FOR-A COMFORT plus G30a User manual

Blood Glucose Monitoring System
Blutzucker-Überwachungsssystem
Système de contrôle de la glycémie
Sistema per il monitoraggio della glicemia
COMFORT plus G30a
Owner’s Manual
Bedienungsanleitung
Manuel de l’utilisateur
Manuale dell’utente


EN-1
Safety Information
Read the following Safety Information thoroughly before using
the device.
• Use this device ONLY for the intended use described in this
manual.
• Do NOT use accessories which are not specied by the
manufacturer.
• Do NOT use the device if it is not working properly or
damaged.
• This device does NOT serve as a cure for any symptoms or
diseases. The data measured is for reference only. Always
consult your doctor to have the results interpreted.
• The blood glucose test strip can NOT be used for the
testing of newborns.
• Before using this device to test blood glucose, read all
instructions thoroughly and practice the test. Carry out all
the quality control checks as directed.
• Keep the device and testing supplies away from young
children. Small items such as the battery cover, batteries,
test strips, lancets and vial caps are choking hazards.
• The use of this instrument in a dry environment, especially
if synthetic materials are present (synthetic clothing,
carpets etc.) may cause damaging static discharges that
may cause erroneous results.
• Do NOT use this instrument in close proximity to sources
of strong electromagnetic radiation, as these may interfere
with the correct operation.
• Proper maintenance as well as timely calibration of the
device together with the control solution is essential in
ensuring the longevity of your device. If you are concerned
about the accuracy of the measurement, please contact
the place of purchase or customer service representative
for assistance.
KEEP THESE INSTRUCTIONS IN A SAFE PLACE
Important Information
• Severe dehydration and excessive water loss may cause
readings which are lower than actual values. If you believe
you are suering from severe dehydration, consult a
healthcare professional immediately.
• If your blood glucose results are lower or higher than usual,
and you do not have symptoms of illness, rst repeat the
test. If you have symptoms or continue to get results higher
or lower than usual, follow the treatment advice of your
healthcare professional.
• Use only fresh whole blood sample to test your blood
glucose. Using other substances will lead to incorrect
results.
• If you are experiencing symptoms that are inconsistent
with your blood glucose test results and you have followed
all instructions described in this owner’s manual, call your
healthcare professional.

EN-2
Introduction
Intended Use
This system is intended for use outside the body (in vitro
diagnostic use) by people with diabetes at home and by
health- care professionals in clinical settings as an aid to
monitor the eectiveness of diabetes control. It is intended
to be used for the quantitative measurement of glucose
(sugar) in fresh capillary (from the nger, palm, forearm, and
upper arm) blood samples. It should not be used for the
diagnosis of diabetes, or testing on newborns.
Use only heparin for anticoagulation of whole blood.
Test Principle
Your system measures the amount of sugar (glucose)
in whole blood. The glucose testing is based on the
measurement of electrical current generated by the reaction
of glucose with the reagent of the strip. The meter measures
the current, calculates the blood glucose level, and displays
the result. The strength of the current produced by the
reaction depends on the amount of glucose in the blood
sample.
• We do not recommend using this product on severely
hypotensive individuals or patients in shock. Readings
which are lower than actual values may occur for
individuals experiencing a hyperglycaemic-hyperosmolar
state, with or without ketosis. Please consult the healthcare
professional before use.
• The measurement unit used for indicating the
concentration of blood glucose can have mg/dL or mmol/L.
The approximate calculation rule for conversion of mg/dL
in mmol/L is:
mg/dL Divided by 18 = mmol/L
mmol/L Times 18 = mg/dL
For example:
1)120 mg/dL ÷ 18 = 6.6 mmol/L
2)7.2 mmol/L x 18 = 129 mg/dL approximately

EN-3
Product Overview
1
3
2
4
5
6
7
1. Test Slot
2. Display Screen
3. MButton
4. CButton
5. Data Port
6. Test Strip Ejector
7. SET Button
Screen Display
12
11
13
14
1
4
3 2
8
6
7
5
910
1. Low Battery Symbol
2. Test Strip Symbol
3. Blood Drop Symbol
4. Test Result
5. Face/Low/High Symbol
6
.
Memory Symbol
7. Alarm Symbol
8. Measuring Mode
9. Date
10.Time
11. Ketone Warning
12. Error Message
13. Day Average
14. Measurement Unit

EN-4
Getting Started
Initial Setup
Before using your meter for the rst time or if you change the
meter battery, you should check and update these settings.
Make sure you complete the steps below and have your
desired settings saved.
►Entering the Setting Mode.
Start with the meter o (no test strip inserted). Press SET to
turn on the meter.
BATTERY
SET
Step 1: Setting the date.
With the year ashing, press M until the correct year appears.
Press SET.
With the month ashing, press M until the correct month
appears. Press SET.
With the day ashing, press M until the correct day appears.
Press SET.
Step 2: Setting the time format.
Press M and release to select the desired time format --- 12h
or 24h. Press SET.
Step 3: Setting the time.
With the hour ashing, press M until the correct hour
appears. Press SET.
With the minute ashing, press M until the correct minute
appears. Press SET.

EN-5
Step 4: Setting the Universal Tone
You may set up Beep On, Universal Tone On and Beep O by
press Mto change it.
When Universal Tone is turned on, the meter guides you
through the blood glucose test using beep tones; it also
outputs the result as a series of beeps.
When Beep is turned o, the alarm function will remain
eective.
Step 5: Deleting the memory.
With“dEL” and“ ”symbol on the display, press Mand select “no”
to keep the results in memory then press SET to skip.
To delete all the results, press Mand select“YES”.
Then press SET button to delete all memory.“OK” is displayed
in the meter, which indicates that all data stored is deleted.
BATTERY
SET
Step 6: Setting the reminder alarm.
You may set up any or all of the reminder alarms (1-4). The
meter displays“On”or “OFF”and “ ” , press Mto turn on or
turn o to set the rst reminder alarm.
Press Mto select“On”, then press SET to set the hour. When
the hour is ashing, press Mto add an hour. Press SET to
conrm and go to minutes, press Mto add one minute. Hold
Mlonger to add faster. Press SET to conrm and go to the
next alarm setting.

EN-6
settings before shutdown.
Congratulations! You have completed all settings!
Note:
• These parameters can ONLY be changed in the setting
mode.
• If the meter is idle for 3 minutes during the setting mode, it
will switch o automatically.
Before Testing
Control Solution Testing
Our Control Solution contains a known amount of glucose
that reacts with test strips and is used to ensure your device
and test strips are working together correctly.
Test strips, control solutions, or sterile lancets may not
be included in the kit (please check the contents on your
product box). They can be purchased separately. Please make
sure you have those items needed for a blood glucose test
beforehand.
Do a control solution test when:
you rst receive the device.
you begin using a new vial of test strips.
you suspect the device or test strips are not working
properly.
your blood glucose test results are not consistent with
BATTERY
SET
If you do not want to set an alarm, press SET to skip this step.
BATTERY
SET
If you want to turn o an alarm, nd the alarm number by
pressing SET in the setting mode, press Mto change from
“ON’to“OFF”.
At the time of your alarm, the meter will beep and
automatically turn on. You can press Mto silence the alarm
and insert a test strip to begin testing. If you do not press M,
the meter will beep for 2 minutes then switch o. If you do
not want to test at this time, press Mto switch o the meter.
Press SET to turn o the meter. The meter repeats all your

EN-7
how you feel, or if you think the results are not accurate.
you have dropped or think you may have damaged the
device.
To perform the control solution test, do the following:
Step 1: Insert the test strip to turn on the meter.
Insert the test strip into the meter. Wait for the meter to
display the test strip and blood drop symbols.
Step 2: Press C to mark this test as a control solution test.
If you press Cagain, the“QC” will disappear and this test is no
longer a control solution test.
Step 3: Apply control solution.
Shake the control solution vial thoroughly before use.
Squeeze out a drop and wipe it o, then squeeze out another
drop and place it on the tip of the vial cap. Hold the meter
to move the absorbent hole of the test strip to touch the
drop. Once the conrmation window lls completely, the
meter will begin counting down. To avoid contaminating the
control solution do not directly apply control solution onto
a strip.
Step 4: Read and compare the result.
After counting down to 0, the control solution test result will
appear on the display. Compare this result with the range
printed on the test strip vial and it should fall within this
range. If not, please read the instructions again and repeat
the control solution test.

EN-8
Testing Your Blood Glucose
Test Strip Appearance
Absorbent Hole
Apply a drop of blood
here. The blood will be
automatically absorbed.
Conrmation Window
This is where you conrm
if enough blood has been
applied to the absorbent hole
in the strip.
Test Strip Handle
Hold this part to insert
the test strip into the slot.
Contact Bars
Insert this end of the test strip
into the meter. Push it in rmly
until it will go no further.
Front
ATTENTION:
The front side of test strip with FORA
Logo should face up when inserting test
strip. Test results might be wrong if the
contact bar is not fully inserted into the
test slot.
CODE
109~165
Note:
• Do NOT test your blood.
• The control solution range printed on the test strip vial or
individual foil pack is for control solution use only. It is not a
recommended range for your blood glucose level.
• Refer to the Maintenance section for important
information about your control solutions.
Out-of-range results:
If you continue to get results that fall outside the range
printed on the test strip vial, it means that the meter and
strips might not be working properly.

EN-9
Inserting a Test Strip
Insert the test strip into its slot.
Important!
The front side of test strip should face up when inserting the
test strip. Test results might be wrong if the contact bar is not
fully inserted into the test slot.
Important!
To reduce the chance of infection:
• Never share a lancet or a lancing device.
• Always use a new, sterile lancet. Lancets are for single use
only.Avoid getting hand lotion, oils, dirt, or debris in or on
the lancets and the lancing device.
Preparing the Lancing Device
1.Remove the cap.
2.Insert a new lancet rmly into the white lancet holder cup.
3.Remove the protective disk on the lancet.
Hold the lancet rmly in place and twist o the protective
disk.
4.Replace the cap until it snaps or clicks into place.
5.Rotate the dial to set the desired lancing depth.
6.Pull the cocking control out until the orange bar appears
on the release button window.
Obtaining the Blood Sample
Please follow the suggestions below before obtaining a drop
of blood:
• Wash and dry your hands before starting.
• Select the puncture site either on your ngertips or other
body parts.

EN-10
• Rub the puncture site for about 20 seconds before
penetration.
Blood from the ngertip
1.Press the lancing device tip rmly against the lower side of
your ngertip.
2.Press the release button to prick your nger. A click
indicates that the puncture is complete.
Blood from sites other than the ngertip
Alternative site testing (AST) is when individuals check their
blood glucose levels using other areas of the body other
than the ngertips. The FORA test strips allow AST to be
performed on sites other than the ngertips. Please consult
your health care professional before you begin AST.
We strongly recommend that you perform AST ONLY at the
following times:
• During a pre-meal or fasting state (more than 2 hours since
the last meal).
• Two hours or more after taking insulin.
• Two hours or more after exercise.
Do NOT use AST if:
• You think your blood glucose is low.
• You may not notice if you are hypoglycemic.
• Your AST results are inconsistent with the way you feel.
• You are testing for hyperglycemia.
• Your routine glucose results often uctuate.
To obtain a blood sample from the alternative sites, please
rub the puncture site for approximately 20 seconds.
1.Replace the lancing device cap with the clear cap.
1
2
2.Pull the cocking control out until the orange bar appears
on the release button window.

EN-11
Important!
• Choose a dierent spot each time you test. Repeated
punctures at the same spot may cause soreness and
calluses.
• Avoid lancing the areas with obvious veins to avoid
excessive bleeding.
• It is recommended to discard the rst drop of blood as it
might contain tissue uid, which may aect the test result.
Make Blood Glucose Test
1.Insert the test strip into the test slot of the device. Wait for
the device to display the blood drop“ “.
2.Select the appropriate measuring mode by pressing M.
• General Tests - any time of day without regard to time
since the last meal.
• AC ( ) - no food intake for at least 8 hours.
• PC ( ) - 2 hours after a meal.
• QC ( ) - testing with control solution.
3.Obtain a blood sample.
Use the pre-set lancing device to puncture your desired
site. After penetration, discard the rst drop of blood with a
clean tissue or cotton. Gently squeeze the punctured area
to obtain another drop of blood. Be careful NOT to smear
the blood sample. The volume of blood sample must be at
least 0.5 microliter (μL) for blood glucose testing.
4.Apply the blood sample.
Move your nger to meet the absorbent hole of the test
strip and the drop will be automatically be drawn onto
the test strip. Remove your nger until the conrmation
window is lled. The meter begins to count down. Do not
remove your nger until you hear a beep sound.
5.Read your result.
The results of your blood glucose test will appear after
the meter counts down to 0. The results will be stored
automatically in the meter memory.

EN-12
Announcement of Test Result by Universal
Tone
The blood glucose results will be broken down into individual
digits and each digit represents the corresponding number
of beeps.
The result is announced three times in succession and each
time is preceded by two quick beeps. So you will hear: 2
quick beeps – results – 2 quick beeps – results – 2 quick
beeps – result.
For mg/dL meters, the hundreds are always announced, even
when the result is below 100.
Examples:
80 mg/dL is announced as 1 long beep (0) – 1 single pause –
8 single beeps (8) – 1 single pause – 1 long beep (0)
182 mg/dL is announced as 1 single beep (1) – 1 single pause
– 8 single beeps (8) – 1 single pause – 2 single beeps (2)
For mmol/L meters, the tens are always announced, even
when the result is below 10. The decimal point is represented
by 1 quick beep.
Examples:
6.0 mmol/L is announced as 1 long beep (0) – 1 single pause
– 6 single beeps (6) – 1 single pause – 1 quick beep (.) – 1
single pause – 1 long beep (0)
Note:
Information or warnings in the form of symbols displayed
together with the results are not announced acoustically.
Eject the Used Test Strip and Remove the
Lancet
To eject the test strip, point the strip at a disposal container
for sharp items. The meter will switch itself o automatically
after the test strip is ejected.
Important!
The used lancet and test strip may be biohazards. Please
consult your health-care provider for proper disposal which
complies with your local regulations.

EN-13
Reviewing Test Results
Your device stores the 450 most recent test results along with
respective dates and times in its memory. To enter the device
memory, start with the device switched o.
To review all test results, do the following:
Step 1: Press M.
“”will appear on the display. Press Magain, and the rst
reading you see is the last blood glucose result along with
date, time and the measuring mode.
Step 2: Press M to recall the test results stored in the meter
each time you press.
Step 3: Exit the meter memory.
After the last test results, press M again and the meter will be
turned o.
To review the day-average test results, do the following:
Step 1: Press M.
When “”appears on the display, keep pressing Mfor 3
seconds until the ashing“ ”appears. Release Mand
then your 7-day average result measured in general mode
will appear on the display.

EN-14
Step 2: Press M to review 14-, 21-, 28-, 60- and 90- day
average results stored in each measuring mode in the order
of Gen, AC, then PC.

EN-15
Step 3: Exit the meter memory.
Keep pressing Mand the meter will turn o after displaying
the last test result.
Note:
• Any time you wish to exit the memory, keep pressing Mfor 5
seconds or leave it without any action for 3 minutes. The meter
will switch o automatically.
• Control solution results are NOT included in the day average.
• If using the meter for the rst time,“---” displays when you recall
the test results or review the average result. It indicates that
there is no test result in the memory.
Announcement of Memory Result by
Universal Tone
Only the most recent result that was saved can be
announced acoustically. If you press M to turn the meter on,
you will rst hear the Long-Beep which stands for power-on
and then the most recent result.
Only the average for the last 7 days is announced
acoustically. If the 7-day average cannot be calculated, three
horizontal bars are displayed. This is signaled acoustically
with 3 long beeps representing 3 zeroes.
Transferring Data
You can use the meter with an interface cable and the Health
Care Software System to view your test results on your
personal computer.
To learn more about the Health Care Software System or to
obtain an interface cable separately, please contact local
customer services or the place of purchase for assistance.
Data Transmission Via RS-232 Cable
Step 1: Obtaining the required cable and installing the
software.
To download the Health Care Software System, please visit
the ForaCare Suisse AG, website at www.foracare.ch.

EN-16
Step 2: Connecting to a personal computer.
Connect the cable to a serial port on your computer. With the
meter switched o, connect the other end of the interface
cable to the meter data port. “PC” will appear on the meter
display, indicating that the meter is in communication mode.
Step 3: Data transmission.
To transmit data, follow the instructions provided with the
software. Results will be transmitted with date and time.
Remove the cable and the meter will automatically switch o.
WARNING!
While the meter is connecting to the PC, it will be unable to
perform a blood glucose test.
Maintenance
Battery
Your meter comes with one 3V CR2032 lithium battery.
►Low Battery Signal
The“ ” symbol appears along with display messages: The
meter is functional and the result remains accurate, but it is
time to change the battery.
The“ ” symbol appears with E-b, Error and low:
The power is not enough to do a test. Please change the
battery immediately.automatically be switched o.
Low Battery Warning:
When the batteries will soon be exhausted, 2 quick beeps
are announced three times in succession. This warning is
sounded when the meter is switched on.

EN-17
►Replacing the Battery
To replace the battery, make sure that the meter is turned
o.
Step 1: Press the edge of the battery cover and lift it up to
remove.
Step 2: Remove the old battery and replace with one 3V
CR2032 lithium battery.
Step 3: Close the battery cover. If the battery is inserted
correctly, you will hear a“beep”afterwards.
Note:
• Replacing the battery does not aect the test results stored in
the memory.
• As with all small batteries, these batteries should be kept
away from small children. If swallowed, promptly seek medical
assistance.
• Battery might leak chemicals if unused for a long time. Remove
the battery if you are not going to use the device for an
extended period (i.e., 3 months or more).
• Properly dispose of the battery according to your local
environmental regulations.
Caring for Your Device
• To clean the exterior of the device, wipe it with a cloth
moistened with tap water or a mild cleaning agent, then
dry the device with a soft dry cloth. Do NOT rinse with
water.
• Do NOT use organic solvents to clean the device.
Device Storage
• Storage condition: -20°C to 60°C (-4°F to 140°F), below 95%
relative humidity.
• Always store or transport the device in its original storage
case.
• Avoid dropping and heavy impact.
• Avoid direct sunlight and high humidity.
Meter Disposal
The used meter should be treated as contaminated and may
carry a risk of infection during measurement. The batteries in
this used meter should be removed and the meter should be
disposed in accordance with local regulations.
The meter falls outside the scope of the European Directive
2002/96/ EC-Directive on waste electrical and electronic
equipment (WEEE).

EN-18
Caring for Your Test Strips
• Storage condition: 4°C to 40°C (39.2°F to 104°F) for glucose
test strip, below 85% relative humidity. Do NOT freeze.
• Store your test strips in their original vial only. Do not
transfer to another container.
• Store test strip packages in a cool and dry place. Keep away
from direct sunlight and heat.
• After removing a test strip from the vial, immediately close
the vial cap tightly.
• Touch the test strip with clean and dry hands.
• Use each test strip immediately after removing it from the
vial.
• Write the opening date on the strip vial label when you rst
open it. Discard remaining test strips after 6 months.
• Do not use test strips beyond the expiry date. This may
cause inaccurate results.
• Do not bend, cut, or alter a test strip in any way.
• Keep the strip vial away from children since the cap and the
test strip may be a choking hazard. If swallowed, promptly
see a doctor for assistance.
For further information, please refer to the test strip package
insert.
Important Control Solution Information
• Use only our control solutions with your device.
• Do not use the control solution beyond the expiry date or
3 months after rst opening. Write the opening date on the
control solution vial and discard the remaining solution
after 3 months.
• It is recommended that the control solution test be done
at room temperature 20°C to 25°C (68°F to 77°F). Make
sure your control solution, device, and test strips are at this
specied temperature range before testing.
• Shake the vial before use, discard the rst drop of control
solution, and wipe o the dispenser tip to ensure a pure
sample and an accurate result.
• Store the control solution tightly closed at temperatures
between 2°C to 30°C (35.6°F to 86°F). Do NOT freeze.
Reference Values
The device provides you with plasma equivalent blood
glucose results.
Time of day Normal plasma glucose range
for people without diabetes
(mg/dL)
Fasting and before
meals
< 100 mg/dL (5.6 mmol/L)
2 hours after meals < 140 mg/dL (7.8 mmol/L)
Source: American Diabetes Association (2012). Clinical Practice
Recommendations. Diabetes Care, 35 (Supplement 1): S1-100.
Please consult your doctor to determine a target range
that works best for you.
Table of contents
Languages:
Other FOR-A Blood Glucose Meter manuals

FOR-A
FOR-A MD Reference guide

FOR-A
FOR-A GD20FM01 Reference guide

FOR-A
FOR-A V10 User manual

FOR-A
FOR-A Diamond PRIMA User manual

FOR-A
FOR-A Premium V10a User manual

FOR-A
FOR-A Diamond Mini User manual

FOR-A
FOR-A DM40 User manual

FOR-A
FOR-A GD20 User manual

FOR-A
FOR-A 6 Connect User manual

FOR-A
FOR-A D10 User manual

FOR-A
FOR-A V30a Instruction sheet

FOR-A
FOR-A GD40 User manual

FOR-A
FOR-A D40 Series User manual

FOR-A
FOR-A G71a User manual

FOR-A
FOR-A Diamond Mini User manual

FOR-A
FOR-A V12 User manual

FOR-A
FOR-A Advanced pro GD40g User manual

FOR-A
FOR-A MD User manual

FOR-A
FOR-A Diamond CUFF BP User manual

FOR-A
FOR-A D15 User manual