TESTO 420 User manual

1 Contents
3
1Contents
1Contents ...................................................................................................3
2Safety and the environment....................................................................4
2.1. About this document........................................................................4
2.2. Ensure safety...................................................................................5
2.3. Protecting the environment..............................................................5
3Specifications ..........................................................................................5
3.1. Use ..................................................................................................6
3.2. Technical data .................................................................................6
3.2.1. Bluetooth module.............................................................................................6
3.2.2. General technical data .....................................................................................6
4Product description...............................................................................10
4.1. Overview........................................................................................10
4.1.1. Measurement setup .......................................................................................10
4.1.2. Overview of testo 420 .................................................................................... 11
5First steps ..............................................................................................13
6Using the product ..................................................................................15
6.1. Switching Bluetooth® on and off....................................................15
6.2. Settings for the measurement........................................................16
6.2.1. Damping (Gliding Average)............................................................................16
6.2.2. Hood Calibration ............................................................................................16
6.2.3. Zeroing Interval (Automatic Zeroing).............................................................. 16
6.3. Volume flow measurement setup ..................................................17
6.4. Measuring......................................................................................20
6.4.1. Volume flow measurement.............................................................................20
6.4.2. Pitot tube measurement.................................................................................20
6.4.3. Differential pressure measurement ................................................................21
6.5. Saving ...........................................................................................21
6.6. Transferring measurement data to the PC ....................................22
7Maintaining the product ........................................................................23
7.1. Cleaning the instrument.................................................................23
8Tips and assistance...............................................................................23
8.1. Questions and answers .................................................................23
8.2. Accessories and spare parts .........................................................24
9Authorizations........................................................................................25

2 Safety and the environment
4
2Safety and the environment
2.1. About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Symbols and writing standards
Representation
Explanation
Warning advice, risk level according to the
signal word:
Warning! Serious physical injury may occur.
Caution! Slight physical injury or damage to
the equipment may occur.
> Implement the specified precautionary
measures.
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be
followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu Elements of the instrument, the instrument
display or the program interface.
[OK] Control keys of the instrument or buttons of
the program interface.
... | ... Functions/paths within a menu.
“...” Example entries

3 Specifications
5
2.2. Ensure safety
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
> Do not perform contact measurements on non-insulated, live
parts.
> Do not store the product together with solvents. Do not use any
desiccants.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts
from Testo.
> Dangers may also arise from the systems being measured or
the measuring environment: Note the safety regulations valid in
your area when performing the measurements.
2.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
3Specifications
The testo 420 is used for volume flow measurements (main
application), Pitot tube measurements and pressure measurements
for air conditioning and ventilation systems. Thanks to its
interchangeable flow hoods, the testo 420 can be used for air inlets
and outlets of various sizes.
With the additional App (Android/iOS), the readings can be
conveniently displayed on a tablet or smartphone, and in addition a
measurement can be started, stopped and saved via the App.

3 Specifications
6
3.1. Use
3.2. Technical data
The use of the wireless module is subject to the regulations and
stipulations of the respective country of use, and the module may only
be used in countries for which a country certification has been granted.
The user and every owner has the obligation to adhere to these
regulations and prerequisites for use, and acknowledges that the re-
sale, export, import etc. in particular in countries without wireless
permits, is his responsibility.
3.2.1. Bluetooth module
Feature Values
Bluetooth Range >20 m (free field)
Bluetooth type LSD Science & Technology Co., Ltd
L Series BLE module (08 May 2013) based
on TI CC254X chip
Qualified Design ID B016552
Bluetooth radio
class
Class 3
Bluetooth company 10274
3.2.2. General technical data
Feature Values
Measurement
parameters
Temperature: °C / °F
Humidity: %RH / %rF / td°C / WB°C
Flow velocity : m/s / ft/min
Volume flow: m³/h / cfm / l/s
Pressure (absolute pressure): hPa / mbar / kPa
Pressure (differential pressure): Pa / hPa /
mbar / mmH2O / inH2O
Measuring cycle 1/sec
Interfaces Probe interface Mini DIN
Micro USB

3 Specifications
7
Measuring ranges Temperature: -20 to +60 °C / -4 to 140 °F
Humidity: 0 to 100% RH
Flow velocity: 0 to 14 m/s /
0 to 2750 ft/min
Volume flow: 40 to 4000 m³/h /
25 to 2300 cfm / 11 to 1100 l/s
Pressure (absolute pressure):
700 to 1100 hPa
Pressure (differential pressure): -
120 to +120 Pa
Resolution Temperature: 0.1 °C / 0.1°F
Humidity: 0.1% RH
Flow velocity: 0.01 m/s
Volume flow: 1 m³/h / 1 cfm
Pressure (absolute pressure): 0.1 hPa /
0.1 mbar / 0.01 kPa
Pressure (differential pressure): 0.001 Pa /
0.00001 hPa / 0.00001 mbar / 0.0001 mm
H2O / 0.000001 inH2O
Accuracy
(nominal temperatur
e 22°C/71.6°F)
Temperature: ±0.5 °C (0 to +70 °C) / ±0.8 °C
(-20 to 0 °C)
Humidity: ±1.8% RH +3% of m.v. at +22 °C
(5 to 80% RH) (longer high humidity application
can result in a temporary sensor drift)
Flow velocity: no accuracy specification since
calculated variable
Volume flow 1: ±3% of m.v. +12 m³/h at
+22 °C, 1013 hPa (85 to 3500 m³/h)
±3% +7 cfm (50 to 2,100 cfm)
Error absolute pressure compensation:
±0.04% of m.v. / hPa deviating from 1013 hPa
Pressure (absolute pressure): ± 3 hPa
Pressure (differential pressure): ±2% of m.v.
+ 0.5 Pa (at 22°C, 1013 hPa)
Error absolute pressure compensation:
±0.04% of m.v. / hPa deviating from 1013 hPa
1All accuracy specifications apply under laboratory conditions or with necessary
compensation (correction factor) with the standard hood 610x610 mm. Minimum
outlet size of 360x360 mm

3 Specifications
8
Temperature
coefficient
Humidity: ± 0.03% RH / K (deviating from
22°C, in the range 0 to 60 °C)
Volume flow: ± 0.02% of m.v. / K
(deviating from 22 °C, in the range 0 to 60 °C)
Pressure (absolute pressure): ± 0.02% of
m.v. / K (deviating from 22 °C, in the range
0 to 60 °C)
Pressure (differential pressure): ± 0.02% of
m.v. / K (deviating from 22 °C, in the range
0 to 60 °C)
Response time t90 Temperature: approx. 45 s
Humidity: approx. 15 s
Flow velocity: approx. 1 s
Volume flow: approx. 1 s
Pressure (absolute pressure): approx. 1 s
Pressure (differential pressure): approx. 1 s
Operating and
ambient conditions
Storage temperature: -20 to +60 °C / -4 to
140 °F
Operating temperature: -5 to +50 °C/+23 -
+122°F
Humidity: 0 to 100% RH
Pressure range: 800 to 1100 hPa
Housing /
measurement setup
Measuring instrument housing material: ABS
Body material: PP
Standard hood material: Nylon
Measuring instrument dimensions:
150x85x35 mm
Body dimensions: 490x970x610 mm
Dimensions of measurement setup with
standard hood: 610x970x610 mm
Weight of entire measurement setup
approx. 2900 g

3 Specifications
9
Power supply 4 x 1.5 V rechargeable/non-rechargeable
batteries
Type AA / alkaline manganese, mignon
Battery lifetime: approx. 40h (eroing interval
10 seconds, display illumination off,
Bluetooth off)
Display Type: Dot matrix
Dimensions: 3.5 inches
Directives,
standards and tests
EU guideline: 2014/30/EU
Warranty Duration: 2 years
Warranty terms: see website
www.testo.com/warranty

4 Product description
10
4Product description
4.1. Overview
4.1.1. Measurement setup
1 Volume flow hood (standard hood 610x610 mm)
2 Actuator for manual measurement
3 Measuring instrument testo 420
4 Measurement base with differential pressure cross
5 Integrated flow straightener

4 Product description
11
4.1.2. Overview of testo 420
1 Battery compartment, on rear of the instrument
2 Display
3 Control keys
4 Probe socket Mini-DIN (only for use on the measurement base)
5 Micro-USB port
6 Connection for pressure measurement
Instrument status icons:
Icon Meaning
Battery capacity
Bluetooth
/ / Measuring mode:
Pressure measurement, Pitot tube, volume
flow (air pressure from above into the hood /
suction outlet)
Actual Actual volume flow:
The current ambient conditions are used to
calculate the volume flow. The actual
barometric pressure is measured with the
internal sensor. In applications with volume
flow hood, the temperature is measured by
the integrated temperature/humidity sensor,
with Pitot tube measurement the actual
temperature needs to be input manually.

4 Product description
12
Icon Meaning
Standard Standard volume flow:
The standard settings for temperature and
barometric pressure (21 °C / 1013 hPa) are
used to calculate the volume flow.
K-factor Factor by which the current reading is
multiplied. Depends on which outlet
measurements are being taken at.
Pitot Tube factor
(P-factor)
The Pitot tube factor for Pitot tubes is
generally the same and must be entered:
•Pitot tubes from Testo: 1.00
•Pitot tubes from other manufacturers: the
Pitot tube factor can be found in the
instruction manual or you should ask your
supplier.
Control keys
Key Function
Menu
[►, ■]Holds / starts / stops a measurement
[ESC] Switches to the previous view / to the
measurement view
Saves the measured values
[►, ▲, ▼,◄] Navigation within the menu
[ ] Confirms a selection
Switch instrument on/off (press and hold
down)
Switch illumination on/off (press briefly)

5 First steps
13
5First steps
Inserting batteries/rechargeable batteries
1. Open the battery compartment.
2. Insert the batteries or rechargeable batteries (scope of delivery
includes 4x 1.5V Type AA/ LR6).
3. Close the battery compartment.
When not in use for a long period of time, take the
batteries/rechargeable batteries out.
Implementing settings
1. Press to access the menu.
2. Select the menu item you require using ►, ▲, ▼, ◄.
Key functions
Display Explanation
►, ▲, ▼,◄ Change parameter, select unit
Confirm input
1. Menu Level 2. Menu Level 3. Menu Level
Application Flow Hood K-factor
Actual/Standard
Pitot tube Duct
Pitot tube factor
Pitot Temperature
Actual/Standard
Pressure only --
Measuring Program
Single Measurment --
Continous
Measurement2
--
2Maximum of 15 minutes

5 First steps
14
1. Menu Level 2. Menu Level 3. Menu Level
Continous/Punctual
measurement (nur für
Staurohr)
measuring duration3
Memory New Folder --
T420 folder --
Display Volume Flow on/off
Diff.Pressure on/off
Temperature on/off
Velocity on/off
Humidity on/off
Abs.Pressure on/off
Device Settings Language English/German/Italia
n/French/Spanish
Backlight Auto Off On/Off
Auto Off On/Of
Bluetooth On/Off
Date&Time Date Format
Time Format
Set Date&Time
Gliding average 5 – 20 sec
Hood adjustment Supply Air
Exhaust Air
Zeroing int 1-20 sec
Reset Device -- --
3Maximum of 25 points and 1 minute per point.

6 Using the product
15
6Using the product
6.1. Switching Bluetooth® on and off
In order to be able to establish a connection via Bluetooth,
you need a tablet or smartphone with the Testo App testo
420 already installed on it.
You can get the App for iOS instruments in the App Store
or for Android instruments in the Play Store.
Information about compatibility can be found in the relevant
App Store.
Measurements can be performed and saved with App into
the instruments memory. Measuring mode and save
function on the instrument itself is not available during
Bluetooth communication is established with a mobile
device.
Switching on Bluetooth
1. Hold ▲down for 3 seconds.
- Once the Bluetooth icon is shown on the display, Bluetooth is
switched on.
- If no connection is established, Bluetooth switches off after 10
mins.
or
1. Press -> Device Settings -> Bluetooth, ►and using ▲/▼
-> select Off. Confirm with .
- Once the Bluetooth icon is shown on the display,
Bluetooth is switched on.
- If no connection is established, Bluetooth switches off after 10
mins.
Display Explanation
flashes There is no Bluetooth connection, or a
potential connection is being searched for.
is permanently
displayed
There is a Bluetooth connection.
is not displayed Bluetooth is disabled.

6 Using the product
16
6.2. Settings for the measurement
6.2.1. Damping (Gliding Average)
If the readings fluctuate widely, it is advisable to damp the
readings. The time range for the damping can be set manually
between 5-20 seconds.
1. Press , then select Device settings and Gliding average.
- The damping can be set between 5-20 seconds.
6.2.2. Hood Calibration
This input is provided for the filing of calibration data by the relevant
calibration laboratory. The flow hood-specific adjustment data can
be input manually for supply air and exhaust air and has a direct
effect on the measurement results. Input option of 0.001-9.999.
1. Press , then select Device settings and Hood adjustment.
- Calibration of the hood can be set for supply air and exhaust air.
6.2.3. Zeroing Interval (Automatic Zeroing)
The pressure sensor carries out automatic zeroing at regular
intervals. These intervals can be set via the automatic zeroing.
1. Press , then select Device settings and Zeroing int.
- The zeroing interval can be set between 1-20 seconds.

6 Using the product
17
6.3. Volume flow measurement setup
Standard hood
(610x610 mm, scope of delivery; 360x360, accessories)
1. Pull the lower end of the hood over the measurement base.
2. Attach the hood at two corners using the snap fasteners.
3. Tighten the closure.
4. Push the support rods through the hood, along the markings
and into the funnel in the measurement base.
5. Push the support rods on the top of the hood into the brackets.
- The hood is installed.

6 Using the product
18
Large hood
(1220x610 and 1220x305, accessories)
1. Install the aluminium frame and stretch the fabric hood over the
framework, so that the elastic band lies in the recess of the
frame. Make sure that the elastic band fits correctly, especially
at the corners.
2. Pull the lower end of the hood over the measurement base.
3. Attach the hood at two corners using the snap fasteners.
4. Tighten the closure.
5. Push the support rods through the hood, along the markings
and into the funnel in the measurement base.
6. Push the support rods on the top of the hood into the brackets.
- The hood is installed.

6 Using the product
19
Attaching the measuring instrument
1. Push the testo 420 completely into the instrument holder, pay
attention to the catch on the right-hand and left-hand side in the
bracket.

6 Using the product
20
6.4. Measuring
6.4.1. Volume flow measurement
✓The volume flow hood is fitted.
1. Switch on the instrument.
2. In the instrument settings, set the application volume flow hood
and also the required measurement program:
Single measurement or Continuous measurement.
3. Press ►, ■ on the testo 420, or the trigger on the measurement
base to hold or start and stop the measurement.
4. Press to save the measurement data. Unsaved
measurement data is lost once you carry out the next
measurement.
- The target folder and file name are displayed, confirm with
to save the measurement data under this name and in the
selected folder.
6.4.2. Pitot tube measurement
1. Remove the testo 420 from the measurement base.
2. Connect the hoses to the testo 420 and to the Pitot tube.
3. Press -> Application -> Pitot tube and there set the duct
geometry, the Pitot tube factor and the temperature, and
choose between actual and standard.
4. Select the required measurement program.
5. Carry out the measurement.
6. Press to save the measurement data. Unsaved
measurement data is lost once you carry out the next
measurement.
- The target folder and file name are displayed, confirm with
to save the measurement data under this name and in the
selected folder.
Other manuals for 420
4
Table of contents
Other TESTO Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Hioki
Hioki 3532-50 instruction manual

Winland Electronics
Winland Electronics EnviroAlert EA800-IP troubleshooting guide

McCrometer
McCrometer Water Specialties ML04 OPERATION AND MAINTENANCE MANUAL PARTS LIST

Vega
Vega VEGAPULS 63 Quick setup guide

Seikom Electronic
Seikom Electronic NLSW45-4 Installation and operating instruction

Gossen MetraWatt
Gossen MetraWatt MINITEST PRO operating instructions