Alarms & Troubleshooting (Continued)
9-volt
Battery Low
9-volt battery is low but
pump is operable.
Change battery soon.
9-volt Battery
Depleted
9-volt battery is depleted and
cannot operate pump;
install fresh battery immediately.
High Pressure Possible blockage downstream (between
pump and patient), clamp or kink in fluid
path. Find cause and remove blockage to
continue. Or, press STOP/START to stop
pump and silence alarm; remove blockage
and restart pump.
Upstream
Occlusion
(Model 6101
only)
Fluid is not flowing from the reservoir
to the pump. Check for a kink, a closed
clamp or an air bubble in tubing between
reservoir and pump. Press STOP/START
to silence alarm for two minutes; remove
obstruction and restart pump.
Error Detected Pump fault occurred. Close tubing clamp,
remove pump from service.
Air In Line
Detected
Air in fluid path. Press NEXT to silence
alarm. Make sure tubing is threaded in Air
Detector properly. If fluid path contains
air, close clamps, disconnect from patient
and prime fluid path.
NOTE: When the pump is powered up in
LLO, the Air Detector will automatically
be turned on. If Air Detector is not
required, turn it off in the Options
Menu if you do not want to use it.
3 Beeps Every
5 Minutes;
Stopped on
Main Screen
Pump is stopped. Start pump
if appropriate.
Dose
scheduled
to start in -
Minutes / Start
the Pump
Pump is stopped, but a dose is scheduled
to be delivered in the number of minutes
indicated. In order for the dose to be
delivered, you must start the pump.
Dose scheduled
to start is now
overdue / Start
the Pump
Pump is stopped, but a dose is overdue
for its scheduled delivery. In order for the
dose to be delivered, you must start the
pump.
Dose Duration
is being
lengthened
You have entered a new volume that would
exceed the maximum rate of 350 ml/hr. The
pump automatically lengthened the dose
duration to accommodate the volume you
entered, but confirmation
or a new duration is required.
Dose Cycle
is being
lengthened
The dose duration you have entered
exceeds the dose cycle that is currently
programmed. The pump automatically
lengthened the dose cycle, but
confirmation or a new cycle is required.