sorb VOC, and the color is mostly brownish instead of
green.
4. Freshly opened tube exposed to acetone. The tube’s color
almost didn’t change, though tube capacitance suffered.
REFERENCES
1 California Code of Regulations,Section 5218:
https://www.dir.ca.gov/title8/5218.html
2 American Conference of Governmental Industrial Hygien-
ists (ACGIH) Threshold Limit Value: https://www.osha.gov/
dts/chemicalsampling/data/CH_220100.html
3 MultiRAE User’s Guide:
http://www.raesystems.com/sites/default/files/content/
resources/
4 MultiRAE Benzene Tube Cartridge Instructions:
http://www.raesystems.com/sites/default/files/content/
resources/MultiRAE%20Benzene%20Tube%20Car-
tridge%20QSG%20-%20A5_Page%20RevA2.pdf
5 MultiRAE Benzene Datasheet:
http://www.raesystems.com/sites/default/files/content/
resources/Datasheet_MultiRAE-%20Benzene_DS-1098-
01.pdf

Other Rae Measuring Instrument manuals

Rae microrae User manual

Rae

Rae microrae User manual

Rae ToxiRAE Pro User manual

Rae

Rae ToxiRAE Pro User manual

Rae Q-RAE User manual

Rae

Rae Q-RAE User manual

Rae QRAE 3 User manual

Rae

Rae QRAE 3 User manual

Rae GammaRAE II Installation guide

Rae

Rae GammaRAE II Installation guide

Rae UltraRAE 3000 Assembly instructions

Rae

Rae UltraRAE 3000 Assembly instructions

Rae QRAE II User manual

Rae

Rae QRAE II User manual

Rae UltraRAE 3000 User manual

Rae

Rae UltraRAE 3000 User manual

Rae AutoRAE Lite User manual

Rae

Rae AutoRAE Lite User manual

Rae IAQRAE PGM-5210 Manual

Rae

Rae IAQRAE PGM-5210 Manual

Rae GammaRAE II User manual

Rae

Rae GammaRAE II User manual

Rae MultiRAE Plus Guide

Rae

Rae MultiRAE Plus Guide

Rae UltraRAE PGM-7200 User manual

Rae

Rae UltraRAE PGM-7200 User manual

Rae DoseRAE 2 User manual

Rae

Rae DoseRAE 2 User manual

Rae microrae User manual

Rae

Rae microrae User manual

Rae DoseRAE 2 User manual

Rae

Rae DoseRAE 2 User manual

Rae ToxiRAE Pro PID User manual

Rae

Rae ToxiRAE Pro PID User manual

Rae GammaRAE II Responder User manual

Rae

Rae GammaRAE II Responder User manual

Rae MiniRAE Lite User manual

Rae

Rae MiniRAE Lite User manual

Rae ppbRAE 3000 User manual

Rae

Rae ppbRAE 3000 User manual

Rae MultiRAE Plus PGM-50/4 User manual

Rae

Rae MultiRAE Plus PGM-50/4 User manual

Rae MiniRAE 2000 PGM-7600 User manual

Rae

Rae MiniRAE 2000 PGM-7600 User manual

Rae NeutronRAE II User manual

Rae

Rae NeutronRAE II User manual

Rae AreaRAE Installation guide

Rae

Rae AreaRAE Installation guide

Popular Measuring Instrument manuals by other brands

VOLTCRAFT DT 2 L / k operating instructions

VOLTCRAFT

VOLTCRAFT DT 2 L / k operating instructions

CTC Union MMX2000 Series product manual

CTC Union

CTC Union MMX2000 Series product manual

Dostmann Electronic P700-EX Operation manual

Dostmann Electronic

Dostmann Electronic P700-EX Operation manual

Reed Instruments R9230 instruction manual

Reed Instruments

Reed Instruments R9230 instruction manual

Mirion Technologies RDS-31 S user manual

Mirion Technologies

Mirion Technologies RDS-31 S user manual

MR MSENSE DGA 9 Installation and operating instructions

MR

MR MSENSE DGA 9 Installation and operating instructions

Agilent Technologies 490 Micro GC user manual

Agilent Technologies

Agilent Technologies 490 Micro GC user manual

YSI ProDSS Calibration and Routine Maintenance

YSI

YSI ProDSS Calibration and Routine Maintenance

Reichert DENSITY4 user guide

Reichert

Reichert DENSITY4 user guide

Triplett CM1000 user manual

Triplett

Triplett CM1000 user manual

SatLab Freyja GNSS RTK user manual

SatLab

SatLab Freyja GNSS RTK user manual

Sonotec SONAPHONE RD operating manual

Sonotec

Sonotec SONAPHONE RD operating manual

Tintometer Lovibond MD100 instruction manual

Tintometer

Tintometer Lovibond MD100 instruction manual

Apera Instruments TN500 instruction manual

Apera Instruments

Apera Instruments TN500 instruction manual

Huvitz CLM-7000 user manual

Huvitz

Huvitz CLM-7000 user manual

SMERI S-VTX instruction manual

SMERI

SMERI S-VTX instruction manual

HP 400E Operating and service manual

HP

HP 400E Operating and service manual

CUBETAPE PRO quick start guide

CUBETAPE

CUBETAPE PRO quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.