Problem Possible Cause Corrective Action
Efficiency reading incorrect NET efficiency selected.
Ambient temperature probe not
plugged in to T2.
Incorrect fuel selected.
Select GROSS efficiency. See page
13.
Plug ambient probe into T2. See
page 14.
Select the proper fuel for the equip-
ment being tested. See page 13.
Readings are erratic when working
on oil fired equipment. Oil filter not installed or installed
incorrectly (716 only). Make sure the optional sulfur filter
(A773) is installed. See Appendix B.
One or all of the following parame-
ters; Ratio, CO air free, excess air,
and efficiency read and print dashes.
Measured values are such that the
calculated values of these parame-
ters are out of range.
Redo combustion test. Since these
are calculated values, the measure
values must be within certain levels
for these to display. If the measured
oxygen level is above 19.9% these
parameters won’t read.
These parameters might not display
or be applicable in some tests.
Pressure prints as “N/A” on my com-
bustion analysis print out. During combustion analysis if a pres-
sure measurement is not being made
this parameter will print as not being
used.
Perform the combustion test and
also connect the manometer and
monitor pressure.
Battery will not charge or hold a
charge. Defective charger or battery. Replace the charger or battery.
Send to TPI for service.
Beeping noise heard during charg-
ing. Defect in charging circuit or shorted
battery. Disconnect from the charger and
contact TPI at 800-368-5719.
Analyzer won’t turn off Oxygen and/or carbon monoxide lev-
els outside limits. Allow the analyzer to purge longer.
Appendix C: ERROR CODES & TROUBLESHOOTING (Continued)
Pressure sensor will not zero. Pressure sensor needs to be reset. Send to TPI for calibration.
42
Charger Socket Connection for 220V/115V charger
T1 Socket Connection for thermocouple plug on flue probe
Connection for any 'K' type thermocouple probe
T2 Socket Connection for ambient 'K' type thermocouple probe
Connection for any 'K' type thermocouple probe
Gas Sample Port Connection for Mini Pump Protection Filter and Flue
Probe
P (+) Port Connections for Pressure Tubing
P (-) Port
USB Port Connection for optional A807 cable for communication
to a PC or connection for A806 combustible gas leak
detection probe.
Top View
7
Charger
Socket
USB Port
P(+)
Port
P(-)
Port
Gas
Sample
Port
T1 Socket
T2 Socket