SYSMEX pocH-100i User manual

AUTOMATED HEMATOLOGY ANALYZER
INSTRUCTIONS FOR USE
SYSMEX CORPORATION
KOBE, JAPAN
Copyright © 2005–2015 by SYSMEX CORPORATION
All rights reserved. No part of this Instructions for use may be
reproduced in any form or by any means whatsoever without
prior written permission of SYSMEX CORPORATION.
Code No. CG184474 en-uk
PRINTED IN JAPAN
Date of Last Revision: April 2015
Software Version: 01-23 and onwards

Overview and Initial settings
• T he poc H-10 0
i
isequippedwitharinsecup,and after aspiratinga sample orcontrolblood, thepiercer
(pipette)isautomaticallycleaned.Itisnotnecessarytowipethesamplepipette(aspiratingpiercer).
•
Operate the instrument as instructed. Reliability of test results cannot be guaranteed if instructions in this
manual are not followed. If the instrument fails to function properly as a result of either the user failing to
operatethesystemasspeciedinthemanualortheuser’sutilizationofaprogramnotspeciedbySysmex,
theproductwarrantywillnotapply.
• We reserve theright of continuous product enhancement. Design and specications may be subject to
change due to further product development.This may result in deviation of actual product properties
againstthepropertiesstatedinthismanual.
• Thecontentsofthescreensillustratedinthismanualmaydifferfromtheactualscreensdisplayedonthe
instrument.
• Datageneratedby thepocH-100
i
donotreplaceprofessionaljudgmentinthedeterminationofadiagno-
sisorinmonitoringpatienttherapy.
•The Sysmex pocH-100
i
is an automated hematology analyzer for in vitro diagnostic use in clinical labo-
ratoriesandphysicianofcelaboratories.ThepocH-100
i
shallonlybeusedforinvitroanalysisofhuman
bloodorarticialcontrolblood.Anyotheruseisregardedasnon-specied.
Onlyreagentsandcleaningsolutionsstatedinthismanualarepermittedforuse.
Observeallcleaningandmaintenanceproceduresattherequiredintervals.
ⅠpocH-100i RevisedApril2015
CE-mark
INSTRUCTIONS FOR USE
Beforeoperatingtheinstrument,carefullyreadthismanual.
Keepthismanualforreference.Forfurtherinformation,pleasecontactyourSysmexrepresent-
ative.
Carefullyread"Safetyinformation"and"Measuresforpersonnel"page.7-2.
Forexplicationofwarningsignsusedinthemanualseechapter"Symbols"page.1.
AUTOMATED HEMATOLOGY ANALYZER
1. SYSTEM OVERVIEW,
TECHNICAL INFORMATION & SETUP
INTRODUCTION
ThesystemdescribedinthismanualismarkedwithaCE-markwhichconrmsthecompliance
withtheessentialrequirementsofthefollowingEuropeanDirectives:
98/79/EC(IVDDirective)
2011/65/EU(RoHSDirective)

Ⅱ
pocH-100i Revised June 2008
Operation
2. ANALYSIS
3. QUALITY CONTROL
Maintenance/Troubleshooting/Safety
4. CLEANING & MAINTENANCE
5. SETTINGS & CALIBRATION
6. TROUBLESHOOTING
7. SAFETY INFORMATION & WASTE DISPOSAL
8. INDEX

1
12345678
INTRODUCTION
INTRODUCTION
pocH-100
i
Revised April 2013
ADDRESSES SYMBOLS
Head Office Japan
SYSMEX CORPORATION
1-5-1 Wakinohama-Kaigandori
Chuo-ku, Kobe 651-0073
JAPAN
Europe
European Representative
SYSMEX EUROPE GmbH
Bornbarch 1
22848 Norderstedt, Germany
Phone: +49 40 5 27 26-0
Fax: +49 40 5 27 26-100
Asia-Pacific
SYSMEX ASIA PACIFIC PTE LTD.
9 Tampines Grande, #06-18,
Singapore 528735
Phone: +65-6221-3629
Fax: +65-6221-3687
Americas
SYSMEX AMERICA, Inc.
577 Aptakisic Road
Lincolnshire, IL 60069, U.S.A.
Phone: +1-224-543-9500
Fax: +1-224-543-9505
• Ordering of supplies and replacement parts
To order supplies or replacement parts, please contact your
local Sysmex representative.
• Service and maintenance
Please contact the Service Department at your local Sysmex
representative.
Symbols to indicate biological risks, warnings and helpful information are presen-
ted throughout this manual to call to attention important information on safety
and handling. Non-compliance with this information compromises the safety fea-
tures incorporated in the analyzer.
Indicates the risk of serious injury or
death by biohazardous materials or
conditions.
Biological risks Warning
If this sign is ignored and the instru-
ment is operated incorrectly, there is a
potentially hazardous situation which
may result in injury of an operator.
If steps indicated by this sign are ig-
nored, analysis results may produce
erroneous data.
Incorrect analysis results
Indicates information which is useful in
operating the instrument.
Note
Indicates that the marked component
can be hot. Do not touch.
Warning, hot
High voltage exceeding 1 kV.
Danger, electric shock
Important
Indicates what you need to know in
order to maintain instrument perform-
ance and to prevent damage.
INTRODUCTION

2
12345678
INTRODUCTION
INTRODUCTION
Abbreviations
•CBC :CompleteBloodcellCount
•LD :LowerDiscriminator
•LL :LowerLimit
•PD :Pre-dilutedmode
•QC :QualityControl
•T1 :TroughDiscriminator1
•T2 :TroughDiscriminator2
•UD :UpperDiscriminator
•UL :UpperLimit
•WB :WholeBloodmode
Unit
•dL :deciliter(0.1liter)
•fL :femtoliter(10-15liter)
•μL :microliter(10-6liter)
•pg :picogram(10-12gram)
Analysis parameters
This instrument provides results for the following parameters:
•WBC:Numberofleukocytes
•RBC :Numberoferythrocytes
•HGB :Hemoglobin concentration
•HCT :Hematocritvalue:Erythrocyteratiooftotalbloodvolume
•MCV:Meancorpuscularvolume
•MCH:Meancorpuscularhemoglobin
•MCHC : Meancorpuscularhemoglobinconcentration
•PLT :Numberofplatelets
•LYM%/W-SCR:%ofsmallleukocytestototalWBC
Theyareassumedtobeequivalenttolymphocytes(whitecell-smallcellratio).
•MXD%/W-MCR:%ofmiddleleukocytestototalWBC
They are assumed to be equivalent to monocytes, eosinophils, and basophils
(whitecell-middlecellratio).
•NEUT%/W-LCR:%oflargeleukocytestototalWBC
Theyareassumedtobeequivalenttoneutrophils(whitecell-largecellratio).
•LYM#/W-SCC:Absolutenumberofsmallleukocytes
Theyareassumedtobeequivalenttolymphocytes(whitecell-smallcellcount).
•MXD#/W-MCC:Absolutenumberofmiddleleukocytes
They are assumed to be equivalent to monocytes, eosinophils, and basophils
(whitecell-middlecellcount).
•NEUT#/W-LCC:Absolutenumberoflargeleukocytes
Theyareassumedtobeequivalenttoneutrophils(whitecell-largecellcount).
•RDW-SD:Calculateddistributionwidthoferythrocytes,standarddeviation
ABBREVIATIONS / UNITS
•RDW-CV:Calculateddistributionwidthoferythrocytes,coefcientofvariation
•PDW :Calculateddistributionwidthofplatelets
•MPV:Meanplateletvolume
•P-LCR:Ratiooflargeplatelets(volumeexceeding12fL)tothetotalnumberof
platelets
When analyzing in the pre-diluted mode, only the following eight parameters are
output: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT.
Sysmex, pocH-100
i
, pocH-pack 65, pocH-pack 65XL and EIGHTCHECK are registered
trademarks of SYSMEX CORPORATION.
•CELLCLEANisatrademarkofSYSMEXCORPORATION.
•CAPIJECTisaregisteredtrademarkofTerumo Corporation.
•ETHERNETisaregisteredtrademarkofXeroxCorporation.
•MICROTAINERisaregisteredtrademarkofBectonDickinsonCompany.
• S A R S T E D T, S-MONOVETTEandMICROVETTEareregisteredtrademarksofWalter
Sarstedt Geräte und Verbrauchsmaterial für Medizin und Wissenschaft.
•ISBT128(InternationalSocietyofBloodTransfusion)iscopyrightedbyandusedunder
License Agreement with ICCBBA, Inc.
•LinuxisaregisteredtrademarkortrademarkofLinusTorvaldsinU.S.A.andothercoun-
tries.
All other trademarks not explicitly mentioned in this manual belong to their respective
owners. Sysmex does not authorize their use.
NAMES
pocH-100i Revised February 2011

3
12345678
pocH-100
i
Revised February 2011
INTRODUCTION
INTRODUCTION
TECHNICAL BASICS
•The calculation of indices is based on international principles in hematology.
•The method for counting blood cells is based on the electric resistance detection prin-
ciple.
• Hemoglobin concentration is determined by a photometric measuring method.
FUNCTIONAL DESCRIPTION
A N A LYS I S
Anexactvolumeofsampleisaspiratedwiththeaspiratingpiercer(pipette).
Thisvolumeofthesampleisthentransferred(togetherwithadenedvolumeofdilu-
ent)tothemixingchamberandfromthereautomaticallytothetransducer.
All parameters are analyzed using the same transducer in the order:
(1)WBC/HGBand(2)R B C / P LT.
ForWBC/HGBanalysis,theWBC/HGBlyseisaddedtothemeasuringchambertopro-
vide further dilution and hemolysis of the sample‘s RBC.
This process takes about ten seconds.
During this reaction period, the erythrocytes are dissolved under the influence of the
lysis, hemoglobin is released and is converted into red methemoglobin. The leukocytes
remain intact.
Thevolumeandnumberoftheleukocytes(WBC)aredeterminedbytheDCdetection
method. In the HGB detector, the hemoglobin concentration is photometrically meas-
ured.
ForRBC/PLTanalysis,thedilutedsampleistransferredfromthemixingchambertothe
transducer, and the volume and blood cell count for erythrocytes and platelets are ana-
lyzed by the DC detection method.
WARRANTY
All Sysmex instruments are warranted against defective material or workmanship for a
period of one year, commencing on the installation date on the customer‘s premises
This warranty does not cover any defect, malfunction or damage due to:
•Accident,neglectorwillfulmistreatmentoftheproduct;
•Failuretouse,operate,serviceormaintaintheproductinaccordancewiththeapplica-
bleSysmexInstructionsforUse;
•Failuretousetheappropriatereagentsandconsumablesspeciedfortheproduct.
INTRODUCTION

4
12345678
INTRODUCTION
INTRODUCTION
pocH-100i Revised June 2008

2345678
1
1-1
1.SYSTEM OVERVIEW,TECHNICAL INFORMATION & SETUP : CONTENTS
1.1. SYSTEM OVERVIEW 1-2
1.1.1. The instrument 1-2
1.1.2. The reagents 1-4
1.2. TECHNICAL INFORMATION 1-6
1.2.1. Performance characteristic - Specifications 1-6
1.2.2. System limitations - Interferences 1-8
1.3. SETUP 1-9
1.3.1. Check parts 1-9
1.3.2. Optional adapters 1-10
1.3.3. Insert paper roll 1-11
1.3.4. Connect the pocH-pack 65/pocH-pack 65XL 1-12
1.3.5. Barcode reader (optional) 1-13
1.3.6. Power cord 1-14
1.3.7. Switch ON 1-14
1.3.8. Set language 1-15
1.3.9. Date & Time 1-16
1.3.10. LCD contrast 1-17
1.3.11. Sample collection tube 1-18
pocH-100
i
Revised February 2011
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP

2345678 1
1-2
Before starting setup and analysis, please read through the following overview to get familiar with the instrument‘s parts and their description.
Open button
System fuses
AC power supply
Float switch connector
for waste bottle
Lysing reagent
inlet nipple
Diluent inlet nipple
Sample position
Adjust LCD screen contrast.
Press to open the sample po-
sition compartment.
AC 10 0 V
AC 240 V
Monitors the level in the waste bottle.
Aspirates lyse reagent.
Aspirates diluent.
Holds the adapter for the
analysis sample tube.
1.1.1.The instrument
1.1. SYSTEM OVERVIEW
Waste outlet nipple
Discharges waste.
Contrast adjustment knob
pocH-100
i
Revised June 2008
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
Color LCD screen

2345678
1
1-3
Main power switch
Serial port
Card slot
Built in thermal printer
Detector
RBC transducer, WBC transducer
and Hemoglobin detector
Rinse cup
Rinses the piercer (pipette) to
avoid sample-to-sample carry-
over.
Aspirates WB/PD blood samples.
Communication with host computer, etc.
9-pin D-SUB serial interface connector
(communication with host computer,
etc.)
Aspirating piercer (pipette)
1.1.SYSTEM OVERVIEW
Inserts a program card specified
by Sysmex.
Socket for Barcode-Reader
Socket for LAN
Tr a p c h a m b e r
In case of malfunction, avoids
reagents’ reflux into the vacuum
pump.
1.1.1.The instrument
pocH-100
i
Revised February 2011
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION

2345678 1
1-4
Purpose & methodology Storage & shelf life Composition
Diluent (blue)
• Diluentforuseinhematology
analyzers.
•Ready-to-usediluentforDC
detectionandwholebloodab-
sorbanceanalysis.
Lysing reagent (purple)
•ReagentforlysingRBCforac-
curateWBCcountandhemo-
globindetermination
•Thereagentiscolorless,trans-
parentandcontainsnocyanide.
•Thisisalysingreagentto
lysetheerythrocytes,allows
aleukocytecountandvolume
distributionanalysisbytheDC
detectionmethod.
•Hemoglobinisreleasedduring
RBClysis,andisconvertedto
theredmethemoglobin.
•OptionalpocH-pack65XLcan
beused.Thepurpose&meth-
odology,Storage&shelflife,
andCompositionaresame
withthepocH-pack65.
• Aportionofthisdilutedsample
istransferredautomatically
tothehemoglobindetector,
wheretheabsorbanceofthe
redpigmentismeasuredto
providethebloodhemoglobin
concentration.
Expirationdateisshownon
theouterpackaging.
•Unopenedpackscanbe
storedfor12months.
Storeat2-30°C.
•Opened,reagentstability
ismax.60days.
Useat15-30°C.
Diluent (blue)
•SodiumChloride 6.38g/L
• Boricacid 1. 0 g/L
• SodiumTetraborate 0.2g/L
• EDTA-2K 0.2g/L
Lysing reagent (purple)
•SodiumChloride 0.6g/L
• Org.quart.ammoniumsalt
8.5g/L
pocH-pack 65
ReplaceanypocH-pack65/pocH-pack65XL
showingsignsofcontaminationorinstability.
Usingitcanresultinincorrectanalysisresults.
Donotuseonceitwasfrozen.
Warning
• UsewholebloodsamplecollectedinEDTAanti-
coagulant.
•
UsepocH-pack65/pocH-pack65XLonlywith
Sysmexreagentsandanalyzers.
•TheperformanceofSysmexinstrumentscannot
beguaranteedifusingotherreagents.
Incorrect analysis results
*ThisisimageofthepocH-pack65.ThepocH-pack65XLdoesnothavewastecontainer.
pocH-100
i
RevisedFebruary2 0 11
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
1.1.2. The r eagents

1- 5
pocH-100
i
Revised April 2015
2345678
1
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION
Purpose & methodology Storage & shelf life Composition
• Strong alkaline detergent.
• Removes lyse reagents/cellular
residuals and blood proteins from
the hydraulic systems, transducer,
whole blood aspiration piercer
(pipette) and the Hgb ow cell.
Expiration date is shown on the
outer packaging.
• Store in a dark place at 1-30 °C.
• Use opened reagent within 60
days.
• Sodium
Hypochlorite 5.00%
Warning
• Avoid contact with skin and eyes.
• In case of skin contact, rinse with water.
• In case of contact with eyes, rinse immediately
with plenty of water and seek medical advice.
If swallowed, seek medical advice and follow
MSDS.
Purpose & methodology Storage & shelf life Composition
Control blood EIGHTCHECK
• 3WP-N (Normal level)
• 3WP-L (Low level)
• 3WP-H (High level)
test the precision of automated/
semi-automatic hematology ana-
lyzers.
Expiration date is shown on the
outer packaging.
• Store at 2-8 °C before and af-
ter opening.
• Once opened, the product is
stable for 7 days if returned to
the refrigerator promptly after
use.
• Control materials provide stable
parameter values after at least
12 hours at room temperature
(25 °C).
• EIGHTCHECK contain stabi-
lized animal erythrocytes,
xed animal white blood cells
and a platelet component in a
medium containing presera-
tives.
Important
Do not use for the calibration of the system!
EIGHTCHECK
3WP
CELLCLEAN
1.1. SYSTEM OVERVIEW
Biological risks
• This product, based on animal blood, should be
considered potentially capable of transmitting
infectious disease.
• Use CELLCLEAN only with Sysmex reagents and
analyzers.
• The performance of Sysmex instruments cannot
be guaranteed if using other detergent.
Incorrect analysis results
• Use EIGHTCHECK only with Sysmex reagents
and analyzers.
• The performance of Sysmex instruments cannot
be guaranteed if using other control material.
Incorrect analysis results
1.1.2.The reagents
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION

2345678 1
1-6
Ambient temperature 15 °C to 30 °C (ideal operating temperature at 23 °C)
(59 °F – 86 °F)
Relative humidity 30% to 85%
Main Unit dimensions
Width: 185 mm (7.3 in)
Depth: 460 mm (18.1 in)
Height: 350 mm (13.8 in)
Weight: approx. 14 kg (30.8 lbs)
Power supply 100-240 VAC (±10%), 50/60 Hz
Power consumption 150 VA or less
Analysis Parameters see Chapter "Abbreviations/Units".
Display range
WBC: 0.0 – 299.9 (× 103/μL)
RBC: 0.00 – 19.99 (× 106/μL)
HGB: 0 – 25.0(g/dL)
PLT: 0 – 1999 (× 103/μL)
Background limits
WBC: 0.3 (×103/μL)
RBC: 0.02 (× 106/μL)
HGB: 0.1 (g/dL)
PLT: 10 (× 10 3/μL)
Analysis time Approx. 148 seconds (After starting an analysis until displaying the analysis report)
Analysis principle
WBC: DC detection method
R B C / P LT: Hydrodynamic Focusing DC detection method
HGB: Non-cyanide HGB method
Required temperature compensation Approx. 512 BTU/h (Approx. 130 kcal/h)
Class of electric shock protection measures Class Ⅰ
Equipment
EMC characteristics Conforms with IEC 61326-1 (Class B, Group 1, Industrial environment)
Safety Conforms with IEC 61010-1 (Overvoltage categoryⅡ,Pollutiondegree2,Portableequipment)
1.2.TECHNICAL INFOR MATION
1.2.1. Performance characteristic - Specifications
pocH-100
i
Revised June 2008
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION

1- 7
pocH-100
i
Revised August 2014
2345678
1
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION
1.2.TECHNICAL INFORMATION
Parameter Whole blood mode Pre-diluted mode
Repeatability
(With 95% reliability limit)
WBC (≥ 4.0× 103/μL)
RBC (≥ 3.00 × 106/μL)
HGB
HCT
MCV
MCH
MCHC
PLT (≥ 100 × 103/μL)
LYM#
MXD# (≥ 1.0 × 103/μL)
NEUT#
LYM%
MXD% (≥ 12%)
NEUT%
RDW-SD or RDW-CV
PDW
MPV
P-L CR
3.5% or less
2.0% or less
1.5% or less
2.0% or less
2.0% or less
2.0% or less
2.0% or less
6.0% or less
15.0% or less
30.0% or less
15.0% or less
15.0% or less
30.0% or less
15.0% or less
4.0% or less
12.0% or less
5.0% or less
20.0% or less
6.0% or less
3.0% or less
2.5% or less
3.0% or less
3.0% or less
3.0% or less
3.0% or less
9.0% or less
—
—
—
—
—
—
—
—
—
—
Accuracy
WBC
RBC
PLT
± 3% or ± 0.2 × 103/μL or less
± 2% or ± 0.03 × 106/μL or less
± 5% or ± 10 × 103/μL or less
± 5% or ± 0.3 ×103/μL or less
± 3% or ± 0.05 ×106/μL or less
± 8% or ± 15 × 103/μL or less
(Excluding the dilution error at sample preparation.)
Linearity in whole blood mode
(applies, when RBC < 7.00 ×106/μL)
WBC 1.0 – 9.9 (×103/μL)
10.0 – 99.9 (×103/μL)
± 0.3 (×103/μL) or less
± 3% or less
RBC 0.3 – 0.99 (×106/μL)
1.00 – 7.00 (×106/μL)
± 0.03 (×106/μL) or less
± 3% or less
HGB 0.1 – 10.0 (g/dL)
10.0 – 25.0 (g/dL)
± 0.2 (g/dL) or less
± 2% or less
HCT 10.0 – 33.3 (HCT%)
33.3 – 60.0 (HCT%)
± 1.0 (HCT%) or less
± 3% or less
PLT 10 – 199 (×103/μL)
200 – 999 (×103/μL)
± 10 (×10 3/μL) or less
± 5% or less
1.2.1. Performance characteristic - Specifications
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP

2345678 1
1-8
1.2.2. System limitations – Interferences
The abnormal sample conditions listed
here are known to affect test results.
The majority of the listed sample condi-
tionsarenotmeasuredquantitatively
because these conditions vary due to
patient population, patient diagnosis,
age, medications, etc. Customers can
perform studies in order to show how
their specific patient populations are af-
fected by various conditions.
Note
Cause Potential Detection
WBC: false high leukocyte count
Lyse resistant erythrocytes Abnormal WBC histogram (WL flag)
Cold agglutinins /
cryoglobulins
Increased MCV, increased MCHC due to decreased HCT and RBC
Platelet aggregation Abnormal WBC histogram (WL flag) and Abnormal PLT histogram (PU flag)
Nucleated erythrocytes Abnormal WBC histogram (WL flag)
RBC: false low erythrocyte count
Cold agglutinins Increased MCV, increased MCHC due to decreased HCT
Microcytosis (severe) Low MCV
Fragmented erythrocytes Abnormal RBC histogram (RL flag) and Abnormal PLT histogram (PU flag)
HGB: false high haemoglobin measure-
ment
Lipaemia MCHC > 36.5 g/dL in severe cases
Abnormal protein MCHC > 36.5 g/dL in severe cases
Leukocytosis WBC >100,000/μL
HCT: false low haematocrit measurement Cold agglutinins Increased MCV and increased MCHC
Fragmented erythrocytes Abnormal RBC histogram (RL flag) and abnormal PLT histogram (PU flag)
HCT: false high haematocrit measurement Leukocytosis Very high leukocyte count with low erythrocyte count present
PLT: false low platelet count Platelet aggregation Abnormal PLT histogram (PU flag)
Giant platelets Abnormal PLT histogram (PU flag)
PLT: false high platelet count Microcytic erythrocytes Low MCV
Fragmented erythrocytes Abnormal RBC histogram (RL flag) and abnormal PLT histogram (PU flag)
Carry-over
WBC 3% or less
RBC 1.5% or less
HGB 1.5% or less
HCT 1.5% or less
PLT 5% or less
Consumables
Reagents: pocH-pack 65/pocH-pack 65XL
Detergent: CELLCLEAN
Controlmaterial:EIGHTCHECK-3WP
Aspirated sample volume in whole blood mode: approx. 15 μL
inpre-dilutedmode: approx.200μLofthedilutedsample;aminimumof20μLcapillarybloodisrequired.
Number of analyses that can be per-
formed with 1 pocH-pack 65/pocH-pack
65XL
pocH-pack 65: Approximately 65
pocH-pack 65XL: Approximately 240
(These numbers are cycle counts.)
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
pocH-100
i
Revised February 2011

1- 9
pocH-100
i
Revised April 2015
2345678
1
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION
1.3. SETUP
1.3.1. Check parts
1x
Container spout kit
No.6
blue (Diluent)
033-0403-1 for pocH-pack 65
033-0405-9 for pocH-pack 65XL
• Move and handle carefully. If the packaging is
damaged,openthepackageinthepresenceofa
representativefromthelogisticscompany.
Pleasestatethedamageclearlyonthereceipt.
•Prerequisites for installation
-Ensurethatpowerisavailable.
-Placeonatablewhich can support the device’s
weight(17.5kg).
-Providesufcientspaceforpropercooling,main-
tenanceorservicework.
The distance from the right side, rear, and top
panelstowallsshouldbeatleast15cm(6In.).
-The instrument must be installed in a dry and
dust-free location.
-Devicemustbeprotectedagainstwater.
•Safety instructions for installation
-
Donotexposetheinstrumenttoexcessive
tem-
peratureuctuationanddirectsunlight.
-Avoidshockandvibrations.
-Theinstallationlocationmustbewellventilated.
-Avoidinstallationneardevicescausingpotential
interference, such as radios, centrifuges, com-
putermonitors,wirelesscommunicationequip-
ment or similar devices.
-Installation of this instrument in places where
chemicals are stored or hazardous gas may be
present is not permitted.
*1Thispartisforopenpipette.
*2The waste bottle assy POCH (C1/WITH FSW
033-0041-1)mightbenecessaryincaseofusing
pocH-pack 65XL (optional).
Important
1x
Sampletubeadapter
green(Controlblood)
442-3051-2
1x
Sampletubeadapter
red(S-MONOVETTE
SARSTEDT2.7mL)
AA155506
1x
Thermalpaper
CL5840
AH042068
1x
Transducerbrush
No. 3
462-3523-6
1x
Tubepolyurethane
4MMID×6MMOD
442-5338-7
2x
MinipipetteNo.10
(1mL)
423-1776-2
1x
Powercord
4622-007-0092(Europe)
265-4 731-5
1x
Wiringcord
No.2188
963-0801-5
1x
CoiltubeSS-10
2 6 6 - 7 110 - 2
1x
Container spout kit
No.7
purple(Lysingreagent)
033-0414-2
1x
Sampletubeadapter
13mmSysmexwhite
CJ296916
1x
Screwdriver(-)
A6012Standard
462-2388-3
•InstructionsforUse
CG184474
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION
1x
ContainerSpoutKit
No.6
red(Waste)
033-0404-5
1x
Wastebottle
assyPOCH
(C1/WITHFSW)
033-0041-1
*2
1x
Container spout kit
No.7
purple(Lysingreagent)
033-0412-5
2x
Fuse02153.15MXP
250VT3.15AH
AL039459
1x
Screwdriver(+)
A6002Philips
462-2387-0
1x
SpacerNo.192
(13 mm)
365-9452-9
*1

1-10
2345678 1
1.3.2. Optional adapters
Adapter name Product code Adapter color
Micro tube adapter
(BD Microtainer tube with EDTA, Product No. 365975) CA392678 Orange
Micro tube adapter
(BD Microtainer tube with EDTA, Product No. 365955) CH515979 Purple
Micro tube adapter
(CAPIJECT (TERUMO)) CE466834 Yellow
Micro tube adapter
(MICROVETTE (SARSTEDT)) BP006581 Blue
Sample tube adapter
(for 15mm diameter tube) AJ501289 Black
•Whenusingmicrotubes,besuretoplacethemin
the correct adapter.
•Besuretoremovethecap.
Important
pocH-100
i
Revised February 2011
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP

1- 11
2345678
1
1
2
3
4
2
1
1.3.3. Insert paper roll
Open the paper holder.
Open paper holder by pushing the knob.
Remove the tape.
Remove the tape.
Insert paper.
Close lid. (*2)
Close the pepar holder.
Insert paper. (*1)
The printer head can get very hot. Do not touch!
Warning, Hot
1.3. SETUP
Static electricity may damage the printer head. Do
not touch.
*1. Insert the paper correctly. If inserted at an an-
gle, the paper may jam.
Important
*2. The printer cover must be closed ("clicking"
sound). If the cover is not closed completely,
an error message will be displayed.
Note
pocH-100
i
Revised June 2008
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION

1-12
2345678 1
1
2
3
4
Set the reagents and remove caps. Set the reagents and remove caps.
Set the reagents and remove caps. Set the reagents and remove caps. *
Insert container spout kits/ float
switch and reapply the caps.
Insert container spout kits/ float
switch and reapply the caps.
Insert container spout kits/ float switch into
the correct bottle and reapply the caps.
Insert container spout kits/ float switch into
the correct bottle and reapply the caps. *
Connect the tubes to the nipples. Connect the tubes to the nipples.
Connect the tubes to the appropriate nip-
ples. Connect the wiring cord.
Connect the tubes to the appropriate nip-
ples. Connect the wiring cord.
1.3.4. Connect the pocH-pack 65/pocH-pack 65XL
pocH-pack 65 pocH-pack 65XL
Fix tubes together.
Fix tubes together.
Biological risks
To avoid infections, wear protective garments
and gloves for cleaning and/or maintenance.
After completion of work, wash hands with
disinfectant.
• Connectthetubestothecorrectnipples;
• Donottotouchthetubesthatentertherea-
gent;
• Preventthereagentsfromspilling.
* The pocH-pack 65XL does not have waste
container.
Important
INTRODUCTION
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
pocH-100
i
Revised February 2011

1-13
2345678
1
1
2
1-13
1.3. SETUP
1.3.5. Barcode reader (optional)
Scans the barcode on the sample collection tube and automatically enters the sample number into the system.
ID Barcode specifications
Barcode labels’ specifications must match the requirements for the barcode reader.
Symbol Check digit Number of digits
ITF None
Modulus 10
Max. 14 digits (sample ID)
Max. 14 digits (sample ID) + 1 digit (check digit) = Max. 15 digits
NW-7 None
Modulus 11-16
Max. 15 digits (sample ID)
Max. 15 digits (sample ID) + 1 digit (check digit) = Max. 16 digits
CODE 39 None
Modulus 43
Max. 15 digits (sample ID)
Max. 15 digits (sample ID) + 1 digit (check digit) = Max. 16 digits
CODE 128 Modulus 103 Max. 15 digits (sample ID) + 1 digit (check digit) = Max. 16 digits
JAN-8 Modulus 10 Max. 7 digits (sample ID) + 1 digit (check digit) = Max. 8 digits
JAN-13 Modulus 10 Max. 12 digits (sample ID) + 1 digit (check digit) = Max. 13 digits
Plug cable into "BR" connection.
Plug cable into "BR" connection.
Turn the main switch ON, press
the trigger switch and check red
LED.
Turn the main switch ON, press the trigger
switch and check red LED.
•Disconnecttheinstrument‘smainsbeforecon-
necting the peripheral device. Else, the two de-
vices may not interchange any data.
•Inidentifyingpatientsamples,maximumintegrity
of data is required.
To avoid barcode identification mistakes, use a
check digit whenever the sample barcode is used.
If a check digit is not used, the chances of the bar-
code being misread increase.
Note
The barcode reader is not included as a standard
accessory. Refer to the barcode reader manual
for detailed information on the connection of the
barcode reader.
Important
pocH-100
i
Revised April 2013
SYSTEM OVERVIEW, TECHNICAL INFORMATION & SETUP
INTRODUCTION
Table of contents
Other SYSMEX Measuring Instrument manuals

SYSMEX
SYSMEX XW-100 User manual

SYSMEX
SYSMEX CyFlow Space User manual

SYSMEX
SYSMEX BK265455 User manual

SYSMEX
SYSMEX CyFlow Ploidy Analyser User manual

SYSMEX
SYSMEX CyFlow Ploidy Analyser User manual

SYSMEX
SYSMEX XW-100 User manual

SYSMEX
SYSMEX CyStain VitalCount User manual

SYSMEX
SYSMEX SF-3000 User manual

SYSMEX
SYSMEX XW-100 Manual

SYSMEX
SYSMEX CyFlow Cube 6 V2m User manual