endolite Smart IP User manual

938249/01-1006
Blatchford Products Limited and ENDOLITE are companies and trademarks of Chas. A. Blatchford and Sons Ltd.
Blatchford Products Ltd
Lister Road, Basingstoke, Hampshire, RG22 4AH, United Kingdom
Telephone: +44 (0)1256 316600 Facsimile: +44 (0)1256 479705
Email: [email protected] www.blatchford.co.uk
938249/01-1006
Blatchford Products Limited and ENDOLITE are companies and trademarks of Chas. A. Blatchford and Sons Ltd.
Blatchford Products Ltd
Lister Road, Basingstoke, Hampshire, RG22 4AH, United Kingdom
Telephone: +44 (0)1256 316600 Facsimile: +44 (0)1256 479705
Email: [email protected] www.blatchford.co.uk
Safety Notes:-
1. Any changes in performance should be reported to the practitioner.
2. If the device becomes immersed in liquid of any kind or exposed to severe
damp, check for function and if necessary report to practitioner immediately.
3. Not to expose the batteries to a naked flame.
4. Dis-assembly and repair of Smart IP must only be carried out by
authorised personnel.
5. Avoid strong magnetic fields, sources of electrical interference, atmospheres
containing powders and liquids and extremes of temperature.
6. Contact Blatchford Products Limited for advice if limb is to be fitted to an
amputee using a pacemaker, hearing aid or any other electronic device.
In normal service an Endolite limb should be inspected annually to check for correct function,
lubrication and security of assembly.
Safety Notes:-
1. Any changes in performance should be reported to the practitioner.
2. If the device becomes immersed in liquid of any kind or exposed to severe
damp, check for function and if necessary report to practitioner immediately.
3. Not to expose the batteries to a naked flame.
4. Dis-assembly and repair of Smart IP must only be carried out by
authorised personnel.
5. Avoid strong magnetic fields, sources of electrical interference, atmospheres
containing powders and liquids and extremes of temperature.
6. Contact Blatchford Products Limited for advice if limb is to be fitted to an
amputee using a pacemaker, hearing aid or any other electronic device.
In normal service an Endolite limb should be inspected annually to check for correct function,
lubrication and security of assembly.
User Instructions
Fine tuning the Smart IP is recommended when a new
cosmesis is fitted, for significant changes in footwear weight
and for a major change in activity.
Please make yourself familiar with the battery failure advice given on the
back of these instructions.
GB
!
Battery failure
When the batteries begin to fail you will receive a warning in the form of ten short beeps from
the limb.The limb will still be usable, and you should seek to replace the batteries at the next
convenient opportunity.
The limb will still be usable, and you should seek to replace the batteries at the next convenient
opportunity.
Upon complete battery failure, the limb will emit one long beep. The limb is still safe to use, as
a medium setting will have been selected, though speed changes will not now occur.
Seek attention from your practitioner to replace the batteries at the earliest opportunity.
Original program settings will not be lost, normal limb function will be resumed upon installation
of new batteries.
User Instructions
Fine tuning the Smart IP is recommended when a new
cosmesis is fitted, for significant changes in footwear weight
and for a major change in activity.
Please make yourself familiar with the battery failure advice given on the
back of these instructions.
GB
!
Battery failure
When the batteries begin to fail you will receive a warning in the form of ten short beeps from
the limb.The limb will still be usable, and you should seek to replace the batteries at the next
convenient opportunity.
The limb will still be usable, and you should seek to replace the batteries at the next convenient
opportunity.
Upon complete battery failure, the limb will emit one long beep. The limb is still safe to use, as
a medium setting will have been selected, though speed changes will not now occur.
Seek attention from your practitioner to replace the batteries at the earliest opportunity.
Original program settings will not be lost, normal limb function will be resumed upon installation
of new batteries.
938249/01-1006
Blatchford Products Limited and ENDOLITE are companies and trademarks of Chas. A. Blatchford and Sons Ltd.
Blatchford Products Ltd
Lister Road, Basingstoke, Hampshire, RG22 4AH, United Kingdom
Telephone: +44 (0)1256 316600 Facsimile: +44 (0)1256 479705
Email: [email protected] www.blatchford.co.uk
938249/01-1006
Blatchford Products Limited and ENDOLITE are companies and trademarks of Chas. A. Blatchford and Sons Ltd.
Blatchford Products Ltd
Lister Road, Basingstoke, Hampshire, RG22 4AH, United Kingdom
Telephone: +44 (0)1256 316600 Facsimile: +44 (0)1256 479705
Email: [email protected] www.blatchford.co.uk
Safety Notes:-
1. Any changes in performance should be reported to the practitioner.
2. If the device becomes immersed in liquid of any kind or exposed to severe
damp, check for function and if necessary report to practitioner immediately.
3. Not to expose the batteries to a naked flame.
4. Dis-assembly and repair of Smart IP must only be carried out by
authorised personnel.
5. Avoid strong magnetic fields, sources of electrical interference, atmospheres
containing powders and liquids and extremes of temperature.
6. Contact Blatchford Products Limited for advice if limb is to be fitted to an
amputee using a pacemaker, hearing aid or any other electronic device.
In normal service an Endolite limb should be inspected annually to check for correct function,
lubrication and security of assembly.
Safety Notes:-
1. Any changes in performance should be reported to the practitioner.
2. If the device becomes immersed in liquid of any kind or exposed to severe
damp, check for function and if necessary report to practitioner immediately.
3. Not to expose the batteries to a naked flame.
4. Dis-assembly and repair of Smart IP must only be carried out by
authorised personnel.
5. Avoid strong magnetic fields, sources of electrical interference, atmospheres
containing powders and liquids and extremes of temperature.
6. Contact Blatchford Products Limited for advice if limb is to be fitted to an
amputee using a pacemaker, hearing aid or any other electronic device.
In normal service an Endolite limb should be inspected annually to check for correct function,
lubrication and security of assembly.
User Instructions
Fine tuning the Smart IP is recommended when a new
cosmesis is fitted, for significant changes in footwear weight
and for a major change in activity.
Please make yourself familiar with the battery failure advice given on the
back of these instructions.
GB
!
Battery failure
When the batteries begin to fail you will receive a warning in the form of ten short beeps from
the limb.The limb will still be usable, and you should seek to replace the batteries at the next
convenient opportunity.
The limb will still be usable, and you should seek to replace the batteries at the next convenient
opportunity.
Upon complete battery failure, the limb will emit one long beep. The limb is still safe to use, as
a medium setting will have been selected, though speed changes will not now occur.
Seek attention from your practitioner to replace the batteries at the earliest opportunity.
Original program settings will not be lost, normal limb function will be resumed upon installation
of new batteries.
User Instructions
Fine tuning the Smart IP is recommended when a new
cosmesis is fitted, for significant changes in footwear weight
and for a major change in activity.
Please make yourself familiar with the battery failure advice given on the
back of these instructions.
GB
!
Battery failure
When the batteries begin to fail you will receive a warning in the form of ten short beeps from
the limb.The limb will still be usable, and you should seek to replace the batteries at the next
convenient opportunity.
The limb will still be usable, and you should seek to replace the batteries at the next convenient
opportunity.
Upon complete battery failure, the limb will emit one long beep. The limb is still safe to use, as
a medium setting will have been selected, though speed changes will not now occur.
Seek attention from your practitioner to replace the batteries at the earliest opportunity.
Original program settings will not be lost, normal limb function will be resumed upon installation
of new batteries.
P
R
O
G
R
A
M
vary speed gradually for approx. 200 steps
Fully flex Fully extendFully flex Fully extendFully extend
HOLD
HOLD
refrain from sitting during programming
Fully extendFlex more than 40°
END
START
SAVE AND EXIT
Programming
Fine Tuning
x2
x2
4 secs
2 secs
maximum 4 seconds
vary speed gradually for approx. 200 steps
Fully flex Fully extendFully flex Fully extendFully extend
HOLD
HOLD
refrain from sitting during programming
Fully extendFlex more than 40°
END
START
SAVE AND EXIT
Programming
Fine Tuning
x2
x2
4 secs
2 secs
maximum 4 seconds

Other endolite Medical Equipment manuals

endolite AvalonK2 User manual

endolite

endolite AvalonK2 User manual

endolite linx User manual

endolite

endolite linx User manual

endolite Comfort Cushion Liner CTTCP22L User manual

endolite

endolite Comfort Cushion Liner CTTCP22L User manual

endolite EchelonVAC User manual

endolite

endolite EchelonVAC User manual

endolite AqualimbTT User manual

endolite

endolite AqualimbTT User manual

endolite orion 3 User manual

endolite

endolite orion 3 User manual

endolite Smart IP System SMARTIPSFPYR User manual

endolite

endolite Smart IP System SMARTIPSFPYR User manual

endolite Avalon K2 User manual

endolite

endolite Avalon K2 User manual

endolite Echelon User manual

endolite

endolite Echelon User manual

endolite aqualimb TF User manual

endolite

endolite aqualimb TF User manual

endolite Navigator Foot User manual

endolite

endolite Navigator Foot User manual

endolite Silcare Breathe User manual

endolite

endolite Silcare Breathe User manual

endolite BladeXT User manual

endolite

endolite BladeXT User manual

endolite 4 Bolt ESK+ 239150 User manual

endolite

endolite 4 Bolt ESK+ 239150 User manual

Popular Medical Equipment manuals by other brands

Medela Clario Instructions for use

Medela

Medela Clario Instructions for use

Fresenius Medical Care DIASAFE 2008 K Technical manual

Fresenius Medical Care

Fresenius Medical Care DIASAFE 2008 K Technical manual

Xstrahl 100 Operator's manual

Xstrahl

Xstrahl 100 Operator's manual

HealthCo Paloma user manual

HealthCo

HealthCo Paloma user manual

ProMedTek Replexa+ instruction manual

ProMedTek

ProMedTek Replexa+ instruction manual

Seca 545 Series Service manual

Seca

Seca 545 Series Service manual

bort medical 104 800 quick guide

bort medical

bort medical 104 800 quick guide

Joerns Hoyer HPL450 manual

Joerns

Joerns Hoyer HPL450 manual

Mamava Solo Assembly guide

Mamava

Mamava Solo Assembly guide

Mizuho MOT-3602X Operator's manual

Mizuho

Mizuho MOT-3602X Operator's manual

Resvent RXiBreeze user manual

Resvent

Resvent RXiBreeze user manual

GCE druva 200 Instructions for use

GCE druva

GCE druva 200 Instructions for use

medi protect.Ankle air Instructions for use

medi

medi protect.Ankle air Instructions for use

Arjo CARENDO Instructions for use

Arjo

Arjo CARENDO Instructions for use

Storz Image1 S H3-Z FI TH 102 Service instructions

Storz

Storz Image1 S H3-Z FI TH 102 Service instructions

ST&G 1321 AP / AL instructions

ST&G

ST&G 1321 AP / AL instructions

Petermann Alpha Caremaster PM-2095 L manual

Petermann

Petermann Alpha Caremaster PM-2095 L manual

Bauerfeind uniprox JT30 instruction manual

Bauerfeind

Bauerfeind uniprox JT30 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.