ECG
SHOW ALL LEADS
• Touch the HR parameter box
• Select Show All tab
CHANGE DISPLAYED LEADS
• Touch HR parameter box
• Next to Leads, touch the currently
displayed lead to be changed
• Touch desired lead to be displayed
•Repeat above to change other channels
displayed on the main screen
NIBP
• Touch NIBP parameter box
• Touch Interval Time
• Select the desired time interval
• Press NIBP Start/Stop button to begin
the interval setting
• To stop a measurement press NIBP
Start/Stop button
SPO2 MONITORING
• Touch SpO2parameter box
• Touch Waveform Size and use the
rotary knob to set appropriately
• Touch Pulse Tone Volume and Tone
Source buttons to activate
ACTIVATE PACEMAKER DETECTION
• Touch HR parameter box
• Touch Settings 2
•Touch Pacer Detection ON
** Note: Pacer Detection is not active when
ESU filter suppression is used **
MASIMO RAINBOW SETTECHNOLOGY
• Touch SpO2 parameter box
• Touch Pulse CO-Ox tab
• Under Parameter Display choose which
value will be the large font by selecting it
in first box. The other two values will be
smaller font
INVASIVE PRESSURES
• Connect hemo pod or cable to M540
• Ensure the transducer adapter cable is
attached to hemo pod
• Attach the transducer to the end of the
adapter cable
• The pressure will automatically appear
when connected
CHANGE IBP LABELS
• Touch the IBP parameter box
• Touch desired IBP tab
•Touch the Label button and select
desired label
ZERO TRANSDUCER
• Turn stoplocks off to patient and open
to air level equal with the phlebostatic
axis (except for ICP monitoring)
• Touch to IBP parameter box
• Touch Zero ->0<- button
CHANGE THE WAVEFORM SCALE
• Touch the IBP parameter box
• Touch desired IBP tab
•Select Scale size desired
ALARM LIMITS
•Touch any Parameter Box on monitor
screen
• Touch the column to be changed
• Utilise the Rotary knob or keyboard
arrows to change
• Press the Rotary knob or Enter to
confirm
•Verify the alarm is On/Off as desired