Pressure Gauge Replacement
Model 464 125 psi
High Quality Groundwater and Surface Water Monitoring Instrumentation
(Page 1 of 2)
Tools and Materials Needed
1. 464 Panel Mount Gauge 1/8" NPTM (Spare) (#111806)
2. 3/32" Allen Key
3. Flat Screwdriver
4. 14 mm Wrench
5. Teflon Tape
Instructions
1. Turn the Control Unit off and remove the batteries.
2. Use the 3/32" Allen key to undo the four hex screws on the sides
of the Control Unit.
Hex
Screws
Hex
Screws
5. Use the flat screwdriver to loosen the two screws from the metal
ring holding the Pressure Gauge in place.
Panel
Pressure Gauge
3. Remove the panel from the Control Unit and flip it over to access
the Pressure Gauge.
®Solinst is a registered trademark of Solinst Canada Ltd.
Pressure Gauge
464 Panel Mount Gauge 1/8" NPTM (Spare) (#111806)
4. Push down on the white “grasping ring” on the elbow fitting and
pull to remove the tubing from the elbow on the Pressure Gauge.
Note the orientation of the elbow; you will reattach it to the new
Pressure Gauge in a similar position.
White”Grasping Ring”
6. Twist to unlock the metal ring from the Pressure Gauge, then
remove the ring and the Pressure Gauge from the panel.
7. Use the 14 mm wrench to remove the elbow fitting from the old
Pressure Gauge.
Teflon is a registered trademark of DuPont.
Note: You may wish to remove other tubes in order to more easily access
the Pressure Gauge. Ensure they are properly reconnected after
the new Pressure Gauge is installed.
8. Wrap Teflon tape around the threads on the new Pressure Gauge
and screw the elbow fitting onto the threads. Note the orientation of
the elbow when tightening; you should match its original position.