Instructions for use - Infinity®Acute Care System - Infinity®Medical Cockpits 11
For your safety and that of your patients
WARNING
Risk of electric shock
Do not connect connectors with an
ESD warning symbol and do not
touch their pins without imple-
menting ESD protective measures. Such pro-
tective measures can include antistatic
clothing and shoes, touching a potential
equalization pin before and during connection
of the pins, or using electrically insulating and
antistatic gloves.
All users concerned must be instructed in
these ESD protective measures.
WARNING
Risk due to modifications
Modifications to the product may lead to mal-
functions and unforeseen risks. This may re-
sult in injury to the patient or the user or in
property damage.
Do not modify this product.
WARNING
No hardware modification is allowed without
the consent of the manufacturer. Hardware
modifications could compromise the safe
electrical and functional use of the product. If
the software of the product is modified
according to these instructions for use,
appropriate inspection and testing must be
conducted to make sure, the equipment is
safe to use.
WARNING
To ensure that the device is properly
grounded, connect the power cable to a hos-
pital-grade outlet.
WARNING
The Infinity Medical Cockpits are not suitable
for use in the presence of oxygen-enriched
atmospheres or flammable anesthetic mix-
tures.
WARNING
Because of the risk of electric shock, never
remove the cover of a device while it is in
operation or connected to a mains power
socket.
WARNING
Do not operate the device in areas such as:
magnetic resonance imaging (MRI) environ-
ments, aircraft, ambulance, home or hyper-
baric chambers.
WARNING
Connect only equipment to the analog and
digital interfaces (signal inputs and outputs)
of which the connected circuit has max.
24 VDC (max. 30 VDC for the system cable)
and is either a touchable SELV circuit accord-
ing to IEC 60950 or double insulated against
primary circuits according to IEC 60601-1.
Connect only passive (not separately ener-
gized) USB devices.
Any devices, or combination of devices, not
complying with the requirements mentioned
in these instructions for use may compromise
the correct functioning of the
Infinity Medical Cockpits. Prior to operating
the Infinity Medical Cockpits, consult the
respective documentation and instructions
for use of all connected devices or combina-
tion of devices.
Anyone connecting additional devices to the
signal input or output may configure a medi-
cal electrical system, and is therefore respon-
sible for ensuring that the system complies
with IEC 60601-1 (3rd edition, clauses 14 and
16) or IEC 60601-1-1 with IEC 60601-1-4, and
IEC 60601-1-2 and IEC 60601-1-8. If you have
any questions, contact Dräger-authorized ser-
vice personnel.