manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. DentalEZ
  6. •
  7. Dental Equipment
  8. •
  9. DentalEZ Star LUBRICATED PROPHY ANGLE User manual

DentalEZ Star LUBRICATED PROPHY ANGLE User manual

Sterilizable up to 135°C
135°C
FEATURES
Rated for use with Star Motors and Attachments
Titan Five Star Prophy Angle:
Star Titan 3 Part #265803
Controlled Speed Reduces Prophy Paste ‘Spin-off’
Field Replaceable Cartridge
Quick, Easy Cleaning
Compatible with Titan 3 and Titan T Systems
Steam Sterilization: 135° C
When angle does not seat on
adapter, gears are not meshed correctly. Remove
angle; spin motor-to-angle adapter drive gear and repeat step one.
INSTALLATION AND OPERATION
PROPHY ANGLE INSTALLATION FOR TITAN 3 OR TITAN T SYSTEMS
1. For the Titan 3 or Titan T systems thread angle onto motor-to-
angle adapter (FIG. 1), Star Titan 3 Part #264073 or Titan T Part #
263970, by hand. Seat completely.
MOTOR-TO-ANGLE
ADAPTER DRIVE GEAR
DISCONNECT
WRENCH
MOTOR-
TO-ANGLE
ADAPTER
MATING INDENTATIONS TITAN
PROPHY ANGLE
2. Fit angle disconnect wrench (FIG. 2), Star
Part #203048, furnished with every angle, to
mating indentations on angle neck; tighten
securely.
SCREW-TO-
BUTTON
ADAPTER
PROPHY CUP INSTALLATION
The prophy angle uses either a screw-on or a
snap-on type polishing cup. When using a
snap-on type cup, use the screw-to-button
adapter, Star Part #258950.
Thread the screw-to-button adapter (FIG. 3), or
polishing cup, with a clockwise rotation until snug
into the hollow cartridge shaft of the prophy angle.
Operate the Titan motor in the forward direction only. If the
Titan motor is operated in reverse direction, the prophy cup
will separate from the motor.
Unthread the adapter, or polishing cup, counterclockwise. If the
screw-to-button adapter is difficult to disengage, remove the prophy cup, and use a
small screwdriver to loosen the screw-to-button adapter.
PROPHY CUP REMOVAL
Before use, verify concentric rotation by operating the handpiece outside the oral cavity in a
safe place.
PLEASE OBSERVE SAFETY INSTRUCTIONS
CAUTION
CAUTION
LUBRICATING
StarDental® Instrument Solutions
RED TIP APPLICATOR RED TIP APPLICATOR
StarDental® Instrument Solutions
1. Lubricate with DentaLube II, Star Part #262539 once a day or before each
sterilization procedure.
2. Remove prophy cup.
3. Place a drop of DentaLube II down the center of the hollow cartridge shaft (allow a few
seconds for the lubricant to drain down the shaft). (See FIG. 4)
4. Place a drop of DentaLube II at the base of the angle. Allow a few seconds for the
lubricant to drain and reach the bronze bushing. (See FIG. 5)
INDICATIONS FOR USE
Under normal operating conditions, air is exhausted from the back of the
head of this handpiece. If air is directed into an open soft tissue wound or
beneath the mucosa or dermis, injury to the patient could result from air
emphysema of air embolism.
This angle has no internal air or water cooling capabilities. External water
for cooling maybe required.
WARNING
WARNING
Motor Accessories Ratio Max Allowable
RPM
5K M TA 1:1 5,000
20K Torque Multiplier & MTA 4:1 5,000
Used intraorally by trained dental professionals for drilling and preparation of dental cavities for
restoration, such as fillings, disking, cavity and crown preps, polishing, post and pin drilling and
pin setting.
Caution: U.S. Federal law restricts this device to sale by or on the order of a
licensed physician.
058952 Rev U Date of Issue: 2022-11-01 ©2022 DENTALEZ, Inc.
LUBRICATED PROPHY ANGLE
DENTALEZ, Inc., StarDental ®Division
1816 Colonial Village Lane
Lancaster, PA 17601 USA
NON-STERILE
TITAN 5 STAR
DO NOT USE FOR ORAL SURGERY
Before operating the handpiece, carefully read and follow all of these instructions. Save
these instructions for future reference and observe all warnings and cautions.
CONTENTS
FEATURES
INSTALLATION AND OPERATION
SAFETY
LUBRICATING
CLEANING AND STERILIZATION
TROUBLESHOOTING
CARTRIDGE REPLACEMENT
REPLACEMENT PART NUMBERS
WARRANTY
FIG. 1
FIG. 2
FIG. 3
SAFETY INSTRUCTIONS
FIG. 4 FIG. 5
It is the responsibility of the user to use only FDA compliant sterilizing
equipment and accessories that are capable of meeting the manufacturers
recommendations below.
Do not exceed 135°C
Do not use:
• dry heat sterilizer • cold sterilizers • ultrasonic cleaners
•immersion techniques • sterilizer bags containing disinfectants
• detergents • surface wipes •surface sprays
• chemical disinfectors • hand soaps • surgical milk
LIMITS ON
REPROCESSING
Repeated processing has minimal effect on these Adapters. End of life is
normally determined by wear and damage due to use.
INSTRUCTIONS
The angle must be cleaned and sterilized before first use and after each
patient use.
CONTAINMENT AND
TRANSPORTATION:
It is recommended that angles are reprocessed as soon as is reasonably
practical following use.
PREPARATION FOR
CLEANING:
Wear puncture resistant gloves to prepare the prophy angle for sterilization.
Remove prophy cup or screw-to-button adapter from prophy angle. Remove
angel from motor-to-angle adaptor.
CLEANING
AUTOMATED:
Washer-disinfectors are not recommended.
CLEANING
MANUAL:
1. Clean and inspect once a day.
2. Clean all external surfaces
(FIG.6) with isopropyl alcohol
and a lint-free
cloth or brush.
3. Use the plastic wrench
to
unthread the end cap by turning clockwise.
(See FIG.7):
4. Remove the cartridge and
inspect as follows:
5. If prophy paste is not found on the
cartridge, internal cleaning is not
necessary. Simply wipe off with
lint-free cloth.
6. If prophy paste is found on the cartridge,
clean off the paste manually. The gear
assembly and end cap may be ultrasonically cleaned (FIG.8).
Do not ultrasonically clean any other components.
WARNING
CAUTION
CAUTION
CLEANING AND STERILIZATION
PLASTIC
WRENCH
UNTHREAD END CAP IN
CLOCKWISE DIRECTION
CARTRIDGE
NOTICE
New Style:
-c
34o and L
Marked On End Cap
-Unthread End Cap in
Clockwise Direction
Old Style:
-No Markings on End Cap
-Unthread End Cap in
CounterClockwise Direction
HOUSING
END CAP
GEAR ASSEMBLY
O-RING
BRONZE BUSHING
FIG. 6
FIG. 7
FIG. 8
MD
TROUBLESHOOTING
REPLACEMENT PART NUMBERS
Titan 3 Five Star Prophy Angle Cartridge Lubricated ........................................... 258948
(Includes cartridge and bronze bushing with O-ring)
Transmission Gear Assembly Lubricated.............................................................. 259008
Prophy Angle End Caps (Package of 2) Lubricated .............................................. 265805
Screw-to-Button Adapter ....................................................................................... 258950
Angle Disconnect Wrench..................................................................................... 203048
O-Rings (Package of 3) ........................................................................................ 258951
DentaLube II ......................................................................................................... 262539
Titan 3 Motor-to-Angle Adapter ............................................................................. 264073
Titan T Motor-to-Angle Adapter ............................................................................. 263970
CARTRIDGE REPLACEMENT
CARTRIDGE
GEAR
ASSEMBLY
BRONZE
BUSHING
1. Remove the end cap using the plastic wrench to turn it clockwise. (See illustration
under Cleaning and Lubricating.)
Note: Do not use pliers on the bronze bushing.
2. Pull bronze bushing to remove cartridge from prophy head as shown below in (FIG.
10). If cartridge is not removable by hand, thread a prophy cup or a screw-to-button
adapter in the cartridge and tug on it.
3. Separate gear assembly from the bronze bushing by pulling apart as shown above.
4. If gears are worn, replace cartridge, Star Part #258948. If not, simply wipe clean
with a lint free cloth.
5. Wipe the inside of the housing with a lint free cloth or cotton swab, barely
dampened with solvent.
Note: Do not immerse head in solvent.
6. Guide the shaft of the gear assembly through the center of the bronze bushing.
Install the assembled cartridge into the angle housing.
7. Position end cap and turn counterclockwise. Hand tighten to secure. Use wrench to
fully seat – 30° maximum.
Do not overtighten as poor performance and damage to angle may result.
PROBLEM: LOW POWER
Cause: Air leak at motor
Correction: Replace gasket;
refer to Titan Motor Instruction Manual.
Cause: Debris build-up
Correction: Refer to Cleaning and Lubricating section.
Cause: Faulty angle
Correction: Replace cartridge.
Note: If a problem persists, contact dealer for further assistance;
or, contact Star technical service at: 1-866-DTE-INFO (1-866-383-4636)
ALWAYS INSIST ON STAR PARTS FOR BEST PERFORMANCE.
DESCRIPTION ITEM NUMBER
NOTICE
NOTICE
CLEANING
MANUAL
CONTINUED:
7. When prophy paste is consistently found in the cartridge, install a new
bronze bushing, or a new cartridge, Star Part #258948, which includes
the bronze bushing. Replacing the bronze bushing renews the seal and
protects the gear and bearing from abrasive paste.
8. Wipe the inside of the housing with a lint-free cloth or cotton swab.
Note: Do not immerse head in solvent.
9. After cleaning, apply one drop of DentaLube II,
Star Part #262539, to the inside
diameter of the O-ring and to the gear as
shown. (See FIG. 9)
10. To reassemble, insert gear assembly into bronze bushing and guide
shaft through the hole in O-ring. Install cartridge into prophy head
angle.
11. Turn end cap counterclockwise and hand tighten to secure. Use
wrench to fully seat – 30° rotation maximum.
Do not overtighten as poor performance and damage to angle may result
LUBRICATION: Lubricate - See ‘Lubrication’ instruction section.
DISINFECTION: Use steam sterilization exclusively for complete sterility as outlined below.
MAINTENANCE: Angle should be routinely maintained, checked for corrosion, worn parts,
etc. and should be replaced on an as needed basis.
INSPECTION &
FUNCTION TESTING:
See Installation, Operating Recommendations, and/or Troubleshooting
sections of this manual.
PACKAGING: Single: A standard sterilization packaging material must be used. Ensure
that the pack is large enough to contain the item without stressing the seals.
WRAPPED
INSTRUMENT
STERILIZATION:
Gravity Type Autoclave: Pre-Vacuum
Autoclave:
• Minimum of 15 minutes at 132° C • Minimum of 4
minutes at 132° C
• Minimum of 30 minutes at 121° C
Do not exceed 135° C
DRYING: Dry the prophy angle for 30 minutes minimum using the sterilization dry
cycle and allow the prophy angle to cool to room temperature in bag prior to
use. If moisture is present, increase your sterilizer dry cycle appropriately.
STORAGE: Store in bag after sterilization until ready to use.
ADDITIONAL
INFORMATION:
Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use
only paper or paper/plastic bags.
MANUFACTURER
CONTACT:
See first or last panels of this manual.
FIG. 9 FIG. 10
LIMITED WARRANTY
O-Rings and Seals are specifically excluded from this warranty.
DENTALEZ warrants the Titan Lubricated Five Star Prophy Angle to be free of defects in material and
workmanship, under normal usage, under the following terms:
Star Products: Warranty Period*:
Lubricated Five Star Prophy Angle 3 Months from date of purchase
Please note the following additional terms of our warranty and return policy:
• Warranty covers manufacturing defects only and does not cover defects resulting from abuse,
accidents, misuse, improper handling, cleaning, care or maintenance or the failure to follow or
observe our operating/use, maintenance and/or installation instructions. Damage resulting from
the use of chemicals, cleaners, disinfectants or sterilization is not covered under this warranty
nor is normal wear and tear. Failure to use authorized parts or an authorized repair facility voids
this warranty.
• Liability is limited to repair or replacement of the defective part(s) or products at our sole discretion.
All other liabilities, in particular liability for damages, including, without limitation, special, exemplary,
consequential or incidental damages are excluded
• THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED
TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
*Provided conditions in warranty are met.
WARRANTY REGISTRATION
Please visit the website below to register your Star product.This will ensure proper registration and prompt
and accurate service should any of your purchases need evaluation or repair during the warranty period.
Star Products: WWW.DENTALEZ.com/warranty
FOR NEW PRODUCTS, MANUALS AND TECHNICAL INFORMATION, CALL
1-866-DTE-INFO OR VISIT OUR WEB SITE @ WWW.DENTALEZ.COM
IMPORTANT NOTICE ABOUT RETURNING PRODUCT TO DENTALEZ INC.
Anyone wishing to return any DENTALEZ Inc. product to a DENTALEZ facility MUST obtain the proper
RETURN AUTHORIZATION NUMBER from the Customer Service Department. The RA Number must
appear clearly marked on both the outside and inside of the returned product carton. When calling for
a return authorization number you will be asked to provide the following information:
To Obtain A
Return Authorization
Number Call
1-866-DTE-INFO
1. Product Name, Model Number or Part Number
2. Serial Number of Product
3. Dealer Return Purchase Order Number
4. Reason For Return
5. Copy of Invoice Showing Purchase
Items will be evaluated. Return Authorization Number is not a guarantee for credit. Your cooperation
with this procedure will help us expedite processing of any returns.

Other DentalEZ Dental Equipment manuals

DentalEZ StarDental 430 SWL Flex User manual

DentalEZ

DentalEZ StarDental 430 SWL Flex User manual

DentalEZ Star TITAN User manual

DentalEZ

DentalEZ Star TITAN User manual

DentalEZ Star User manual

DentalEZ

DentalEZ Star User manual

DentalEZ J-Chair J-3 User manual

DentalEZ

DentalEZ J-Chair J-3 User manual

DentalEZ Aeras Ramvac C3 User manual

DentalEZ

DentalEZ Aeras Ramvac C3 User manual

DentalEZ Star Aeras 500 Elite User manual

DentalEZ

DentalEZ Star Aeras 500 Elite User manual

DentalEZ Aeras 400 Flex User manual

DentalEZ

DentalEZ Aeras 400 Flex User manual

DentalEZ everLight EVLCE User manual

DentalEZ

DentalEZ everLight EVLCE User manual

DentalEZ Heritage User manual

DentalEZ

DentalEZ Heritage User manual

DentalEZ Forest User manual

DentalEZ

DentalEZ Forest User manual

DentalEZ StarDental Blis-sonic K User manual

DentalEZ

DentalEZ StarDental Blis-sonic K User manual

DentalEZ StarDental 430 SWL RA User manual

DentalEZ

DentalEZ StarDental 430 SWL RA User manual

DentalEZ Simplicity LED Light Head Manual

DentalEZ

DentalEZ Simplicity LED Light Head Manual

DentalEZ Aeras Star 500 Elite Handpice User manual

DentalEZ

DentalEZ Aeras Star 500 Elite Handpice User manual

DentalEZ StarDental Concentrix FX User manual

DentalEZ

DentalEZ StarDental Concentrix FX User manual

DentalEZ TruSim User manual

DentalEZ

DentalEZ TruSim User manual

DentalEZ Heritage Simplicity LED Light User manual

DentalEZ

DentalEZ Heritage Simplicity LED Light User manual

DentalEZ Star User manual

DentalEZ

DentalEZ Star User manual

DentalEZ AERAS Star User manual

DentalEZ

DentalEZ AERAS Star User manual

DentalEZ Star Titan Low-Speed User manual

DentalEZ

DentalEZ Star Titan Low-Speed User manual

DentalEZ aXcs Operatory Light User manual

DentalEZ

DentalEZ aXcs Operatory Light User manual

DentalEZ StarDental 430 SWL Torque User manual

DentalEZ

DentalEZ StarDental 430 SWL Torque User manual

DentalEZ RAMVAC Utility User manual

DentalEZ

DentalEZ RAMVAC Utility User manual

DentalEZ Aeras Ramvac RV4-12 User manual

DentalEZ

DentalEZ Aeras Ramvac RV4-12 User manual

Popular Dental Equipment manuals by other brands

Shining 3D Aoralscan 2 user manual

Shining 3D

Shining 3D Aoralscan 2 user manual

Bien Air CA 1:2.5 L MS Instructions for use

Bien Air

Bien Air CA 1:2.5 L MS Instructions for use

VERDER Carbolite Gero CDF 15/1b Installation, operation & maintenance instructions

VERDER

VERDER Carbolite Gero CDF 15/1b Installation, operation & maintenance instructions

Dürr Dental XR 24 Easy Installation and operating instructions

Dürr Dental

Dürr Dental XR 24 Easy Installation and operating instructions

DiO PROBO cure user guide

DiO

DiO PROBO cure user guide

Renfert Kulzer Dynamix speed 2 quick start guide

Renfert

Renfert Kulzer Dynamix speed 2 quick start guide

Woodpecker Endo Smart+ instruction manual

Woodpecker

Woodpecker Endo Smart+ instruction manual

mectron OzoActive Assembly

mectron

mectron OzoActive Assembly

3TECH LED-3200 user guide

3TECH

3TECH LED-3200 user guide

Planmeca Proline XC user manual

Planmeca

Planmeca Proline XC user manual

Ivoclar Vivadent Programat P700/G2 operating instructions

Ivoclar Vivadent

Ivoclar Vivadent Programat P700/G2 operating instructions

Tecno-gaz PEARL user manual

Tecno-gaz

Tecno-gaz PEARL user manual

SCHÜTZ MICROCLEAN instruction manual

SCHÜTZ

SCHÜTZ MICROCLEAN instruction manual

Yeti IQ Waxer NT light 230 V Instructions of use

Yeti

Yeti IQ Waxer NT light 230 V Instructions of use

SILMET SECURApplier Instructions for use

SILMET

SILMET SECURApplier Instructions for use

EMS AIRFLOW Prophylaxis Master quick guide

EMS

EMS AIRFLOW Prophylaxis Master quick guide

DentLight FUSION Instructions for use

DentLight

DentLight FUSION Instructions for use

Ivoclar Vivadent Programat CS2 Short instructions

Ivoclar Vivadent

Ivoclar Vivadent Programat CS2 Short instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.