English
5
8Caution: Portable HF telecommunication devices (radio equipment,
mobile telephones) (including their accessories such as e.g. antenna cables
and external antennas) should not be used within a distance less than 30cm
(12 inches) to the UROstim 2 (including its accessories). Non-compliance
may lead to inferior device performance or faulty operation
9Caution: The use of this device directly next to other devices or stacked on
top of them should be avoided because it may result in malfunction. If the
use as previously described is nevertheless necessary, this device and the
other devices should be monitored to establish that they are working
properly.
10 Applying the electrodes:
a. The product should only be connected to one patient at a time.
b. Before applying the electrodes, clean the skin surface where the electrodes
will be applied. Otherwise, faulty operation cannot be ruled out.
c. Make sure that no metallic objects such as jewellery or piercings come
into contact with the electrodes during stimulation, as this may result
in localised burns.
d. Tattoo inks can contain metallic pigments which, under the influence
of current, can in rare cases cause high current densities and skin
damage. -> Wherever possible, stimulation in areas of the body with
tattoos should be avoided. If it is not possible, stimulation in these
areas of the body should be observed with closer attention and ended
immediately in the event of an emergency.
e. Current densities above 2 mA/cm2on all electrode surfaces may require
increased operator attention or painful eects may occur.
f. Caution: Place the electrodes on the skin so that the electrode surface
makes full and even contact with the skin. Make sure additionally that
the distance between the electrodes is at least 2 cm. Otherwise, too
high current densities can occur and thus painful skin lesions can occur
on the skin.
g. In patients with metal implants who have sensory disturbances in the
area of the metal, special caution should be taken when stimulating
and placing electrodes in this area. The sensation disturbance may
tempt users to increase the setting of the stimulation intensity and lead
to skin irritation with reddening of the skin or pain in the area of the
metal. In this case, it is necessary to stop the stimulation.
11 Do not drop the product or handle it improperly.
Apply only at temperatures of 10°C - 40°C, a relative air humidity of 30%-
75% and air pressure between 70 kPa and 106 kPa). Therefore do not use
the product e.g. in the bathroom or in a similarly humid environment.
Warning: Do not operate the device in the vicinity of potentially
explosive and/or flammable substances or fumes!