Protected by patents: US 4944310, US 5199424, US 5522382, US 6705315. Patents and designs pending.
Tango, Mirage, Mirage Activa, MirageVista, Ultra Mirage, Mirage Swift, and Meridian are trademarks of ResMed Ltd. Mirage, Mirage Activa, Mirage
Vista, Ultra Mirage, Mirage Swift and Meridian are registered in US Patent andTrademark Office. Papillon and Silent Papillon are trademarks of MAP
Medizin-Technologie GmbH.
©2006 ResMed Ltd. 34807/1 06 06
ResMed Ltd (Manufacturer) 1 Elizabeth MacArthur Drive BellaVista NSW 2153 Australia • ResMed Corp (US Designated Agent) 14040 Danielson
Street Poway CA 92064-6857 USA • ResMed (UK) Ltd (EU Authorized Representative) 65 Milton Park Abingdon Oxfordshire OX14 4RX UK •
ResMed Offices Australia, Austria, Brazil, Finland, France, Germany, Hong Kong, Japan, Malaysia, Netherlands, New Zealand, Singapore, Spain,
Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).
5. The Tango Interface
6. Setting up the Tango
Standby mode When the Tango is powered, the power indicator is lit and the run hours are
displayed on the LCD. All changes to altitude and ramp settings are made in
standby mode.
Ramp Setting a ramp time enables therapy to start at a low pressure and build up
to full pressure during the ramp time (see “Set ramp time”).
Altitude Changes in altitude affect the pressure delivered by the Tango. The altitude
feature compensates for pressure changes related to changes in altitude
(see “Set altitude”).
Run hours Run hours are displayed when theTango is in standby mode.
Humidifier The humidifer buttons are used when using the C-Series Heated Humidifier
and when setting the altitude. The heating indicator lights when the
humidifier is on.
Power indicator
Heating
indicator
CAUTION
•Be careful to place the device where it cannot be bumped, and where no one will trip over
the power cord.
•If you put the device on the floor, make sure the area is free from dust and clear of
bedding, clothes or other objects that could block the air inlet.
• Make sure the area around the device is dry and clean.
WARNING
•Make sure the power cord and plug are in good condition and the equipment is not
damaged.
•Only ResMed air tubing should be used with the device. A different type of air tubing may
alter the pressure you actually receive and reduce the effectiveness of your treatment.
•Blocking the hose and/or air inlet of the device while in operation could lead to
overheating of the device.
7. Using theTango
1 Connect power cord to the rear of theTango.
2 Plug the other end into a power outlet.
3 Connect air tubing to air outlet of Tango.
4 Connect assembled mask system to free end of air tubing.
Start therapy Press when the Tango is in standby mode. A circulating dash symbol
displays on the LCD.
If ramp is set, the ramp time displays on the LCD. When ramping is
finished, therapy continues at the prescribed pressure.
If no ramp is set, therapy starts immediately at the prescribed pressure.
Stop therapy Press .
Set ramp time 1. Repeatedly press until the LCD displays the desired time.
2. Press to accept changes and start therapy. (The Tango returns to
standby mode after 5 seconds if no button is pressed.)
The ramp settings are (in minutes):
Off, 10, 20, 45.
Set altitude 1. Hold down for 3 seconds until the current altitude setting is
displayed.
2. Press or to change the altitude setting.
3. Press to accept changes and return to standby mode. (The Tango
returns to standby mode after 10 seconds if no button is pressed.)
The altitude settings are:
1 (0–2000’/0–609 m),
2 (2001–4000’/610–1219 m),
3 (4001–6000’/ 1220–1828 m),
4 (6001–8500’/ 1829–2591 m).
8. Cleaning and Maintenance
9. Changing the Air Filter
1. With the power cord and air tubing disconnected, turn the Tango upside down so the air inlet is
facing you.The air filter is inside the air inlet.
2. Pull the old air filter out of the air inlet.
3. Put the new air filter into the air inlet in the direction shown by the arrow, with the blue side
facing towards you.
WARNING
Beware of electric shock. Do not immerse the flow generator or power cord in water. Always
unplug the flow generator before cleaning and make sure it is dry before reconnecting.
CAUTION
• Do not hang the air tubing in direct sunlight as it may harden over time and eventually
crack.
• Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturizing or
antibacterial soaps or scented oils to clean the air tubing or theTango.These solutions
may cause hardening and reduce the life of the product.
Daily
Disconnect the air tubing and hang it in a clean, dry place until next use.
Weekly
1. Remove the air tubing and wash it in warm water and mild detergent. Rinse thoroughly, hang
and allow to dry.
2. Before next use, reassemble the mask and headgear.
3. Reconnect the air tubing.
Monthly
• Clean the exterior of the Tango with a damp cloth.
• Check the air filter for holes and blockage by dirt. Replace it every six months, or more often in a
dusty environment.
For information on cleaning your mask or humidifier, refer to the manual supplied with your mask or
humidifier.
Air inlet
Air filter
CAUTION
Do not wash the air filter.
10. Troubleshooting
11.Technical Specifications
12. SymbolsWhich Appear on theTango
Problem Possible Cause Solution
Tango won’t start Power cord not connected properly Check power cord.
Faulty power outlet Try another power outlet.
No/low air flow Air tubing not connected properly Adjust air tubing connection.
Air tubing is blocked, pinched or
punctured Unblock or free air tubing. Check air
tubing for punctures.
Treatment pressure seems either
low or too high Altitude setting incorrect Adjust altitude setting.
LCD displays an error message
and treatment stops or can’t be
started
Problem with motor Contact your supplier.
Operating pressure range 4 to 20 cm H2O
Dimensions (L x W x H) 8.5” x 7.4” x 4.5” (215 mm x 189 mm x 114 mm)
Weight 2.4 lb (1.1 kg)
Power Supply Input range 110V, 60Hz, 20VA/10W (typical power consumption), 36VA/15W
(maximum power consumption)
Operating Temperature +41°F to +104°F (+5°C to +40°C)
Operating Humidity 10%–95% non-condensing
Storage and Transport Temperature -4°F to +140°F (-20°C to +60°C)
Storage and Transport Humidity 10%–95% non-condensing
Atmospheric Pressure Range 1060 hPa (sea level) to 680 hPa (8500’/ 2591 m)
Electromagnetic
Compatibility Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments
Air Filter Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing Flexible plastic, 6’6” (2 m)
Air Outlet The 0.87” (22 mm) conical air outlet complies with EN 1281-1
IEC 60601-1 Classifications Class II (double insulation), Type CF
The manufacturer reserves the right to change these specifications without notice.
Attention. Consult
accompanying documents
Type CF equipment Class II equipment
Canadian Standards Association
Drip proof
* Ports may be incorporated into the mask or in connectors that are near the mask.
A warning alerts you to possible injury.
• Read the entire guide before using theTango.
• The advice in this guide should not supersede instructions given by the prescribing physician.
• The Tango should only be used with masks (and connectors*) recommended by ResMed, or by a
physician or respiratory therapist. A mask should not be used unless the Tango is turned on and
operating properly.The vent hole or holes associated with the mask should never be blocked.
Explanation: The Tango is intended for use with special masks (or connectors*) with vent holes
allowing a continuous flow of air out of the mask.
• When the device is turned on and functioning properly, new air from the device flushes the
exhaled air out through the mask vent holes. If the device is not operating, insufficient fresh air is
provided through the mask, and the exhaled air may be rebreathed. Rebreathing exhaled air for
longer than several minutes can sometimes lead to suffocation. If there is a power failure or the
device machine malfunctions, remove the mask.
• The Tango can be set to deliver pressures up to 20 cm H2O. In the unlikely event of certain fault
conditions, pressures up to 30 cm H2O are possible.
• If oxygen is used with this device, the oxygen flow must be turned off when the device is not
operating. If oxygen has been left on while the device is not operating, disconnect the oxygen
and wait 30 minutes before turning on the flow generator again.
Explanation: When the CPAP device is not in operation and the oxygen flow is left on, oxygen
delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and
create a risk of fire.
• Oxygen supports combustion. Do not use oxygen while you are smoking or in the presence of an
open flame.
• Always ensure air is flowing from the device before turning on the oxygen supply.
• Always turn the oxygen supply off before stopping the airflow from the device.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will vary,
depending on where the oxygen is introduced, pressure settings, patient breathing pattern, mask
selection and leak rate.
• Do not use the Tango if there are obvious external defects or unexplained changes in
performance.
• Do not open the Tango case.There are no user serviceable parts inside. Repairs and internal
servicing should only be performed by an authorized service agent.
A caution explains special measures for safe and effective use of the device.
• At low pressures, the flow through the exhalation ports of your mask may not clear all exhaled
gas from the tubing. Some rebreathing may occur.
• The airflow for breathing produced by this device can be as much as 11°F (6°C) higher than the
temperature of the room. Be careful if the room temperature is warmer than 90°F (32°C).
13. General Warnings and Cautions
14. Servicing
The Tango should be inspected by an authorized ResMed service center 5 years from the date of
manufacture. Before this, the device is intended to provide safe and reliable operation if it is
operated and maintained according to the instructions provided by ResMed. Warranty details are
provided with the device at the time of original supply. As with all electrical devices, if any
irregularity becomes apparent, you should have the device inspected by an authorized ResMed
service center.
15. Limited Warranty
Note: Some models are not available in all regions.
If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective
product or any of its components. This Limited Warranty does not cover:
a) any damage caused as a result of improper use, abuse, modification or alteration of the product;
b) repairs carried out by any service organization that has not been expressly authorized by ResMed to perform
such repairs;
c) any damage or contamination due to cigarette, pipe, cigar or other smoke;
d) any damage caused by water being spilled on or into a flow generator.
Warranty is void on product sold, or resold, outside the region of original purchase. Warranty claims on defective
product must be made by the initial consumer at the point of purchase.
This warranty is in lieu of all other express or implied warranties, including any implied warranty of
merchantability or fitness for a particular purpose. Some regions or states do not allow limitations on how long
an implied warranty lasts, so the above limitation may not apply to you.
ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a
result of the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion
or limitation of incidental or consequential damages, so the above limitation may not apply to you. This warranty
gives you specific legal rights, and you may also have other rights which vary from region to region.
For further information on your warranty rights, contact your local ResMed dealer or ResMed office.
ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the
period specified below from the date of purchase by the initial consumer.This warranty is not transferable.
Product Warranty Period
ResMed humidifiers, ResControl™, ResLink™, ResTraxx™ 1 Year
ResMed flow generators 2 Years
Accessories, water chambers and mask systems (including mask frame, cushion, headgear and tubing).
Excludes single-use devices. 90 Days
Global leaders in sleep and respiratory medicine www.resmed.com