WUNDER RS300 User manual

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 1
www.wunder.it
ELECTRONIC HANGING SICK LIFT SCALE MOD. RS300
USER MANUAL
Read this manual carefully before using the instrument
INDEX
2. GENERAL RULES..................................................................................................................................... 2
3. SAFETY.................................................................................................................................................... 3
4. FEATURES ............................................................................................................................................... 7
5. KEY PANNEL............................................................................................................................................ 7
6. WEIGHING OPERATION ......................................................................................................................... 8
7. FUNCTIONS SETTING............................................................................................................................. 9
8. OPERATION FLOW ................................................................................................................................. 9
9. SETTING UP YOUR SCALE STEP BY STEP ............................................................................................. 10
10. INSTRUCTION FOR BATTERY INSTALLATION..................................................................................... 10
11. WARNING INFORMATION ................................................................................................................. 11
12. CLEANING & MAINTENANCE ............................................................................................................ 11
13. DISPOSAL OF THE SCALE ................................................................................................................... 11
14. WARRANTY-LIABILITY ........................................................................................................................ 12
15. IDENTIFICATION PLATES.................................................................................................................... 12

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 2
By choosing the WUNDER mod.RS300 professional electronic hanging scale, you have purchased a high
precision instrument.
Since over 40 years Wunder has placed its experience at the service of health.
This instrument is compliant with national standards in hospitals and clinics with medical class I with
measurement function and is calibrated in conformity
with accuracy class III.
The instrument is equipped with large LCD electronic terminal with function Hold weight.
2. GENERAL RULES
Carefully read this manual before using the instrument as it supplies
important indications concerning OPERATING SAFETY AND MAINTENANCE.
The descriptions and illustrations supplied in this manual are not binding.
WUNDER reserves the right to modify its products as deemed convenient in order to approve them without
being committed to update this publication.
Conventions:
The following symbols have been used in this manual:
0476 EC TYPE APROVAL FOR MEDICAL USE
METROLOGICAL CERTIFICATION AND APPROVAL
ACCURACY CLASS
ATTENTION!
PLACED BEFORE DETERMINING PROCEDURES. COMPLIANCE FAILURE CAN HARM THE OPERATOR
OR PATIENT OR DAMAGE THE PRODUCT
WASTE DISPOSAL
TYPE B PARTS SUPPLIED
BATTERY POWER
INDICATION OF WEIGHT FUNCTIONALITY
O
INDICATION OF STABLE WEIGHT
INTERFERENCE COULD OCCUR NEAR THE APPLIANCES
DUAL INSULATION (CLASS II)

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 3
3. SAFETY
ATTENTION!
Before putting the device into use, please read with care the information given in the Operating
Instructions. They contain important instructions for installation, proper use and maintenance of
the device. The manufacturer shall not be liable for damages arising out of failure to heed the
following instructions:
WARNING!
IMPORTANT DIRECTIONS
The assembling of RS300 dynamometer must be effected by skilled operators only and
before the use be sure that all hanging sick lift elements, dynamometer and sling are
properly assembled, to weight the patient on safety conditions. A faulty assembling of
these elements could cause the risk of falling with serious consequences for the patient.
After assembling phase and before weighing operations, the operator must be sure that:
•RS300 dynamometer and the sling are on VERTICAL POSITION
•During weighing operation the hanging sick lift scale is not moving.
•The operator must ALWAYS follow and hold the sling of the patient during lifting
phase, to avoid torsion and sudden movement of both dynamometer and sling that
could cause breakages and irreparable damages, with the risk of serious
consequences for the patient.
•DO NOT MOVE the hanging sick lift scale when the patient is leaned.
During weighing phase by means of RS300 dynamometer, the operator must be sure that
the patient remains still, to avoid any instruments torsion and the following uncorrect
weighing.
After weighing phase the operator must ALWAYS follow and hold the sling of the patient,
to avoid torsion and sudden movement of both dynamometer and sling that could cause
breakages and irreparable damages, with the risk of serious consequences for the patient.

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 4
-When using electrical components under increased safety requirements, always comply with the
appropriate regulations.
- Improper installation will render the warranty null and void.
- Ensure the voltage marked on the power supply unit matches your mains power supply.
- This device is designed for use indoors.
- Observe the permissible ambient temperatures for use
- The device meets the requirements for electromagnetic compatibility. Do not exceed the maximum values
specified in the applicable standards.
- Ensure that the patient does not lean against the device risk of falling! If you have any problem, contact
your local service partner.
-The maximum scale loading must not be exceded. The device beeps when loading exceeds its maximum
capacity.
3.1 Manufacturer's guide and declaration –electromagnetic immunity
Guide and Statement of manufacturer –Electromagnetic emissions
The electronic scales RS300 model is scheduled for operation in the electromagnetic environment specified
below.
the customer and the user should ensure that it is used in that environment
Emission test
Conformity
Guide to electromagnetic environment
RF Emission CISPR11
Group 1 Class B
The scales RS300 cable model uses RF energy only for its
internal function. The refore its RF emissions are very low and
probably not cause no interference in electrical equipment.
RF Emission CISPR11
Group 1, Class B
The scales RS300 is suitable for use in all sanitary and hospital
buildings, connected to the low voltage public power supply
network.
Harmonic emission
Class A
Voltage fluctuations/
flicker emission
Compliant
Guide and Statement of manufacturer-Electromagnetic Immunity
The electronic scales RS300 model is scheduled for operation in the electromagnetic environment
specified below.
the customer and the user should ensure that it is used in that environment.
Immunity test
Compliance
Guide to electromagnetic environment
Electrostatic
discharges (ESD)
IEC/EN61000 - 4 -2
6kV in contact
8kV in air
The floors should be made of wood, concrete or ceramic. If
the floors are covered in synthetic material, the relative
humidity should be maximum 30%.
Rapid
transients/burst
2kV power supply
The power supply should be of the type used typically in
commercial or hospital environments.

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 5
IEC/EN61000 - 4 - 4
Surge
IEC/EN61000 - 4 - 5
1kV differential
mode
The power supply should be of the type used typically in
commercial or hospital environments.
Voltage dips, short
interruption and
voltage variation
IEC/EN61000 - 4 - 11
<5%UT for 0.5 cycle
40%UT for 05 cycle
70%UT for 25 cycle
<5%UT for 5 sec
The power supply should be of the type used typically in
commercial or hospital environments. Note=UT is the value
of the voltage of the feed.
Power frequency
Magnetic field
IEC/EN61000 - 4 - 8
3A/m
NOTE: UT is the value of the power supply voltage.
CAUTION
The medical device requires particular electromagnetic compatibility precautions and must be installed and
used according to the information provided in the accompanying documents.
Manufacturer's guide and declaration - Electromagnetic emissions
The electronic scales RS300 model can be used in the electromagnetic environment specified below. The
customer or the user of the seat must ensure that it is used in compliance with the specifications below.
Emissions test
Compliance
Guide to electromagnetic environment
Radiated/
conducted /
emissions
IEC/EN61000-4-6
3Vrms 150kHz to
80MHz(for
appliances that are
not life supporting)
Equipment in RF communications and mobile laptops should ot
be used near any part of the tooth, including cables, except
when they meet the recommended separation distance
calculated by the frequency applied to the transmitter.
Recommended separation distance d = 1.2 √ P d = 1.2 √ P from
80 MHz to 800 MHz d = 2.3 √ P from 800 MHz to 2.5 GHz where
the P and maximum rated power output of the transmitter in
Watts (W) according to the manufacturer of the transmitter and
the recommended separation distance in meters (m).
The intensity of the field of a fixed RF transmitters, as
determined by an electromagnetic site, it could be lower than
the level of compliance of each range of frequency b. You can
check interference in proximity of equipment marked with the
following symbol:
Radiated/
conducted
/emissions
IEC/EN61000-4-3
3Vrms 150kHz to
80MHz(for
appliances that are
not life equipment)
With 80 MHz and 800 MHz is applied the higher frequency range. These guidelines may not apply in all
situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 6
people.
a) The intensity of the field for fixed transmitters such as base stations for radiotelephone mobile and
cordless phones) and land radio mobile, amateur radio equipment, radio transmitters in the AM and FM and
TV transmitters cannot be made theoretically and with precision. To establish an electromagnetic
environment due to fixed RF transmitters, it should consider the electromagnetic survey of the site. If the
field strength measured at the place where you use the instrument exceeds the applicable level of
compliance of the above, it should be placed under observation normal operation of the tooth divice. If you
notice abnormal performance, it may take additional measures as a different orientation or position of
instrument.
of frequencies from 150 kHz to 80 MHz should be less than 3 V / m.
Recommended separation distances between portable radio equipment unity.
The scales RS300 model is expected to operate in an electromagnetic environment in which they control the
noise radiated RF. The customer or the tooth may be of help to prevent electromagnetic interference by
ensuring a minimum distance between mobile communication devices and portable RF (transmitters) and
unity, as recommended below, with respect to the maximum power output of equipment radio.
Output power rating
of the transmitter W
Distance separating the frequency of the transmitter m
150 kHz a 80 MHz d =
1,2 √P
80 MHz a 800 MHz d =
1,2 √P
800 MHz a 2,5 GHz d =
2,3 √P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters with maximum rated power output not reported above, the recommended separation
distance d in meters (m) can be calculated using the equation applicable to the frequency of the transmitter,
where P and the maximum rated power output of the transmitter in Watt (W) according to the manufacturer
of the transmitter.
Note: With 80 MHz and 800 MHz is applied the higher frequency range.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 7
4. FEATURES
Model
RS 300
Factory
Wunder sa.bi. srl- Trezzo s/Adda- Milano Italia
Max Division
Max 300kg, e=100g
OIML
Class III
Unit Weight
Kg (OIML)
Display
LCD 25mm con 5 digits
Dimension mm
120 x70x 160
Key & Functions
ON/ZERO/OFF, HOLD, TARE,
Power
6 alkalin batteries mini-stilo code AAA
Temperatura operativa
0°C / 40°C
ATTENTION!
Only the unit kg is approved by OIML (International Organization of Legal Metrology).
5. KEY PANNEL
FUNZIONI RS300:
HOLD = Indicatore blocco
peso
NET = peso netto
GROSS = peso lordo
kg = unità Kg
1. ON/OFF/ZERO/TARE:
1. Switch on/off the scale
2. Press this key to zero the scale. The zero range is +/- 2% of full capacity.
3. Press this key to tare when loads over zero range.
4. In tare mode, press this key again to clear the tarred value when there is no load on the platform.
5. Press this key without release 3 seconds to power off the scale.
2. HOLD:Press this key to lock weight value while weighing. Disable the hold function, press Hold key again.
3. NET :”Net” function indicator
4. HOLD:”Hold” function indicator
Tasti funzione
On/Off/Zero/Tare
Hold (blocco)
3. HOLD
4. NET
5. GROSS

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 8
6. WEIGHING OPERATION
Before reading detailed instructions on how to use all the weighing functions that are built into your scale,
please read the following important guidelines:
-Always be sure that the display shows `Zero` before use, if it does not then please press the ZERO key.
- The Professional Medical Hoist scale is designed to detect when a stable weight is achieved, your reading
should be taken at this point.
ATTENTION!
Tilt angle over 3° will cause false reading.
During weighing phase by means of RS dynamometer, the operator must be sure that the
patient remains still, to avoid any instruments torsion and the following uncorrect
weighing.
We recommend to not use any lift or Hoist that hanger bar do not swivel 360 degrees.
After assembly at the lifter, the dynamometer must not be rotated horizontally. Rotation is
only allowed in connection with 360 degree swivel bearings installed in the lift system.
ATTENTION!
For measuring the patient ALWAYS SAFE
The operator MUST place a pillow
under the patient raised
WEIGHING THE PATIENT MUST BE STRICTLY OBSERVED
INSTRUCTIONS PARAGRAPH NUMBER 3

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 9
7. FUNCTIONS SETTING
Auto-Off Time Adjustment
This feature affects the period of non-operation. Once exceed the setting time, the device will shut off
automatically.
Auto off time: 120 sec/180 sec/240 sec/300 sec/off
Buzzer Adjustment
This function affects the buzzer sound ON and OFF by user’s preference.
Buzzer: On/Off
8. OPERATION FLOW
A.OFF BEEP End
On Off
120 180 240 300 Off
:
Setting the time for auto time-off - 120/180/240/300/off (seconds)
Buzzer ON(beep on)/OFF(beep off)
Save the settings
Press and hold Hold key for 3 seconds.,
Display shows “Set”.
Notice:
Hold means ”select” “
” jump to another selection
horizontally.
On/Off/Zero/Taremeans “ enter” “
” jump to another
selection vertically.

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 10
9. SETTING UP YOUR SCALE STEP BY STEP
For Example: Set up 180s AUTO-OFF time and Buzzer OFF
Step 1. Press HOLD for 3 seconds to enter setup mode.
Step 2. Press ZERO to enter A.OFF mode when A.OFF is displayed
Step 3. Press HOLD to select 180 S which means AUTO-OFF TIME is 180 seconds.
Step 4. Press ZERO to decide and return to A.OFF mode and press HOLD to shift to BEEP setup mode.
Step 5. Press ZERO to enter BEEP setup mode and press HOLD again to select between ON and OFF .
Step 6. Press ZERO to decide and return to BEEP setup mode.
Step 7. Press HOLD to select END and press ZERO to finish setting.
10. INSTRUCTION FOR BATTERY INSTALLATION
RS uses six AAA size alkaline batteries. Please read the following instruction before using the scale.
1.Find the battery cover at 2.Remove the battery cover.
the backside of scale
3. Take out the battery case 4.Installing new AAA size batteries.
5. Reinstalling the battery case. 6. Installing the battery cover.

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 11
11. WARNING INFORMATION
1.Low battery indication This warning show that the voltage of battery is very low, please change
a new one.
2. The loading is over limit when power on, please reduce the loading.
3. The loading is under limit when power on, please increasing the loading.
4. Over load or Counting error–
The loading is over limit when power on, please reduce the loading and try again. If the
trouble is still exist, please call the service.
5. Counting error (too low)–
The loading is lower than limit when power on, please increase the loading and try again. If
the trouble is still exist, please call the service.。
6.EEPROM error--
The programs of the scale is error, please call the service.
12. CLEANING & MAINTENANCE
We would recommend using alcohol based wipes or similar when cleaning the scales. Please do not use
large amounts of water when cleaning the scales as this will cause damage to the scales electronics, you
should also refrain from using corrosive liquids or high pressure washers. Always disconnect the scales
from the mains power supply before cleaning.
The machine is sold approved with first metrological test passed (plate with green M). If required, the first
periodic test (green label) can be performed by the manufacturer. At the expiry date indicated on the green
label, the periodic test must be performed on the machine. Subsequent calibration is always necessary, if
one or more safety markings are damaged or the display shows abnormal weights. Always have maintenance
performed by qualified personnel. The WUNDER customer technical support service is at your disposal.
ATTENTION!
In some countries, calibration can be performed only by an authorised/qualified agent. Contact your dealer for
further
13. DISPOSAL OF THE SCALE
EU 2002/96/EC
This product complies with EU Directive 2002/96/EC. The symbol of the barred waste basket shown on the
appliance indicates that this product, which, must be treated separately from domestic refuse at the end of
its useful lifetime, must be taken to a differentiated collection centre for electrical and electronic devices, or
returned to the dealer when a new equivalent device is purchased. The user is responsible for taking the
appliance to the appropriate collection facility at the end of its lifetime. Appropriate differentiated collection
for submitting the used appliance to environmentally compatible recycling, treatment, and disposal helps to

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 12
prevent potential negative effects on the environment and on human health, and favours the recycling of
the materials of which the product is composed.
For more detailed information regarding the collection systems available, contact your local waste disposal
service or the shop where you purchased the scale.
As a consumer, you are obliged by law to return used or discharged batteries. You can deposit your old
batteries in the public collection points of your city, or you can deposit them at any retail facility selling
batteries of any type that has dedicated collection receptacles. Also in the case of the “demolition” of
electrical and electronic appliances, they must be removed and deposited in the dedicated collection
receptacles.
NOTE: The following symbols indicate the presence of harmful substances in the batteries:
Pb Pb = batteries that contain lead
Cd Cd = batteries that contain cadmium
Hg Hg = batteries that contain mercury
ATTENTION!
Do not throw the electrical parts or batteries into the household waste. Dispose of the batteries at the
collection centres in your area.
14. WARRANTY-LIABILITY
WUNDER products are guaranteed under warranty.
The warranty is valid for faults or malfunctions of the instrument due to manufacturing or material defects,
and WUNDER reserves the right to repair or replace the article.
In the event that the repairs of the fault or the delivery of the replacement unit do not have a positive
outcome, legal measures are to be considered valid. The term of guarantee is two years, taking effect on the
date of purchase.
The manufacturer shall not be held liable for damage deriving from any of the following causes: improper or
unsuitable storage or use; improper installation or placement in service by the owner or a third party; normal
wear and tear from use; changes or modifications; unsuitable or negligent manipulation; excessive use;
chemical, electrochemical or electrical interference or interference from humidity; unless such causes may
be attributed to negligence on the part of Wunder
15. IDENTIFICATION PLATES

WUNDER SA.BI.SRL EN USER MANUAL
RS 300 REV.01 10/01/2017 13
Table of contents
Other WUNDER Scale manuals

WUNDER
WUNDER ASEP-6 User manual

WUNDER
WUNDER VT30 User manual

WUNDER
WUNDER C201 Specification sheet

WUNDER
WUNDER AGS-1500 User manual

WUNDER
WUNDER CT20 User manual

WUNDER
WUNDER RA User manual

WUNDER
WUNDER C201 User manual

WUNDER
WUNDER RT 10 User manual

WUNDER
WUNDER BABY02 User manual

WUNDER
WUNDER RS User manual