manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Viavi
  6. •
  7. Test Equipment
  8. •
  9. Viavi T-BERD 6000A V2 Parts list manual

Viavi T-BERD 6000A V2 Parts list manual

Other manuals for T-BERD 6000A V2

2

Other Viavi Test Equipment manuals

Viavi OneAdvisor 800 Series User manual

Viavi

Viavi OneAdvisor 800 Series User manual

Viavi T-BERD/MTS-5800 Parts list manual

Viavi

Viavi T-BERD/MTS-5800 Parts list manual

Viavi SmartClass FBP-OLP-82 User manual

Viavi

Viavi SmartClass FBP-OLP-82 User manual

Viavi PSD90-3 User manual

Viavi

Viavi PSD90-3 User manual

Viavi T-BERD MTS 5800 Parts list manual

Viavi

Viavi T-BERD MTS 5800 Parts list manual

Viavi OneAdvisor ONA-800 User manual

Viavi

Viavi OneAdvisor ONA-800 User manual

Viavi T-BERD/MTS 8000 User manual

Viavi

Viavi T-BERD/MTS 8000 User manual

Viavi T-BERD MTS 5800 User manual

Viavi

Viavi T-BERD MTS 5800 User manual

Viavi HST-3000 User manual

Viavi

Viavi HST-3000 User manual

Viavi ALT-9000 User manual

Viavi

Viavi ALT-9000 User manual

Viavi IFR 6000 User manual

Viavi

Viavi IFR 6000 User manual

Viavi OneAdvisor-800 User manual

Viavi

Viavi OneAdvisor-800 User manual

Viavi T-BERD MTS 5800 Parts list manual

Viavi

Viavi T-BERD MTS 5800 Parts list manual

Viavi T-BERD 5800 Parts list manual

Viavi

Viavi T-BERD 5800 Parts list manual

Viavi T-BERD MTS 5800 Parts list manual

Viavi

Viavi T-BERD MTS 5800 Parts list manual

Viavi T-BERD/MTS 5800 User manual

Viavi

Viavi T-BERD/MTS 5800 User manual

Viavi COSA-4055 Parts list manual

Viavi

Viavi COSA-4055 Parts list manual

Viavi 4100 Series Parts list manual

Viavi

Viavi 4100 Series Parts list manual

Viavi T-BERD/MTS-5800 OCC-4056C Parts list manual

Viavi

Viavi T-BERD/MTS-5800 OCC-4056C Parts list manual

Viavi SmartClass OLP-85 Series User manual

Viavi

Viavi SmartClass OLP-85 Series User manual

Viavi OneAdvisor ONA-800 Technical manual

Viavi

Viavi OneAdvisor ONA-800 Technical manual

Viavi CT-X User manual

Viavi

Viavi CT-X User manual

Viavi 3900 Series Owner's manual

Viavi

Viavi 3900 Series Owner's manual

Viavi T-BERD/MTS 5800 User manual

Viavi

Viavi T-BERD/MTS 5800 User manual

Popular Test Equipment manuals by other brands

Ambit P.F. Series instruction manual

Ambit

Ambit P.F. Series instruction manual

Megger BMM500 Series user guide

Megger

Megger BMM500 Series user guide

RJG Lynx Sensor Tester operating instructions

RJG

RJG Lynx Sensor Tester operating instructions

Mark-10 TT02 Series user guide

Mark-10

Mark-10 TT02 Series user guide

Martel 3001 Operator's manual

Martel

Martel 3001 Operator's manual

ElproSys DiagProg4 instruction manual

ElproSys

ElproSys DiagProg4 instruction manual

Redtech TRAILERteck T05 user manual

Redtech

Redtech TRAILERteck T05 user manual

Venmar AVS Constructo 1.0 HRV user guide

Venmar

Venmar AVS Constructo 1.0 HRV user guide

Test Instrument Solutions SafetyPAT operating manual

Test Instrument Solutions

Test Instrument Solutions SafetyPAT operating manual

Hanna Instruments HI 38078 instruction manual

Hanna Instruments

Hanna Instruments HI 38078 instruction manual

Kistler 5495C Series instruction manual

Kistler

Kistler 5495C Series instruction manual

Waygate Technologies DM5E Basic quick start guide

Waygate Technologies

Waygate Technologies DM5E Basic quick start guide

StoneL DeviceNet CK464002A manual

StoneL

StoneL DeviceNet CK464002A manual

Seica RAPID 220 Site preparation guide

Seica

Seica RAPID 220 Site preparation guide

Kingfisher KI7400 Series Training manual

Kingfisher

Kingfisher KI7400 Series Training manual

Kurth Electronic CCTS-03 operating manual

Kurth Electronic

Kurth Electronic CCTS-03 operating manual

SMART KANAAD SBT XTREME 3G Series user manual

SMART

SMART KANAAD SBT XTREME 3G Series user manual

Agilent Technologies BERT Serial Getting started

Agilent Technologies

Agilent Technologies BERT Serial Getting started

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

VIAVI Solutions
Quick Card
T-BERD 6000Av2 Network Tester
Datacom Bit Error Rate Testing (BERT)
This quick card describes how to configure the T-BERD 6000A as Data Terminal Equipment (DTE) and
run a Bit Error Rate Test on a Datacom interface using typical configuration settings. Please refer to
the MSAM Data Communication and Diphase Testing Manual for an explanation of all settings.
Equipment Requirements:
•T-BERD 6000Av2 equipped with the following:
oBERT software release V26.3 or greater
oTest options:
▪High Speed Datacom test option
(VIAVI Part# CTHSDATA)
•Multi-Services Application Module
(VIAVI Part# C1010V2)
•Datacom Physical Interface Module (PIM)
(VIAVI Part# CPHSDATAV2)
•One of the following Datacom cables to connect the
Datacom PIM to the line under test:
oRS-232/V.24, MIL-188c, EIA-530 Cable
(VIAVI Part# CB-44385)
oRS-449/V.36 Cable (VIAVI Part# CB-44388)
oV.35 Cable (VIAVI Part# CB-44390)
Figure 1: Equipment Requirements
The following information is required to complete the test:
•Interface (RS-232/V.24, MIL-188c, EIA-530, RS-449/V.36, or V.35)
•Signal Mode (Balanced or Unbalanced)
•Timing Mode (Synchronous or Asynchronous)
•Rx Timing Source (Internal, Interface, or Recovered)
•Tx Timing Source (Internal or Interface)
•Out of Band Flow Control (On or Off)
•Test Patterns(s)
•BER Pass/Fail Threshold
Connect to Line Under Test:
•Connect the desired Datacom cable to the Datacom PIM.
•Use the connector labeled “To DCE”to connect to the line under test.
VIAVI Solutions
Launch and Configure Test:
1. Press the Power button to turn on the
test set.
2. Press the Home button and tap the
BERT icon to launch the MSAM.
3. Using the Select Test menu or Quick
Launch menu, launch the HS Datacom►
HS Datacom BERT►Terminate test.
Figure 2: Home screen
Figure 3: Launch Screen
4. Tap the tools menu and select
.
5. Press to continue.
Figure 4: Reset Test to Defaults
VIAVI Solutions
6. Press the Setup Soft Key, , on the top right side of the screen. Select the indicated tabs and
configure your test as follows. Leave all other values at default, unless specified in the work order.
Folder
Option
Value(s)
Interface
Interface
RS-232/V.24, EIA-530, RS-449/V.36, etc.
Equipment Type
DTE
Signal Mode
Balanced or Unbalanced
Timing
Timing Mode
Synchronous or Asynchronous
Rx Timing Source
Recovered or Interface (RT). If unknown, select
Interface (RT).
Tx Timing Source
Internal (Synth), External (BNC), or Interface (ST).
If unknown, select Interface (ST) for Synchronous timing,
select Internal (Synth) for Asynchronous timing.
Synthesizer Frequency
Enter frequency in kHz; i.e. 9.6 kHz for 9600 bps.
Pattern
Pattern Mode
ANSI
Pattern
Enter the first Pattern in your test plan (QBF, QRSS, etc.)
7. Press the Results Soft Key to view the
Test Results screen.
8. Tap the Signal Lead tab at the bottom of
the screen and tap the RTS and DTR
buttons.
9. Press the Restart soft key .
Figure 5: Test Results Screen, Signal Leads
VIAVI Solutions
Contact Us +1 844 GO VIAVI)
(+1 844 468 4284)
To reach the VIAVI office nearest
you, visit viavisolutions.com/contacts.
© 2018 VIAVI Solutions Inc.
Product specifications and descriptions in this
document are subject to change without notice.
10. Using the drop-down menus, select
“HS Datacom/BERT” for the right results
display.
11. Allow the test to run for desired duration
and verify the following:
•Summary LED is green.
•Pattern Sync LED is green.
•Bit Error Rate result does not exceed
your required threshold. (0.00E+00 if
pass/fail threshold unknown)
Figure 6: Test Results Screen, HS Datacom/BERT
12. In the T-BERD’s Quick Config menu,
change “Pattern” to the next value in the
test plan.
13. Press the Restart soft key to reset
results.
14. Allow test to run for desired duration and
verify the following:
•Pattern Sync LED is green.
•Bit Error Rate does not exceed your
required threshold. (0.00E+00 if
pass/fail threshold unknown)
Figure 7: Test Plan
15. Repeat steps 12 through 14 for all Patterns in the test plan. Patterns may include:
•QRSS
•QBF
•R-Trip Delay
Simulates live traffic
Quick Brown Fox message
Measures Round Trip Delay (RTD) instead of Bit Errors (RTD values
are shown instead of BER in the “Payload/BERT” results display)