8. Unscrew the brass fitting from the Solenoid Valve Assembly,
and remove the magnetic plunger.
®Solinst is a registered trademark of Solinst Canada Ltd.
Note: The photo above shows the tubing fittings and Shuttle Valve
Assembly fitting removed from the Solenoid Valve Assembly,
you do not need to remove these.
12.Screw the Solenoid Valve Assembly into the Shuttle Valve
Assembly. Screw the two fittings back into the Shuttle Valve
Assembly (see top left photo for proper connections).
13. Plug the Solenoid Connector back into the Circuit Board,
and apply a small amount of adhesive to the outside of the
Connector to help secure it to the Circuit Board.
14. Slide the Shuttle Valve Assembly back into place under the
bracket and secure it using the three screws.
15. Push the three black tubes back into the fittings on the Shuttle
Valve Assembly. Use the photo on Page 1 to ensure correct
connections.
16. Push the clear tubes into the fittings on the Solenoid Valve
Assembly. Use the photo on Page 1 to ensure correct
connections.
17.Place the panel in the Control Unit case and reinstall the four
hex screws.
18.Reinstall the batteries. Test the Control Unit to ensure proper
connections were made and the Solenoid is now functioning.
Solenoid Valve
Assembly
Shuttle Valve
Assembly
9. Inspect the small rubber circular pad on the end of the plunger.
If this pad is damaged, dented, etc. the Solenoid will not seal
properly, therefore you may need a new Solenoid Valve
Assembly. (Contact Solinst). See photo below comparing a
damaged pad to a pad in good condition.
11.Ensure the o-ring is properly seated inside the brass fitting.
Insert the magnetic plunger back into the Solenoid Valve
Assembly, and screw the brass fitting back onto the Assembly.
Magnetic
Plunger
Steel Fitting Brass
Fitting
O-ring
Magnetic
Plunger
10.To clean the Assembly, use a cotton swab and distilled water
to clear any debris and dirt from inside the brass fitting, around
the o-ring, steel fitting and from the magnetic plunger.