WIKA 232.30.100 Operating instructions

EN
DE
FR
ES
Additional operating instructions
Zusatz-Betriebsanleitung
Mode d’emploi complémentaires
Manual de instrucciones adicional
Additional operating instructions for EMICOgauge
Zusatz-Betriebsanleitung für EMICOgauge
Mode d’emploi complémentaires pour EMICOgauge
Manual de instrucciones adicional para EMICOgauge
EMICOgauge examples

EN
DE
DEFR
ES
2
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
Additional operating instructions for EMICOgauge
Page
3 - 18
Zusatz-Betriebsanleitung für EMICOgauge
Seite
19 - 34
Mode d’emploi complémentaires pour EMICOgauge
Page
35 - 50
Manual de instrucciones adicional para EMICOgauge
Página
51 - 66
© 04/2022 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions and additional operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung und Zusatz-Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d‘emploi et le Mode d’emploi complémentaires avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones y el manual de instrucciones adicional antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!

EN
Contents
Contents
Declarations of conformity can be found online at www.wika.com.
1. General information 4
2. Design and function 5
2.1 Description . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . 5
3. Safety 6
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . . . 6
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . . . 6
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . . . 6
3.4 Responsibility of the operator . . . . . . . . . . . . . . . . . . 7
3.5 Personnel qualification . . . . . . . . . . . . . . . . . . . . 7
3.6 Labelling / Safety marking . . . . . . . . . . . . . . . . . . . 7
4. Transport, packaging and storage 8
4.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . . 8
4.2 Packaging and storage . . . . . . . . . . . . . . . . . . . . 8
5. Commissioning, operation 9
5.1 Mounting at the process connection . . . . . . . . . . . . . . . 9
5.2 Commissioning and operation . . . . . . . . . . . . . . . . 10
6. Faults 10
7. Maintenance and cleaning 12
7.1 Maintenance of pressure gauge . . . . . . . . . . . . . . . . 12
7.2 Separation at the EMICOgauge connection . . . . . . . . . . . . 12
7.3 Maintenance of instrumentation valve . . . . . . . . . . . . . . 13
7.4 Repair of instrumentation valve . . . . . . . . . . . . . . . . 14
7.5 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . 14
8. Dismounting, return and disposal 15
8.1 Dismounting the EMICOgauge instrument hook-up . . . . . . . . . 15
8.2 Return. . . . . . . . . . . . . . . . . . . . . . . . . 16
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . 16
9. Specifications 17
3
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge

EN
1. General information
4
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
Supplementary documentation
These additional operating instructions apply in connection with the following documents:
▶
Operating instructions of the mounted pressure gauge, e.g. “Mechanical pressure
measuring instruments”, article number 2408976
▶
Operating instructions of the mounted instrumentation valve, e.g. “Shut-o valves and
valve manifolds”, article number 14294094
▶
Documents of the components used
1. General information
■
The instrument described in the operating instructions has been designed and manufac-
tured using state-of-the-art technology. All components are subject to stringent quality
and environmental criteria during production. Our management systems are certied to
ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument.
Working safely requires that all safety instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and general safety regula-
tions for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions on to the next operator or owner of the instrument.
■
Skilled personnel must have carefully read and understood all operating instructions
supplied, prior to beginning any work.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modications.
■
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with inter-
national standards.

EN
2. Design and function
5
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
2. Design and function
2.1 Description
The EMICOgauge instrument hook-up, consisting of pressure gauge and instrumentation
valve, minimises the number of leakage points and thus reduces the risk of media escaping
into the environment.
The valve bonnets of the EMICOgauges are marked with an orange ring in addition to the
coloured dust cap.
Pressure gauge, model 232.30.100
EMICOgauge connection with 360° swivel connection
Instrumentation valve, model IVM, double-block-and-bleed, instrument version
Cone for metal-to-metal seat
O-ring sealing
Support ring
The 360° swivel adapter connection of the EMICOgauge enables quick alignment of the
pressure gauge with simultaneous pressure sealing. In addition, this makes it easy and
safe to disassemble the EMICOgauge into pressure gauge and valve. See chapter 7.2
“Separation at the EMICOgauge connection”.
2.2 Scope of delivery
Cross-check scope of delivery with delivery note.
EMICOgauge example
Sectional illustration of

EN
3. Safety
6
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
3. Safety
3.1 Explanation of symbols
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury
or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or
damage to property or the environment, if not avoided.
WARNING!
... indicates a potentially dangerous situation that can result in burns, caused
by hot surfaces or liquids, if not avoided.
Information
... points out useful tips, recommendations and information for ecient and
trouble-free operation.
3.2 Intended use
The EMICOgauge instrument hook-up is used to measure pressure in the process indus-
try. Using the shut-o and venting function(s) of the instrumentation valve, the measuring
instrument can be safely separated from the process. Due to conformity with TA-Luft
(VDI 2440), there is a signicantly reduced risk of fugitive emissions. This is also ensured
by the EMICOgauge connection, a 360° swivel connection (swivel adapter) with a special
pressure sealing.This connection enables easy replacement and the positioning of the
measuring instruments at the measuring location.
The EMICOgauge is only suitable for gaseous and liquid aggressive media that are not
highly viscous or crystallising.
The maximum permissible operating pressure of the EMICOgauge is determined by the
component(s) having the weakest performance data.
3.3 Improper use
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
▶
Refrain from unauthorised modications to the instrument.
▶
Do not use the instrument within hazardous areas without suitable appro-
val.
▶
Do not use the instrument with abrasive or viscous media.

EN
3. Safety
7
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
Any use beyond or dierent to the intended use is considered as improper use.
Do not use this instrument in safety or emergency shutdown devices.
3.4 Responsibility of the operator
Due to the redundant pressure sealing of the EMICOgauge connection, up to 20 mounting
and dismounting operations of the pressure gauge can be carried out. See chapter 7.2
“Separation at the EMICOgauge connection”. After that, the operator is obliged to use a
new EMICOgauge instrument hook-up, so that the leak tightness of the overall system can
be guaranteed.
Systematically record mounting and dismounting operations at measuring
locations with EMICOgauges in order to be able to plan replacement of the
instrument in good time.
The legibility of the marking must be observed during time in use but at least during inspec-
tion periods of three years. If any harm of the legibility is found please contact the manufac-
turer to renew the marking.
3.5 Personnel qualification
WARNING!
Risk of injury should qualication be insucient!
Improper handling can result in considerable injury and damage to property.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualications described
below.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowled-
ge of country-specic regulations, current standards and directives, are capable of carrying
out the work described and independently recognising potential hazards.
3.6 Labelling / Safety marking
Pressure gauge and instrumentation valve are marked in accordance with the documenta-
tion supplied, depending on the version.The text “EMICOgauge” is shown on the dial.

EN
4. Transport, packaging and storage
8
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
4. Transport, packaging and storage
4.1 Transport
Check the instrument for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
▶
When unloading packed goods upon delivery as well as during internal
transport, proceed carefully and observe the symbols on the packaging.
▶
With internal transport, observe the instructions in chapter 4.2 “Packaging
and storage”.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in
installation site, sending for repair).
Permissible conditions at the place of storage:
■
Storage temperature: -40 ... +70 °C [-40 ... +158 °F]
■
Protect the measuring instruments from humidity and dust!
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Hazardous environments, ammable atmospheres
WARNING!
Before storing the instrument (following operation), remove any residual
media.This is of particular importance if the medium is hazardous to health,
e.g. caustic, toxic, carcinogenic, radioactive, etc.

EN
5. Commissioning, operation
9
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
5. Commissioning, operation
Only use original parts.
WARNING!
Physical injuries and damage to property and the environment caused
by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, ammable
or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), and also with refrigeration plants and compressors, there is a
danger of physical injuries and damage to property and the environment.
Should a failure occur, aggressive media with extremely high temperature
and under high pressure or vacuum may be present at the instrument.
▶
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
▶
Wear the requisite protective equipment
5.1 Mounting at the process connection
■
When unpacking, check all components for any external damage.
■
Ensure that all unused connections are closed. Plug screws for vent connections are
included in delivery, though not pretted.
■
The EMICOgauge must only be carried by the valve body, and not by the valve handles
nor the pressure gauge case.
■
Check the product marking for correctness (see chapter 3.6 “Labelling, safety marking”).
■
Before mounting, ensure that the process line has been depressurised using the main
valve.
■
Remove the protective cap of the process connection.
■
Ensure that the sealing faces are clean and do not show any mechanical damage.
■
The correct sealing element must be used for the respective connection.
Threaded connection
To screw in threaded connections the appropriate tool must only be applied through the
spanner ats provided for this purpose.
The tightening torque is dependent on the process connection and sealing element used.
Flange connection
The ange standard (e.g. ANSI B16.5 or EN 1092-1) for mating anges and mounting
elements (sealings and stud bolts) must match the EMICOgauge supplied. Before
mounting, the mating anges and mounting elements must be checked for correct size and
nominal pressure rating.

EN
5. Commissioning, operation / 6. Faults
10
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
5.2 Commissioning and operation
To close the valve, turn the handle in a clockwise direction as far as the stop.To open the
valve, turn the handle anticlockwise as far as the stop. It will take approximately 4 turns
from open to closed and vice versa.
The ow can be controlled by not totally opening/closing the valve.
Please note that the handle may present a slight play and spin freely up to a
quarter turn when in the depressurised state.This is due to the constructive
design.
■
The valve, especially the handle, must not be subjected to any external loading (e.g. use
as a climbing aid, support for objects).
■
The handle may only be used by hand; the use of tools is not accepted.
■
The torque on reaching the valve stop must only be hand-tight.
■
Ensure that all valves of the product are closed prior to opening the main valve of the
process line.
6. Faults
The following table contains the most frequent causes of faults and the
necessary countermeasures.
Faults Causes Measures
Fault at the pressure
gauge
Various See the chapter on faults in the operating
instructions for the mounted pressure
gauge.
Leakage at the
EMICOgauge
connection
Error at the sealing point Disassemble the product as described in
the chapter “Separation at the EMICOgauge
connection”.
If leakage cannot be eliminated, replace
product.
Dicult valve handle
operation
Lubrication problem
due to unchanged valve
position for a long time
period
Perform a functional test as described in
chapter 7.3 “Maintenance of instrumentation
valve”and shorten inspection interval as
appropriate.
Defective parts due
to corrosion, process
conditions, age
Replace product.
Overtightened gland nut,
defective sealing packing
Replace product.

EN
6. Faults
11
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
Faults Causes Measures
Leakage at the
spindle
Leaking sealing packing See chapter 7.4 “Repair of instrumentation
valve”
No ow or restricted
ow of media
Valve closed or partially
open
Ensure that that valve is open
Incorrect mounting Ensure that that valve is correctly mounted
Clogging by unsuitable
media
Check compatibility of the media
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the product
must be taken out of operation immediately.
▶
Ensure that there is no longer any pressure present and protect against
being put into operation accidentally.
▶
Contact the supplier.
▶
If a return is needed, please follow the instructions given in chapter 8.2
“Return”.
WARNING!
Physical injuries and damage to property and the environment caused
by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, ammable or
toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic, radioac-
tive), and also with refrigeration plants and compressors, there is a danger
of physical injuries and damage to property and the environment. Should a
failure occur, aggressive media with extremely high temperature and under
high pressure or vacuum may be present at the product.
▶
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
▶
Wear the requisite protective equipment (see chapter 3.6 “Personal
protective equipment”).

EN
7. Maintenance and cleaning
12
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
7. Maintenance and cleaning
CAUTION!
Physical injuries and damage to property and the environment
If the EMICOgauge has already been commissioned, before maintenance
and cleaning, make sure that there is no pressure at the product and no
risk of burns. See chapter 8.1 “Dismounting the EMICOgauge instrument
hook-up”.
7.1 Maintenance of pressure gauge
The pressure gauges are maintenance-free. Regular checks should be carried out to
ensure the measurement accuracy. In the case of pressure gauges with blow-out devices
or safety pressure gauges, the safety elements (e.g. blow-out plug or blow-out back) must
be replaced after 10 years.The replacement may only be carried out by skilled personnel
using original parts and after authorisation of the plant manager.
7.2 Separation at the EMICOgauge connection
Here, the pressure gauge can be easily separated from the instrumentation valve.
Tools: SW 22 and 27 open-ended spanner; SW 27, 60 Nm torque spanner
Pressure gauge dismounting
1. With the SW 22 open-ended spanner, at the spanner ats , secure the pressure
gauge against turning and unscrew the 360° swivel connection anticlockwise at the
spanner ats using the SW 27 open-ended spanner.
2. Carefully pull the pressure gauge o the valve. Thus, ensure that the metal sealing
faces are not damaged.
3. Remove O-ring sealing and support ring.
4. Clean the EMICOgauge connection inside and outside, in particular, the old high-tem-
perature ceramic paste must be completely removed. With cleaning, take care that no
sealing faces are damaged.
Spanner ats (SW 22) for positioning the
pressure gauge
Spanner ats (SW 27) of the 360° swivel
connection for dismounting or mounting the
pressure gauge
Threads of the 360° swivel connection
Cone for metal-to-metal seat
O-ring sealing
Support ring

EN
7. Maintenance and cleaning
13
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
Pressure gauge mounting
Before each pressure gauge mounting, the following must be ensured or be on-site:
■
Clean working surface
■
All sealing faces and threads are clean and undamaged
■
1 new sealing set, consisting of O-ring und support ring (5-piece sealing set, article
number 14525042)
■
High-temperature ceramic paste, WS-600-030 (tube 30 gram, article number 14525042)
Mounting instructions
1. Carefully slide the support ring included in the sealing set and then the O-ring over the
sealing cone and insert it neatly into the groove provided.
2. Apply the high-temperature ceramic paste to the thread of the 360° swivel connection
and the cone .
3. Before screwing the pressure gauge tight, align the pressure gauge in the 360° swivel
connection (swivel adapter) with the valve by holding the spanner flats and turning
the spanner flats and screw it hand-tight. Afterwards, the pressure gauge must no
longer be twisted in the 360° swivel connection.
4. Fix the instrumentation valve so that it cannot turn with the pressure gauge as it is
screwed tight.
5. Secure the pressure gauge against turning using the SW 22 open-ended spanner on
the spanner flats and tighten the EMICOgauge connection again with the SW 27
torque spanner set to 60 Nm ±5 Nm.
The pressure gauge mounting is now complete. After the last step, the pressure gauge
can no longer be positioned without loosening the EMICOgauge connection again. After
alignment, step 5 must be carried out again.
7.3 Maintenance of instrumentation valve
Valves must be checked within the context of regular maintenance.
Inspection interval
Depending on the process, the ambient conditions and the number of operations, the
required inspection interval may vary from daily to several weeks.
Recommended inspection interval: ≤ 3 months
Inspection checklist
1. Leak test of the process and instrument connections
2. Leak test between bonnet and valve body
3. Function test of correct opening and closing
In case the leak test of item 2 fails, proceed with the adjustment of the sealing packing
described below.

EN
7. Maintenance and cleaning
14
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
7.4 Repair of instrumentation valve
Adjustment of the sealing packing
This adjustment is required when a leak can be detected at the valve spindle, directly
below the handle, or when during operation there is no torque or resistance when turning
the handle.
1. Loosen the counter nut
2. Increase the packing compression through the gland nut applying a torque of
≥ 13 … ≤ 18 Nm [18 … 25 lbs ft]
3. Fix the gland nut with the counter nut
If after the adjustment of the sealing packing the leakage problem persists, the packing
must be returned for repair.
Perfect functioning of the product can only be guaranteed when original
accessories and spare parts are used.
7.5 Cleaning
CAUTION!
Physical injuries and damage to property and the environment
Improper cleaning may lead to physical injuries and damage to property and
the environment. Residual media in the dismounted product can result in a
risk to persons, the environment and equipment.
▶
Rinse or clean the dismounted product.
▶
Take sucient precautionary measures.
1. Prior to cleaning, properly disconnect the product from the process.
2. Clean the product carefully with a moist cloth.
CAUTION!
Damage to property
Improper cleaning may lead to damage to the product!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any hard or pointed objects for cleaning.

EN
8. Dismounting, return and disposal
15
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
8. Dismounting, return and disposal
8.1 Dismounting the EMICOgauge instrument hook-up
Before starting the dismounting, ensure that the main valve of the process line is closed.
WARNING!
Risk of burns
During dismounting there is a risk of dangerously hot media escaping.
▶
Let the product cool down suciently before dismounting it!
WARNING!
Physical injury
When dismounting, there is a danger from aggressive media and high
pressures.
▶
Wear the requisite protective equipment
▶
Observe the information in the material safety data sheet for the corres-
ponding medium.
▶
Only disconnect the valve once the system has been depressurised.
WARNING!
Physical injuries and damage to property and the environment throu-
gh residual media
▶
Upon contact with hazardous media (e.g. oxygen, acetylene, ammable
or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), and also with refrigeration plants and compressors, there is a
danger of physical injuries and damage to property and the environment.
▶
Wear the requisite protective equipment
▶
Observe the information in the material safety data sheet for the corres-
ponding medium.
▶
Before storage of the dismounted product (following use) wash or clean it,
in order to protect persons and the environment from exposure to residual
media.
1. To avoid pressure surges, open the shut-o valve (blue) slowly.
2. In case the valve is equipped with a vent connection, open this connection with a suita-
ble tool. Remove plug screw/vent screw (if available).
The valve can be dismounted now.

EN
8. Dismounting, return and disposal
16
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
8.2 Return
Strictly observe the following when shipping the product:
All products delivered to WIKA must be free from any kind of hazardous substances (acids,
bases, solutions, etc.) and must therefore be cleaned before being returned.
WARNING!
Physical injuries and damage to property and the environment
through residual media
Residual media in the dismounted product can result in a risk to persons, the
environment and equipment.
▶
With hazardous substances, include the material safety data sheet for the
corresponding medium.
▶
Clean the instrument, see chapter 7.5 “Cleaning”.
When returning the product, use the original packaging or a suitable transport packaging.
Information on returns can be found under the heading “Service” on our local
website.
8.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of product components and packaging materials in an environmentally compatible
way and in accordance with the country-specic waste disposal regulations.

EN
9. Specifications
17
14525888.01 04/2022 EN/DE/FR/ES
WIKA additional operating instructions for EMICOgauge
9. Specifications
Basic information for pressure gauges
Scale range Between 0 ... 0.6 bar and 0 ... 420 bar
[0 ... 10 psi and 0 ... 6,000 psi]
As well as corresponding measuring spans for other units and
+/- scale ranges
Connection location Lower mount (radial)
Assembly
Mounting type EMICOgauge connection: 360° swivel connection (swivel
adapter) with special pressure sealing
Spare part: 5 x sealing set consisting of O-ring and support
ring; article no. 14525042
Instrumentation valve
■
For instrument version of IVM → See data sheet AC 09.17
■
Model IV10, IV11 or IV20 → See data sheets AC 09.22 and
AC 09.19
Operating conditions
Medium temperature
With unlled pressure gauge -20 ... +150 °C [-4 ... +302 °F]
With lled pressure gauge -20 ... +100 °C [-4 ... +212 °F]
Ambient temperature -20 ... +60 °C [-4 ... +140 °F]
Leak tightness of the overall
system
Fullls the fugitive emission requirements
per TA Luft (VDI 2440), helium tested,
specic leakage rate: < 1∙10-4 mbar l/m s

WIKA additional operating instructions for EMICOgauge18
14525888.01 04/2022 EN/DE/FR/ES

DE
Inhalt
1. Allgemeines 20
2. Aufbau und Funktion 21
2.1 Beschreibung . . . . . . . . . . . . . . . . . . . . . . 21
2.2 Lieferumfang . . . . . . . . . . . . . . . . . . . . . . 21
3. Sicherheit 22
3.1 Symbolerklärung . . . . . . . . . . . . . . . . . . . . . 22
3.2 Bestimmungsgemäße Verwendung. . . . . . . . . . . . . . . 22
3.3 Fehlgebrauch . . . . . . . . . . . . . . . . . . . . . . 22
3.4 Verantwortung des Betreibers. . . . . . . . . . . . . . . . . 23
3.5 Personalqualifikation . . . . . . . . . . . . . . . . . . . . 23
3.6 Beschilderung / Sicherheitskennzeichnung . . . . . . . . . . . . 23
4. Transport, Verpackung und Lagerung 24
4.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . 24
4.2 Verpackung und Lagerung . . . . . . . . . . . . . . . . . . 24
5. Inbetriebnahme, Betrieb 25
5.1 Montage am Prozessanschluss . . . . . . . . . . . . . . . . 25
5.2 Inbetriebnahme und Betrieb . . . . . . . . . . . . . . . . . 26
6. Störungen 26
7. Wartung und Reinigung 28
7.1 Wartung des Manometers . . . . . . . . . . . . . . . . . . 28
7.2 Trennung am EMICOgauge-Anschluss . . . . . . . . . . . . . 28
7.3 Wartung des Instrumentierungsventils . . . . . . . . . . . . . . 29
7.4 Instandsetzung des Instrumentierungsventils . . . . . . . . . . . 30
7.5 Reinigung . . . . . . . . . . . . . . . . . . . . . . . 30
8. Demontage, Rücksendung und Entsorgung 31
8.1 Demontage des EMICOgauge-Geräte-Hook-ups . . . . . . . . . . 31
8.2 Rücksendung . . . . . . . . . . . . . . . . . . . . . . 32
8.3 Entsorgung . . . . . . . . . . . . . . . . . . . . . . . 32
9. Technische Daten 33
Inhalt
Konformitätserklärungen nden Sie online unter www.wika.de.
WIKA-Zusatz-Betriebsanleitung für EMICOgauge 19
14525888.01 04/2022 EN/DE/FR/ES

DE
1. Allgemeines
WIKA-Zusatz-Betriebsanleitung für EMICOgauge20
14525888.01 04/2022 EN/DE/FR/ES
Ergänzende Dokumentation
Diese Zusatz-Betriebsanleitung gilt im Zusammenhang mit folgenden Dokumenten:
▶
Betriebsanleitung des angebauten Manometers, z. B. „Mechanische Druckmessgerä-
te“, Artikelnummer 2408976
▶
Betriebsanleitung des angebauten Instrumentierungsventiles, z. B. „Absperrventile
und Ventilblöcke“, Artikelnummer 14294094
▶
Dokumente der verwendeten Bauteile
1. Allgemeines
■
Das in der Betriebsanleitung beschriebene Gerät wird nach dem aktuellen Stand der
Technik konstruiert und gefertigt. Alle Bauteile unterliegen während der Fertigung stren-
gen Qualitäts- und Umweltkriterien. Unsere Managementsysteme sind nach ISO 9001
und ISO 14001 zertiziert.
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät. Voraus-
setzung für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshinweise
und Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvorschrif-
ten und allgemeinen Sicherheitsbestimmungen einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des
Gerätes für das Fachpersonal jederzeit zugänglich aufbewahrt werden. Betriebsanlei-
tung an nachfolgende Benutzer oder Besitzer des Gerätes weitergeben.
■
Das Fachpersonal muss alle mitgelieferten Betriebsanleitungen vor Beginn aller Arbei-
ten sorgfältig durchgelesen und verstanden haben.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
■
Technische Änderungen vorbehalten.
■
Werkskalibrierungen / DKD/DAkkS-Kalibrierungen erfolgen nach internationalen
Normen.
This manual suits for next models
6
Table of contents
Languages:
Other WIKA Measuring Instrument manuals