manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Dräger
  6. •
  7. Respiratory Product
  8. •
  9. Dräger PSS 7000 Series User manual

Dräger PSS 7000 Series User manual

3356233 (A3-D-P) Page 1 of 4
PSS®7000 Series
Self contained breathing apparatus with Sentinel 7000 Instructions for Use
i
For Your Safety
General safety information for the Dräger PSS® 7000 Series of self-contained breathing
apparatus (SCBA):
Use of this apparatus requires relevant equipment training; observance of this user
instruction (UI); and compliance with national regulations, laws and standards governing
the use of respiratory apparatus.
 UseofthisapparatusmustbeonlyforthepurposespeciedinthisUI,orasconrmed
in writing by Dräger.
 Onlytrainedandcompetentpersonnelshouldinspectandservicetheapparatusatspecied
intervals. Records of inspections and servicing are to be maintained in line with NFPA
1852 – Standard on Selection, Care, and Maintenance of Open-Circuit Self-Contained
Breathing Apparatus.
Use of the apparatus should be consistent with NFPA 1500 – Standard on Fire Department
Occupational Safety and Health Program.
The automatic alarm function of the Sentinel 7000 detects movement of the wearer and
movement or vibration to which the wearer may be subjected.
Do not use any form of chemical marking or paint on the equipment.
Dräger recommend that a service contract be obtained from your Dräger branch or agent.
Contact Dräger for details of service contracts and training courses.
Use only original Dräger spare parts and test equipment for service, repair and maintenance.
Notify Dräger if there is component fault or failure.
 All approved respiratory equipment shall be selected, tted, used, and maintained in
accordance with Mine Safety and Health Administration (MSHA), Occupational Safety
and Health Administration (OSHA), and other applicable regulations.
Before occupational use of this respirator a written respiratory protection program must
be implemented meeting all the local government requirements. In the United States
employers must comply with OSHA 29 CFR 1910.134 which includes medical evaluation,
training,andttesting.
DenitionsofAlertIcons
The following alert icons are used throughout this document to provide and highlight
areasoftheassociatedtextthatrequireagreaterawarenessbytheuser.Adenitionof
the meaning of each icon is as follows:
This icon indicates additional information on how to avoid inconveniences
when carrying out the instructions described.
This icon indicates a potentially hazardous situation which, if not avoided,
could result in physical injury to the user, or damage to the product. It
may also highlight an alert against unsafe practices.
This icon indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury to the user.
This icon indicates an imminently hazardous situation which, if not avoided,
could result in death or serious injury to the user.
Warranty and Liability Statements
Terms and conditions of warranty for the Dräger PSS® 7000 Series of SCBA and
associated accessories can be found in the original Dräger invoice document, or can be
obtained from Dräger on request.
The Dräger guarantee is void if the original sealing caps on the apparatus are tampered
with, removed or broken
Responsibility for the reliable function of the apparatus transfers to the owner or
operator when it is serviced or repaired by personnel not authorised by Dräger, or when
it is used in a manner not conforming to its intended use.
This device has been tested and complies with the FCC rules. Changes or
modications not expressly approved by the manufacturer will void your authority
to operate this product.
Technical Data
High-Pressure Connections
The following SCBA high-pressure (HP) connections are available for compressed air
cylinders:
2216 psi connection to CGA 346
4500 psi connection to CGA 347
Quick-connect cylinder coupling (2216 psi or 4500 psi).
Power Supplies
Main battery – 7.5 V
Back-up battery – 3 V
Head-up display battery – 3 V.
RIC UAC Connection
2216 psi or 4500 psi male, quick-release coupling with pressure relief valve (PRV).
LDR to Face mask Connection
Dräger push-in connector.
Pressure and Flow Details
Medium pressure – 87 psi to 130 psi
Airow–Inexcessof1000liters/minute
Airowat290psi–Inexcessof500liters/minute.
EOST Alarms
Activation commencement range (mechanical and electronic):
2216 psi cylinder – 600 psi to 510 psi
4500 psi cylinder – 1215 psi to 1035 psi.
General Information
Batteries
Danger of explosion. Do not change the batteries in an explosive or
ammable atmosphere.
Do not dispose of batteries in a re. Batteries must be disposed of in
line with local regulations.
Do not attempt to recharge any non-rechargeable battery.
Battery information:
Main battery – 7.5 V (5 x 1.5 V AA alkaline batteries)
Back-up battery – 3 V (CR123 lithium battery)
HUD battery – 3 V (CR123 lithium battery).
Use only the following approved battery types:
Procell by Duracell LR6 (1.5 V)
Duracell Plus LR6 (1.5 V)
PanasonicCR123AL/1BP(3V).
Batteries are supplied with the equipment but are not tted. Dräger
recommend that the batteries be removed when the system is not used
for long periods as a small amount of discharge still occurs. The normal
operating life of the batteries is dependent on usage time, frequency of
alarms, backlight illumination and ambient temperature.
When the SCBA is in operational use; should the display screen switch o
and the functionality of the HUD be powered by the back-up battery, then after
use of the SCBA Dräger recommend that the back-up battery is replaced.
The back-up battery will only supply power when the main battery is
disconnected or discharged. When this occurs, the back-up battery will only
supply power for HUD functions.
Preparation for Use
Equipmentcongurationsfornon-CBRNusearedetailedinaNIOSHApprovalTable
(refertoinstructionno.3356261).CongurationsforCBRNusearedetailedlaterinthis
instruction (refer to CBRN Use).
Installing the Back-Up Battery
Orientate the SCBA to access the battery compartment.
Unscrew and remove the battery cap (1, Fig 2), using a suitable coin.
 Insertthebattery,+veterminalendrst,intothebatterycompartment.
 Retandsecurethebatterycap.Do not overtighten.
Installing the Main Battery
Inspect the sealing rim around the battery terminals. Ensure that the terminals of the
battery, and the pressure transmitter module, are clean and undamaged.
Lift and turn over the SCBA to access the pressure transmitter module.
Insert the battery pack into the backplate recess. Refer to Figure 3.
 Positionthumbsontopofthetwoscrews,pushdownrmlytolockthebatterypack.
Refer to Figure 3 and 4.
 Whilstpushingdown,conrmthetwoslidinglocksmovetotheirlockedposition
viewed through the two keyholes as illustrated. Refer to Figure 5.
The Sentinel 7000 will emit a single tone and will commence the self-check
sequence (refer to Sentinel 7000 Self Check).
Press and hold the RH and LH buttons of the user interface until the display clears,
then immediately release the buttons.
A battery check is performed during the self-check sequence. If the
battery voltage is below a preset voltage the backlight will illuminate; the
low battery icon (Fig 6) will be displayed; the Sentinel 7000 will emit a
series of tones for approximately four seconds; and then the display will
switch o. When this occurs replace the battery.
To remove the battery insert and press the two pronged key (supplied
with the SCBA) into the two keyholes at the base of the battery. This will
open the locking latch allowing the battery to be removed.
Installing the Cylinder
The following instructions are for a screw-on cylinder coupling. Installation
of a quick-connect cylinder coupling is detailed in a separate UI (refer to
instruction no. 3356260).
Set the carrying-frame backplate to the short (S) position (refer to Adjusting the
Backplate Height).
Inspect and check the following:
The external thread of the cylinder valve port.
Theinternalthreadoftherst-stageregulatorhandwheel.
The O-ring seal in the pressure connector is in position and not damaged.
Theboretothesinteredlterinthe HPconnectoroftherst-stageregulatoris
clean and free from dirt and contamination.
Lay the carrying frame horizontal and fully extend the cylinder strap.
6
2580
2804
3
2797
MPa
psi
bar
9
2191
8
2006
2186
7
D
D
D
Take care to prevent impact damage as the cylinder valve is aligned with therst-stage
regulator handwheel.
 Insertthecylinder(valveendrst)throughthecylinderstrap,toalignthecylinder
valve with the handwheel.
Lift the cylinder and backplate into the vertical position (supported on the end of the
cylinder).
Align and fully tighten the handwheel (clockwise). Do not use tools or overtighten.
Place the unit back into the horizontal position.
Take up the slack of the cylinder strap and activate the cam-lock mechanism by
pulling the free end of the strap back over the cylinder (Fig 7).
Secure the strap using the Velcro fastening.
Adjusting the Backplate Height
Lift the SCBA into the vertical position.
Simultaneously press the two spring-loaded buttons (Fig 8) of the locking catch.
Slide the yoke to the required position (short (S), medium (M) or long (L)) and
release the spring-loaded buttons.
 Grasptheyokeandtheframeandattempttoraiseandlowertheyoketoconrm
that the locking catch is fully engaged.
Connecting the LDR and HUD
Insert and push the male coupling of the LDR hose into the quick-release coupling
of the SCBA medium-pressure hose until it latches into position. Check the security
of the connection (do not connect the LDR to the face mask at this stage).
Install a serviceable HUD into the face mask (refer to instruction no. 3356234).
i
i
NOTICE
i
i
NOTICE
!
CAUTION
i
i
NOTICE
i
i
NOTICE
!
CAUTION
!
DANGER
i
i
NOTICE
!
WARNING
Description and Intended Use
ThisUIdescribestheDrägerPSS®7000SeriesofSCBAttedwiththefollowing:
A Dräger Sentinel 7000 electronic monitoring system, with integral personal alert
safety system (PASS).
A Dräger wireless head-up display (HUD) unit.
The Dräger PSS® 7000 is a compact and lightweight apparatus that provides the wearer
withrespiratoryprotectionwhenworkingincontaminatedoroxygen-decientconditions.
The system can be used as a self-contained system or with an independent air supply
toincreaseeectiveoperationtime.Thesystemisfullycompatiblewithawiderangeof
compressed-air cylinders, face masks and lung demand regulators (LDRs).
The carrying system has a height-adjustable articulating backplate (6, Fig 1) that provides
improved comfort and ease of use, resulting in increased manoeuvrability and reduced
fatigue. The carbon-composite backplate incorporates a sliding height-adjustable yoke
thathasthreepresetpositions.Aexiblepivotjointlocatedatthebaseofthebackplate
(in line with the hip of the wearer) moves in response to the twisting and bending of the
wearer, improving the weight distribution and freedom of movement. The system includes
adjustable shoulder straps (2) and waist belt (11).
The Sentinel 7000 electronic monitoring system is a multi-function system that provides
continuous monitoring of the SCBA status including remaining cylinder pressure,
movement of the wearer, main battery condition, end-of-service time (EOST) and PASS.
It provides visual indications of system status and audible and visual alarms in warning
conditions. User control and monitoring of the system is through a user interface (1)
that incorporates switches, LEDs, a liquid crystal display (LCD) display screen and an
alarm sounder. A backlight illuminates the user interface display screen when required.
Additional alarm sounders (5), with warning LEDs top and bottom, are mounted on the
backplate and operate only during PASS alarms. The operating settings are preset and
non-adjustable by the user.
The PASS function of Sentinel 7000 electronic monitoring system is an alarm system
that can be activated manually or automatically. The automatic alarm uses a motion
sensor to detect movement and activate a pre-alarm and main alarm at timed intervals
when no movement is sensed. The manual alarm is activated by a press button on the
user interface. A limitation of the PASS is that it detects movement or vibration to which
the wearer is subjected.
Power supplies include a main battery located in the backplate, a back-up battery
located beneath the pressure module and a HUD battery located in the HUD. The
system is switched on by pressing the buttons on the user interface or by cylinder
pressure felt at a pressure module, with cylinder pressure transmitted to the pressure
module through a high-pressure hose (7). At switch on, a self test is performed and then
the system adopts the active mode where the PASS function is operational.
Allvariantsutilisethesamehigh-performance,rst-stage,regulator(9)ttedwithanEOST
mechanical whistle (10). The pressure regulator supplies medium-pressure breathing
air to a quick-release coupling (4) via a medium-pressure hose (3). Incorporated in the
rst-stageregulatorisarapidinterventioncrewuniversalairconnection(RICUAC)(8).
The RIC UAC is a male coupling that allows emergency replenishment of breathing air
to the air cylinder while wearer is breathing from the apparatus.
A wireless HUD unit (refer to instruction no. 3356234) located in the face mask
provides visual indications to the wearer of a number of system conditions. Signals from
theSentinel7000totheHUDarefromapressuretransducer/transmitterincorporated
in the pressure module.
!
WARNING
!
WARNING
1
3775
1
2
3
5
6
78
9
10
4
11
5
3
4
2798
5
2743
!
CAUTION
2
1
i
i
NOTICE
i
i
NOTICE
PSS®is a registered trade mark of Dräger
3356233 (A3-D-P) Page 2 of 4
PSS®7000 Series
Self contained breathing apparatus with Sentinel 7000 Instructions for Use
i
 Pressthefrontbutton(2,Fig15)oftheLDRandcheckthatairowisdeliveredinto
the face mask.
 Presstheresetbutton(1)tostoptheairow.
Close the cylinder valve and breathe normally to empty the system of pressure.
When the system is empty, inhale and hold breath for approximately 8 seconds. Readjust
and retest until the face mask contracts and holds on to the face with no air leaks.
Open the cylinder valve slowly, but fully, to pressurise system.
During Use
Do not commence any operation (including supplied-air respirator (SAR)
operations) using a cylinder that is less than 90% full. The eective working
duration of the apparatus is dependent on the capacity of the air cylinder
and the breathing rate of the wearer.
On activation if the unit fails to operate, or a low battery alarm activates,
then the relevant batteries should be renewed.
Regularly observe warning and display signals, and note the following:
For emergency help or assistance, press the yellow button in the center of the user
interface to activate the PASS alarm (see below).
To illuminate the display backlight, press and release the LH or RH button of the
user interface.
Remaining cylinder pressure is indicated on the user interface and the HUD LEDs.
The user interface shows the cylinder pressure digitally (numeric) and as an analog
segment display (Fig 19). The HUD shows the cylinder pressure with LEDs (refer
to the Table 1). The user interface digital display is the most accurate indicator.
Remaining time to EOST alarms is indicated on the user interface as a digital (numeric)
display.
React to the following alarm and warning signals as necessary:
EOST – The user interface will emit an audible alarm tone, and red and blue LEDs
willash;theredLEDontheHUDwillash;themechanicalwhistleontherst-
stage regulator will sound.
PASS pre-alarm – A repeating audible alarm tone
will be emitted from the
user
interface
sounder and the
additional alarm sounders. Move the user interface to
cancelthealarm(donotattempttousethebuttonstoswitchothepre-alarm.
PASS alarm – A
high-level sweeping alarm will be emitted from the
user interface
sounder and the
additional alarm sounders
; red and blue LEDs on the user interface
(Fig 10)
and top and bottom
of the
additional alarm sounders
(Fig13) will ash
intermittently; the user interface will show the alarm icon (Fig 20).
Simultaneously
press and hold the RH and LH buttons of the user interface to cancel the alarm.
Low main battery – A low battery icon (Fig 6) will be displayed on the user interface
ortheG/YbatteryLED(Fig16)willashyellow.
LowHUDbattery–TheG/YbatteryLEDwillashgreen.
LossofHUDcommunication–TheHUDblueLEDash
Additionalairowcanbedeliveredintothefacemaskasfollows:
Press and release the bypass button (3, Fig 15) to deliver a single jet of air into the
face mask.
Additionalairowrequired–Pressandrotatethebypassbuttontodeliverasustained
airsupply(85to130liters/minute)intothefacemask
Excessiveorlossofairow–Closethecylindervalvethenimmediatelybeginto
slowly reopen the valve. Use the cylinder valve as a regulating valve to set the air
owtomeettheuserrequirement.Thisprocedurecanalsobeusedwithratchet-type
cylinder valves.
The air ow procedures above are emergency measures that may greatly
reduce the operating duration of the air supply. When activated the
user must immediately evacuate to a safe area. The reason for using the
procedure must be investigated and repaired before reusing the apparatus.
Table 1 Head-up display LEDs
LEDs
Cylinder Contents R A G G
Full to 3/4 l l l l
3/4 to 1/2 l l l
1/2 (for 20 secs once) l ¬
1/2 to 1/4 l l
1/4 to approx. 100 psi ¬
Key: lOn
¬Flashing
R Red
A Amber
G Green
Personal Identity Information
Personal identity information (e.g. user name, brigade name, station number) can be
stored in the user interface memory. Once stored, the information is scrolled across the
screen by pressing the RH button of the user interface during use.
The personal identity information can be pre-programmed or uploaded from a personal
ID card during the self-check sequence. The procedure to upload information from a
personal ID card into the system memory is:
Locate and hold the personal ID card on to the back of the user interface, directly
behind the screen.
Press the LH + RHbutton of the user interface to start the self-check sequence.
When the personal identity icon (Fig 21) displays, immediately press the LH button.
Thescreenwill brieyilluminateandthe radialsegmentswillbegintoswitcho
clockwise.
The system will read, upload and store the data from the transponder embedded in
the card to the memory in the user interface. When the full data upload is complete,
the screen will display the tick icon (Fig 9).
Press and release the RH button of the user interface to check that the uploaded data
is correct. The user details will scroll from right to left across the screen (Fig 22).
Repeat the procedure if the information is incomplete or inaccurate.
 Ifrequired,switchobypressingandholdingtheLHandRHbuttonsoftheuser
interface until the display clears, then immediately release the buttons.
Further options for personal identity information and additional scrolling options are
available using the Dräger PSS 7000 PC Link Module. The Link Module is available as
an optional accessory that allows wireless reading and reprogramming of the system
operating settings, enabling additional functions, and downloading a record of events
(datalog). Contact Dräger for full details.
!
WARNING
i
i
NOTICE
!
WARNING
Fully loosen the head straps of the face mask and place the neck strap over the back of
the neck.
 Presstheresetbutton(1,Fig15)oftheLDRtoswitchothepositivepressure.
Open the cylinder valve slowly, but fully, to pressurise system. The Sentinel 7000 and
HUD systems will activate.
Putting on the Face Mask
Use only the mask sizes that have been conrmed by a quantitative t
test (QNFT).
Select the correct size of mask and inner mask.
Put on the face mask (for non-CBRN use, refer to instruction no. 9021649; for CBRN
use, refer to instruction no. 3356259).
!
WARNING
Pre-Operational Checks
Sentinel 7000 Self Check
The self check can be performed with or without a pressurised cylinder
tted.
The sentinel 7000 can be programmed to include a personal identity
option that allows reading user details from a personal ID card into the
system memory. When this option is enabled, the personal identity icon
(Fig 21) appears during the self-check sequence. Refer to the Personal
Identity Information section for further details.
Press the LH + RH button of the user interface, or open the cylinder valve of a pressurised
cylinder. The following self-check sequence will commence:
First stage – The Sentinel 7000 will emit a single tone from the user interface and
the additional alarm sounders, and the display backlight will illuminate.
Second stage – The display will show a tick symbol (Fig 9); the blue, red and green
LEDs (Fig 10) will illuminate.
Third stage – The display will show the cylinder type (Fig 11) (2216 psi or 4500 psi).
Fourth stage – The display will show the normal operating screen (Fig 12); the blue, red
and green LEDs (Fig 10) will illuminate; the blue and red LEDs (1, Fig 13) will illuminate.
End of sequence – Two ‘trill’ alarms will sound; the display will show the normal
operatingscreen(Fig12);thegreenLED(Fig10)willashatapproximatelyone
secondintervalstoconrmthattheSentinel7000haspassedtheselfcheckandis
in the active mode.
If a cross icon (Fig 14) with a fault code is displayed, the Sentinel 7000
has failed the self check. Contact Dräger for repair.
Press and hold the RH and LH buttons of the user interface until the display clears,
then immediately release the buttons.
High-Pressure Leak and EOST Warning Test
The following is not intended as a test of HUD functionality. If the HUD fails
to operate as described, refer to the HUD UI (instruction no. 3356234).
Place the face mask next to the SCBA (within three feet of the pressure module).
 Presstheresetbutton(1,Fig15)oftheLDRtoswitchothepositivepressure.
Open the cylinder valve slowly, but fully, to pressurise system. The Sentinel 7000
and HUD systems will activate as follows:
The Sentinel 7000 will emit a single tone and will commence a self-check sequence
(refer to Sentinel 7000 Self Check).
Up to approximately 45 seconds after the start of the self check, all six HUD LEDs
(Fig 16)willashtwicetoindicatethattheSentinel7000iscommunicatingwiththe
HUD.
Dependingoncylinderpressure,someofthelineoffourHUDLEDs(red/amber/
green/green)willash(onfor15seconds/ofor45seconds).
Close the cylinder valve and observe the user interface display. The pressure reading
shall not decrease more than 300 psi in 20 seconds.
If the SCBA fails this test, or an immediate leak is evident, do not continue
with the check. Vent the pressure, switch o the system and contact
Dräger for repair.
Cover the outlet of the LDR with the ball of the hand and press the front button (2, Fig
15).
Carefully lift the ball of the hand to slowly vent the system – observe the pressure
displayed on the user interface.
Note the pressures for the electronic EOST activation (indicated by an audible alarm
tone; ashing red and blue LEDs on the user interface; ashing red LED on the
HUD) and the mechanical whistle operation. Electronic and Mechanical activation
does not need to be simultaneous but both must be within the acceptable range (refer
to Technical Data – EOST Alarms). Refer to the Service Manual if the pressures are
outside the acceptable range.
Allow the system to vent to zero pressure. The display will show zero pressure and all
oftheuserinterfaceLEDs(Fig10)willashatapproximatelyonesecondintervals.
The green LED indicates that the system is still in active mode.
Press and hold the RH and LH buttons of the user interface until the display clears,
then immediately release the buttons.
 TheHUDredLED(Fig16)willashintermittently.
 Afterapproximately180secondsallsixHUDLEDswillashtwicetoindicatethatthe
unithasloggedofromtheSentinel7000.
 PresstheresetbuttonoftheLDRtoswitchothepositivepressure.
Check that the face mask port and the connector O-ring of the LDR are clean and
undamaged.
Align and push the LDR into face mask port until it latches in position.
Check the attachment by attempting to pull the LDR away from the face mask.
Normal Operation
All safety information, including the preliminary safety details, must be read
and understood before commencing any user task. The procedures detailed
in this UI may only be carried out by trained and competent SCBA users.
Before use checks
Carry out a visual inspection of the complete SCBA including any ancillary equipment that
will be used during the operation. The tasks detailed in the preparation for use and the
pre-operational checks sections must also be completed satisfactorily before release of
the SCBA for operational use. Components that are worn or damaged must be repaired
or replaced before proceeding. Do not use the equipment if any of the following is found:
TheSCBAfailstomeetanyofthetestspecications.
Any casing is cracked or broken.
The LCD display of the user interface cannot be read, or one or more of the display
segments is inoperative.
 Audibleand/orvisualalarmsarenotfunctioning.
Any fault icon shown.
Putting on the Apparatus
This apparatus may only be worn by trained and competent personnel.
The user must ensure that any accessories, ancillary equipment, turnout
gear and other clothing items do not interfere with the SCBA and do not
create a safety hazard.
If necessary, adjust the backplate height (refer to Adjust the Backplate Height).
Loosen the shoulder straps and waist belt and put on the SCBA.
Check that the shoulder pads are not twisted and take the weight of the system on the
shoulders by pulling the shoulder straps. Do not fully tighten at this stage.
Close the waist belt buckle and pull the ends of the waist belt forward until the strap
paddingtssecurelyandcomfortablyoverthehips(Fig17).Tuckthebeltendsbehind
the hip pads.
Pull the shoulder straps until the SCBA rests securely and comfortably on the hips. Do
not overtighten. Pull the strap retainers down to secure the strap ends (Fig 18).
i
i
NOTICE
i
i
NOTICE
i
i
NOTICE
i
i
NOTICE
!
WARNING
!
WARNING
!
WARNING
12
2192
psi
10
2200
BR
G
R B
2210
13
1
14
2201
17
2193
20
2198
psi
19
2202
psi
16
2189
R
A
G
G
B
G/Y
LED
11
2190
psi
15
2117
1
2
3
22
3387
21
2373
18
3088
psi
3356233 (A3-D-P) Page 3 of 4
PSS®7000 Series
Self contained breathing apparatus with Sentinel 7000 Instructions for Use
i
Carrying harness – Thorough cleaning
Do not immerse pneumatic or electronic components in cleaning
solutions or water.
Remove the harness from the carrying system and clean using one of the following
methods:
a. Clean manually in a bath containing recommended cleaning or disinfecting agents.
b. Machine wash (at 86 °F (30 °C)) using a suitable cleaning agent.
Rinse the harness thoroughly with clean water to remove all cleaning and disinfecting
agents.
Dry all components including internal parts.
Storage (Ready for Use)
Fully extend the shoulder straps, waist belt and head harness straps.
Store apparatus in a cool dry environment, free from dust and dirt (do not expose to
direct sunlight).
 WhentheSCBAisstoredwiththecylindertted,theapparatusmustbesupportedby
the cylinder and not the backplate.
Remove the batteries if the equipment is not used for long periods (a small amount of
discharge occurs during storage).
Storage preparation
Extend the shoulder straps, waist belt and the straps of the face mask.
For storage, place the face mask in a protective bag (contact Dräger for supply of a
suitable bag).
Route rubber hoses in such a way that the bend radius is not too acute and the hose
is not stretched, compressed or twisted.
 Withthesystemswitchedo,asmallamountofbatterypowerisconsumed.Ifthe
system is not to be used for a long period, remove the batteries.
Storage conditions
Store the equipment between 5 °F and 77 °F (-15 °C and +25 °C). Ensure that the
environment is dry, free from dust and dirt, and does not subject the equipment to wear
or damage due to abrasion. Do not store the equipment in direct sunlight.
Fix the breathing apparatus securely to any raised mounting point to prevent it from
falling.
Special instructions
Use of an Independent Air Supply (SAR Airline Connection)
Air quality must conform to the statutory requirements.
Use of an airline connection by a second person (buddy-breather) voids
NIOSH approval.
The time required for the wearer to escape to a safe area must be within the
remaining breathing time of the cylinder, taking into account the remaining
air content in the cylinder and the breathing rate of the wearer.
Independent air supplies must meet the following standards:
 Type-1gaseousairasdenedin:CGACommoditySpecicationforAir,G-71(grade
D or higher)
NFPA 1989 Standard on Breathing Air Quality for Fire and Emergency Services
Respiratory Protection
Air supply pressure – 87 psi to 125 psi
Airline hose length – 5 feet to 300 feet (maximum working hose length must not
exceed 12 individual hose lengths)
 Airlineowrate–550liters/minute
Approved for use at temperatures above minus 25 degrees Fahrenheit (minus 31.7
degrees Centigrade).
Sentinel 7000 alarms and warning signals will operate as normal.
The user interface display will show cylinder pressure when the cylinder
valve is open.
Turn on the independent air supply.
Connect the independent air supply coupling to the secondary supply hose (refer to
instruction no. 3355853) and breathe normally.
Close the cylinder valve (if the EOST warning signal activates, silence by taking several
short deep breaths or momentarily operating the LDR bypass button (3, Fig 15).
If any air supply problems are encountered, proceed as follows:
Open the cylinder valve to return to breathing from the attached cylinder.
Disconnect the independent air supply coupling.
Leave the hazardous area by the shortest and safest escape route, if necessary.
Cautions and Limitations
D – Air-line respirators can be used only when the respirators are supplied with respirable
air meeting the requirements of CGA G – 7.1, Grade D or higher quality.
E–UseonlythepressurerangesandhoselengthsspeciedintheUser'sInstructions.
I–Containselectricalpartsthatmaycauseanignitioninammableorexplosiveatmospheres.
J – Failure to properly use and maintain this product could result in injury or death.
M–Allapprovedrespiratorsshallbeselected,tted,used,andmaintainedinaccordance
with MSHA, OSHA, and other applicable regulations.
N – Never substitute, modify, add, or omit parts. Use only exact replacement parts in the
congurationasspeciedbythemanufacturer.
O–RefertoUser'sInstructionsand/ormaintenancemanualsforinformationonuseand
maintenance of these respirators.
S–SpecialorcriticalUser'sInstructionsand/orspecicuselimitationsapply.Referto
User'sInstructionsbeforedonning.
CBRN Use
TheDrägerPSS®7000SeriesiscertiedbyNationalInstituteforOccupationalSafetyand
Health (NIOSH), for limited chemical, biological, radiological or nuclear (CBRN) use and by
the Safety Equipment Institute (SEI) to meet the requirements of NFPA 1981. Approvals are
only valid when the apparatus is used with compressed air cylinders approved by NIOSH.
EquipmentcongurationsforCBRNusearedetailedinaCBRNApprovalLabel(referto
instruction no. 3356226).
IfanFPSNFPALDR,Drägerrecommendthataquantitativettest(QNFT)beperformed
onthefacemaskbeforeuseinaCBRNenvironment.Thettestmustbeconducted
strictly in accordance with the requirements outlined in the OSHA Respiratory Protection
Standard 29 CFR, Section 1910.134.
Q – Use in conjunction with personal protective ensembles that provide appropriate levels
of protection against dermal hazards.
R–SomeCBRNagents maynot presentimmediateeectsfromexposure,but can
result in delayed impairment, illness or death.
T – Direct contact with CBRN agents requires proper handling of the SCBA after each
use and between multiple entries during the same use. Decontamination and disposal
procedures must be followed. If contaminated with liquid chemical warfare agents,
dispose of the SCBA after decontamination.
U – The respirator should not be used beyond 6 hours after initial exposure to chemical
warfare agents to avoid possibility of agent permeation.
!
CAUTION
!
WARNING
i
i
NOTICE
After Use
Do not remove the SCBA until in a safe area. Do not drop or throw down
the SCBA as damage could occur.
Release and loosen all of the face mask straps.
Lift and remove the face mask.
Fully extend all of the straps of the head harness.
Close the cylinder valve.
Release the waist belt buckle. Lift the shoulder strap ends to release the strap
retainers (Fig 18) and then lift the shoulder strap buckles to loosen the straps.
Press the front button (2, Fig 15) of the LDR to vent system.
 Presstheresetbutton(1)toswitchothepositivepressure.
If the LDR has been set to bypass, press and rotate the bypass button (3) to switch
othebypass.
Press and hold the RH and LH buttons of the user interface until the display clears,
then immediately release the buttons. After approximately 180 seconds, all six HUD
LEDswillashtwicetoindicatethattheunithasloggedo.
Carry out a visual inspection of the complete SCBA including any ancillary equipment
used during the operation.
If required, lightly lubricate the LDR O-ring (recommended lubricant is Molykote 111).
 PasstheSCBAtotheservicedepartment,withdetailsofanyfaults/damagethat
occurred during use.
Removing the Cylinder
Do not attempt to remove the cylinder with the system pressurised.
Close the cylinder valve and fully vent the system.
Lift the free end of the cylinder strap from the Velcro fastening and lift the strap
against the cam-lock mechanism to release the strap tension. Loosen the strap.
 Unscrewthehandwheeloftherst-stageregulatorfromcylindervalve.
 Liftandcarefullyslidethecylinderawayfromrst-stageregulatortowardsthetopof
the backplate.
Remove the cylinder from the apparatus.
Inspect and recharge the cylinder (refer to Charging the Cylinder).
Charging the Cylinder
Air quality for compressed air cylinders must conform to the minimum
grade requirements for Type 1 gaseous air as dened in the CGA
Commodity Specication for Air, G-7.1 (Grade D or higher quality)
and where appropriate be in accordance with: NFPA 1989 Standard on
Breathing Air Quality for Emergency Services Respiratory Protection.
Recharge cylinders to the indicated working pressure marked on the
cylinder.
If the moisture content exceeds the recommended levels, ice particles
can form, reducing or blocking the airow.
Cleaning the SCBA
Cleaning agents and disinfectants listed below are not manufactured
by Dräger and have been reviewed only for compatibility when used to
clean or disinfect the subject Dräger product(s). Read and comply with
all instructions for use provided by the manufacturers of such agents
and disinfectants. Dräger expressly disclaims all responsibility for any
damage, personal injury or loss resulting from the use of such agents
or disinfectants.
Do not exceed 86 °F (30 °C) for washing, disinfecting and rinsing
solutions. Do not exceed 140 °F (60 °C) for drying, and remove
components from the drying facility immediately when dry. Drying time
in a heated dryer must not exceed 30 minutes.
Do not immerse pneumatic or electronic components in cleaning
solutions or water.
If water is trapped and then freezes inside the pneumatic system of the
breathing apparatus (such as the lung demand regulator), operation
will be impaired. Prevent any liquid from entering, and thoroughly dry
the breathing apparatus after cleaning to prevent this from occurring.
Refer also to the Instructions for Use for the lung demand regulator, face mask and other
associated equipment.
Manual cleaning of the breathing apparatus (USA)
Cleaning and disinfecting materials:
Cleaning agent – 1008 Green Liquid Hand Dish Wash
Disinfecting agent – 800 Spur-Tex Disinfectant Cleaner-deodorant (concentration: 1.6 %
(2ozpergallon))
Use only clean lint-free cloths
1. Prepare cleaning solution as per manufacturer’s instructions. Clean the breathing
apparatus manually using a cloth moistened with cleaning solution to remove excess
dirt.
2. Prepare disinfecting solution as per manufacturer’s instructions. Apply to all internal
and external surfaces, ensuring that all surfaces remain visibly wet for 15 minutes.
3. Rinse all components thoroughly with clean water to remove all cleaning and disinfecting
agents.
4. Dry all components using a dry cloth, in a heated dryer or in air.
5. Contact service personnel or Dräger if disassembly of pneumatic or electronic
components is required.
Manual cleaning of the breathing apparatus (Canada)
Cleaning and disinfecting materials:
Cleaning agent – mild soap solution
 Disinfectingagent–NeutralDisinfectantCleaner(concentration:0.5%(0.5ozper
gallon or 15 ml per 3.785 liters))
Use only clean lint-free cloths
1. Prepare cleaning solution as per manufacturer’s instructions. Clean the breathing apparatus
manually using a cloth moistened with cleaning solution to remove excess dirt.
2. Prepare disinfecting solution as per manufacturer’s instructions. Apply to all internal
and external surfaces, ensuring that all surfaces remain visibly wet for 10 minutes.
3. Rinse all components thoroughly with clean water to remove all cleaning and disinfecting
agents.
4. Dry all components using a dry cloth, in a heated dryer or in air.
5. Contact service personnel or Dräger if disassembly of pneumatic or electronic
components is required.
!
WARNING
!
WARNING
!
WARNING
!
WARNING
!
WARNING
!
CAUTION
!
CAUTION
!
CAUTION
!
CAUTION
S – Special or Critical Users’ Instructions
Approved for use at temperatures above minus 25 degrees Fahrenheit (minus 31.7
degrees Centigrade).
 Whenusedasacombinationsupplied-airrespirator/self-containedbreathingapparatus
(SAR/SCBA),notmorethan20percentoftheairsupplycanbeusedduringentry.
During supplied air use, the cylinder valve must remain closed. If the supplied air fails,
open the cylinder valve and immediately proceed to fresh air.
 Suppliedairsourcemustmeetthefollowingcriteria:pressure87–125psi,airowrate
atleast550liters/minute.
Important Note: If it is decided to exit the working area with the airline disconnected
or, in an emergency, if the air supply fails, breathe normally and immediately proceed
as follows:
Open the cylinder valve (counterclockwise) slowly, but fully and breathe normally.
Disconnect the hose of the independent air supply from the male coupling of the airline
hose connection. Breathe normally and immediately leave the hazardous area by the
shortest and safest route.
Safety Warning: The remaining duration begins from the time of opening the
cylinder valve and disconnecting the independent air supply. The time required to
allow the wearer to escape to a safe area must be within the remaining air capacity
(volume) of the cylinder taking into account the breathing rate of the wearer.
RIC UAC Use
Use of the RIC UAC should be by trained and competent personnel only.
 TheRICUACllinghoseisarequiredcomponentoftheNFPA1981certication.
OnlyuseallinghosewhichhasbeencertiedtoNFPA1981foruseinimmediately
dangerous to life or health (IDLH) atmospheres.
Do not use the RIC UAC connection for second person (buddy-breather).
Do not use the RIC UAC support pack to transfer air from one compressed air
breathing apparatus to another.
Do not allow oil, grease or other contaminants to contact the RIC UAC connection.
Do not attempt to disassemble or repair the RIC UAC connection.
Caution: The secondary air supply pressure to the RIC UAC must not exceed
maximumratedworkingpressureofthecylinder(s)beinglled.
Caution: If the pressure relief valve of the RIC UAC is activated, the SCBA must be
returned to the nearest Dräger branch or agent.
Caution:Ifaleakisdetectedwhilerellinginacontaminatedoroxygen-decient
gaseousatmosphere,stoprellingandimmediatelyleavethehazardousarea.
Contact Details
Any issues with the equipment, including damage, malfunction, or failure of the
breathing apparatus that may present a hazard to the user should be reported to
Dräger US Customer Service – Phone: 1-800-437-2437.
Contactwiththecerticationorganisationsmaybereachedat:
NIOSH, NPPTL
Phone: 1-412-386-4000
SEI (NFPA)
1307 Dolley Madison Blvd
Suite 3A, McLean
VA 22101
Phone: 1-703-442-5732
Equipment Maintenance Procedures
TasksidentiedinTables2or3asmaintenancetasksaretobecarriedoutbytrained
and competent personnel only (authorised by Dräger). Refer to the maintenance manual
for further information.
Equipment servicing
Service and test the breathing apparatus, including out-of-use apparatus, in accordance
with the Table 2. Record all service details and testing. Refer also to the Instructions for
Use for the lung demand regulator, face mask and other associated equipment.
Additional inspection and testing may be required in the country of use to ensure
compliance with national regulations.
3356233 (A3-D-P) Page 4 of 4
PSS®7000 Series
Self contained breathing apparatus with Sentinel 7000 Instructions for Use
i
3356233
© Dräger Safety UK Limited
Edition 20 – July 2017 (Edition 01 – March 2007)
Subject to alteration
Draeger Safety UK Limited
Ullswater Close Tel +44 1670 352891
Blyth, NE24 4RG Fax +44 1670 356266
United Kingdom www.draeger.com
Table 2
Equipment servicing
Component/
System
Task After use Every
month
Every year Every
10
years
Complete
apparatus
Clean and disinfect 
Visual inspection  
Functional testing  
Breathing cycle and static tests (see Note 1) 
Lung demand
regulator
Clean and disinfect (see Note 2) 
First-stage
regulator
Medium-pressure check (see Note 1) 
Inspectthesinteredlter(seeNote1andNote3) 
Inspect the high-pressure connector O-ring
(see Note 1 and Note 4)

Overhaul. Contact Dräger for the Repair Exchange (REX) service
(see Note 5)

Cylinder Charge cylinder to correct working pressure 
Check charged pressure (stored cylinders only) 
Check test date of cylinder (carbon composite cylinders over 15
years old must be retired)

Recertication According to national regulations in the
country of use
Cylinder valve Overhaul Atthetimeofcylinderrecertication
Notes
Dräger recommendations
1 These maintenance tasks may only be carried out by Dräger or trained service personnel. Details of the tests are contained in the
Technical Manual which is issued to service personnel that have attended a relevant Dräger maintenance course.
2 Lightly lubricate the O-ring of the lung demand regulator as required (recommended lubricant is Dow Corning 111 Valve Lubricant
and Sealant). Products other than the recommended lubricant are not tested and may damage the equipment.
3 Replacethesinteredlterifadropinrst-stageregulatorperformanceisobservedduringaowcheckorifitisvisiblydamaged.
4 Replace the high-pressure connector O-ring if it is found to leak during functional testing or if the O-ring is visibly damaged.
5 Where the breathing apparatus is subjected to a high level of use (in training establishments etc.), reduce the overhaul period
fortherst-stageregulator.Inthesecircumstances,Drägerrecommendthattheoverhaulfrequencyshouldbelessthan5,000
applicationsofuse.Anapplicationofuseisdenedasasingleuseofthefullyassembledbreathingapparatus,wheretheuser
breathes from the air cylinder. It does not include system pressurization for pre-operational checks.
Symptom Fault Remedy
Face mask air leak O-ring at LDR to face mask connection
Headstraps not tight
Exhalation valve leaking
Speech diaphragm defective
Renew or lubricate O-ring
Tighten
Maintenance task
Maintenance task
Unsatisfactory communication Speech diaphragm defective Maintenance task
High-pressure air leak Loose connections
Faulty or missing hose seals
Maintenance task
Maintenance task
Safety relief valve venting First-stage regulator defective Maintenance task
LDRallowingconstantairowinto
the fask mask
Bypass button engaged
Internal fault
Turnothebypassbutton(3,Fig15)
Maintenance task
Whistle not sounding correctly Whistle dirty or faulty Maintenance task
Whistle sounding continuously Damaged sealing on HP capillary
Defective activation mechanism
Maintenance task
Maintenance task
Low battery indication on the user
interface (Fig 6)
Low main battery Renew the main battery
Fault code indication on the user
interface (Fig 14)
Sentinel 7000 failure Maintenance task
Sentinel 7000 will not switch on Low main battery
Low cylinder pressure (below 145psi)
Unknown
Renew the main battery
Recharge cylinder to maximum working
pressure
Maintenance task
HUDbatteryLEDashinggreen Low battery - HUD Renew the HUD battery
HUDbatteryLEDashingyellow Low main battery Renew the main battery
HUD will not log on to the
Sentinel 7000
HUD not close enough
Low HUD battery
Low cylinder pressure (below 100 psi)
Unknown
Move the HUD to within three feet of the
pressure module
Renew the HUD battery
Recharge cylinder to maximum working
pressure
Maintenance task
Battery pack sliding locks not
moving to the locked position.
Dirty, damaged or faulty components. Attempt remedy actions in the order shown
below:
1.Pushdownrmlyonthebatterypack.
2. Remove and clean the battery pack and
backplate recess and retry.
3. Replace the battery pack and retry.
4. Contact Dräger.
Table 3
Troubleshooting



Other manuals for PSS 7000 Series

5

Other Dräger Respiratory Product manuals

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger X-plore Air FFP2 NR User manual

Dräger

Dräger X-plore Air FFP2 NR User manual

Dräger saver CF Series User manual

Dräger

Dräger saver CF Series User manual

Dräger PAS AirPack 1 User manual

Dräger

Dräger PAS AirPack 1 User manual

Dräger X-plore 4300 User manual

Dräger

Dräger X-plore 4300 User manual

Dräger FPS-COM 7000 User manual

Dräger

Dräger FPS-COM 7000 User manual

Dräger X-plore 9500 User manual

Dräger

Dräger X-plore 9500 User manual

Dräger X-plore 2100 User manual

Dräger

Dräger X-plore 2100 User manual

Dräger X-plore 9500 User manual

Dräger

Dräger X-plore 9500 User manual

Dräger X-plore 6000 User manual

Dräger

Dräger X-plore 6000 User manual

Dräger PARAT C box User manual

Dräger

Dräger PARAT C box User manual

Dräger X-plore 8000 User manual

Dräger

Dräger X-plore 8000 User manual

Dräger X-plore 6300 User manual

Dräger

Dräger X-plore 6300 User manual

Dräger Panorama Nova S User manual

Dräger

Dräger Panorama Nova S User manual

Dräger PAS Micro CFH User manual

Dräger

Dräger PAS Micro CFH User manual

Dräger X-plore 1700 User manual

Dräger

Dräger X-plore 1700 User manual

Dräger PAS Micro Series User manual

Dräger

Dräger PAS Micro Series User manual

Dräger Panorama Nova S-RP User manual

Dräger

Dräger Panorama Nova S-RP User manual

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger PSS 5000 series User manual

Dräger

Dräger PSS 5000 series User manual

Dräger Panorama Nova P - sideport - RA User manual

Dräger

Dräger Panorama Nova P - sideport - RA User manual

Dräger Oxy 3000 MK II User manual

Dräger

Dräger Oxy 3000 MK II User manual

Dräger FPS 7000 User manual

Dräger

Dräger FPS 7000 User manual

Dräger X-plore Series User manual

Dräger

Dräger X-plore Series User manual

Popular Respiratory Product manuals by other brands

3M 6000 instruction sheet

3M

3M 6000 instruction sheet

Bullard DC50 Series instruction sheet

Bullard

Bullard DC50 Series instruction sheet

Urrea USR05 User's Manual and Warrantly

Urrea

Urrea USR05 User's Manual and Warrantly

SATA Air star C operating instructions

SATA

SATA Air star C operating instructions

Venus P100 Fit and Use Instructions

Venus

Venus P100 Fit and Use Instructions

Trudell Medical International AeroChamber Plus Flow-Vu Chamber quick start guide

Trudell Medical International

Trudell Medical International AeroChamber Plus Flow-Vu Chamber quick start guide

SAFETY WORKS 9900 Instructions for use and maintenance

SAFETY WORKS

SAFETY WORKS 9900 Instructions for use and maintenance

Prizma Profi Sonic A manual

Prizma

Prizma Profi Sonic A manual

MSA FireHawk Ultra Elite 10048814 Maintenance and Repair

MSA

MSA FireHawk Ultra Elite 10048814 Maintenance and Repair

RPB Nova 3 instruction manual

RPB

RPB Nova 3 instruction manual

Honeywell NORTH 4000 Series Instructions for use

Honeywell

Honeywell NORTH 4000 Series Instructions for use

MSA Phalanx Alpha Instructions-parts list

MSA

MSA Phalanx Alpha Instructions-parts list

Pari Vios PRO Instructions for use

Pari

Pari Vios PRO Instructions for use

cleanAIR CHEMICAL 2F Plus user manual

cleanAIR

cleanAIR CHEMICAL 2F Plus user manual

cleanAIR Asbest user manual

cleanAIR

cleanAIR Asbest user manual

Rtech SPIRITUS PRO owner's manual

Rtech

Rtech SPIRITUS PRO owner's manual

INTERSPIRO Spirocom user manual

INTERSPIRO

INTERSPIRO Spirocom user manual

Omron C801KD instruction manual

Omron

Omron C801KD instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.