Series 611T/612T Isolator User's Manual Universal DC Voltage/Current Input
___________________________________________________________________________________________
- 6 -
CAUTION: Do not insert sharp or oversized objects into the
push-button switch openings as this may damage the unit. When
depressing the push-buttons, use a blunt tipped object and apply
pressure gradually until you feel or hear the tactile response.
1. Connect a voltage or current source to the input, as required
(refer to Electrical Connections Drawing 4501-732).
2. Apply power and the module’s green “Run” LED will light.
3. If the yellow “ST” LED is also ON, then the reconfiguration
push-buttons must first be unlocked by simultaneously
pressing the “MODE” and “DOWN” button for one second.
The yellow “ST” LED should turn OFF (See Table 6). The
push-buttons may now be used to reconfigure the module.
4. Press and hold the “MODE” push button until the green
“RUN” LED turns OFF and the yellow “CH1” LED flashes. In
this mode, the unit is ready to accept a full-scale input signal
for transmitter input channel 1. If you do not wish to change
the full-scale parameter, skip to step 8.
5. Adjust the input source to the “full-scale” input signal level
according to your input range. The full-scale input signal
must be within the input range limits of the module. The
absolute value of this level must be greater than the
absolute value of the opposite endpoint—otherwise, see
Note 4.
6. Press the “UP” or “DOWN” push-button one time to cause
the module to auto-adjust the corresponding output level to
full-scale. If the output is not exactly at the desired full-scale
level, then each successive depression of the “UP” or
“DOWN” switch will increment or decrement the output
signal by a small amount. Holding the switch depressed will
increase the amount of increment or decrement. Note that a
reverse acting output response can be easily obtained by
using the push-buttons to adjust the output level accordingly.
For a reverse acting output, you would adjust the output
level to a lower level here in response to the full-scale input,
and a higher level in response to a zero input in Step 10.
See Note 4.
7. Press the “SET” push-button to accept the full-scale I/O
value. Note that every time the “SET” button is pressed, the
yellow “ST” LED will flash once and the full-scale I/O point
will be captured.
8. Press the “MODE” push button one more time. The green
“RUN” LED will still be OFF and the yellow “CH1” LED will
be ON, indicating that the unit is ready to accept the zero
value for input channel 1. If you do not wish to change the
channel 1 zero parameter, skip to step 12.
9. Adjust the input source to the “zero” input signal according
to your input range. The “zero” input signal must be within
the input range limits for the module. The module uses the
full-scale and zero levels to set the input span (span = full-
scale - zero). The absolute value of this level must be
less than the absolute value of the opposite endpoint—
otherwise, see Note 4.
10. Press the “UP” or “DOWN” push-button one time to cause
the module to auto-adjust its output level to a zero output
level. If the output is not exactly at the desired zero level,
then each successive depression of the “UP” or “DOWN”
switch will increment or decrement the output signal by a
small amount. Holding the switch depressed will increase
the amount of increment or decrement. Note that a reverse
acting output response can be easily obtained by using the
push-buttons to adjust the output level accordingly.
For a reverse acting output, you would adjust the output
level to a higher level in response to the zero input here, and
a lower level in response to a full-scale input in Step 6. See
Note 4.
11. Press the “SET” push-button to accept the zero I/O value.
Note that every time the “SET” button is pressed, the yellow
“ST” LED will flash once and the zero I/O point will be
captured.
12. Press the “MODE” push button a third time and the
program sequence for channel 1 will be completed.
Model 611T units will return to run mode at this point and
calibration is complete (the green “RUN” LED will turn ON).
You may skip ahead to step 14 to optionally lock-out
reconfiguration mode.
On 612T models, the green “RUN” LED will still be OFF, the
yellow “CH1” LED will turn OFF, and the yellow “CH2” LED
will flash. In this mode, the unit is ready to accept a full-
scale input signal for transmitter input channel 2. To
calibrate the second I/O channel, repeat steps 4-11
substituting channel 2 for channel 1. If you do not wish to
program channel 2, skip to step 13.
13. Press the “MODE” push button a fourth time on 612T units
and the module will return to run mode. The green “RUN”
LED will turn ON, the status LED is OFF, and the channel
LED’s will now reflect over/under range status. Calibration
is complete at this point. Note that during configuration, until
“SET” is pressed, pushing “MODE” will only toggle between
the full-scale & zero I/O sequence, without affecting their
values. Further, if no push-buttons are depressed within 2
minutes, then the module will automatically revert to run
mode (green “Run” LED will light) and no changes will be
made to the original zero or full-scale I/O settings.
14. OPTIONAL CONFIGURATION LOCK-OUT - At any time,
you may optionally lock-out the push-buttons to prevent
inadvertent reconfiguration or tampering in the field. The
unit may be locked out of configuration mode by pressing
the MODE and UP push-buttons simultaneously for 1
second. The status LED will turn ON to indicate that the
lock feature is enabled. You may unlock the configuration
mode by pressing the MODE and DOWN buttons
simultaneously for one second. The status LED will be
turned OFF to indicate the lock is disabled.
Notes (Configuration Procedure):
1. In configuration mode, the module’s LED’s are used to
indicate the program sequence as illustrated in Table 5:
Table 5: Configuration Mode LED Indication
LED
INDICATOR
LED ON
LED OFF
LED
FLASHING
Green “RUN” Run Mode Config Mode NA
Yellow “CH1” Calibrate
CH1 Zero
Calibrate
CH2
Cal CH1
Full-Scale
Yellow “CH2” Calibrate
CH2 Zero
Calibrate
CH1
Cal CH2
Full-Scale
Yellow “ST” NA NA One Flash as
SET pressed
2. The reconfiguration push-buttons may be alternately locked
out, or unlocked as illustrated in Table 6: