USH COMPACT HIGH-CAPACITY COMPRESSION TYPE LOADCELL OPERATION MANUAL 01AUG2020REV.1.06
1.Product outline
USH series are compact compression type loadcells with high-capacity.
There are suitable for built-in applications.
2.Handling prohibitions and precautions
3.Installation procedure
(1) For loadcell mounting, be sure to use proper screws by the specifications. Screw insertion length should
be around 1.5times longer than diameter of a screw. Screws must be tightened at instructed torque. Use
screw locking adhesive if needed.
Recommended tightening torque.....M2: 0.32 N·m, M4: 2.7 N·m
(2) The base plate must be made of steel having enough rigidity. Also its flatness should be 0.01mm or less
and surface roughness should be less than Ra3.2.
(3) The load plate must be made of steel having enough rigidity and hardness. Also its flatness should be
0.01mm or less, surface roughness should be less than Ra3.2, and hardness should be more than HRC50.
The loadcell characteristics may be changed with the flatness of the base.
4.Wiring color of loadcell cables
5.Indicator connection
Check the signal name and wiring color before connection.
(1) The cable of USH loadcell has 6 bare wire ends. Therefore USH should be used with an indicator that
accepts 6-wire input. In case USH is used with a 4-wire input indicator, both +EXC and +S wires must be
inserted to +EXC terminal of an indicator, also both -EXC and -S wires must be inserted to -EXC
terminal of an indicator. In this case, sensitivity may change a little.
Please connect SHIELD wire to SHIELD terminal of indicator.
Since load cell main body is insulated from SHIELD cable, make sure main body and indicator ground
point is connected in single-point grounding.
(2) Apply the recommended voltage for the excitation. If the applied voltage exceeds the maximum
excitation voltage, USH may get permanent damage.
(3) To extend loadcell cable, make sure to use a cable with the same diameter or bigger diameter than
standard cable. A voltage drop may occur when small diameter cable (having high resistance) is used,
and it may affect the result of measurement.
Re-calibration with an standard weight could be needed if USH is used with a 4-wire input indicator.
Please consult our sales representative for cable extension.
Prohibitions
Do not disassemble this product.
Avoid dropping or giving a shock to the product.
Precautions
●Do not install the product in the following environments:
- In an atmosphere with corrosive or combustible gases;
- Locations where temperature or humidity exceeds
specifications;
- Locations subject to drastic temperature fluctuations or
icing and condensation;
- Locations exposed to direct sunlight;
- Locations subject to direct vibration or shock
●Consult us if this product is to be u sed under a special
environment.
●Provide sufficient strength to the installation location.
●Perform adequate isolation if the product is used in the
following locations:
- Locations subject to noise such as static electricity;
- Locations subject to strong electric field and intense
magnetic field.
●Perform electric welding before mounting the loadcell.
Take caution not to make current flow in the load cell if
the welding needs to be performed after mounting.
●Handle this product as industrial waste when disposing.
Core wire Signal name / wiring color
+EXC +S -S -EXC +SIG -SIG SHIELD
Six wires Red Yellow Orange Black Green White Blue
Prohibited actions while operating the product
(Prohibited actions)
Events that may cause injury to personnel or material damage
in case of misuse.
Bolt
Load direction
Load plate
Loadcell
Base plate
There is a hollow at the
bottom of load cell.
For installation, make sure
no load is applied on that area.
6.Calibration
If calibration is needed, do it on indicator. For more information, refer to the manual of indicator.
■Actual load calibration
In this calibration method, actual load is applied to a loadcell, and the value of the actual load is input using
the keys. Accurate calibration with minimal error can be performed.
Prepare a reference item such as a weight in advance.
■Equivalent input calibration
In this calibration method, only the rated output value and the rated capacity value of the loadcell are input
using the keys. This method does not involve actual load.
The values are stated on the attached data sheet upon the purchase of loadcell.
7.Trouble shooting
Determine whether the error is occurring on the loadcell side or the indicator side.
Check the damage of USH in the following ways.
(1) In case a load is applied larger than safe overload or a load in unexpected direction is applied to the USH,
make sure to check if it didn’t get damaged after recalibration.
(2) When the reading on indicator is unstable or odd, do an inspection as follows after checking wiring and
precautions.
①Check the input resistance of loadcell under no-load condition. (between +EXC and -EXC)
②Check the output resistance of loadcell under no-load condition.(between +SIG and -SIG)
③Check the signal output of loadcell (Zero-balance) under no-load condition
In case measured values above are out-of-specification, please consult us.
If no problem found on USH, check the indicator in the following ways.
(1) Disconnect the USH from indicator. Check if the reading on indicator is near 0, while short the input
terminals(+SIG and -SIG).
(2) Measure the voltage of the output terminals(+EXC and -EXC) of indicator, and verify if the excitation
voltage is applied properly.
For more information, refer to the manual of indicator.
■Six-wire type loadcell → Six-wire type indicator ■Six-wire type loadcell → Four-wire type indicator
+EXC
−S
+SIG
−SIG
SHIELD
+S
−EXC
+EXC
−S
+SIG
−SIG
SHIELD
+S
−EXC
Loadcell side Indicator side Loadcell side Indicator side
㸩IN
㸫IN
㸫OUT 㸩OUT
+EXC
+SIG
−SIG
SHIELD
−EXC
+EXC
−S
+SIG
−SIG
SHIELD
+S
−EXC
㸩IN
㸫IN
㸫OUT 㸩OUT
* SHIELD wire is not connected to load cell main body.
7HVWHU
قٔكقٕك
ق5HGك
ق:KLWHك
ق%ODFNك
ق*UHHQك
ق%OXHك
ق2UDQJHك
ق<HOORZك
㸩EXC
㸫S
㸩SIG
㸫SIG
SHIELD
㸩S
㸫EXC
㸩IN
㸫IN
㸫OUT 㸩OUT
Disconnect the USH from the
indicator. And verify if the input
resistance of loadcell (between +EXC
and -EXC) is within the specification
range under no-load condition.
7HVWHU
قٔكقٕك
ق5HGك
ق:KLWHك
ق%ODFNك
ق*UHHQك
ق%OXHك
ق2UDQJHك
ق<HOORZك
㸩EXC
㸫S
㸩SIG
㸫SIG
SHIELD
㸩S
㸫EXC
㸩IN
㸫IN
㸫OUT 㸩OUT
Disconnect the USH from the
indicator. And verify if the output
resistance of loadcell (between +SIG
and -SIG) is within the specification
range under no-load condition.
7HVWHU
قٔكقٕك
,QGLFDWRU
ق5HGك
ق:KLWHك
ق%ODFNك
ق*UHHQك
ق%OXHك
ق2UDQJHك
ق<HOORZك
㸩EXC
㸫S
㸩SIG
㸫SIG
SHIELD
㸩S
㸫EXC
㸩EXC
㸫S
㸩SIG
㸫SIG
㸩S
㸫EXC
㸩IN
㸫IN
㸫OUT 㸩OUT
Connect the USH with an indicator.
And measure the output voltage
(Vout[mV/V]) under no-load
condition, while an excitation
voltage (Vin[V]) is applied. Verify if
the signal output of loadcell
( [mV/V]) is within the
specification range. (Zero-balance)
Vout
Vin
8.External dimension
9.Specifications
Model USH-10KN USH-20KN USH-50KN
Rated capacity 10 20 50 kN
Rated output 1±20% mV/V
Safe overload 200 %R.C.
Zero balance ±20 %R.O.
Non-linearity 0.2 %R.O.
Hysteresis 0.15 %R.O.
Repeatability 0.1 %R.O.
Compensated temperature range -10 to +70 ℃
Safe temperature range -30 to +85 ℃
Temperature effect on zero 0.2 %R.O./10℃
Temperature effect on span 0.2 %R.O./10℃
Input resistance Approx. 1000 Ω
Output resistance Approx. 1000 Ω
Recommended excitation voltage 5 V
Maximum excitation voltage 7.5 V
Insulation resistance (DC50V) 1000 or more MΩ
Cable φ2.5 6-conductor flexible cable (3m)
Cable end: 7 wires
Deflection at rated capacity 0.06 0.07 mm
Natural frequency 20 15 kHz
Loadcell material Stainless steel
Weight 0.1 0.18 kg
■USH-10KN, USH-20KN ■USH-50KN
8QLWPP
Load direction

25
22
ȭ40
6.1
ȭ13
SR60
P.C.D.32
120°
3-M4 Depth 6
ȭ2.5 6-conductor
flexible cable (3m)
Cable end: 7 wires
Load direction

ȭ26
13
12
SR40
ȭ8
4.1
P.C.D.22.75
120°
3-M2 Depth 3
ȭ2.5 6-conductor
flexible cable (3m)
Cable end: 7 wires
International Sales Department
9-11 Nihonbashi Hisamatsu-cho, Chuo-ku, Tokyo 103-0005
Tel: +81-3-3639-6120 Fax: +81-3-3639-6130
Unipulse Corporation

This manual suits for next models

2

Other Unipulse Measuring Instrument manuals

Unipulse F701+ User manual

Unipulse

Unipulse F701+ User manual

Unipulse F377A User manual

Unipulse

Unipulse F377A User manual

Unipulse UPM100 User manual

Unipulse

Unipulse UPM100 User manual

Unipulse TM500 User manual

Unipulse

Unipulse TM500 User manual

Unipulse F325 User manual

Unipulse

Unipulse F325 User manual

Unipulse F370 User manual

Unipulse

Unipulse F370 User manual

Unipulse UTM II-20Nm User manual

Unipulse

Unipulse UTM II-20Nm User manual

Unipulse F372A User manual

Unipulse

Unipulse F372A User manual

Unipulse TM400 User manual

Unipulse

Unipulse TM400 User manual

Unipulse F320 User manual

Unipulse

Unipulse F320 User manual

Unipulse F340A User manual

Unipulse

Unipulse F340A User manual

Unipulse TM201 User manual

Unipulse

Unipulse TM201 User manual

Popular Measuring Instrument manuals by other brands

Hanna Instruments HI 95771C instruction manual

Hanna Instruments

Hanna Instruments HI 95771C instruction manual

Lang Radar Touch RT500 manual

Lang

Lang Radar Touch RT500 manual

Vanguard Instruments Company ACCU-TRANS II Series operating instructions

Vanguard Instruments Company

Vanguard Instruments Company ACCU-TRANS II Series operating instructions

Golf Buddy LR4 instruction manual

Golf Buddy

Golf Buddy LR4 instruction manual

PlantCare Moistick digital manual

PlantCare

PlantCare Moistick digital manual

Diamond Antenna SX-200 Instruction leaflet

Diamond Antenna

Diamond Antenna SX-200 Instruction leaflet

Seojin Instech SLT Series Operation manual

Seojin Instech

Seojin Instech SLT Series Operation manual

Beluk EMM5 Reference

Beluk

Beluk EMM5 Reference

MEYTEC GT-200RX manual

MEYTEC

MEYTEC GT-200RX manual

Endress+Hauser Levelflex M FMP43 operating instructions

Endress+Hauser

Endress+Hauser Levelflex M FMP43 operating instructions

Leupold RX operating instructions

Leupold

Leupold RX operating instructions

Tektronix 1L30 instruction manual

Tektronix

Tektronix 1L30 instruction manual

EG&G ORTEC 455 Operating and service manual

EG&G

EG&G ORTEC 455 Operating and service manual

MADDALENA microCLIMA U Installation and operating instructions

MADDALENA

MADDALENA microCLIMA U Installation and operating instructions

AFTI WATCHDOG 5002-001 quick start guide

AFTI

AFTI WATCHDOG 5002-001 quick start guide

Advanced Energy Industries LumaSense Technologies E2T Quasar 2 Series quick start guide

Advanced Energy Industries

Advanced Energy Industries LumaSense Technologies E2T Quasar 2 Series quick start guide

Hach APA 6000 Installation

Hach

Hach APA 6000 Installation

AEA Technology, Inc. Bravo MRI II operating manual

AEA Technology, Inc.

AEA Technology, Inc. Bravo MRI II operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.