manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Dräger
  6. •
  7. Respiratory Product
  8. •
  9. Dräger PAS Colt PPH User manual

Dräger PAS Colt PPH User manual

PAS®Colt PPH
Combination escape/ChargAir respiratory protection equipment Instructions for Use
i
PAS®is a trademark of Dräger 3359701 (A3-D-P) Page 1 of 2
3359701
© Dräger Safety UK Limited
Edition 06 – August 2018 (Edition 01 – April 2011)
Subject to alteration
Draeger Safety UK Limited
Ullswater Close Tel +44 1670 352 891
Blyth, NE24 4RG Fax +44 1670 356 266
United Kingdom www.draeger.com
D
r
ä
g
e
r
'
220
200
150
50
100
bar
'
1
3653
72
4
3
6
5
1
3654
2
1
2
3394
31
3656
41
3657
5
3658
6
3669
7
bar
'
200 220
150
100
50
3661
8
3659
9
1 For your safety
1.1 General safety statements
●Before using this product, carefully read the Instructions for Use.
●Strictly follow the Instructions for Use. The user must fully understand
and strictly observe the instructions. Use the product only for the
purposes specified in the Intended Use section of this document.
●Do not dispose of the Instructions for Use. Ensure that they are
retained and appropriately used by the product user.
●Only fully trained and competent users are permitted to use this
product.
●Comply with all local and national rules and regulations associated
with this product.
●Only trained and competent personnel are permitted to inspect, repair
and service the product. Dräger recommends a Dräger service
contract for all maintenance activities and that all repairs are carried
out by Dräger.
●Properly trained service personnel must inspect and service this
product as detailed in the maintenance section of this document.
●Use only genuine Dräger spare parts and accessories, or the proper
functioning of the product may be impaired.
●Do not use a faulty or incomplete product, and do not modify the
product.
●Notify Dräger in the event of any component fault or failure.
1.2 Definitions of alert icons
Alert icons are used in this document to provide and highlight text that
requires a greater awareness by the user. A definition of the meaning of
each icon is as follows:
WARNING
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided,
could result in physical injury or damage to the product or
environment. It may also be used to alert against unsafe practices.
NOTICE
Indicates additional information on how to use the product.
2 Description
2.1 Product overview
The Dräger PAS Colt PPH (positive-pressure hood) provides respiratory
protection for use in a contaminated or oxygen-deficient environment to
escape to a safe breathing environment or to a designated holding area.
In the designated holding area the PAS Colt PPH air cylinder can be
recharged, and then kept fully charged, by a charging air source
(ChargAir). This allows the wearer to breathe normally in the holding area,
and then to escape a safe breathing environment when a full evacuation is
possible.
The main features of the product are:
●The carrying harness is a bandolier shoulder harness and waist belt.
●The air cylinder is a carbon-composite cylinder fitted in a hip-mounted
cylinder holster. The cylinder pressure is shown on the gauge (Fig 1,
Item 4), and when the cylinder is full it provides a nominal duration of
15 minutes.
●The cylinder valve has an automatic activation device that is operated
by removing a two-pronged locking clip (Fig 1, Item 1). The clip is
connected by a strap to a sprung-gate clip (Fig 1, Item 6) that
automatically pulls the locking clip as the escape hood is removed from
its storage pouch.
●The pressure reducer (Fig 1, Item 2) reduces the cylinder pressure to
a medium pressure, which is the pressure required at the lung demand
valve (Fig 1, Item 5).
●The ChargAir connector (Fig 1, Item 7) is a male coupling that is used
with a charging air source to rapidly refill the air cylinder while the
wearer is breathing from the equipment. The ChargAir connector
contains a pressure-relief valve (Fig 1, Item 3) that prevents over
charging and possible cylinder damage by opening to relieve pressure
at a set level above 200 bar. If the pressure-relief valve activates, the
breathing apparatus must be returned to Dräger for inspection and
repair for safety reasons.
●The ChargAir outputs in the holding area must be female Aeroquip
couplings (FD17-1003-04-04).
●The escape hood (Fig 2) has a rubber ring that seals around the neck,
and an internal nose cup that fits over the nose and mouth to supply
air to the wearer.
●The lung demand valve regulates the breathing air supply to the
wearer during use. The valve is switched on by the inhalation of the
wearer (first-breath activation), and can be switched off by pressing a
reset button (Fig 2, Item 1) when required.
2.2 Intended use
When this product is used with an approved air cylinder it provides the
wearer with respiratory protection for: escaping from contaminated or
oxygen-deficient conditions; or for remaining in a holding area with a
compatible charging air source.
The air cylinder used with this product must be a certified Dräger cylinder;
otherwise the operation of the product may be impaired. Contact Dräger for
further information.
Operating temperature range: -15 °C to 60 °C.
2.3 Approvals
The European standards, guidelines, and directives according to which
this product is approved are specified in the declaration of conformity (see
declaration of conformity or www.draeger.com/product-certificates).
!
!
i
i
3 Use
WARNING
The time required to allow the wearer to escape to a safe area
must be within the capacity of the equipment. When selecting the
type and duration of escape equipment it is essential to consider
potential hazards, storage location and escape route.
3.1 Preparation for use
WARNING
On receipt of the PAS Colt PPH from Dräger, the product is not
configured for immediate operational use. The cylinder automatic
activation device is not connected, and the air cylinder may be
discharged. Carry out the following procedure to prepare the product
for use.
Immediately following removal of the equipment from its packaging:
1. If the air cylinder is supplied discharged, see Section 4.2.5 for cylinder
charging instructions.
2. Fit the air cylinder and connect the automatic activation device (see
Section 4.2.2).
3.2 Putting on the carrying harness
Refer also to Fig 1.
1. Check that the pointer of the cylinder pressure gauge is inside the green
area, and the cylinder automatic activation device is properly
connected (Fig 1, Items 1 and 6).
2. Open the waist belt buckle and fully extend the waist and shoulder
strap.
3. Place the left arm through the shoulder harness, taking the harness
over the head and on to the right shoulder, positioning the strap
diagonally across the body with the cylinder positioned against the left
hip.
4. Loop the waist belt around the waist and fasten the buckle – do not
tighten.
!
!
5. Hold the cylinder valve with the left hand and lift until the waist belt is
in line with the waist. Then tighten the waist belt strap until the
equipment is secure and comfortable on the waist. Pull down to adjust
the shoulder strap.
3.3 Putting on the hood and escape procedure
NOTICE
Remove protective head gear before putting on the hood.
1. Remove the hood from the pouch. This action removes the locking clip
(Fig 1, Item 1), automatically activating the cylinder valve.
2. Place both hands inside the neck seal and stretch the hood over the
head.
○Wearers with spectacles – take care when stretching the neck seal
over the spectacles.
○Wearers with long hair – tuck the hair inside the hood.
3. Position the nose cup over the nose and mouth, and inhale to activate the
breathing air supply. The duration of the cylinder air begins from the time of
the first-breath activation of the lung demand valve.
NOTICE
The actual breathing time available from the cylinder depends on the
wearer’s air-use rate. Higher breathing rates associated with
increased physical effort would reduce the available escape time.
The wearer must be able to reach the required area before the air
cylinder is empty.
4. Breathe normally and immediately proceed to one of the following
areas:
○A safe breathing environment (see Section 3.3.1).
○A designated holding area with a ChargAir breathing air supply
(see Section 3.3.2).
3.3.1 Safe breathing environment
WARNING
Do not remove the hood until in a safe breathing environment.
CAUTION
Do not drop or throw down equipment as damage could occur.
Once in a safe breathing environment:
1. Remove the hood and continue to breathe normally.
2. Press the reset button (Fig 2, Item 1) to switch off the air flow through
the lung demand valve.
3. Carefully roll the escape hood to ensure that the visor is not creased.
Insert the hood into the pouch with the visor in the lower portion of the
pouch as shown in Fig 1.
3.3.2 Designated holding area (ChargAir use)
WARNING
The ChargAir source shall meet the requirements for breathing air
according to EN 12021.
If the ChargAir system is used in a potentially contaminated
environment, ensure that all couplings are free of contaminants
before connecting the couplings.
CAUTION
The ChargAir source pressure must not exceed 200 bar. Using a
ChargAir source of higher pressure could activate the ChargAir
pressure-relief valve and damage the valve seat. If the ChargAir
pressure-relief valve activates, the breathing apparatus must be
returned to the nearest Dräger branch or agent for inspection and
repair at the first opportunity.
1. Ensure that the charging source outlet connector is compatible with the
ChargAir connector on the PAS Colt PPH (Fig 1, Item 7).
2. Remove the protection caps from the ChargAir connectors.
3. Ensure that the hoses are not bent or kinked and push the ChargAir
connectors together until a click is heard and the ring (Fig 3, Item 1) is
visible behind the sleeve on the female coupling.
○Filling the cylinder begins immediately.
○The cylinder is full when the pointer of the cylinder pressure gauge
is inside the green area. This can be within 60 seconds, although
the filling time and final pressure are dependent on the pressure
and volume of air in the cylinder and the charging air source.
○Due to rapid compression of cylinder air, a small temperature
increase occurs during refilling. The system will top-up the cylinder
as it cools if it remains connected to the charging air source.
4. Keep the ChargAir system connected and breathe normally until a full
evacuation is possible.
5. When required, disconnect the quick coupling by pulling the sleeve of
the female coupling away from the male coupling (see the arrow in
Fig 3). As the coupling separates, a hiss or pop may be heard as the
shut-off valves in the couplings close. The duration of the cylinder air
begins from the time of disconnecting from the ChargAir connector.
6. Immediately proceed to a safe breathing environment (see Section 3.3.1).
3.4 After use
After any use, the product must be checked and then returned to the ready-
for-use condition. The after use maintenance tasks (see Section 4.1.2)
must be carried out by Dräger or trained service personnel to prepare the
product for use.
i
i
i
i
!
!
!
!
PAS®Colt PPH
Combination escape/ChargAir respiratory protection equipment Instructions for Use
i
Dow Corning®and Molykote®are registered trademarks of Dow Corning Corporation. 3359701 (A3-D-P) Page 2 of 2
4 Preparing and maintaining the
PAS Colt PPH
The tasks in this section must be carried out to prepare and maintain the
PAS Colt PPH in a ready-for-use condition.
4.1 Maintenance periods
4.1.1 Daily checks
It is essential that escape apparatus is ready for use at all times, and
Dräger therefore recommend a daily check. If the customer’s on-site risk
assessment concludes that less regular checks are acceptable, this can be
extended to a maximum of one month. It is the customer’s responsibility to
ensure that the equipment is ready for use at all times.
●Check that the pointer of the cylinder pressure gauge is inside the
green area. Charge the cylinder if it is in the red area (see Section
4.2.5).
●Ensure that the automatic activation device of the cylinder valve is
properly connected (Fig 1, Items 1 and 6).
4.1.2 Maintenance table
Dräger recommend that regular inspection, testing and servicing of the
breathing apparatus is carried out in accordance with the table below. The
table applies also to out-of-use (stored) equipment.
Additional inspection and testing may be required in the country of use to
ensure compliance with national regulations governing the use,
maintenance, examination and testing of this product.
Notes
Dräger recommendations
1. Clean the equipment if it is dirty. If it the equipment has been exposed
to contaminants, disinfect any components that come into direct and
prolonged contact with the skin.
2. As a guide, lubricant should be felt on the fingers but not seen. If
relubrication is required, lightly apply Dow Corning®Molykote®111
(other lubricants are not tested and may damage the equipment).
3. These maintenance tasks may only be carried out by Dräger or trained
service personnel. Details of the tasks are contained in the technical
manual which is issued to service personnel that have attended a
relevant Dräger maintenance course.
4.2 Maintenance tasks
4.2.1 Visual inspection
Carry out a visual inspection, checking the full breathing apparatus
including all component parts and accessories. Check that the equipment
is clean and undamaged, paying particular attention to pneumatic
components, hoses and connectors. Typical signs of damage that may
affect the operation of the breathing apparatus include impact, abrasion,
cutting, corrosion and discoloration. Report dirty or damaged equipment to
Dräger or trained service personnel, and do not use until faults are rectified
and the product is cleaned.
4.2.2 Fitting the cylinder
CAUTION
Fitting a cylinder that is charged above 200 bar could activate the
ChargAir pressure-relief valve and damage the valve seat. If the
ChargAir pressure-relief valve activates, the breathing apparatus
must be returned to the nearest Dräger branch or agent for
inspection and repair at the first opportunity.
1. Ensure that the cylinder is fully charged, with the pointer of the cylinder
pressure gauge inside the green area.
2. Check that the threads of the valve port and the pressure reducer
handwheel are undamaged, and the O-ring (Fig 4, Item 1) is in position
and undamaged.
3. Fully insert the cylinder into the carrying holster. To prevent damage,
ensure that the cylinder remains clear of the handwheel of the
pressure reducer while inserting the cylinder.
4. Align the cylinder with the pressure reducer and tighten the handwheel
(anticlockwise) hand tight (Fig 5).
5. Press the reset button (Fig 2, Item 1) to ensure that the lung demand
valve is switched off.
6. If the escape hood is not in its pouch, carefully roll the hood to ensure
that the visor is not creased. Insert the hood into the pouch with the
visor in the lower portion of the pouch as shown in Fig 1.
7. Connect the sprung-gate clip of the automatic activation device to the
hood (Fig 1, Item 6).
Component
/System
Task After
use
Every
month
Every
year
Every
10
years
Complete
equipment
Visual inspection (see
Note 1 and
Section 4.2.1)

Leak test (see
Section 4.2.6)

Functional tests (see
Note 3)

Lung
demand
valve O-ring
Check and lubricate if
necessary (see Note 2)

Pressure
reducer
Basic overhaul. Contact
Dräger for the Repair
Exchange (REX) service

Cylinder
valve
Basic overhaul. Contact
Dräger for the Repair
Exchange (REX) service

Cylinder Charge to correct
pressure (see
Section 4.2.5)

Cylinder pressure test
and recertification; and
check the test date on
the cylinder (composite
cylinders over 15 years
old must be retired)
Carry out in line with national
regulations
!
4.2.3 Removing the cylinder
WARNING
High-pressure air release may cause injury to the user or other
personnel near the breathing apparatus. Ensure that the locking
clip is fitted in the cylinder valve and fully vent the system before
attempting to disconnect the air cylinder.
When the cylinder valve outlet is open to atmosphere (not connected
to the pressure reducer or a charging adaptor), do not remove the
locking clip unless the cylinder is completely discharged. Removing
the locking clip would immediately exhaust high-pressure cylinder
air. To fully discharge the cylinder, see Section 4.2.4.
1. Ensure that the locking clip is fitted in the cylinder valve (Fig 1, Item 1),
and disconnect the sprung-gate clip (Fig 1, Item 6) from the hood.
2. Press the front button (Fig 2, Item 2) to fully vent the system.
3. Disconnect the pressure reducer from the cylinder valve.
4. Carefully remove the cylinder from the holster. To prevent damage, ensure
that the handwheel of the pressure reducer remains clear of the cylinder.
4.2.4 Fully discharging the air cylinder
1. Ensure that the pressure reducer is connected to the cylinder valve.
2. Remove the locking clip from the cylinder valve (Fig 6).
3. Press the release button and remove the lung demand valve from the
hood (Fig 7).
4. Press the front button (Fig 2, Item 2) of the lung demand valve and
allow the cylinder to fully vent. Do not direct the air toward the face,
eyes or skin.
5. When the cylinder is empty, reconnect the lung demand valve to the
hood. Check the security of the attachment by gently attempting to pull
the coupling apart.
4.2.5 Air cylinder charging
WARNING
Air quality for compressed-air cylinders shall meet the
requirements for breathing air according to EN 12021.
NOTICE
Do not use the ChargAir system for normal cylinder filling. The
ChargAir system is intended for emergency use only.
●The charging connector is a G5/8 connector as per EN 144-2:1999.
●Refer also to the instructions supplied with the cylinder and the
charging unit for recharging the cylinder.
●Only charge compressed-air cylinders which:
○Conform to national standards.
○Feature the original manufacturer’s test date and test mark.
○Have not exceeded the test date indicated on the cylinder by the
last testing station.
○Are not damaged.
●Dräger recommend a charge rate of 27 bar/minute (rapid charging will
increase the temperature resulting in an incomplete charge).
●To prevent over charging of the cylinder, Dräger recommend using a
pressure-limiting device on the charging compressor.
1. Remove the cylinder (see Section 4.2.3).
2. Connect the charging adaptor to the cylinder valve (Fig 8).
3. If the cylinder is still pressurized, approximately match the charging
line pressure to the cylinder pressure.
4. Remove the locking clip from the cylinder valve (Fig 6).
5. Recharge the cylinder to 200 bar (the pointer of the cylinder pressure
gauge inside the green area).
○Compression of cylinder air can cause a small temperature
increase during refilling, resulting in an incomplete charge. If
necessary, top-up the cylinder when it cools.
6. When the cylinder is fully charged, refit the locking clip: press the
button down against the spring and insert the locking clip into the slot
(Fig 9), ensuring that the button remains in the locked (down) position.
7. Vent pressure from the charging hose and then remove the charging
adaptor from the valve.
8. Refit the cylinder (see Section 4.2.2).
4.2.6 Leak test
WARNING
If breathing apparatus fails to meet any of the standards or
parameters described in the leak test, or if an immediate leak is
evident, there is a system fault. Report the fault to trained service
personnel or contact Dräger. Do not use the breathing apparatus
until the fault condition is rectified.
1. Ensure that the pointer of the cylinder pressure gauge is inside the
green area.
2. Press the reset button (Fig 2, Item 1) to ensure the lung demand valve
is switched off.
3. Remove the locking clip from the cylinder valve (Fig 6) to pressurize
the system and then refit the locking clip: press the button down
against the spring and insert the locking clip into the slot (Fig 9),
ensuring that the button remains in the locked (down) position.
○There should be no audible leak. If there is any leak, investigate
and repair the leak before use. If necessary, use a soapy solution
to locate the leak.
4. Press the front button (Fig 2, Item 2). A small amount of air will vent
from the system.
5. Press the reset button (Fig 2, Item 1) to switch off the air flow through
the lung demand valve.
4.3 Cleaning and disinfecting
CAUTION
Do not exceed 60 °C for drying, and remove components from the
drying facility immediately when dry. Drying time in a heated dryer
must not exceed 30 minutes.
Do not immerse pneumatic or electronic components in cleaning
solutions or water.
If water is trapped and then freezes inside the pneumatic system
of the breathing apparatus (such as the lung demand valve),
operation will be impaired. Prevent any liquid from entering, and
thoroughly dry the breathing apparatus after cleaning to prevent
this from occurring.
Do not lubricate or allow oil, grease or other contaminants to
contact ChargAir fittings or hoses.
For information about suitable cleaning and disinfecting agents
and their specifications refer to document 9100081 on
www.draeger.com/IFU.
!
!
i
i
!
!
Refer also to the Instructions for Use for the lung demand valve, face mask
and other associated equipment.
●Use only clean lint-free cloths
1. Clean the breathing apparatus manually using a cloth moistened with
cleaning solution to remove excess dirt.
2. Apply disinfecting solution to all internal and external surfaces.
3. Rinse all components thoroughly with clean water to remove all
cleaning and disinfecting agents.
4. Dry all components using a dry cloth, in a heated dryer or in air.
5. Contact service personnel or Dräger if disassembly of pneumatic or
electronic components is required.
5 Troubleshooting
There is no user troubleshooting on the PAS Colt PPH. Contact Dräger or
trained service personnel to report any issues with the product.
6 Storage
Dräger recommend that the PAS Colt PPH is stored in a cool dry
environment, in a location where it is not subject to wear or damage due to
abrasion. Note also the following:
●Extend the shoulder strap and the waist belt.
●Route rubber hoses in such a way that the bend radius is not too acute
and the hose is not stretched, compressed or twisted.
●Fix the apparatus securely to any raised mounting point to prevent it
from falling.
7 Disposal
When required, dispose of the PAS Colt PPH in accordance with national
or local regulations for waste disposal.
8 Order list
Description Quantity Order code
Dow Corning®
Molykote®111
100 grams 3331247

Other Dräger Respiratory Product manuals

Dräger X-plore 6000 User manual

Dräger

Dräger X-plore 6000 User manual

Dräger EPAS User manual

Dräger

Dräger EPAS User manual

Dräger Oxy 3000 MK II User manual

Dräger

Dräger Oxy 3000 MK II User manual

Dräger X-plore 8000 User manual

Dräger

Dräger X-plore 8000 User manual

Dräger CPS 7800 User manual

Dräger

Dräger CPS 7800 User manual

Dräger Panorama BelAir GC-RA User manual

Dräger

Dräger Panorama BelAir GC-RA User manual

Dräger X-plore 4700 User manual

Dräger

Dräger X-plore 4700 User manual

Dräger Panorama Nova PE User manual

Dräger

Dräger Panorama Nova PE User manual

Dräger PAS Micro Series User manual

Dräger

Dräger PAS Micro Series User manual

Dräger Panorama Nova S - PE/ESA User manual

Dräger

Dräger Panorama Nova S - PE/ESA User manual

Dräger X-plore 8000 User manual

Dräger

Dräger X-plore 8000 User manual

Dräger PSS 5000 Series User manual

Dräger

Dräger PSS 5000 Series User manual

Dräger X-plore 4390 User manual

Dräger

Dräger X-plore 4390 User manual

Dräger RBD 5000 User manual

Dräger

Dräger RBD 5000 User manual

Dräger FPS-COM 7000 User manual

Dräger

Dräger FPS-COM 7000 User manual

Dräger X-plore 8000 User manual

Dräger

Dräger X-plore 8000 User manual

Dräger Panorama Nova S-RP User manual

Dräger

Dräger Panorama Nova S-RP User manual

Dräger X-plore 1700 User manual

Dräger

Dräger X-plore 1700 User manual

Dräger Panorama Nova S - ESA User manual

Dräger

Dräger Panorama Nova S - ESA User manual

Dräger PAS Colt Series User manual

Dräger

Dräger PAS Colt Series User manual

Dräger Panorama Nova RP User manual

Dräger

Dräger Panorama Nova RP User manual

Dräger PAS Micro CFH User manual

Dräger

Dräger PAS Micro CFH User manual

Dräger X-plore Bayonett User manual

Dräger

Dräger X-plore Bayonett User manual

Dräger Panorama Nova Dive User manual

Dräger

Dräger Panorama Nova Dive User manual

Popular Respiratory Product manuals by other brands

Moretti KYARA PUFFY instruction manual

Moretti

Moretti KYARA PUFFY instruction manual

Bullard EVA Series user manual

Bullard

Bullard EVA Series user manual

MSA Advantage 400 Series instructions

MSA

MSA Advantage 400 Series instructions

Bal Supply 05144 User instruction manual

Bal Supply

Bal Supply 05144 User instruction manual

Rysam SN99V User instructions

Rysam

Rysam SN99V User instructions

Roscoe Medical 50040 Guide book

Roscoe Medical

Roscoe Medical 50040 Guide book

Medisana IN 525 instruction manual

Medisana

Medisana IN 525 instruction manual

Little Doctor LD-210C instruction manual

Little Doctor

Little Doctor LD-210C instruction manual

MSA alpha Series operating manual

MSA

MSA alpha Series operating manual

CleanSpace HALO quick start guide

CleanSpace

CleanSpace HALO quick start guide

ResMed VPAP S brochure

ResMed

ResMed VPAP S brochure

LiveFree N95 User instructions

LiveFree

LiveFree N95 User instructions

Air Liquide Alnest F1 Instructions for use

Air Liquide

Air Liquide Alnest F1 Instructions for use

MD feellife Air Pro VIII user manual

MD

MD feellife Air Pro VIII user manual

ESAB AIR 140 User instructions

ESAB

ESAB AIR 140 User instructions

Circadiance SleepWeaver 3D quick start guide

Circadiance

Circadiance SleepWeaver 3D quick start guide

SUNMED Nebutech 8660 user guide

SUNMED

SUNMED Nebutech 8660 user guide

Nortev Flexineb C1 Instructions for use

Nortev

Nortev Flexineb C1 Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.