manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Smiths Medical
  6. •
  7. Medical Equipment
  8. •
  9. Smiths Medical CADD-Prizm VIP User manual

Smiths Medical CADD-Prizm VIP User manual

AGreen indicator light: blinks when pump is running
BAmber indicator light: blinks when pump is stopped
or alarming
CPolemount bracket recess (back of pump)
DPower jack (for power pack or AC adapter)
EData in/out jack (for printer or modem cable)
FBattery compartment (side of pump)
GCassette (part of reservoir or administration set
that attaches to pump)
HAir Detector
Keypad
STOP/START Stops and starts the pump
LOCK Displays or changes lock level (security level)
? Help key: displays more information about a screen
ENTER Sets new value; selects choice from menu;
clears screen
NEXT Advances to next programming screen; backs
out of an option or menu
DOSE Inoperable in Intermittent delivery mode
OPTIONS Displays the Options Menu, pages through options
´Answers yes; increases values on
programming screen; pages through menus;
pages through event log
ÎAnswers no; decreases values on
programming screen; pages through menus;
pages through event log
Using the Å(Help) Key
1. For information about screen or alarm, press ? key.
2. If?→appears in lower right corner, press ? to see more help.
3. Press NEXT to exit help.
ALARMS & TROUBLESHOOTING CONTINUED
Air In Line
Detected
3 Beeps Every
5 Minutes;
Stopped
on Main
Screen
Dose
Scheduled
to Start In —
Minutes / Start
the Pump
Dose
Scheduled to
Start is Now
Overdue/Start
the Pump
Air in fluid path. Press NEXT to silence alarm.
Make sure tubing is threaded in Air Detector
properly. If fluid path contains air, close
clamps, disconnect from patient, and prime
fluid path.
Pump is stopped. Start pump if appropriate.
Pump is stopped, but a dose is scheduled to be
delivered in the number of minutes indicated.
In order for the dose to be delivered, you must
start the pump.
Pump is stopped, but a dose is overdue for its
scheduled delivery. In order for the dose to be
delivered, you must start the pump.
ALARMS & TROUBLESHOOTING
The CADD-Prizm pump signals with audible alarms and
messages on the display when a condition requires your
attention. In addition to referring to this table, use the ? key
while the alarm is occurring for information about the alarm and
corrective action.
Reservoir
Volume Low
Reservoir
Volume Zero
9-volt
Battery Low
9-volt Battery
Depleted
High Pressure
Upstream
Occlusion
(Model 6101
only)
Error Detected
Level of fluid in the reservoir is low. Prepare to
install a new reservoir.
Reservoir Volume is 0.0 ml. Press NEXT to stop
alarm. Install new reservoir if appropriate.
9-volt battery is low but pump is operable.
Change battery soon.
9-volt battery is depleted and cannot operate
pump; install fresh battery immediately.
Possible blockage downstream (between pump
and patient), clamp, or kink in fluid path. Find
cause and remove blockage to continue. Or,
press STOP/START to stop pump and silence
alarm; remove blockage and restart pump.
Fluid is not flowing from the reservoir to the
pump. Check for a kink, a closed clamp or an
air bubble in tubing between reservoir and
pump. Press STOP/START to silence alarm for
two minutes; remove obstruction and restart
pump.
Pump fault occurred. Close tubing clamp,
remove pump from service.
RECORDKEEPING FUNCTIONS
1. Press OPTIONS.
2. Press
´or Îuntil Event Log option
appears.
3. Press ENTER to display Event Log.
4. Press
Îto page back through Event Log
entries or ´to page forward.
5. When finished, press NEXT.
1. Stop the pump.
2. Press NEXT repeatedly until Milliliters
Given screen appears.
3. Press ENTER to clear and update date
and time.
1. Turn printer on.
2. Connect printer cable to data in/out jack
and to back of printer.
3. If necessary, press ´until
Printing appears.
4. Press ENTER.
5. Press
´or Îto select desired report.
6. Press ENTER.
7. When report is finished, detach printer
cable from pump.
To View
Event Log
To Clear
Milliliters Given
To Print Reports
• Rx Setting
• Event Log
The pump may be running or stopped to view functions or to print.
WARNING: This Quick Reference Card should be used by
clinicians only. Do not permit patients to have access to this card,
as the information would allow access to all programming and
operating functions.
Smiths Medical ASD, Inc.
St. Paul, MN 55112, USA
Phone: 1-214-618-0218
Toll-Free USA: 1-800-258-5361
www.smiths-medical.com
Smiths Medical is part of the global technology business Smiths Group plc.
EC Authorized Representative
Smiths Medical International Ltd.
WD24 4LG, UK
Phone: +44 (0) 1923 246434
Product(s) described may not be licensed or available for sale in Canada or other
countries outside of the United States.
Smiths Medical Canada Ltd.
Markham, Ontario, Canada, L3R 4Y8
Phone: 905-477-2000
Toll-Free: 1-800-387-4346
CADD, CADD-Prizm, Medication Cassette Reservoir Design and the Smiths Medical
and CADD design marks are trademarks of Smiths Medical. The symbol ® indicates
the trademark is registered in the U.S. Patent and Trademark Office and certain other
countries. ©2010 Smiths Medical. All rights reserved. 9/10 IN19812
Quick Reference Card for Clinicians
Intermittent Lock Level 1
Power DataIn/Out
START
STOP
Y
N
LOCK
DOSE
NEXT
ENTER OPTIONS
?
D
E
C
F
G
B
H
A
®
CADD-Prizm®VIP
Ambulatory Infusion Pump
Model 6100 and 6101
Please read the Instructions for Use or Operator’s Manual supplied with the product
for detailed instructions, warnings, and cautions on the safe use of the product.
IN19812.indd 1 10/26/11 1:27 PM
Time
Duration
KVO KVO
Delay
(using Next Dose
Start Time)
Dose
Volume
Cycle
Duration
DoseDose
PROGRAMMING SCREENS PUMP OPERATIONS
To View Current
Lock Level
To Start or
Stop Pump
To View Time
Remaining
To View Next
Dose Start Time
To Change
Battery
1. Stop the pump.
2. Press LOCK.
3. Press NEXT to return to original screen.
1. Press STOP/START.
2. Press
´.
If the pump is stopped during a dose, the
subsequent doses are delayed by the
amount of time the pump is stopped. If the
pump is stopped during delivery at the KVO
rate, the start times of subsequent doses
are not affected.
Time Remaining is the amount of time
remaining to deliver the current dose and
dose cycle.
1. Press OPTIONS.
2. Press
´or Îuntil Time Remaining
appears.
3. Press ENTER.
NOTE: If dashes appear in place of times,
the dose has been delayed by the Next Dose
Start Time.
1. Press NEXT until Next Dose Start Time
appears. The time will be displayed.
1. Stop the pump.
2. Press button on the battery door and
slide battery door forward.
3. Remove and discard used battery.
4. Match + and - markings on new battery
with markings on pump.
5. Insert battery; the pump should beep.
6. Replace the battery door.
7. The pump will power up.
Reservoir Volume Volume of fluid in filled reservoir
Dose Volume Volume of dose in milliliters
Dose Duration Time period over which dose will
be delivered
Dose Cycle Time from the start of one dose to the
start time of the next dose
KVO Rate Keep Vein Open rate allows delivery of
a minimal amount of drug to maintain
catheter patency
Next Dose Start Time Date and time designated to start delivery
of next dose
Dose Rate Rate at which dose will be delivered
(automatically calculated)
Milliliters Given Displays total amount delivered since date
and time shown
Air Detector Status Shows whether Air Detector is required,
turned on or off (appears if Air Detector is
attached)
«Prizm Match
Battery»
«Prizm Open
Door w/Button»
CHANGING THE CASSETTE
1. Stop the pump.
2. Close all tubing clamps.
3. If necessary, use key to unlock
used cassette.
4. Use side of key or coin to unlatch
used cassette.
5. Disconnect tubing from patient.
6. Remove and discard used cassette.
1. Close tubing clamp.
2. Insert new cassette hooks into pump’s
hinge pins.
3. Place pump with cassette on firm surface
and push down on top of pump.
4. Use coin or side of key to latch cassette.
WARNING: Attach the cassette properly. A
detached or improperly attached cassette
could result in unregulated gravity infusion
of medication from the fluid container or
a reux of blood. If you are using a CADD®
administration set or CADD™medication
cassette reservoir that does not have the flow
stop feature, you must use a CADD® extension
set with anti-siphon valve or a CADD®
administration set with either an integral or
add-on anti-siphon valve to protect against
unregulated gravity infusion that can result
from an improperly attached cassette.
5. Gently twist and pull on cassette to make
sure it is attached properly.
6. If
Reset Reservoir to — ml? appears,
press ´if appropriate.
7. When
Prime Tubing? appears, open
tubing clamp and press ´if appropriate.
Follow priming instructions on pump’s
screen. When primed, close tubing clamp.
8. If applicable, thread tubing through
Air Detector.
9. When
Start the Pump? appears, connect
new tubing to patient’s access site, open
tubing clamp, and press ´.
To Remove
Used Cassette
To Attach
New Cassette
D. Zurn 05/16/96
Prizm Latched, Unlocked
Priming the
Fluid Path
Using the
Prime Option
The priming feature is available either:
a) when you attach a cassette (above), or
b) in the Options Menu. To access the
Prime option in the Options Menu:
Pump must be in LL0 or LL1.
1. Stop the pump.
2. Press OPTIONS.
3. Prime should be displayed; press ENTER.
4. Follow priming instructions on
pump’s screen.
To Reset
Reservoir
Volume
Without
Changing
Cassette
1. Stop the pump.
2. Press NEXT to display Reservoir
Volume screen.
3. Press ENTER.
4. If message appears asking you to reset the
Reservoir Volume, press ´.
Intermittent Infusion Cycle
IN19812.indd 2 10/26/11 1:27 PM

Other manuals for CADD-Prizm VIP

4

This manual suits for next models

2

Other Smiths Medical Medical Equipment manuals

Smiths Medical CADD-Legacy PCA User manual

Smiths Medical

Smiths Medical CADD-Legacy PCA User manual

Smiths Medical acapella User manual

Smiths Medical

Smiths Medical acapella User manual

Smiths Medical CADD-Legacy PLUS Reference guide

Smiths Medical

Smiths Medical CADD-Legacy PLUS Reference guide

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical Oxi-Pulse 30 WW1030 User manual

Smiths Medical

Smiths Medical Oxi-Pulse 30 WW1030 User manual

Smiths Medical acapella portex User manual

Smiths Medical

Smiths Medical acapella portex User manual

Smiths Medical CADD-Solis User manual

Smiths Medical

Smiths Medical CADD-Solis User manual

Smiths Medical Pneupac babyPAC 100 User manual

Smiths Medical

Smiths Medical Pneupac babyPAC 100 User manual

Smiths Medical Surgivet Advisor User manual

Smiths Medical

Smiths Medical Surgivet Advisor User manual

Smiths Medical CADD-legacy 1 User manual

Smiths Medical

Smiths Medical CADD-legacy 1 User manual

Smiths Medical Pneupac babyPAC B100 Assembly instructions

Smiths Medical

Smiths Medical Pneupac babyPAC B100 Assembly instructions

Smiths Medical acapella duet Vibratory PEP Therapy System User manual

Smiths Medical

Smiths Medical acapella duet Vibratory PEP Therapy System User manual

Smiths Medical CADD-MS 3 User manual

Smiths Medical

Smiths Medical CADD-MS 3 User manual

Smiths Medical 1884SS User manual

Smiths Medical

Smiths Medical 1884SS User manual

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP User manual

Smiths Medical Level 1 H-1200 User manual

Smiths Medical

Smiths Medical Level 1 H-1200 User manual

Smiths Medical BCI 3304 User manual

Smiths Medical

Smiths Medical BCI 3304 User manual

Smiths Medical CADD-Solis User manual

Smiths Medical

Smiths Medical CADD-Solis User manual

Smiths Medical CADD-Solis Reference guide

Smiths Medical

Smiths Medical CADD-Solis Reference guide

Smiths Medical level 1 HOTLINE HL-90 User manual

Smiths Medical

Smiths Medical level 1 HOTLINE HL-90 User manual

Smiths Medical Equator User manual

Smiths Medical

Smiths Medical Equator User manual

Smiths Medical 1850 User manual

Smiths Medical

Smiths Medical 1850 User manual

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical

Smiths Medical CADD-Prizm VIP 6100 User manual

Smiths Medical PORTEX acapella choice User manual

Smiths Medical

Smiths Medical PORTEX acapella choice User manual

Popular Medical Equipment manuals by other brands

Huntleigh CONTOURA 980B Instructions for use

Huntleigh

Huntleigh CONTOURA 980B Instructions for use

inhealth Blom-Singer Instructions for use

inhealth

inhealth Blom-Singer Instructions for use

Arjo Sorrento Floor Bed Instructions for use

Arjo

Arjo Sorrento Floor Bed Instructions for use

Cardiac Science 5300 instruction manual

Cardiac Science

Cardiac Science 5300 instruction manual

lopital Amfora Instructions for use

lopital

lopital Amfora Instructions for use

eitan medical avoset quick start guide

eitan medical

eitan medical avoset quick start guide

BodyCAP e-Celsius Performance quick start guide

BodyCAP

BodyCAP e-Celsius Performance quick start guide

AGFA Valory user manual

AGFA

AGFA Valory user manual

Viztek VIZION DR + installation guide

Viztek

Viztek VIZION DR + installation guide

Otto Bock 4030 Manu Direxa Basic manual

Otto Bock

Otto Bock 4030 Manu Direxa Basic manual

Abbott e i-STAT Alinity Quick reference guide

Abbott

Abbott e i-STAT Alinity Quick reference guide

DentalEZ Simplicity Installation, operation & care manual

DentalEZ

DentalEZ Simplicity Installation, operation & care manual

IRUDEK ECO instruction manual

IRUDEK

IRUDEK ECO instruction manual

Covidien Kangaroo 924 Directions for use

Covidien

Covidien Kangaroo 924 Directions for use

ECOPOSTURAL C3523 instructions

ECOPOSTURAL

ECOPOSTURAL C3523 instructions

Summit to Sea GRAND DIVE PRO manual

Summit to Sea

Summit to Sea GRAND DIVE PRO manual

Reison nordiska 10-296 user manual

Reison

Reison nordiska 10-296 user manual

BIOTRONIK ProMRI Quick reference guide

BIOTRONIK

BIOTRONIK ProMRI Quick reference guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.