manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Thermo Scientific
  6. •
  7. Measuring Instrument
  8. •
  9. Thermo Scientific Orion AQUAfast II User manual

Thermo Scientific Orion AQUAfast II User manual

Other Thermo Scientific Measuring Instrument manuals

Thermo Scientific Orion Star User manual

Thermo Scientific

Thermo Scientific Orion Star User manual

Thermo Scientific Oakton pH 450 User manual

Thermo Scientific

Thermo Scientific Oakton pH 450 User manual

Thermo Scientific NanoDrop Lite Instruction Manual

Thermo Scientific

Thermo Scientific NanoDrop Lite Instruction Manual

Thermo Scientific Orion 320 User manual

Thermo Scientific

Thermo Scientific Orion 320 User manual

Thermo Scientific RadEye B20 User manual

Thermo Scientific

Thermo Scientific RadEye B20 User manual

Thermo Scientific NITON XL3t 500 User manual

Thermo Scientific

Thermo Scientific NITON XL3t 500 User manual

Thermo Scientific NanoDrop One User manual

Thermo Scientific

Thermo Scientific NanoDrop One User manual

Thermo Scientific Orion Star A211 User manual

Thermo Scientific

Thermo Scientific Orion Star A211 User manual

Thermo Scientific 48i User manual

Thermo Scientific

Thermo Scientific 48i User manual

Thermo Scientific SOLA II Flare User manual

Thermo Scientific

Thermo Scientific SOLA II Flare User manual

Thermo Scientific SENTINEL RadEye G User manual

Thermo Scientific

Thermo Scientific SENTINEL RadEye G User manual

Thermo Scientific AutoPILOT PRO User manual

Thermo Scientific

Thermo Scientific AutoPILOT PRO User manual

Thermo Scientific HAAKE User manual

Thermo Scientific

Thermo Scientific HAAKE User manual

Thermo Scientific Star A221 User manual

Thermo Scientific

Thermo Scientific Star A221 User manual

Thermo Scientific EUTECH INSTRUMENTS OAKTON 450 Series User manual

Thermo Scientific

Thermo Scientific EUTECH INSTRUMENTS OAKTON 450 Series User manual

Thermo Scientific Orbitrap Fusion User manual

Thermo Scientific

Thermo Scientific Orbitrap Fusion User manual

Thermo Scientific RadEye B20 How to use

Thermo Scientific

Thermo Scientific RadEye B20 How to use

Thermo Scientific 42i User manual

Thermo Scientific

Thermo Scientific 42i User manual

Thermo Scientific Orion Star User manual

Thermo Scientific

Thermo Scientific Orion Star User manual

Thermo Scientific RadEye B20 User manual

Thermo Scientific

Thermo Scientific RadEye B20 User manual

Thermo Scientific 2210AM User manual

Thermo Scientific

Thermo Scientific 2210AM User manual

Thermo Scientific Evolution 300 User manual

Thermo Scientific

Thermo Scientific Evolution 300 User manual

Thermo Scientific Orion Star User manual

Thermo Scientific

Thermo Scientific Orion Star User manual

Thermo Scientific Orion Versa Star Pro User manual

Thermo Scientific

Thermo Scientific Orion Versa Star Pro User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Phosphate LR 0.05 - 4.0 mg/l Method Preparation
For USEPA regulatory use, valid range 0.15 - 4.0 mg/l PO4
0.0.0
RESULT
RESULTRESULT
PLr
Perform zero calibration (see “Operation”).
Add one PHOSPHATE LR No.1 tablet straight from the
foil to the 10 ml sample, and crush using a clean stir rod.
Add one PHOSPHATE LR No.2 tablet straight from the
foil to the same sample and crush using a clean stir rod.
Allow to dissolve completely, cap the vial, and align the
∇and ∆marks.
Wait for a color reaction time of 10 minutes!
Press the "zero/test" key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l PO4.
Measuring tolerance: ±0.1 mg/l PO4
Method notes
Observe application options, analysis regulations and matrix effects of
methods. Reagent tablets are designed for use in chemical analysis
only and should be kept well out of the reach of children.
If necessary, request material safety data sheets.
Ensure proper disposal of reagent solutions.
1. Always adhere to the sequence of tablet addition.
2. The aqueous sample solution should have a pH of between 6 and 7.
Only phosphate (V) ions react.
3. Interferences:
High concentrations of Cu, Ni, Cr(III), V(V) and W(V) causes interfering
color. Silicates do not interfere with the analysis (masked by citric acid
in the tablet).
Notes
Conversions
mg/l PO4x 0.75 = mg/l P2O5
mg/l PO4x 0.33 = mg/l P
Optics: LED: λ= 660 nm
Battery: 9 V block battery (life = approx. 600 tests)
Auto-OFF: Auto unit switch-off occur approx. 15 minutes
after a key was last pressed.
Ambient conditions
:
5-40°C
30-90% rel. humidity (non-condensing)
Compliance:
DIN EN 55 022, 61 000-4-2, 61 000-4-8,
50 082-2, 50 081-1, DIN V ENV 50 140, 50 204
FCC Part 15 Class A
ICES – 003 Issue 2
Technical data
Operation
User messages
METHOD
METHODMETHOD
METHOD
METHODMETHOD
0.0.0
RESULT
RESULTRESULT
÷Err
EOI
-Err
PLr
Switch the unit on using the "power" switch
This display shows the method. Press the "mode" key
until the desired method is displayed.
Fill a clean vial with the sample up to the 10 ml mark,
screw the cap on, and place in the sample chamber with
the ∇vial mark aligned with the ∆housing mark.
Press the "zero/test" key.
The method symbol flashes for approx. 3 seconds.
Confirms zero calibration.
After zero calibration is completed, remove the vial from
the sample chamber.
The characteristic color starts to appear after the addition
of the reagent tablet(s) (see "Method Preparation").
Cap the vial again and place in the sample chamber with
the ∇and ∆marks aligned.
Press the "zero/test" key.
The method symbol flashes for approx. 3 seconds.
The result appears in the display.
Repeat the analysis:
Press the "zero/test" key once again.
New zero calibration:
Press the "mode" key until the desired method symbol
appears in the display again.
Light absorption too great. Reason - e.g. soiled lens.
Measuring range exceeded or excessive turbidity.
Result below measuring range limit.
Replace 9 V battery immediately; no further analysis are
possible.
Avoiding errors in photometric measurements
1. Thoroughly clean vials, caps and stir rod after each analysis in order
to prevent carry-over errors. Even minute reagent residues lead to
incorrect measurements. Use the supplied brush for cleaning.
2. Ensure that the outer walls of the vials are dry and clean before performing
the analysis. Fingerprints or water droplets on the light entry surfaces of
the vials lead to incorrect measurements.
3. “Zero calibration” and “Test” must be performed using the same vial,
since different vials can possess slightly different tolerances.
4. For “Zero calibration” and “Test”, ensure that the vial is always positioned
in the sample chamber in such a way that the graduation with the white
triangle points toward the marking on the housing.
5. Always perform “Zero calibration” and “Test” with closed vial.
6. Bubbles on the inside walls of the vial can lead to incorrect
measurements.
To prevent this, cap the vial and remove the bubbles by swirling the vial
before performing the test.
7. You must prevent water from penetrating into the sample chamber. The
entry of water into the housing of the photometer can destroy electronic
components and lead to corrosion damage.
8. Soiling of the lens (LED and photosensor) in the sample chamber leads
to incorrect measurements.
Check - and if necessary clean - the light entry surfaces of the sample
chamber at regular intervals. Clean using a moist cloth and cotton balls.
9. Always add the reagent tablets to the sample straight from the foil without
touching them with your fingers.
10. Major temperature differentials between the photometer and the
environment can lead to incorrect measurements - e.g. due to the
formation of condensate in the area of the lens or on the vial.
Specified tolerances at T = 20 °C.
11. For best results pipette samples.
Calibration Standards
Standards for calibration should be prepared similar to samples.
LO BAT
zero
test
power
zero
test
zero
test
Correct filling of the vial
correct wrong
10 ml 10 ml
Orion
AQUAfast
®
II Photometer Phosphate LR
The unit can be reset to the factory calibration as follows:
Press and hold both "mode" and "zero/test" together.
Switch the unit on using the "power" key. Release "mode"
and "zero/test" keys after approx. 1 second.
The following messages will alternate in the display.
The unit is reset to factory settings.
(SEL stands for Select)
or:
The unit operates with a calibration performed by the
user. (If the user calibration is to be retained, switch the
unit off using the "power" key.)
Factory calibration is activated by pressing the "mode"
key. The following messages will alternate in the display:
Switch the unit off using the "power" key.
User calibration : cAL
Factory calibration : CAL
CAL
SEL
SEL
CAL
cAL
SEL
User notes
E 10
E 70
E 71
Calibration mode
CAL
PLr
CAL
RESULT + x
RESULT + xRESULT + x
0.0.0
RESULT
RESULTRESULT
CAL
: :
Notes
CAL
cAL
METHOD
METHODMETHOD
METHOD
METHODMETHOD
Recommended calibration values
Phosphate: between 1 and 3 mg/l PO4
CAL
Technical changes without notice.
Printed in Germany.
Calibration factor “out of range”
Factory calibration incorrect / deleted
User calibration incorrect / deleted
Press and hold "mode" key.
Switch unit on using "power" key.
Release "mode" key after approx. 1 second.
These messages will alternate in the display.
If necessary, press "mode" key until the desired method
alternates with CAL.
Perform zero calibration as described.
Press the "zero/test" key.
The method symbol flashes for approx. 3 seconds.
These messages will alternate in the display.
Place the standard to be used in the sample chamber
with the ∇and ∆marks aligned (see "Method
Preparation"). Press the "zero/test" key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display alternating with CAL.
If the result corresponds to the value of the standard used
(within the allowed tolerance), exit calibration mode by
pressing the "power" key.
Pressing the "mode" key once increases the displayed
result by 1 digit.
Pressing the "zero/test" key once decreases the displayed
result by 1 digit.
Continue pressing the keys until the displayed result
corresponds to the value of the standard used.
If you press the "power" key twice, the new correction
factor is calculated and stored in the user calibration level.
Confirms calibration (3 seconds).
Factory calibration active.
Calibration has been set by the user.
233314-001 Rev.E
zero
test
power
mode
zero
test
mode
zero
test
power
mode zero
test
power
power
mode