
15
INIT-Initialize Memory
The INIT menu allows the operator to selectively clear data stored for program, calibration and/or clock functions.
To select one of the functions, simply press the SET key when the upper display shows the INIT menu. Next, use
the UP/Down arrow keys to select which data to clear. The choices are as follows:
1. VAR Miscellaneous Variables
2. PRG Program Data
3. TIM Time Data
4. CBR Calibration Data
Finally, with the desired data type displayed in the lower display, press the SET key again to reset the data to the
factory set values. The lower display will show CLR indicating the data has been cleared.
TCCB-Thermocouple Calibration
Before beginning the temperature calibration, place a NIST calibrated thermometer in the center of the vacuum
oven chamber and close the door. Press the MENU key until the upper display shows TCCB, then press the SET
key. The unit enters the thermocouple calibration mode and displays the first temperature set-point (flashing).
Use the UP/Down arrow keys to select the desired temperature. Press the SET key to accept the first desired
calibration point. Allow the unit ample time to reach each set point. A 50°C calibration point will take
approximately 2 hours to stabilize. When the chamber temperature has completely stabilized, record the
temperature of the thermometer in the chamber center. Using that value as the reference, use the UP/DOWN
keys to match the oven display temperature to the reference temperature. Press the SET key. The controller will
WAIT and SAMPLE values for the low end calibration, then displays the second temperature set-point (flashing).
Use the UP/DOWN arrow keys to adjust the desired second set-point. Press the SET key to enter the second set-
point. Allow the unit time to heat and stabilize at the second set-point. When the unit is stable, record the
temperature of the reference thermometer. Adjust the display to match the reference thermometer using the
UP/DOWN arrow keys. Press the SET key. The controller will then SAMPLE this data. When the controller is
finished, the next menu will appear (VCMC).
VCMC-Vacuum Calibration
Before calibrating the vacuum readout system, it is necessary that the barometric or atmospheric pressure be
known. One way of doing this is to contact the U.S. Weather Bureau in the installation area. The bureau will
provide the barometric pressure corrected to sea level, this pressure must then be corrected to the altitude where
the calibration will be performed. This can be done by interpolating the data from Table 9. Once a pressure for
the calibration altitude is obtained, the value is subtracted from the sea level value on Table 9 to obtain the
correction factor. The correction factor will then be subtracted from the value obtained from the bureau.
Obviously, the corrected value obtained will be only as accurate as the accuracy to which the exact elevation is
known.
Example: The barometric pressure corrected to sea level, as reported by the U.S. Weather Bureau on June 13,
1983, was 30.19 In. of Hg, in Pittsburgh, PA. The elevation at the calibration site, also in Pittsburgh, is 850 ft.
above sea level. Interpolating the data from Table the pressure at 850 ft. is 29.02 In. of Hg; subtracting this value
from 29.92 (sea level pressure) gives a correction factor of 0.9. The actual pressure at the calibration site is:
30.19 -0.9 = 29.29 In of Hg.
Another method is to read the pressure directly from a temperature compensated barometer that can be adjusted
for altitude. This method will give the pressure at the elevation of the calibration site without the need for any
additional calculations or corrections. Generally, the pressure value measured should fall within the 29 to 30
inches-of-mercury range. In fact, measurements outside this range should be suspect.
Slight variations will exist usually because of pressure variations due to weather conditions. Therefore, the most
accurate way to calibrate the Model 282A vacuum readout system is to first obtain the atmospheric pressure using
one of the methods above.
NOTE: The procedure below for calibrating the vacuum display assumes the vacuum pump is being controlled
by the Auxiliary Contacts, Tabs 3 & 4. The vacuum display can be calibrated with manual pump operation,
as well.
To calibrate the vacuum display perform the following:
1. While in the VCMC calibration mode, open the vacuum (full-CCW) valve to expose the chamber to atmospheric
pressure.
2. Connect a suitable vacuum pump to the PURGE port along with a vacuum gauge as shown in Figure 5.
Figure 5: Vacuum Pump Hook-Up