
MS 32/1/B R 03/03 8
OPERATION AND ADJUSTMENT HF 32 AND HF 85
Adjustment of the safety trip point
Essential conditions : - The load cell must be correctly installed and wired.
- The hook of the lifting system should be free of any load.
- A successfull operating test
Equipment required : - A load (“Pm”) equivalent to 100 % of the working load limit
of the lifting system .
- An additional load equivalent to 10 % of « Pm ».
ACTION EXPLANATION
1Lift the two loads
(100% + 10% of Pm)
by +/- 10 cm.
The additional 10 % load represents the overload value permitted under the
European directives.
If an overload situation is detected in the wire rope ( without lifting the loads)
the preadjustment of the load cell in the factory is too low for your lifting
system.
Solution: release the tare screw (F - see fig. 4) one turn at a time and restart
at the beginning of point 1.
2To adjust the safety
trip point.
Use the tare screw (F).
Warning : incorrect adjustment may damage the circuit – breaker.
Do not forget the locking nut (G). The switch is a normally closed contact
(NC), requiring the opening of the switch to set off an overload condition.
Turn clockwise to reduce the value of the trip point.
Turn anticlockwise to increase the value of the trip point
3Lower the loads
4Remove the
additional load (10%
of « Pm »)
5Lift load « Pm »
using slow speed
lifting
No overload condition should be detected.
However, should the overload condition be tripped, see point 7 : Dynamic
effects filter.
6Lift load « Pm »
using high speed
lifting.
No overload condition should be detected.
However, should the overload condition be tripped, see point 7 : Dynamic
effects filter.
7Dynamic effects
filter using
potentiometer
« EFF. DYN »
Under certain conditions, the dynamic effects may create an overload
condition which will set off the safety system. The potentiometer on the front
of the monitor acts as a filter for this phenomenon by slowing down the
reaction of the relays in relation to the opening of the microswitch. The
potentiometer « EFF. DYN » has an adjustment range of 0 seconds (-) to ± 3
seconds (+).