ResMed AirFit N30 User manual

English 1
Disinfection guide Clinical use only
This guide is intended for multi-patient use of the AirFit™N30 nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning instructions. The
processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.
Mask component1
Cleaning
Validated number of cycles
Mild Alkaline, Anionic Detergent e.g. Alconox Neodisher Mediclean Forte
Thermal Disinfection
Manual Washer-Disinfector
• Cushion 30
• Standard frame 30
• QuietAir™ frame2– – –
• Split-band headgear 30
1May not be available in all regions. For full details regarding the correct use of this mask, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com
2Not intended for multi-patient re-use.
A Split-band headgear
1 Buckle
2 Buttonhole
3 End cap
B Cushion
C QuietAir™ / Standard frame
4 Stabiliser
5 QuietAir vent
6 Vent
7 Mask tubing
English 2
Thermal disinfection procedure
1. Disassembly Disassemble the mask according to the instructions in the User Guide.
2. Cleaning
1. Make a solution of a mild alkaline, anionic detergent and water1as per manufacturer’s instructions. ResMed has validated the following:
• Alconox 1% concentration (10 g/L) 30-55°C
2. Soak all mask components for 5 to 10 minutes. Ensure there are no air bubbles.
3. After the soaking time, while keeping the components in the solution, clean the inside and outside with a soft bristle brush as follows. Pay particular attention to the crevices and cavities.
• Cushion / frame / headgear 1 minute brushing
4. Rinse the components as follows under running water1(≤60°C) directly contacting all surfaces.
• Cushion / frame 30-60 seconds rinsing
• Headgear 1 minute rinsing, while squeezing repeatedly
3. Inspection and drying
1. Inspect and if required, repeat all steps in section [2. Cleaning] until visually clean.2
2. Squeeze the headgear to remove excess water.
3. Allow the components to air dry out of direct sunlight.3
4. Disinfection
1. Soak the mask components in a hot water1bath using one of the following temperature-time combinations:
• 75°C for 30 minutes
• 80°C for 10 minutes
• 90-93°C for 1 to 10 minutes.
2. Squeeze the headgear to remove excess water.
3. Allow the components to air dry out of direct sunlight.
5. Inspection Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears etc), the mask component should be discarded and replaced. Slight discolouration
of the silicone components may occur and is acceptable.
6. Reassembly Reassemble the mask according to the instructions in the User Guide.
7. Storage and transportation For storage and transportation details, check the User Guide.
1For all cleaning, rinsing and disinfection steps, use drinking quality water. Where local regulation impose water quality requirements for medical device reprocessing, these requirements must be met.
2Failure to clean the mask component as indicated may result in inadequate disinfection.
3Drying is not required if thermal disinfection is continued immediately.
English 3
Thermal disinfection procedures - Washer-Disinfector
1. Disassembly Disassemble the mask according to the instructions in the User Guide.
2. Washer-Disinfector1
1. Soak the mask components in a 5 litre rinse bath of cold tap water1for 1 minute. Ensure there are no air bubbles on the components.
2. Rinse each component under cold running tap water1for 30 seconds. Ensure internal surfaces are flushed.
3. Transfer the components into an ISO 15883 series compliant washer-disinfector. For the standard frame component, place on a spray nozzle3to facilitate water flow through the mask tubing. Components should be
orientated to ensure effective drainage.
4. Set the washer-disinfector using the following settings:
Phase/Stage Recirculation time (minutes) Temperature (°C) Detergent
Pre-wash 4 Cold –
Wash 10 43-60 2-10 mL/L (0.2-1%) Neodisher Mediclean Forte
Neutralisation 6 43-60 1-2 mL/L (0.1-0.2%) Neodisher Z
Final Rinse 3 43-60 –
Thermal Disinfection 5-10 90-93 –
3. Inspection and drying
1. Inspect and if required, repeat all steps in section [2. Washer-Disinfector] until visually clean2.
2. Squeeze the headgear to remove excess water.
3. Allow the components to air dry out of direct sunlight.
4. Inspection Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears etc), the mask component should be discarded and replaced. Slight discolouration
of the silicone components may occur and is acceptable.
5. Reassembly Reassemble the mask according to the instructions in the User Guide.
6. Storage and transportation For storage and transportation details, check the User Guide.
1For all cleaning, rinsing and disinfection steps, use drinking quality water. Where local regulations impose water quality requirements for medical device reprocessing, these requirements shall be met.
2Failure to clean the mask component as indicated may result in inadequate disinfection.
3Use a washer-disinfector load carrier/rack containing spray/injector nozzles.
GENERAL WARNINGS AND CAUTIONS
• Always follow cleaning and disinfection instructions. Some cleaning products may damage the mask, its parts and their function, or leave harmful residual vapours. Any deviations from the
procedures listed in this guide, including exceeding the specified number of maximum cycles, can have an adverse effect on the mask and consequently the safety or the quality of therapy.
• Mask components should not be subjected to autoclave or ethylene-oxide gas sterilisation.
• When using detergents, disinfectants or equipment, always follow the manufacturer’s instructions. In the event of conflict, this guide takes precedence.
• Always follow safe operating practices, including the use of appropriate personal protective equipment (PPE), as required. Refer to the manufacturer’s safety instructions for more details.
ResMed Pty Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide. AirFit is a trademark and/or registered trademark of the ResMed family of companies. Alconox is a trademark of Alconox Inc. Neodisher is a trademark of
Chemische Fabric Dr Weigert. For patent and other intellectual property information, see ResMed.com/ip. © 2020 ResMed. 608575/3 2020-01
ResMed.com

Other manuals for AirFit N30

6

Other ResMed Personal Care Product manuals

ResMed Swift LT User manual

ResMed

ResMed Swift LT User manual

ResMed C-Series Tango User manual

ResMed

ResMed C-Series Tango User manual

ResMed AIRCurve 10 ST User manual

ResMed

ResMed AIRCurve 10 ST User manual

ResMed Swift LT-F User manual

ResMed

ResMed Swift LT-F User manual

ResMed ApneaLink User manual

ResMed

ResMed ApneaLink User manual

ResMed Quattro FX User manual

ResMed

ResMed Quattro FX User manual

ResMed S9 V-Auto Technical manual

ResMed

ResMed S9 V-Auto Technical manual

ResMed Swift FX Bella User manual

ResMed

ResMed Swift FX Bella User manual

ResMed Narval CC User manual

ResMed

ResMed Narval CC User manual

ResMed AirCurve10 ASV User manual

ResMed

ResMed AirCurve10 ASV User manual

ResMed AcuCare Series User manual

ResMed

ResMed AcuCare Series User manual

ResMed AIRMINI Installation guide

ResMed

ResMed AIRMINI Installation guide

ResMed AirFit N20 User manual

ResMed

ResMed AirFit N20 User manual

ResMed S9 AutoSet CS Technical manual

ResMed

ResMed S9 AutoSet CS Technical manual

ResMed AIRCurve 10 Vauto User manual

ResMed

ResMed AIRCurve 10 Vauto User manual

ResMed Swift LT-F User manual

ResMed

ResMed Swift LT-F User manual

ResMed Power Station II User manual

ResMed

ResMed Power Station II User manual

ResMed ResLink User manual

ResMed

ResMed ResLink User manual

ResMed S9 Series Operating and safety instructions

ResMed

ResMed S9 Series Operating and safety instructions

ResMed S9 Series User manual

ResMed

ResMed S9 Series User manual

ResMed ApneaLink Plus User manual

ResMed

ResMed ApneaLink Plus User manual

ResMed RCM User manual

ResMed

ResMed RCM User manual

ResMed Mirage SoftGel User manual

ResMed

ResMed Mirage SoftGel User manual

ResMed S9 Series Safety guide

ResMed

ResMed S9 Series Safety guide

Popular Personal Care Product manuals by other brands

HairMax LaserComb Ultima 9 user manual

HairMax

HairMax LaserComb Ultima 9 user manual

BresoTEC URU user guide

BresoTEC

BresoTEC URU user guide

Carmen FC1800 manual

Carmen

Carmen FC1800 manual

Nuface trinity user manual

Nuface

Nuface trinity user manual

Andis NT-2 instructions

Andis

Andis NT-2 instructions

SOLAC CT8665 Instructions for use

SOLAC

SOLAC CT8665 Instructions for use

Darco APB Instructions for use

Darco

Darco APB Instructions for use

Cosmopolitan COSPA09 instruction manual

Cosmopolitan

Cosmopolitan COSPA09 instruction manual

RAEWHA SC Laser MX-830 manual

RAEWHA

RAEWHA SC Laser MX-830 manual

Orliman actius ACE807 quick start guide

Orliman

Orliman actius ACE807 quick start guide

Vitek VT-2231 VT Manual instruction

Vitek

Vitek VT-2231 VT Manual instruction

Innova PB570 owner's manual

Innova

Innova PB570 owner's manual

BaByliss SLIM ST326E user manual

BaByliss

BaByliss SLIM ST326E user manual

Clarisonic MIA 2 user guide

Clarisonic

Clarisonic MIA 2 user guide

Juzo Compression segments Instructions for use

Juzo

Juzo Compression segments Instructions for use

Handicare OneWaySlide quick start guide

Handicare

Handicare OneWaySlide quick start guide

Adler AD 2178 user manual

Adler

Adler AD 2178 user manual

Aura Allure Instruction manual and users guide

Aura

Aura Allure Instruction manual and users guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.