
2. Fit angle disconnect wrench,
StarDental Part #203048, furnished
with every angle, to mating
indentations on angle neck; tighten
securely(FIG. 3).
1. Loosen knurled ring enough to disengage splines.
2. Pull head assembly up; turn to desired position.
3. Carefully mesh the splines on the knuckle with the
splines of the head (see FIG.1).
4. Tighten knurled ring onto knuckle threads until snug.
1. Open latch (FIG.5).
2. Insert a concentric .0925" diameter (2.350 mm) shank RA (standard
length for Contra Angle) bur the full length of chuck.
3. Engage latch lock firmly in latch of instrument.
4. Tug on bur to verify proper seating and retention.
INSTALLATION INTO THE MANUAL LATCH ANGLE
NOTE: A concentric .0925" diameter (2.350 mm) groove & flat shank (ISO 1797-1.Type 1) bur should be used.
NOTE: Before installing bur, install motor to the motor-to-angle adapter; then install latch angle to
motor-to-angle adapter as described in Angle Installation.
INSTALLATION INTO THE AUTO LATCH ANGLE
1. Grasp angle firmly so that your thumb touches the cap; push on center of cap (FIG. 4).
2. While continuing to hold cap down, insert a concentric .0925" diameter (2.350 mm) shank RA
(standard length for Contra Angle) bur the full length of chuck.
3. Rotate bur until fully engaged in latching mechanism and release thumb.
4. Tug on bur to verify proper seating and retention.
TITAN BALL BEARING
AUTO LATCH ANGLE HEAD
KNURLED RING
SPLINES
KNUCKLE
UNLOCKED
POSITION
DO NOT USE FOR ORAL SURGERY
Lubricate the Transmission Gear Daily
1. To disassemble angle: unthread knurled ring. Remove
knuckle and transmission gear from angle head.
2. Place a drop of DentaLube II in the two transmission gear
(FIG. 6) joints between the gear and gear housing and a
drop of DentaLube II in the hole in the center of the gear
housing. Rotate the gears to allow DentaLube II to work
into the bearings. Apply a light coat of DentaLube II to the
gear teeth.
3. To reassemble angle, insert transmission gear into angle
head. Engage splines of knuckle with mating splines of
head to achieve proper alignment.
4. Tighten knurled ring onto knuckle threads until snug.
5. Place angle on motor, connect to handpiece tubing; insert bur and run for 20 seconds to evenly distribute
lubricant and disperse any residual lubricant.
6. Proceed to Cleaning & Sterilization procedures before use.
REMOVAL FROM AUTO LATCH
Grasp angle firmly so that your thumb touches the cap; push on center of cap and hold cap down while
pulling bur straight out.
REMOVAL FROM MANUAL LATCH
Open latch, pull bur straight out.
Before operating the angle, carefully read and follow all of these instructions. Save these instructions for future
reference and observe all warnings and cautions.
MOTOR-TO-ANGLE
ADAPTER DRIVE GEAR
NOTE: If angle does not seat on adapter, gears are not meshed
correctly.Remove angle;spin motor-to-angle adapter drive
gear and repeat Step 1.
1. Thread angle onto the motor-to-angle adapter,
StarDentalTitan 3 Part #264073 or Titan T Part # 263970,
by hand. Seat completely.
ANGLE INSTALLATION
ANGLE ADJUSTMENT
BUR INSTALLATION AND REMOVAL
LUBRICATION - For Lubricated Angles ONLY
DISCONNECT
WRENCH
MATING
INDENTATIONS
BALL BEARING
MANUAL LATCH
ANGLE
MOTOR-TO-
ANGLE ADAPTER
Motor
MOTOR-TO-ANGLE ADAPTER
TITAN BALL BEARING
AUTO LATCH ANGLE
PUSH ON
CENTER OF END CAP
StarDental® Instrument Solutions
RED TIP APPLICATOR
KNUCKLE TRANSMISSION
GEAR KNURLED
RING
REPLACEMENT OF THE AUTO LATCH CARTRIDGE
NOTE: Remove transmission gear first before removing cartridge from head (FIG.8.)
1. Unthread knurled ring.
2. Remove knuckle and transmission gear from head.
3. Unthread end cap using end cap wrench, StarDental Part #261339 (FIG. 7). Note
the order and orientation of components when removed from head. Reverse the
order when reassembling cartridge.
End Cap
Wrench
CARTRIDGE REPLACEMENT
CAUTION
TO MINIMIZE RISK OF SERIOUS INJURY:
• Always insert bur the full length of the chuck. If you attempt to extend it from the chuck, the bur may
dislodge during use and possibly injure you or your patient.
• Use only recommended burs to prevent damage to the autochuck and in some cases to prevent burs from
improper or incorrect placement. An improperly clamped bur may dislodge during use, causing injury to you
or your patient. A sudden high-pitched sound or increased vibration while handpiece is operating alerts the
operator to the possibility that the chuck is not latched properly to retain the bur or a bent bur is inserted.
TO MINIMIZE RISK OF INJURY OR DAMAGE TO ANGLE:
• Replace cartridge when handpiece does not run at full speed or runs noisily.
• Follow these safety instructions:
- Confirm bur is secure by tugging on it after it is loaded
- Verify concentric rotation by operating handpiece outside oral cavity. Do not use if it does not
run concentrically or runs with excessive noise.
- Do not operate angle without an installed bur.
- Always use burs with overall length of less than 24mm. Shank length must be at least 7.3mm.
- When using auto chuck angle, never depress end cap while bur is rotating.
NOTICE LubeFree Angles denoted by having 3 rings on neck - see image on first panel - are lubricated
for life. Never clean ultrasonically or immerse in solvent that would dilute the lubricant.
Motor 5K 5K 20K 20K 20K
Accessories M TA 16:1 Endodontic
Motor to Angle
Adapter
M TA Torque Multiplier
& MTA
Torque Multiplier &
16:1 Endodontic Motor
to Angle Adapter
Ratio 1:1 16:1 1:1 4:1 64:1
Max Allowable RPM 5,000 300 20,000 5,000 300
4. Pushing from the front of the head, remove the cartridge by using a 3/32" pin (FIG. 10).
5. Clear housing of all dirt, debris, and/or remaining parts of previous cartridge. Install new wave
washer - supplied with new cartridge. Insert new cartridge by installing gear and small bearing end
first. Slide cartridge into housing with alignment tab of cartridge mating with groove in head. Install
end cap.
6. To reassemble angle, insert transmission gear into angle head. Engage splines of knuckle with
mating splines of head to properly align.
7. Tighten knurled ring onto knuckle threads until snug.
REPLACEMENT OF THE MANUAL LATCH CARTRIDGE
1. Unthread knurled ring.
2. Remove knuckle and transmission gear from head.
3. Unthread end cap using plastic wrench (FIG. 9).
4. Position latch arm in the unlocked position.
5. At rear of head, push out old cartridge using a 3/32" pin
(FIG. 10).
6.
Clear housing of all dirt, debris, and/or remaining parts
of previous cartridge.
Insert new cartridge with alignment
tab of cartridge
and gears toward end cap and align with mating
groove in head. Install
end cap (FIG. 11).
7. To reassemble angle, insert transmission gear into angle head. Engage
splines of knuckle with mating splines of head to achieve proper alignment.
8. Tighten knurled ring onto knuckle threads until snug.
PLASTIC WRENCH
GEARS
END
CAP
HEAD
3/32"
PIN
MATING
GROOVE
ALIGNMENT
TAB
NOTE: Remove transmission gear first before removing
cartridge from head.
Used intraorally by trained dental professionals for drilling and preparation of dental cavities for restoration,
such as fillings, disking, cavity and crown preps, polishing, post and pin drilling and pin setting.
062732 Rev P Date of Issue 2023-01-30 ©2023 DENTALEZ, Inc.
TITAN LOW-SPEED
BALL BEARING LATCH ANGLES
TITAN 3 LUBEFREE BB AUTO LATCH
TITAN 3 LUBRICATED BB AUTO LATCH
TITAN-T LUBEFREE BB AUTO LATCH
TITAN 3 LUBEFREE BB MANUAL LATCH
TITAN 3 LUBRICATED BB MANUAL LATCH
TITAN-T LUBEFREE BB MANUAL LATCH
TITAN-T LUBRICATED BB MANUAL LATCH
INDICATIONS FOR USE
Caution: U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
Under normal operating conditions, air is exhausted from the back of the head of this
handpiece. If air is directed into an open soft tissue wound or beneath the mucosa or
dermis, injury to the patient could result from air emphysema of air embolism.
This angle has no internal air or water cooling capabilities. External water for cooling maybe
required.
WARNING
WARNING
Sterilizable up to 135°C
135°C
CONTENTS
FEATURES
ANGLE ADJUSTMENT
ANGLE INSTALLATION
BUR INSTALLATION AND REMOVAL
SAFETY INSTRUCTIONS
LUBRICATION
CARTRIDGE REPLACEMENT
CLEANING AND STERILIZATION
TROUBLESHOOTING
LATCH REPLACEMENT
REPLACEMENT PART NUMBERS
WARRANTY
Rated for use with StarDental Motors andAttachments
Auto Latch or Manual Bur Latch Contra Angles
Smooth Running, Low Friction Performance
Field Replaceable Cartridges and Transmission Gears
Easy Maintenance
Long Term Reliability
Steam sterilization: 275° F (135° C) max.
FEATURES
FIG.1
FIG.2
FIG.3
FIG.4
FIG.5
FIG.6
WARNING
SAFETY INSTRUCTIONS
FIG.7
FIG.9
FIG.11
END CAP
GEAR AND
SMALL BEARING
CARTRIDGE ALIGNMENT TAB
MATING
GROOVE
WAVE
WASHER FIG.8
FIG.10
NON-STERILE
DENTALEZ, Inc., StarDental ®Division
1816 Colonial Village Lane
Lancaster, PA 17601 USA