Section 1 -2 2 4 6 1Y nd 2246 Mod A Service
SPECIFICATION
INTRODUCTION
The 2246 1Y and 2246 Mod A are 100 MHz, four-
hannel, dual-sweep, portable os illos opes for
general-purpose use. A mi ropro essor-based
operating system ontrols most of the fun tions in
ea h instrument, in luding a fully integrated menu-
driven voltage and time measurement system with
SmartCursors®. A menu-driven servi e mode
provides for onfiguring of ertain menu and readout
displays, internal alibration, and servi ing
diagnosti s. The instruments in lude additional EMI
shielding. (See Table 1-2, Environmental Chara ter
isti s, for more information.)
The verti al defle tion system has four input han
nels. Two hannels have 11 basi defle tion fa tors
from 2 mV to 5 V per division, and two hannels have
two basi defle tion fa tors of 0.1 V and 0.5 V per
division. Basi defle tion fa tors an be extended
with attenuator probes. VOLTS/DIV readouts are
swit hed to display the orre t verti al s ale fa tors
when properly oded probes are onne ted to the
verti al input onne tors.
The horizontal defle tion system provides single,
dual, or delayed sweeps from 0.5 s to 20 ns per divi
sion (delayed sweep, 5 ms to 20 ns per division). The
trigger system provides stable triggering over the full
bandwidth of the verti al defle tion system.
Alphanumeri rt readouts of the verti al and hori
zontal s ale fa tors are displayed at the bottom of the
s reen. On-s reen verti al and horizontal ursors
provide a urate voltage, time, frequen y, and
phase measurements; measurement values are dis
played at the top of the rt.
The measurement features in lude voltage measure
ments for+Peak, -Peak, Peak-to-Peak, and average
DC, or positionable ursors for measuring voltage dif
feren e, time differen e, frequen y, and phase.
SmartCursors® that visually tra k voltage measure
ments, trigger levels, and ground an be pla ed on
displayed waveforms. Delay-time and delta-delay
measurements for time, frequen y, and phase are
available in ALT and B Horizontal Modes.
STANDARD ACCESSORIES
The following items are standard a essories shipped
with the 2246 1Y and 2246 Mod A instrument:
2 Probes, 10X, 2 meter, with a essories
1 Probe, IX, 2 meter, with a essories
1 Power ord
1 Power ord lamp
1 CRT implosion shield, blue plasti (installed)
1 Fuse, 2A, 250 V, slow-blow
1 Atta hing a essory pou h
1 A essory pou h, ziplo
1 Front over
The following items are standard a essories shipped
with the 2246 1Y instrument:
2 Operators manuals
2 Servi e manuals
2 Referen e guides
The following items are standard a essories shipped
with the 2246 Mod A instrument:
1 Operators manual
1 Referen e guide
RECOMMENDED RECALIBRATION
SCHEDULE
To ensure a urate measurements, he k the per
forman e of this instrument every 2000 hours of
operation, or, if used infrequently, on e ea h year.
Repla ement of omponents in the instrument may
also ne essitate readjustment of the affe ted
ir uits.
NOTE
The silver-oxide batteries used in the
2246 1Y (instruments serial numbered
B714999 and below) M ST BE REPLACED
AT LEAST EVERY 12 MONTHS to ensure
proper operation of the instrument.
REV OCT 1991 1-1