Smiths Medical CADD-Legacy PLUS Reference guide

ADisplay shows programmed values
and messages
BAC Adapter Jack used to plug in AC Adapter
CAccessory Jack (not currently in use)
DIndicator Light indicates AC power in use
EAir Detector
FCassette (part of reservoir or administration set
that attaches to pump)
GThreaded Mounting Hole for use with Polemount
Bracket Adapter (back of pump)
HBattery Compartment (back of pump)
ICassette Latch attaches cassette to pump (side of pump)
KEYPAD
STOP/START Stops and starts the infusion
ENTER/CLEAR Enters or clears displayed value
PRIME Fills tubing with fluid
LOCK Displays or changes Lock Level (security level)
NEXT Advances to next programming screen
Increases or decreases displayed values
or scrolls through menu items
ON/OFF Turns the pump on or off (low power)
´
Î
Warning: This Quick Reference Card should be used by clinicians
only. Do not permit patients to have access to this card, as the
information would allow access to all programming and operating
functions. Assistance with the CADD-Legacy®PLUS pump is
available to clinicians 24-hours-a-day by calling 1-800-258-5361.
Read the entire Operator’s Manual before operating the CADD-Legacy®
ambulatory infusion pump. Failure to properly follow warnings, cautions
and instructions could result in death or serious injury to the patient.
ALARMS & TROUBLESHOOTING CONTINUED
Rate is in
ml/hr
Upstream
Occlusion
Two-tone alarm
Air In Line
Detected
Two-tone alarm
Stopped
Three beeps
every five minutes
Programming
Incomplete
Value Not Saved
Error
Two-tone alarm
No Message
Displayed
Two-tone alarm
No Disposable,
Clamp Tubing
Two-beep
(long-short) alarm
Service Due
Two-tone alarm
Screen Displays
Current Pump
Status
Two-beep
(long-short) alarm
Fluid is not flowing from the reservoir to the
pump. Check for a kink or air bubble in the
tubing between the reservoir and pump. Remove
the occlusion to resume operation or press
STOP/START or NEXT to stop the pump and
silence alarm for two minutes, then remove the
occlusion and restart the pump.
The cassette is not aligned with the pump,
or is damaged, or a malfunction of the pump
sensor(s) is occuring. Reposition the pump to
silence the alarm.If repositioning the pump
does not silence the alarm within two minutes, the
pump will display, “No Disposable, Clamp Tubing.”
Air in fluid path. Press NEXT or STOP/START to
silence alarm. Make sure tubing is threaded in
air detector properly. If the fluid path contains
air, close clamps, disconnect from patient and
prime fluid path.
Pump is stopped. Start pump, if appropriate.
Verify all programming screens by pressing
ENTER/CLEAR before moving to next screen or
starting pump.
A value was not saved by pressing ENTER/
CLEAR. Press NEXT to resume programming.
An error has occurred. Remove pump
from service.
The pump was running when batteries were
removed or, the batteries were removed within
15 seconds after stopping the pump. Batteries
must be reinstalled.
IMPORTANT: Always stop pump before
removing batteries.
The disposable was removed, or the cassette is
not aligned with the pump or is damaged, or a
malfunction of the pump sensor(s) occurring.
Clamp the tubing immediately. Press STOP/
START or NEXT to silence alarm.
The pump is due for service.
CADD, CADD-Legacy, the Medication Cassette Reservoir design and the Smiths Medical
and CADD design marks are trademarks of Smiths Medical. The symbol ® indicates
the trademark is registered in the U.S. Patent and Trademark Office and certain other
countries. ©2010 Smiths Medical. All rights reserved. 10/10 IN19893
Smiths Medical ASD, Inc.
St. Paul, MN 55112, USA
Phone: 1-214-618-0218
Toll-Free USA: 1-800-258-5361
www.smiths-medical.com
Smiths Medical is part of the global technology business Smiths Group plc.
EC Authorized Representative
Smiths Medical International Ltd.
TN25 4BF, UK
Phone: +44 (0) 1233 722100
Product(s) described may not be licensed or available for sale in Canada or other
countries outside of the United States.
Smiths Medical Canada Ltd.
Markham, Ontario, Canada, L3R 4Y8
Phone: 905-477-2000
Toll-Free: 1-800-387-4346
CADD-Legacy®PLUS
Infusion Pump
Model 6500
Quick Reference Card for Clinicians
Lock Level 1 (LL1)
®
®
A
B
C
D
E
F
G
H
I
CHANGING THE CASSETTE
To Remove
Used
Cassette
To Attach
New
Cassette
ALARMS & TROUBLESHOOTING
The CADD-Legacy®PLUS pump signals with audible alarms and
messages that appear on the display when a condition requires
your attention.
Level of fluid in the reservoir is low. Prepare to
install new reservoir.
Reservoir volume has reached 0.0 ml. Press
NEXT or STOP/START to silence alarm. Install
a new reservoir, if appropriate.
Batteries are low, but pump is operable. Change
batteries soon.
Batteries are depleted and cannot operate pump;
install new batteries immediately.
With AC adapter attached, batteries have been
removed while pump was running, or you have
tried to start pump with depleted batteries.
Install new batteries.
High pressure is being caused by an
occlusion in the fluid path between the pump
and the patient. Remove the occlusion to resume
operation. Or press STOP/START or NEXT to stop
the pump and silence the alarm for two minutes,
then remove occlusion and restart pump.
Res Vol Low
Three beeps
Reservoir
Volume Empty
Two-tone alarm
Low Bat
3 two-tone beeps
every five minutes
Battery Depleted
Two-tone alarm
Battery Removed
Pump Won’t Run
Two-tone alarm
High Pressure
Two-tone alarm
1. Stop the pump.
2. Close all tubing clamps.
3. Disconnect tubing from patient.
4. Use coin to unlatch used cassette.
5. Remove and discard used cassette.
1. Close all tubing clamps.
2. Insert new cassette hooks into pump’s hinge pins.
3. Place pump with cassette on firm surface and
push down on top of pump so cassette fits tightly
against the pump.
4. Insert a coin into the lock, push in and turn until
the line on the lock lines up with the arrow on the
side of the pump.
5. Gently twist and pull on cassette to make sure it
is attached properly.
6. Press ENTER/CLEAR to reset reservoir volume, if
appropriate.
7. If the Air Detector is in use, make a small loop
of tubing underneath the air detector and hold
it with your thumb. Place the tubing over the
groove in the air detector and tuck it under the
catch. Seat the tubing into the groove by gently
pulling the tube upward until it is under the
retention nubs and flat in the groove.
WARNING: Attach the cassette (the part of
the Medication Cassette reservoir or CADD®
administration set that attaches to the pump)
properly. An improperly attached or detached
cassette could result in unregulated gravity infusion
of medication from the fluid container or a reflux of
blood. If you are using a CADD®administration set or
Medication Cassette reservoir that does not have the
flow stop feature (catalog number does not start with
21-73xx): You must use a CADD®extension set with an
integral anti-siphon valve or a CADD®administration
set with either and integral or add on anti-siphon
valve to protect against unregulated gravity infusion
that can result from an improperly attached cassette.
The CADD-Legacy®PLUS pump signals with audible alarms and
messages that appear on the display when a condition requires
your attention.
IN19893.indd 1 26-10-2011 15:36:20
PROGRAMMING / SET-UP SEQUENCE
CONTINUOUS DELIVERY
Batteries must be installed; pump must be stopped and in LL1.
NOTE: Value Not Saved is displayed if a value is scrolled and ENTER/
CLEAR is not pressed. Press NEXT to continue programming.
WARNING: If the pump is dropped or hit, inspect it for damage. Do not
use a pump that is damaged or is not functioning properly.
View Dose Cycle
PUMP OPERATIONS
1. Stop the pump.
2. Press LOCK.
3. View lock level.
4. Press LOCK or ENTER/CLEAR two more times.
1. Press and hold STOP/START until
(– – – – – – – – – – – –) appears on the display.
2. Release STOP/START key. STOPPED will appear
on the display when the pump is stopped.
1. Press and hold STOP/START until
(– – – – – – – – – – – –) disappears from the display.
2. Release STOP/START key. RUN will appear on the
display when the pump is running.
Pump must be stopped and in LL0 or LL1.
Warning: Do not prime the fluid path with the
tubing connected to a patient as this could result in
over-delivery of medication or air embolism.
1. Press and hold PRIME until the word PRIME
appears on the display, along with (––––––––––––).
2. Release the PRIME key.
3. Press and hold PRIME. Priming appears on the
display. Continue priming until the fluid path is
free of air.
4. Press NEXT to return to the main screen.
1. Stop the pump.
2. Press NEXT to display the reservoir volume screen.
3. Press ENTER/CLEAR to reset the value to previously
programmed amount.
1. Press and hold ON/OFF until pump beeps
and powers up.
1.PressandholdON/OFFuntil(••••••••••••)
appears on the display.
2. Release ON/OFF key.
3. The screen will go blank as the pump goes into
a lower power state.
1. Stop the pump.
2. Push down and hold the arrow button on the
battery door while sliding the door off.
Remove and discard old batteries.
IMPORTANT: Always stop pump before
removing batteries.
3. Install new batteries, matching polarities shown
on pump. Replace battery door and close.
4. Start the pump.
Warning: If a gap is present anywhere between the
battery door and the pump housing, the door is not
properly latched. If the battery door becomes
detached or loose, the batteries will not be properly
secured which could result in loss of power or
non-delivery of medication.
Warning: Do not use rechargeable NiCad or nickel
metal hydride (NiMH) batteries. Do not use carbon
zinc (“heavy duty”) batteries.
View Reservoir
Volume
Enter Continuous
Rate
Clear Milliliters
Given
Verify Air
Detector Status
Verify Upstream
Sensor Status
Verify
Programming
Main Screen
View Reservoir
Volume
View Dose Volume
View Dose Duration
Main Screen
View KVO Rate
Verify Dose
Remaining
View the
Lock
Level
Stop the
Pump
Start the
Pump
Press NEXT.
1. Press ENTER/CLEAR to reset.
2. Press NEXT.
3. Press´orÎto select desired Continuous
Rate in ml/hr.
4. Press ENTER/CLEAR.
5. Press NEXT.
Warning: This pump delivers medication at a
continuous rate or intermittent rate in milliliters
per hour (ml/hr). Programming the pump at a
delivery rate other than what is prescribed will
cause over- or under-delivery of medication.
6. Press ENTER/CLEAR to clear the given
value. The display will show 0.00.
7. Press NEXT.
8. Verify the setting is correct.
9. Press NEXT.
10.Verify the setting is correct.
11.Press NEXT.
12.Press NEXT repeatedly to review program.
INTERMITTENT DELIVERY
Batteries must be installed; pump must be stopped and in LL1.
Press NEXT.
1. Press ENTER/CLEAR to reset.
2. Press NEXT.
3. Press NEXT.
4. Press NEXT.
5. Press NEXT.
6. Press NEXT.
Note: Appears only during dosing or if dose is
interrupted.
7. Press NEXT.
8. Press´orÎto select amount of time
before next dose. Select IMMEDIATE if
dose should begin when pump is started.
9. Press ENTER/CLEAR.
10. Press NEXT.
11. Press ENTER/CLEAR to clear the given
value. The display will show 0.00.
12. Press NEXT.
13. Verify the setting is correct.
14. Press NEXT.
15. Verify the setting is correct.
16. Press NEXT.
17. Press NEXT repeatedly to review program.
Enter Dose
Starts In
Clear Milliliters
Given
Prime the
Fluid Path
Reset the
Reservoir
Volume
Turn the
Pump On
Turn the
Pump Off
Change the
Batteries
Verify Air Detector
Status
Verify Upstream
Sensor Status
Verify
Programming
PROGRAMMING SCREENS
Main Screen Displays status of pump
(STOPPED or RUN)
CONTINUOUS MODE
Reservoir Volume Volume of fluid in reservoir
Continuous Rate Continuous rate of infusion (ml/hr)
Given Total amount of medication delivered since
display was last cleared
Air Detector Indicates if air detector is turned on or off
(status only) and high or low sensitivity level
Upstream Sensor Indicates if turned on or off
(status only)
INTERMITTENT MODE
Reservoir Volume Volume of fluid in reservoir
Dose Volume Volume of an individual dose
Dose Duration Amount of time over which an individual
dose is infused
Dose Cycle Time from start of one dose to start
of next dose
KVO Rate Keep Vein Open rate in ml/hr delivered
between doses
Dose Remaining Time remaining for delivery of current
dose (only appears during dosing or
if dose is interrupted)
Dose Starts In Amount of time before start of next dose
Given Total amount of medication delivered since
display was last cleared
Air Detector Indicates if air detector is turned on or
(status only) off and high or low sensitivity level
Upstream Sensor Indicates if turned on or off
(status only)
IN19893.indd 2 26-10-2011 15:36:21

This manual suits for next models

1

Other Smiths Medical Medical Equipment manuals

Smiths Medical CADD-Legacy Reference guide

Smiths Medical

Smiths Medical CADD-Legacy Reference guide

Smiths Medical PORTEX acapella choice User manual

Smiths Medical

Smiths Medical PORTEX acapella choice User manual

Smiths Medical BCI SPECTRO2 30 User manual

Smiths Medical

Smiths Medical BCI SPECTRO2 30 User manual

Smiths Medical Portex Blue Line Ultra User manual

Smiths Medical

Smiths Medical Portex Blue Line Ultra User manual

Smiths Medical Surgivet Advisor User manual

Smiths Medical

Smiths Medical Surgivet Advisor User manual

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical acapella portex User manual

Smiths Medical

Smiths Medical acapella portex User manual

Smiths Medical Graseby 1200 User manual

Smiths Medical

Smiths Medical Graseby 1200 User manual

Smiths Medical acapella duet Vibratory PEP Therapy System User manual

Smiths Medical

Smiths Medical acapella duet Vibratory PEP Therapy System User manual

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical CADD-Legacy PLUS Pump User manual

Smiths Medical

Smiths Medical CADD-Legacy PLUS Pump User manual

Smiths Medical CADD-Solis User manual

Smiths Medical

Smiths Medical CADD-Solis User manual

Smiths Medical acapella User manual

Smiths Medical

Smiths Medical acapella User manual

Smiths Medical 1884SS User manual

Smiths Medical

Smiths Medical 1884SS User manual

Smiths Medical CADD-Solis User manual

Smiths Medical

Smiths Medical CADD-Solis User manual

Smiths Medical CDS 9000 User manual

Smiths Medical

Smiths Medical CDS 9000 User manual

Smiths Medical BCI 3304 User manual

Smiths Medical

Smiths Medical BCI 3304 User manual

Smiths Medical Pneupac babyPAC 100 User manual

Smiths Medical

Smiths Medical Pneupac babyPAC 100 User manual

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical CADD-Prizm VIP 6100 Reference guide

Smiths Medical

Smiths Medical CADD-Prizm VIP 6100 Reference guide

Smiths Medical SurgiVet AES User manual

Smiths Medical

Smiths Medical SurgiVet AES User manual

Smiths Medical Medfusion 3500 User manual

Smiths Medical

Smiths Medical Medfusion 3500 User manual

Smiths Medical Oxi-Pulse 30 WW1030 User manual

Smiths Medical

Smiths Medical Oxi-Pulse 30 WW1030 User manual

Popular Medical Equipment manuals by other brands

CU Medical Systems iPAD SP1 Operation & Maintenance Policy

CU Medical Systems

CU Medical Systems iPAD SP1 Operation & Maintenance Policy

GCE S 400 Instructions for use

GCE

GCE S 400 Instructions for use

RotoBed Care manual

RotoBed

RotoBed Care manual

ulrich medical Pezo PEEK cage Series Assembly and disassembly instructions with special cleaning instructions

ulrich medical

ulrich medical Pezo PEEK cage Series Assembly and disassembly instructions with special cleaning instructions

FIOR & GENTZ NEURO CLASSIC-SWING Instructions for use

FIOR & GENTZ

FIOR & GENTZ NEURO CLASSIC-SWING Instructions for use

Braun Aesculap MACS II Instructions for use/Technical description

Braun

Braun Aesculap MACS II Instructions for use/Technical description

Promeba STRETCHER Series user guide

Promeba

Promeba STRETCHER Series user guide

Pure Processing FlexiPump user manual

Pure Processing

Pure Processing FlexiPump user manual

Dräger Babylog 1 HF Instructions for use

Dräger

Dräger Babylog 1 HF Instructions for use

Shenzhen Mindray Bio-Medical Electronics DC-80S Operator's manual

Shenzhen Mindray Bio-Medical Electronics

Shenzhen Mindray Bio-Medical Electronics DC-80S Operator's manual

ST&G 1321 AP / AL instructions

ST&G

ST&G 1321 AP / AL instructions

Flaem Suction Pump Instructions for use manual

Flaem

Flaem Suction Pump Instructions for use manual

Gima PARKER manual

Gima

Gima PARKER manual

Masimo iSpO2 Quick reference guide

Masimo

Masimo iSpO2 Quick reference guide

VENUS CONCEPT MP2 user manual

VENUS CONCEPT

VENUS CONCEPT MP2 user manual

KaVo SURGmatic S11 L Instructions for use

KaVo

KaVo SURGmatic S11 L Instructions for use

MedRx AVANT Series Training manual

MedRx

MedRx AVANT Series Training manual

Chison SonoEye P5 user manual

Chison

Chison SonoEye P5 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.