129591 03 2019-01-30 8
•Always make sure to position the power cord and other connected cables in a
secure way to prevent the risk of personnel falling.
•To prevent the calibrator from tipping over and causing damage to equipment or
systems, please avoid pulling the cables connected to the calibrator.
•Before switching on the calibrator make sure that it is set to the voltage of the
mains electricity supply.
•Always position the calibrator to enable easy and quick disconnection of the
power source (mains inlet socket).
•The connectors, on the front panel of the calibrator, must NEVER be connected
to a voltage source.
•Thermostats connected to the switch test input must not be connected to any
other voltage source during a test.
•Never use heat transfer fluids such as silicone, oil, paste, etc. in the dry-block
calibrators. These fluids may penetrate the calibrator and cause electrical
hazard, damage or create poisonous fumes.
•The calibrator must be kept clear within an area of 20 cm on all sides and 1
metre above the calibrator due to fire hazard.
•Use only insulation plugs supplied by AMETEK Denmark A/S.
•Never try to modify the insulation plugs to make them fit the sensor (CTC-1205
only)
Caution – Hot surface
This symbol is visible on the grid plate.
•Do not touch the grid plate, the well or the insertion tube as the calibrator is
heating up – they may be very hot and cause burns.
•
Do not touch the handle of the calibrator during use – it may be very hot and
.
When setting up the calibrator, you must… (cf. figure next page):
1. Place the calibrator on an even horizontal surface where you intend to use it and away
from all draughts. The calibrator must be kept clear within an area of 20 cm on all sides
and 1 metre above the calibrator due to fire hazard.
Caution…
Do not use the instrument if the internal fan is out of order
.
2. Ensure a free supply of air to the internal fan located at the bottom of the instrument
(pos. 1). The area around the calibrator should be free of draught, dirt, flammable
substances etc.
3. Check that the fuse size corresponds to the applied voltage on (pos. 2). The fuse is
contained in the power control switch (on/off switch (230V/115V)).
Warning
The two main fuses must have the specified current and voltage rating and be of the
specified type. The use of makeshift fuses and the short-circuiting of fuse holders
are prohibited and may cause a hazard.