SPECIFICATIONS
Range: –58 to 1999 °F (–50 to 1300 °C)
Resolution: 0.1° or 1° (selectable)
Accuracy: ±0.3% + 1°C
Sampling Rate: 2.5 times per second
Display: 3½ digit LCD, ⅝ inches high (1.5cm)
Fail-sa e: Low battery indicator
Power: 9 volt alkaline battery
Accessories: Type-K thermocouple, rubberized case and a flip-out stand
FRONT PANEL QUICK REFERENCE
ON/OFF: Turns power on and off
MAX: ecords and updates maximum values
HOLD: “Freezes” the display
0.1°: Selects resolution to 0.1°
1°: Selects resolution to 1°
°C/°F: Selects the temperature to be displayed in either Fahrenheit or Celsius
OPERATION
1. Insert the probe plug into the receptacle located on the front of the thermometer.
The plug and receptacle are keyed with a small and a large pin. Make certain
that the plug is inserted properly.
2. Press the ON/OFF button to turn the unit on.
3. Press the °C/°F button to display the temperature in Fahrenheit (°F will appear
on the display) or Celsius (°C will appear on the display).
4. Set the thermometer to desired resolution 0.1° or 1°.
5. Place the probe tip in contact with the material to be measured and read the
temperature on the display.
MAX MODE
In the MAX mode, the thermometer will display the highest temperature reading
achieved. It does not display the current temperature if it falls below the MAX reading.
CURRENT TEMPERATURE
DISPLAY READS
42° 42°
60° 60°
40° 60°
Example: In the MAX mode
Press the MAX button (MAX will appear on the display). The thermometer will display
the maximum temperature reading on the LCD display. Press the MAX button again to
exit MAX recording mode.
HOLD
Press the HOLD button to “freeze” the display at the current reading (HOLD will
appear in the display). Press the HOLD button again to return to the current reading.
PROBES
The probe supplied are ultra-fast response Type K thermocouple naked bead probes
with 4’ leads and teflon insulation. Maximum operating temperature range for this
thermocouple is 500°F (260°C).
An accessory stainless steel triple purpose probe with handle is available. An
accessory surface probe is also available (see accessories).
Longer leads and extension may be used. The lead effect for 164 feet (50 meters) is
generally less the 0.2°C with a limit of error of 2.2°C. Type-K wire and Type-K
connectors must be used.
RECORDER OUTPUT
A voltage proportional to the display reading appears at the recorder OUTPUT socket
located on the right side of the unit. Output is via a standard 3-pole 3.5 mm jack.
Output speci ications are as ollows:
1mV output per degree in the 0.1° display mode
0.1 mV output per degree in the 1° display mode
Output examples:
Mode Readings Output
0.1° 26.5° 26.5 mV
1° 1500° 150.0 mV
OFFSET ADJUSTMENT (Fine tuning)
The offset controls are set at the factory to allow for the variations found in standard
thermocouples. By adjusting the offset controls, found on the side of the unit, you can
optimize measurement accuracy for a particular thermocouple at a particular
temperature.
This unit has been calibrated Traceable
®
to standards provided by the National
Institute of Standards and Technology (NIST). While offset adjustments are possible,
they will interfere with the factory calibration. Traceable
®
recalibration must be made
by the factory (see ecalibration).
BENCH STAND/WALL MOUNT
This unit is supplied with a bench stand that is on the back of the case. Locate the
rectangular slots on the raised portion of the back of the unit. With your thumb and
forefinger, lift the stand out. To close the stand, simply snap it shut.
This unit may be wall mounted by using a screw. Set the screw into the wall at the
location desired, the head of the screw will need to slip into the receptacle on the back
of the unit, so do not set the screw flush to the wall. Once the screw has been properly
set, hang the unit in place by sliding the receptacle on the back of the unit over the
head of the screw.
ACCESSORIES
Control Cat. No. 4014—
Stainless Steel Triple Purpose Probe, probe diameter
1
/
8
" (0.32cm), probe length 6"
(15.2cm), overall length 9" (22.9cm), cable length 50" (127cm).
Control Cat. No. 4008—
Surface Probe, flat disk on end has a diameter of 0.39" (1cm), overall length 9"
(23cm), cable length 36" (91.4cm).
Control Cat. No. 4028—
Ultra-fast response, naked bead thermocouple. (identical to probe supplied with unit)
cable length 48" (122cm).
Control Cat. No. 8039—
Low-Temperature Probe - Stainless-steel with handle; triple purpose (liquids, air/gas,
and semi-solids), Dimensions: diameter 0.17 inch; stem length 12 inches; overall
length 17 inches.
Control Cat. No. 8613—
High-Temperature Probe - Ten-foot-long 0.19-inch diameter braided metal wire cable
with smooth tip measures -73 to 982°C continuous or 1093°C short-term use.
DISPLAY MESSAGES
OL indicates that no probe is present indicates that the battery is low and needs
replacement (see batter replacement).
ALL OPERATIONAL DIFFICULTIES
If this thermometer does not function properly for any reason, please replace the
battery with a new high quality battery (see “Battery eplacement” section). Low
battery power can occasionally cause any number of “apparent” operational difficulties.
eplacing the battery with a new fresh battery will solve most difficulties.
TRACEABLE
®
DIGITAL THERMOMETER WITH RECORDER OUTPUT INSTRUCTIONS
Traceable
®
is a registered trademark of Control Company ©2008 Control Company. 92-4134-00 ev. 1 052008
BATTERY REPLACEMENT:
A “ ” symbol will appear on the LCD display when the battery needs replacement.
Before replacing the battery turn the unit off and unplug the probe. emove the three
screws from the back of the thermometer and lift off the back of the case. (NOTE:
When the top middle screw is removed, the 2-piece stand with come off the case).
emove the exhausted battery and replace it with a new 9-colt alkaline battery.
eplace the back of the thermometer, replace the 2-piece stand, reinstall the three
screws and tighten securely.
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E- ail s[email protected]
www.control3.co
Control Co pany is ISO 9001 Quality-
Certified by DNV and ISO 17025 accredited
as a Calibration Laboratory by A2LA.

Other Traceable Thermometer manuals

Traceable RTD PLATINUM 4132 User manual

Traceable

Traceable RTD PLATINUM 4132 User manual

Traceable 4351 User manual

Traceable

Traceable 4351 User manual

Traceable POCKET ULTRA User manual

Traceable

Traceable POCKET ULTRA User manual

Traceable 4045 User manual

Traceable

Traceable 4045 User manual

Traceable WORKHORSE 4425 User manual

Traceable

Traceable WORKHORSE 4425 User manual

Traceable 4800 User manual

Traceable

Traceable 4800 User manual

Traceable 4770 User manual

Traceable

Traceable 4770 User manual

Traceable 4241 User manual

Traceable

Traceable 4241 User manual

Traceable 90002-03 User manual

Traceable

Traceable 90002-03 User manual

Traceable 4247 User manual

Traceable

Traceable 4247 User manual

Traceable 4482 User manual

Traceable

Traceable 4482 User manual

Traceable 9446075 User manual

Traceable

Traceable 9446075 User manual

Traceable 98767-45 User manual

Traceable

Traceable 98767-45 User manual

Traceable 4470 User manual

Traceable

Traceable 4470 User manual

Traceable 4344 User manual

Traceable

Traceable 4344 User manual

Traceable 5651TR User manual

Traceable

Traceable 5651TR User manual

Traceable 4330 User manual

Traceable

Traceable 4330 User manual

Traceable RTD PLATINUM User manual

Traceable

Traceable RTD PLATINUM User manual

Traceable 90000-39 User manual

Traceable

Traceable 90000-39 User manual

Traceable ULTRA User manual

Traceable

Traceable ULTRA User manual

Traceable LOLLIPOP 4371 User manual

Traceable

Traceable LOLLIPOP 4371 User manual

Traceable 4480 User manual

Traceable

Traceable 4480 User manual

Traceable 4482 User manual

Traceable

Traceable 4482 User manual

Traceable 4351 User manual

Traceable

Traceable 4351 User manual

Popular Thermometer manuals by other brands

Xylem ebro TFI-54 operating instructions

Xylem

Xylem ebro TFI-54 operating instructions

Dwyer Instruments IR2 series manual

Dwyer Instruments

Dwyer Instruments IR2 series manual

Hama TC210 operating instructions

Hama

Hama TC210 operating instructions

Kaz V906G-CAN instructions

Kaz

Kaz V906G-CAN instructions

Fluke VT02 user manual

Fluke

Fluke VT02 user manual

ADTEMP 419 quick start guide

ADTEMP

ADTEMP 419 quick start guide

Jet 310016 Operation manual

Jet

Jet 310016 Operation manual

Fluke 54 Series II user manual

Fluke

Fluke 54 Series II user manual

geo-FENNEL FT 1300-1 user manual

geo-FENNEL

geo-FENNEL FT 1300-1 user manual

Extech Instruments RH200W user manual

Extech Instruments

Extech Instruments RH200W user manual

AVYCON AVT BT130MT-B manual

AVYCON

AVYCON AVT BT130MT-B manual

BSi BT-61 Operation manual

BSi

BSi BT-61 Operation manual

Kinsa QuickCare Thermometer Instructions for use

Kinsa

Kinsa QuickCare Thermometer Instructions for use

Topcom DIABLO 100 user guide

Topcom

Topcom DIABLO 100 user guide

Radio Shack Wireless BBQ thermometer quick guide

Radio Shack

Radio Shack Wireless BBQ thermometer quick guide

AccuMac AM1950 user guide

AccuMac

AccuMac AM1950 user guide

Cosori CMT-R161S user manual

Cosori

Cosori CMT-R161S user manual

REED R2020 instruction manual

REED

REED R2020 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.