Best practices for using high performance oscilloscopes
Adhere to a regular schedule of calibration of the instrument. A Tektronix
Calibration Service plan is the most efficient way to ensure specified performance
over the lifetime of your Tektronix instrument. The service plan also guarantees
continuous compliance with international quality and traceability standards.
Contact your Tektronix Representative or refer to the www.tektronix.com Web
site for instrument calibration services.
Instrument temperature
and clearance
The instrument must have 3 inches of clearance on both the left and right sides for
proper ventilation. The air temperature around the instrument must be between
+5 °C and +45 °C (+41 °F and +113 °F).
DPO70000SX
environmental conditions
For measurements impacted by sub-picosecond skew, leave the instrument
powered on for at least 6 hours to allow the instrument to stabilize to environmental
conditions before running SPC, deskew, or taking precision measurements.
Deskew channels Use the Deskew/Attenuation control window to compensate for propagation
delays of different length probes and to set the input/output ratio of any external
attenuation or gain between the signal and input channels. Deskew relates directly
to the vertical settings, so it is important to deskew at the vertical settings at which
measurements will be taken or within the same amplifier range. Refer to the
Deskew/Attenuation Control Window topic in the Online Help of the instrument
for information on the deskew and attenuation procedures.
Operating system restore The instrument contains an operating system restore file on a separate partition of
the hard drive. The preferred method to restore the operating system is to use the
hard disk restore file. For more information, see the instrument user manual.
Updates to firmware and
applications
You should regularly check for updates to the instrument firmware and
applications of your Tektronix products. Access the www.tektronix.com/software
Web page and search for the model number of your instrument (or application) to
check for updates.
6 MSO70000C/DX, DPO70000C/D/DX/SX, DSA70000C/D, DPO7000C Best Practices