
Maintenance and Troubleshooting
© Rice Lake Weighing Systems ● All Rights Reserved 15
5.0 Maintenance and Troubleshooting
5.1 Maintenance
The following section provides instructions for maintaining and cleaning the Rice Lake line of scales.
Do not immerse the scale in cleaning or other liquid solutions.
Do not use Isopropyl alcohol or other solutions to clean the indicator display surface.
5.1.1 Basic Maintenance
Before using the scale and after periods of non-use, check the scale for proper operation and function. If the scale does not
operate correctly, contact a qualified service personnel.
Basic maintenance includes:
• Check the overall appearance of the entire scale for any obvious signs of damage.
• Inspect the condition of the AC adapter cord for cracking, fraying or broken/bent prongs.
• Using an adapter that has been used elsewhere may cause issues with the scale.
• Change batteries on a regular interval. Leaving in for long periods of time can cause batteries to leak.
5.1.2 Cleaning
Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Before beginning the cleaning
process, disconnect the scale from the AC power source.
• Clean all external surfaces with a clean, damp cloth or tissue.
• Mild soap and water solution may be used.
• Standing platform may be cleaned with Isopropyl alcohol
• Dry with a clean soft cloth.
5.2 Troubleshooting and Testing
Refer to Table 5-1 to diagnose and correct failure prior to contacting service personnel.
Symptom Possible Cause Corrective Action
Scale does not turn on Dead battery Connect the scale to a power source
Faulty electrical outlet Use a different electrical outlet
Bad power supply Replace adapter
Questionable weight or
the scale does not zero
External object is interfering with scale Remove interfering object from scale
0.0 was not displayed prior weighing Have the patient off scale, zero scale and begin weighing process again
Scale is not level Level scale and begin weighing process again
Scale is out of calibration Check weight with a known weight value
E messages displayed
E06 Identifier - ADC AD too high
E07 AD too low
E10 Overload Scale is overloaded; remove load from scale
E4L BAT Battery low but usable – 1 bar on indicator display.
E4U Battery low and unstable – no bars on indicator display.
E11 CAL Calibration Error – recalibrate the scale
Err 2 Low saturation state (low A/D) Load cell is not connected properly; check cables and mechanical connections;
if the problem persists, replace load cells.
Err 3 High saturation state (high A/D) See Err 2
Err 6 Unstable weight; cannot calibrate. Check the load cells mechanical surroundings to ensure nothing is touching them
and the cables are properly welded
SAT Damaged load cell cable Replace load cell cable
Table 5-1. Troubleshooting Table for the Rice Lake Scale Line