uebe Domotherm Rapid User manual

Digital clinical thermometer for oral, rectal or axillary
self-measurement of human body temperature, direct mode
0830
PZN-03025437
Instructions for use
0123
Flexible tip
SAFETY INSTRUCTIONS
MEASURING THE TEMPERATURE
MEASURING POINTS
In the rectum (rectal measurement)
In the mouth (oral/sublingual measurement)
Under the arm (axillary measurement)
Display On/off button
Measuring
sensor
Battery compart-
ment cover
Carefully insert the tip 1-2 cm into the anus. Minimum
measurement time: 30 seconds
Do not consume hot or cold food or drinks for at least 10
minutes before measuring body temperature. Insert the
tip into one of the two heat pockets under the tongue,
on the left- or right-hand side of the root of the tongue.
Close the mouth and breathe calmly through the nose.
Minimum measurement time: 60 seconds
The skin must be completely dry. Place the tip under the
arm, so that it is fully in contact with the skin. The ther-
mometer should be at right angles to the body. Place the
arm across the chest, so that the tip is tightly covered.
Minimum measurement time: 3 minutes.
and the thermometer switches to measuring mode. Place
the thermometer at the desired measurement point.
During measurement the current temperature is shown
on the display with a ashing “°C” symbol. If the increase
in temperature over a period of 16 seconds is less than
0.10 °C, the thermometer will beep. If the temperature
is being measured rectally, this beep means that the ap-
proximate nal temperature is now being displayed; the
temperature will only change marginally from this point
on. If the temperature is being measured orally or under
the arm, it is important to stick to the intended minimum
measurement, irrespective of whether the device beeps.
At temperatures below 37.50°C, the normal signal tone
will sound (10 long, even tones). At temperatures of
37.50°C and higher, the fever alarm will sound (30 short
tones in sets of three).Read the temperature from the dis-
play. The thermometer will switch itself off automatically
after approx. 10 minutes.You can also turn it off earlier by
pressing the on/off button.The temperature will be saved
until the next one is measured.
Press the on/off button to switch on the thermometer.
A short signal tone conrms that the thermometer is
switched on. At the same time, all signals on the display
are illuminated as a visual functionality check. The last
measurement is then displayed for approx. 2 seconds
together with a small “M” symbol (M = memory). An in-
ternal test value of 37.00°C ± 0.01°C is then displayed
• Do not drop the thermometer.It is not resistant to shock
or impact.
• The thermometer contains small parts (battery etc.)
which children could swallow. Do not let children use
the device unsupervised.
• Protect the device from high temperatures and direct
sunlight.
• Do not open the casing (except to change the battery).
• Do not walk, run or speak while using the thermometer.
• Keep the thermometer in the plastic case.
• Using the thermometer other than as intended can lead
to inaccurate measurements.
• The use of this device in the vicinity of mobile phones,
microwaves or other devices with strong electromag-
netic elds can cause malfunctions. Maintain a distance
of at least three metres from such devices when using
the thermometer.
CLASSIFICATION OF THE TEMPERATURES MEASURED
Temperatures measured rectally are generally 0.5 °C
higher than those measured orally, while temperatures
measured under the arm are 0.5 °C lower.
Warning: Always consult a doctor if the reading shows
a high fever.
37.5 °C to 37.9 °C: elevated temperature
38.0 °C to 38.9 °C: moderate fever
39.0 °C and above: high fever
CHANGING THE BATTERY
If the symbol appears after all signals have been illu-
minated in the functionality check, a battery replacement
is required.
• Do not use a metal object to push the battery out of
its holder.
• Carefully pull the battery holder about 1 cm out of the
casing.
• Remove the battery compartment cover from the ther-
mometer.
• Insert a new battery (LR41 or SR41, 1.55V) with the +
sign towards the top.
• Push the batteryholder back into the casing and replace
the battery compartment cover.Take care not to damage
or misalign the seal when doing this.

METROLOGICAL INSPECTION
TECHNICAL DETAILS
WARRANTY
DISPOSAL
EXPLANATION OF SYMBOLS
CLEANING
Basically,a metrological inspection is recommended at in-
tervals of 2 years.However, professional users in Germany
have to comply with the aforementioned according to
“Regulation for Operators of Medical Devices”. This can be
implemented either by UEBE Medical GmbH, an authori-
ty responsible for metrology, or authorised maintenance
services.For this, please observe your national provisions.
© 2017 UEBE Medical GmbH. Not to be reproduced,either
in whole or in part. Subject to technical modications. 7 0830 214 BB 2020-05 EN
Type: Maximum thermometer, direct
mode
Measuring points/
body reference points: Rectal, oral, axillary
Range: 32.00 °C to 42.99 °C
Accuracy: ± 0.10 °C between 35.50 °C and
42.00 °C
± 0.20 °C in other temperature
ranges
Operating conditions: Ambient temperature 10 to 40
°C, relative humidity 30 to 85 %
Storage and transport
conditions: Ambient temperature -10 to 60
°C, relative humidity 25 to 90 %
Battery: LR41 or SR41, 1.55 V
Battery life: 2 years
IP rating: IP 22: protection from solid for-
eign bodies with diameters of
12.5 mm and above; protection
against water droplets
Batteries and technical appliances must not be
disposed of with domestic waste, but should be
handed in at the appropriate collection and dis-
posal points.
Clean the thermometer before and after every use, using
a soft cloth and isopropyl alcohol diluted with water, or
cold soapy water.
The device can be immersed in water or disinfectant solu-
tion when cleaning. Do not immerse the device for longer
than 30 minutes. Do not use boiling water, gas or a steam
autoclave to sterilise the device.
This product complies with Council Directive
93/42/EEC of 5 September 2007 concerning
medical devices and bears the mark, CE 0123
(TÜV SÜD Product Service GmbH).
Degree of protection against electric shock:
TYP BF
Please observe operating instructions
Lot number
Manufacturer
During the warranty period of 2 years from the date of
purchase, we remedy defects at our discretion and at our
expense in our plant through repair work or replacement
delivery of a defect-free device. The cost of returning the
device to our factory shall be borne by the sender. Com-
plaints that are sent back without prepayment will not be
accepted by UEBE.
The warranty does not cover normal wear and tear of
wearing parts or damage caused by failure to observe
the instructions for use,improper handling (e.g. breakage,
leaking batteries) or disassembly of the device by the buy-
er. Furthermore, the warranty does not constitute grounds
for asserting claims for damages against us.
Warranty claims can only be advanced in the warranty
period and by presenting proof of purchase. In the event
of a warranty claim, the unit must be sent to the follow-
ing address together with the proof of purchase and a
description of the complaint: UEBE Medical GmbH, Zum
Schlag 18, 97877 Wertheim, Germany.
The statutory claims and rights of the buyer against the
seller (for example, defect claims, producer liability) are
not restricted by this warranty.
UEBE Medical GmbH
Bürgermeister-Kuhn-Straße 22
97900 Külsheim
Germany
www.uebe.com
ERROR MESSAGES
Measured temperature is below 32.00°C and
is therefore outside the measuring range.
Measured temperature is above 42.99°C and
is therefore outside the measuring range.
Electronic fault. If the fault occurs again,
contact the UEBE customer service centre.
0123
This manual suits for next models
1
Other uebe Thermometer manuals