TESTO 330 LX User manual

testo 330 LX · Flue gas analyzer
Instruction manual

2

1 Contents
3
1Contents
1Contents ...................................................................................................3
2Safety and the environment....................................................................6
2.1. About this document........................................................................6
2.2. Ensure safety...................................................................................7
2.3. Protecting the environment..............................................................8
3Specifications ..........................................................................................9
3.1. Use ..................................................................................................9
3.2. Technical data ...............................................................................10
3.2.1. Examinations and licenses ............................................................................10
3.2.2. Bluetooth®module (option) ............................................................................10
3.2.3. Declaration of Conformity ..............................................................................12
3.2.4. Measuring ranges and resolution...................................................................13
3.2.5. Accuracy and response time..........................................................................14
3.2.6. Other instrument data ....................................................................................15
4Product description...............................................................................17
4.1. Case 0516 3300 (accessory).........................................................17
4.1.1. Bottom level view...........................................................................................17
4.1.2. Top level view................................................................................................18
4.2. Case 0516 3301 (accessory).........................................................19
4.2.1. Bottom level view...........................................................................................19
4.2.2. Middle level view............................................................................................20
4.2.3. Top level view................................................................................................21
4.3. Measuring instrument ....................................................................22
4.3.1. Overview........................................................................................................22
4.3.2. Keypad ..........................................................................................................23
4.3.3. Display...........................................................................................................24
4.3.4. Device connections........................................................................................25
4.3.5. Interfaces.......................................................................................................25
4.3.6. Components ..................................................................................................26
4.3.7. Carrying strap (0440 1001) ............................................................................27
4.4. Modular flue gas probe..................................................................28

1 Contents
4
5First steps.............................................................................................. 29
5.1. Commissioning ............................................................................. 29
5.2. Getting to know the product .......................................................... 29
5.2.1. Mains unit / rechargeable battery................................................................... 29
5.2.1.1. Changing the battery ....................................................................... 29
5.2.1.2. Charging batteries ........................................................................... 30
5.2.1.3. Mains operation............................................................................... 30
5.2.2. Connecting probes / sensors ......................................................................... 31
5.2.3. Switching on .................................................................................................. 32
5.2.4. Calling up a function ...................................................................................... 32
5.2.5. Entering values.............................................................................................. 32
5.2.6. Show graphic................................................................................................. 33
5.2.7. Printing / saving data ..................................................................................... 34
5.2.8. Remembering data (clipboard) ...................................................................... 34
5.2.9. Confirming an error message ........................................................................ 34
5.2.10. Switching off .................................................................................................. 35
5.3. Address/Location .......................................................................... 35
5.4. Measurement records ................................................................... 37
5.5. Instrument diagnosis..................................................................... 38
6Using the product ................................................................................. 39
6.1. Performing settings ....................................................................... 39
6.1.1. Assigning the right function key ..................................................................... 39
6.1.2. Instrument settings ........................................................................................ 39
6.1.2.1. Readings display............................................................................. 39
6.1.2.2. Alarm limits...................................................................................... 41
6.1.2.3. Units................................................................................................ 41
6.1.2.4. Date / time....................................................................................... 41
6.1.2.5. Energy management ....................................................................... 42
6.1.2.6. Display brightness ........................................................................... 42
6.1.2.7. Choose measurement type .............................................................42
6.1.2.8. Printer ............................................................................................. 43
6.1.2.9. Bluetooth®....................................................................................... 43
6.1.2.10. Language ........................................................................................ 43
6.1.2.11. Country version ............................................................................... 44
6.1.2.12. Password protection........................................................................ 44
6.1.3. Sensor settings.............................................................................................. 45
6.1.3.1. NO2addition.................................................................................... 45
6.1.3.2. O2reference .................................................................................... 45
6.1.3.3. Sensor protection ............................................................................ 46
6.1.3.4. Recalibration / adjustment ............................................................... 46
6.1.4. Fuels.............................................................................................................. 47
6.1.5. Programs....................................................................................................... 48

1 Contents
5
6.2. Measuring......................................................................................49
6.2.1. Preparing for measurement ...........................................................................49
6.2.1.1. Zeroing phases................................................................................49
6.2.1.2. Using the modular flue gas probe....................................................50
6.2.1.3. Configuring the reading display .......................................................51
6.2.1.4. Setting location and fuel ..................................................................51
6.2.2. Flue gas.........................................................................................................51
6.2.3. Draught-Measurement...................................................................................53
6.2.4. Micro pressure probe.....................................................................................54
6.2.5. BImSchV (testo 330-2 LX) .............................................................................55
6.2.6. CO undiluted..................................................................................................56
6.2.7. Smoke No. / HCT...........................................................................................56
6.2.8. Differential pressure.......................................................................................57
6.2.9. Differential temperature .................................................................................58
6.2.10. O2 air.............................................................................................................58
6.2.11. Gas flow.........................................................................................................59
6.2.12. Oil flow...........................................................................................................60
6.2.13. CO ambient....................................................................................................60
6.2.14. CO2 ambient..................................................................................................61
6.2.15. Automatic furnaces ........................................................................................62
6.2.16. Solid fuel measurement .................................................................................63
6.2.17. Gas pipe tests................................................................................................64
6.2.17.1. Tightness test 1 ............................................................................... 64
6.2.17.2. Tightness test 2 ............................................................................... 65
6.2.17.3. Let By Test ......................................................................................66
6.2.17.4. Leak detection .................................................................................67
6.3. Transferring data ...........................................................................68
6.3.1. Report printer.................................................................................................68
6.3.2. PC / Pocket PC..............................................................................................68
7Maintaining the product ........................................................................69
7.1. Cleaning the measuring instrument ...............................................69
7.2. Replacing sensors .........................................................................69
7.3. Recalibrating / adjusting sensors...................................................70
7.4. Replacing additional filter ..............................................................70
7.5. Cleaning the modular flue gas probe.............................................71
7.6. Replacing the probe module..........................................................71
7.7. Changing the thermocouple...........................................................71
7.8. Condensate container ...................................................................73
7.9. Checking / replacing the particle filter............................................74
8Tips and assistance...............................................................................75
8.1. Questions and answers .................................................................75
8.2. Accessories and spare parts .........................................................76
8.3. Updating the instrument software..................................................80

2 Safety and the environment
6
2Safety and the environment
2.1. About this document
This document describes the products testo 330-1 LX and
testo 330-2 LX with the instrument setting Country version |
Germany.
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Symbols and writing standards
Representation Explanation
Warning advice, risk level according to the
signal word:
Warning! Serious physical injury may occur.
Caution! Minor physical injury or damage to the
equipment may occur.
> Apply the specified precautionary measures.
Note: Basic or further information.
testo 330-1 LX The description only applies for the specified
instrument model testo 330-1 LX or
testo 330-2 LX
1. ...
2. ...
Action: several steps, the sequence must be
followed.
> ... Action: one step or optional step.
- ... Result of an action.
Menu Elements of equipment , equipment display or
program interface.
[OK] Control keys on equipment or control buttons in
program interface.

2 Safety and the environment
7
Representation Explanation
... | ... Functions / paths within a menu.
“...” Example entries
2.2. Ensure safety
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
> Do not perform contact measurements on non-insulated, live
parts.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts
from Testo.
> Any further or additional work must only be carried out by
authorised personnel. Testo will otherwise refuse to accept
responsibility for the proper functioning of the measuring
instrument after repair and for the validity of certifications.
> Only use the device in closed, dry rooms and protect it from rain
and moisture.
> Temperatures given on probes/sensors relate only to the
measuring range of the sensors. Do not expose handles and
feed lines to any temperatures in excess of 70 °C unless they
are expressly permitted for higher temperatures.
> The testo 330 LX must be checked before commissioning for
any visible damage. Do not commission the testo 330 LX if
there are signs of damage on the housing, mains unit or supply
lines. Electrical risk.

2 Safety and the environment
8
> Dangers may also arise from the systems being measured or
the measuring environment: Note the safety regulations valid in
your area when performing the measurements.
Use distilled water, or alternatively mild solvents such as
isopropanol to clean the flue gas analyzer. If using
isopropanol, please refer to the instruction leaflet for the
product. Isopropanol fumes have a slight narcotic effect,
and typically cause irritation of the eyes and sensitive
mucous membranes. When using it, please ensure that
there is adequate ventilation.
Do not store any objects that have come into contact with
solvents and/or degreasers (e.g. isopropanol) in the case.
Evaporating or leaking solvents and/or degreasers may
cause damage to the instrument and to the sensors.
The use of strong or harsh alcohol or brake cleaner can
result in damage to the instrument.
For products with Bluetooth®(optional)
Changes or modifications that have been made without the explicit
consent of the responsible approval authority, may cause the
retraction of the type approval.
Data transfer may be disturbed by equipment that uses the same
ISM-band, e.g. WLAN, microwave ovens, ZigBee.
The use of radio communication links is not permitted, among
others, in aeroplanes and hospitals. For this reason the following
points must be ensured before entering:
> Switch off the device:
> Isolate the device from any external power sources (mains
cable, external rechargeable batteries, ...).
2.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
> The button cell used in the instrument contains 1,2-
Dimethoxyethane (CAS 110-71-4). See EC Regulation No.
1907/2006 (REACH) Art. 33.

3 Specifications
9
3Specifications
3.1. Use
The testo 330 LX is a handheld measuring device for the
professional flue gas analysis of furnace systems:
• Small furnaces (burning oil, gas, wood, coal)
The solid fuel measurement adapter (0600 9765) is
required for measurements on solid fuel systems. The
adapter protects the measuring instrument from harmful
substances (dust, organic compounds, etc.).
• Low-temperature and condensing boilers
• Gas water heaters
These systems can be adjusted using the testo 330 LX and
checked for compliance with the applicable limit values.
The following tasks can also be carried out with the testo 330 LX:
• Regulating the O2-, CO- and CO2-, NO-, NOx- values in
furnace systems for the purpose of ensuring optimal operation.
• Draught measurement.
• Measuring and regulating the gas flow pressure in gas water
heaters.
• Measuring and optimising the flow and return temperatures of
heating systems.
• CO and CO2 environment measurement.
• Detection of CH4 (methane) and C3H8 (propane).
• The testo 330 LX can be used for measurements on CHP
stations in accordance with the first German Federal Immission
Control Ordinance (BImschV).
• In principal, the CO sensor can also be used for
measurements on CHP stations. If you should carry out
more than 50 measurements on CHP stations per year,
please contact your nearest testo service centre or send the
testo 330 for checking to testo Service.
A worn NOx filter of the CO sensor can be ordered as a
spare part (order no. 0554 4150) and replaced.
Testo guarantees the functionality of its products when
used in accordance with their intended purpose. This
guarantee does not apply to features of Testo products in
combination with unauthorised third-party products.
Competitor products are not authorised by Testo.
As is common practice, Testo generally excludes support,
warranty or guarantee claims relating to functionality that

3 Specifications
10
has not been guaranteed by Testo as part of the product
offered. Claims shall also be excluded in the event of
improper use or handling of the products, e.g. in
combination with unauthorised third-party products.
testo 330 LX must not be used:
• as a safety (alarm) instrument
testo 330 LX with Bluetooth®option:
The use of the wireless module is subject to the
regulations and stipulations of the respective country of
use, and the module may only be used in countries for
which a country certification has been granted. The user
and every owner has the obligation to adhere to these
regulations and prerequisites for use, and
acknowledges that the re-sale, export, import etc. in
particular in countries without wireless permits, is his
responsibility.
3.2. Technical data
3.2.1. Examinations and licenses
As declared in the certificate of conformity, this product complies
with Directive 2014/30/EC.
3.2.2. Bluetooth®module (option)
Country Comments
Canada Contains FCC ID: 4957A-MSR
Product IC ID: 6127B-2016T330
IC Warnings
USA Contains FCC ID: RFRMS
FCC ID: WAF-2016T330
FCC Warnings

3 Specifications
11
Europa + EFTA
EU countries:
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany
(DE), Estonia (EE), Finland (FI), France (FR), Greece
(GR), Ireland (IE), Italy (IT), Latvia (LV), Lithuania (LT),
Luxembourg (LU), Malta (MT), Netherlands (NL), Austria
(AT), Poland (PL), Portugal (PT), Romania (RO),
Sweden (SE), Slovakia (SK), Slovenia (SI), Spain (ES),
Czech Republic (CZ), Hungary (HU), United Kingdom
(GB), Republic of Cyprus (CY).
EFTA countries:
Iceland, Liechtenstein, Norway, Switzerland
Japan
Japan Information
Bluetooth SIG
List
TÜV TÜV by RgG 310
TÜV approved:
EN50379-1, -2, -3 / VDI4206-1
Warnings
This instrument complies with Part 15B of the FCC Rules and Industry Canada RSS-210
(revision 8). Commissioning is subject to the following two conditions:
(1) This instrument must not cause any harmful interference and
(2) this instrument must be able to cope with interference, even ifthis has undesirable
effects on operation.
Cet appareil satisfait à la partie 15B des directives FCC et au standard Industrie Canada
RSS-210 (révision 8). Sa mise en service est soumise aux deux conditions suivantes :
(1) cet appareil ne doit causer aucune interférence dangereuse et

3 Specifications
12
(2) cet appareil doit supporter toute interférence, y compris des interférences qui
provoquerait des opérations indésirables.
FCC Warnings
Information from the FCC (Federal Communications Commission)
For your own safety
Shielded cables should be used for a composite interface. This is to ensure continued
protection against radio frequency interference.
FCC warning statement
This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This
equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference by one or more of the following
measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver
is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Caution
Changes or modifications not expressly approved by the party responsible for compliance
could void the user's authority to operate the equipment. Shielded interface cable must be
used in order to comply with the emission limits.
Warning
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may
cause undesired operation.
Japan Information
当該機器には電波法に基づく、技術基準適合証明等を受けた特定無線設備を装
着している。
3.2.3. Declaration of Conformity
You can find the EU declaration of conformity on the Testo
homepage www.testo.com under the product-specific downloads.

3 Specifications
13
3.2.4. Measuring ranges and resolution
Parameter Measuring range Resolution
O20...21 Vol.% 0.1 vol.%
CO 0...4000 ppm 1 ppm
CO, H2-comp.1 0...8000 ppm 1 ppm
AmbCO
through flue gas
probe
0...2000 ppm 1 ppm
AmbCO with
probe 0632 3331
0...500 ppm 1 ppm
NO 0...3000 ppm 1 ppm
Draught -9.99...40 hPa 0.01 hPa
∆P 0...300 hPa 0.1 hPa
Temperature -40...1200 °C 0.1°C (-40.0...999.9 °C)
1°C (rest of range)
Efficiency net 0...120 % 0.1 %
Flue gas loss 0...99.9 % 0.1 %
AmbCO2 with
probe 0632 1240
0...1 vol.
0...10000 ppm
-
Gas leak testing
with probe
0632 3330
0...10000 ppm
CH4 / C3H8
-
1above the sensor protection threshold: Resolution 1 ppm (up to max. 30,000
ppm)

3 Specifications
14
3.2.5. Accuracy and response time
Parameter Accuracy Response
time
O2±0.2 vol.% < 20s (t90)
CO ±20 ppm (0...400 ppm)
±5% of mv (401...2000 ppm)
±10% of mv (2001...4000 ppm)
< 60s (t90)
CO, H2-comp. ±10 ppm or ±10 % of mv2
(0...200 ppm)
±20 ppm or ±5 % of mv2
(201...2000 ppm)
±10% of mv (2001...8000 ppm)
only testo 330-2:
8000...30000 ppm
(automatic dilution)
< 60s (t90)
AmbCO through
flue gas probe
±10 ppm (0...100 ppm)
±10 % of mv (101….2000 ppm)
< 35s (t90)
AmbCO with
0632 3331
±5 ppm (0...100 ppm)3
±5 % of mv ( >101 ppm)
approx. 35s
(t90)
NO ±2 ppm (0…39,9 ppm)
±5 % of mv (40….2000 ppm)
±10 % of mv (2001...3000 ppm)
< 30s (t90)
Draught4±0.02 ppm or ±5% of mv2
(-0.50...0.60 hPa)
± 0.03 hPa (0.61...3.00 hPa)
±1.5 % of mv (3.01...40.00 hPa)
-
∆P ± 0.5 hPa (0.0...50.0 hPa)
±1 % of mv (50.1...100.0 hPa)
±1.5 % of mv (rest of range)
-
Temperature ± 0.5 °C (0.0...100.0 °C)
±0.5 % of mv (rest of range)
probe
dependent
Efficiency - -
2higher value is valid
3at 10...30 °C, outside this range additionally ±0.2 % of mv / °C
4with fine draught measurement option: Measuring range 0...100, 0Pa,
resolution 0.1Pa

3 Specifications
15
Parameter Accuracy Response
time
Flue gas loss - -
AmbCO2,
through
0632 1240
±75 ppm + 3 % of mv
(0...5000 ppm)
±150 ppm + 5 % of mv
(5001...10000 ppm)
approx. 35s
(t90)
Gas leak testing
with 0632 3330
- < 2s (t90)
3.2.6. Other instrument data
Flue gas analyser
Characteristic Values
Storage / and
transport
temperature
-20...50 °C
Operating
temperature
-5...45 °C
Ambient humidity 0…90 % rH, not condensing
Power supply Battery: 3.7 V / 2.6 Ah
Mains unit: 6 V/1.2 A
Protection class IP40
Weight 600 g (excluding battery)
Dimensions 270 x 90 x 65 mm
Memory 500,000 readings
Display Graphic colour display, 240 x 320 pixels
Flue gas
overpressure
max. 50 mbar
Negative pressure max. 80 mbar
Gas leak testing
probe
visual indication (LED)
audible indication by buzzer
Storage
temperature battery
±0...35 °C
Battery charge time approx. 5-6 h

3 Specifications
16
Characteristic Values
Battery operation
time
6h (pump on, 20 °C ambient temperature)
Bluetooth® (option) Range < 10 m
Warranty Measuring instrument: 48 months
LX-sensors O2, CO: 60 months
Other sensors: 24 months
Flue gas probe: 48 months
Thermocouple: 12 months
Battery: 12 months
Further warranty terms: see website
www.testo.com/warranty

4 Product description
17
4Product description
4.1. Case 0516 3300 (accessory)
Recommended for stowing away the measuring instrument and
accessories (example)
4.1.1. Bottom level view
1 Sealing clip
2 testo 330-1 /-2 LX flue gas analyzer
3 Repository for printer accessories
• Spare batteries for IRDA printer
• 1 roll of spare thermal paper (0554 0568)

4 Product description
18
4 Repository for printer
• IRDA printer (0554 0549)
• Bluetooth®/IRDA printer (0554 0620)
5. Instruction manual
6 Lock
7 Probes
• Flue gas probe (e.g. 0600 9741)
• Pitot tube for heating check (0635 2050)
8 Large storage compartment
• Mains unit for testo 330-1 /-2 LX (0554 1096)
• Differential temperature set (0554 1208)
• Spare dirt filter (0554 0040)
9Round storage compartment
• Hose connection set with pressure adapter (0554 1203)
4.1.2. Top level view
1 Soot pump set (0554 0307)
2 Storage compartment
• Fine pressure probe (0638 0330)

4 Product description
19
3 Storage compartment
• Capillary hose set for fine pressure probe (0554 1215)
• Connecting cable for surface probe (0430 1215)
4 Combustion air temperature probe (0600 9787)
5. Surface temperature probe Type K (0604 0994)
4.2. Case 0516 3301 (accessory)
Recommended for stowing away the measuring instrument and
accessories (example)
4.2.1. Bottom level view
1 Fine pressure probe (0638 0330)
2 testo 308 smoke tester (0632 0308)

4 Product description
20
4.2.2. Middle level view
1 Sealing clip
2 testo 330-1 /-2 LX flue gas analyzer
3 Repository for printer accessories
• Spare batteries for IRDA printer
• 1 roll of spare thermal paper (0554 0568)
4 Repository for printer
• IRDA printer (0554 0549)
• Bluetooth®/IRDA printer (0554 0620)
5. Instruction manual
6 Lock
7 Probes
• Flue gas probe (e.g. 0600 9741)
• Pitot tube for heating check (0635 2050)
Table of contents
Other TESTO Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Ametek
Ametek Chandler Engineering 4207D operating manual

Chauvin Arnoux
Chauvin Arnoux F23 user manual

SPM
SPM intellinova parallel MB INS MB 4V Installation and operation instruction

Tenmars
Tenmars TM-740 user manual

ELKHART BRASS
ELKHART BRASS Cobra Max 81910 Installation, operating, & maintenance instructions

Eaton
Eaton PXM4K Series quick start guide