3
*Avanti J-25 I and Avanti J-25, 50-Hz models
at 4°C (at 24°C ambient) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 000 rpm (69 700 ×g)
at 7°C (at 24°C ambient) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 000 rpm (75 600 ×g)
NOTE: At ambient temperatures above 24°C, the temperatures listed above may not be
achievable.
†The critical speed range is the range of speeds over which the rotor shifts so as to rotate about
its center of mass. Passing through the critical speed range is characterized by some vibration.
‡Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius
and speed (r
ω
2) to the standard acceleration of gravity (g) according to the following formula:
where ris the radius in millimeters,
ω
is the angular velocity in radians per second
(2
π
RPM /60), and gis the standard acceleration of gravity (9807 mm/s2). After substitution:
RCF rω2
g
---------=
RCF 1.12rRPM
1000
------------
⎝⎠
⎛⎞
2
=
JA-25.50 Rotor
JA-25.50 FIXED ANGLE ROTOR
SPECIFICATIONS
Maximum speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 000 rpm*
Critical speed range†. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 600 to 800 rpm
Density rating at maximum speed. . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 g/mL
Maximum allowable imbalance of opposing loads. . . . . . . . . . . . . 2 grams
Relative Centrifugal Field‡at maximum speed
at rmax (108 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 600 ×g
at rav (73.2 mm). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 200 × g
at rmin (38.5 mm). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 000 ×g
kfactor at maximum speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 418
Conditions requiring speed reduction . . . . . . . . . . . . . . . . .see RUN SPEEDS
Number of tube cavities (round-bottom). . . . . . . . . . . . . . . . . . . . . . . . . . 8
Available bottles and tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . see Table 3
Nominal tube dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 ×104 mm
Nominal tube capacity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 mL
Nominal rotor capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 400 mL
Approximate acceleration time to maximum speed
(rotor fully loaded)
in an Avanti®J centrifuge (to 25 000 rpm) . . . . . . . . . . . . . . 3 1/2min
in a J2 series centrifuge (to 21 000 rpm) . . . . . . . . . . . . . . . . 1 3/4min
Approximate deceleration time from maximum speed
(rotor fully loaded, max. brake)
in an Avanti®J centrifuge (from 25 000 rpm). . . . . . . . . . . . 1 3/4min
in a J2 series centrifuge (from 21 000 rpm). . . . . . . . . . . . . . 2 1/2min
Rotor weight, empty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1 kg (17.8 lb)
Rotor weight, maximum sample mass . . . . . . . . . . . . . . . . 8.7 kg (19.1 lb)
Rotor material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . aluminum
Rotor entry code for microprocessor-controlled J2 series
instruments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25.5
rmin
rav
rmax
34°
Axis of Rotation